K112303 · Boston Scientific · DQY · Nov 9, 2011 · Cardiovascular
Device Facts
Record ID
K112303
Device Name
RUBICON SUPPORT CATHETER
Applicant
Boston Scientific
Product Code
DQY · Cardiovascular
Decision Date
Nov 9, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Rubicon Support Catheter is intended to facilitate placement and support of guidewires and other interventional devices within the peripheral vasculature and to allow for exchange of guidewires, and provide a conduit for the delivery of saline or contrast solutions.
Device Story
Rubicon Support Catheter is a multipurpose intravascular device used in peripheral vasculature. It features an ultra-low profile tip, hydrophilic coating on the distal 40 cm, and 3 radiopaque markers for positioning. The device includes a proximal female luer-lock port for guidewire entry and fluid injection. It is compatible with 0.014-inch guidewires. Used by physicians in clinical settings, the catheter is advanced over a pre-positioned guidewire or through an introducer sheath to a lesion site. It supports guidewires or other CTO devices during procedures, facilitates wire exchanges, and allows for saline or contrast injection. By providing a stable conduit, it assists in navigating complex vascular anatomy, potentially improving procedural success in interventional peripheral vascular cases.
Clinical Evidence
Bench testing only. Performance testing included biocompatibility (MEM elution, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity, hemolysis, hemocompatibility, complement activation), effective length, sheath insertion/withdrawal force, shaft burst pressure, tensile strength, kink resistance, torque strength, radiopacity, and coating integrity. No clinical data was required.
Technological Characteristics
Intravascular catheter; single lumen; hydrophilic coating (distal 40 cm); 3 radiopaque markers; proximal luer-lock port. Compatible with 0.014-inch guidewires. Max infusion pressure 300 psi. Materials and manufacturing processes are consistent with predicate devices. Sterilization method not specified.
Indications for Use
Indicated for use in the peripheral vasculature to facilitate placement and support of guidewires and interventional devices, enable guidewire exchange, and serve as a conduit for saline or contrast delivery.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Spectranetics Quick-Cross Support Catheter (K033678)
Submission Summary (Full Text)
{0}------------------------------------------------
Traditional 510(k) Submission Rubicon™ Support Catheter
0. 1 >& 3
NOV - 9 2011
# 510(k) Summary
per 21 CFR §807.92
| Submitter's Name and Address | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name and Information | Glenn Jacques<br>Principal, Regulatory Affairs<br>Phone: 763-494-1152<br>Fax: 763-494-2222<br>e-mail: jacquesg@bsci.com |
| Date Prepared | 10 August 2011 |
| Proprietary Name | Boston Scientific Rubicon™ Support Catheter |
| Common Name | Percutaneous Catheter |
| Product Code | DQY |
| Classification | Class II, 21 CFR Part 870.1250 |
| Predicate Device | Spectranetics<br>Quick-Cross Support Catheter<br>K033678 23 February 2004 |
| Device Description | The Boston Scientific Rubicon Support Catheters are multipurpose intravascular devices. The catheters feature an ultra low profile tip, a lubricious hydrophilic coating that is applied to the surface of the distal 40 cm of the catheter, and 3 radiopaque markers spaced equally along the distal shaft which aid in estimating geometry within the vascular system. The distal radiopaque marker is positioned approximately 2mm away from the distal catheter tip. The proximal portion of the catheter includes one female luer-lock port connected to the proximal end of the catheter for guidewire entry and fluid injection. The Rubicon Support Catheters are available in 135cm and 150cm lengths. The shafts have varying stiffness and diameters. The device has a proximal shaft diameter of 3.4 Fr. tapering to a distal shaft diameter of 2.1Fr. The catheter is compatible with 0.014 in (0.36 mm) guidewires. |
| Intended Use of Device | The Rubicon Support Catheters are multipurpose intravascular devices that can be used for wire exchanges, saline, contrast injection and to support a guidewire or other CTO (Chronic Total Occlusion) devices. The Support Catheter can be back loaded over a pre-positioned guidewire or may be introduced through a previously positioned appropriately sized introducer sheath and advanced to the targeted area of the lesion. The guidewire is advanced through the lesion and the support catheter is advanced over the wire until the guidewire exits the lesion and the Support Catheter reaches the patent lumen of the vessel. |
| Indications for Use | The Rubicon Support Catheter is intended to facilitate placement and support of guidewires and other interventional devices within the peripheral vasculature and to allow for exchange of guidewires, and provide a conduit for the delivery of saline or contrast solutions. |
{1}------------------------------------------------
### Comparison of Technological Characteristics
The Rubicon™ Support Catheter incorporates substantially equivalent device design and materials, packaging design and materials, fundamental technology, manufacturing processes, sterilization process as predicate Boston Scientific devices and intended use as those featured in the predicate device, Spectranetics Quick-Cross Support Catheter (K033678).
## Comparison to Predicate Device Characteristics
| Comparison to Predicate Device Characteristics | | | | | | | | |
|------------------------------------------------------|------------------------------|--|------------------------------------------------------------|---------|-----------------|-------------------------------------------|---------|-------------------------|
| | | | Proposed Device<br>Rubicon TM Support Catheter | | | Predicate Device<br>Quick-Cross (K033678) | | |
| Infusion Flow<br>Rate | | | Sterile Saline | | Contrast* | Sterile Saline | | Contrast* |
| | Length<br>(cm) | | 150 psi | 300 psi | 150 psi 300 psi | Length<br>(cm) | 150 psi | 300 psi 150 psi 300 psi |
| | 135 | | 1.3 | 2.3 | 0.4 0.9 | 135 | 1.1 | 1.6 0.4 1.0 |
| | 150 | | 1.3 | 2.2 | 0.4 0.9 | 150 | 1.0 | 1.5 0.4 0.7 |
| Effective Lengths | 135 cm and 150 cm | | | | | 90 cm, 65 cm, 135 cm and 150 cm | | |
| Radiopaque<br>Markers | 3 - equidistant (15 mm) | | | | | 3 - equidistant (15 mm, 50 mm) | | |
| Distal<br>Radiopaque<br>Marker | 2 mm from distal tip | | | | | 3 mm from distal tip | | |
| Hydrophilic<br>coating | Distal 40 cm | | | | | Distal 40 cm | | |
| Recommended<br>Introducer<br>Sheath<br>Compatibility | 4F | | | | | 4F, 5F | | |
| Recommended<br>Guidewire | 0.014 inches | | | | | 0.014, 0.018, 0.035 inches | | |
| Lumen | Single lumen | | | | | Single lumen | | |
| Proximal Shaft<br>Outer Diameter | 0.044 inches | | | | | 0.039, 0.044, 0.063 inches | | |
| Distal Shaft Outer<br>Diameter | 0.027 inches | | | | | 0.026, 0.030, 0.050 inches | | |
| Maximum<br>Infusion Pressure | 300 psi | | | | | 300 psi | | |
| Classification | Class II per 21 CFR 870.1250 | | | | | Class II per 21 CFR 870.1250 | | |
*Tested with 75% Omnipaque™, 25% Saline Solution
{2}------------------------------------------------
K112333
p. 3 of 3
#### Performance Data
Bench testing and leveraged/new biocompatibility testing for the Rubicon Support Catheter were performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing.
Biocompatibility was leveraged from predicate BSC testing.
| MEM Elution Cytotoxicity | Hemolysis Assay Indirect Extraction |
|-----------------------------------------|-------------------------------------|
| Guinea Pig (Maximization) Sensitization | Partial Thromboplastin Time |
| Intracutaneous Reactivity | In Vitro Hemocompatibility Assay |
| Systemic Toxicity (Acute) | Complement Activation |
| Materials Mediated Rabbit<br>Pyrogen | USP Physicochemical |
| Hemolysis Assay Direct<br>Contact | Natural Rubber Latex |
The following screening biocompatibility tests were completed on the Rubicon™ Support Catheter:
| MEM Elution Cytotoxicity |
|--------------------------|
| USP Physicochemical |
| Latex |
Hemolysis Assay Direct Contact Hemolysis Assay Extract
Latex
The following in-vitro performance tests were completed for the Rubicon™ Support Catheter:
| Effective Length | Sheath Insertion and Withdrawal<br>Force |
|---------------------------------|------------------------------------------|
| Inner Diameter - Distal Shaft | Catheter Shaft Burst Pressure |
| Outer Diameter - Proximal Shaft | Catheter Tensile |
| Outer Diameter - Distal Shaft | Shaft Kink Resistance |
| Marker Band Spacing | Torque Strength |
| Contrast Flow Rate | Radiopacity |
| Flow rates for DFU labeling | Coating Integrity |
#### Conclusion
Based on the indications for use, technological characteristics, safety and performance testing, the Rubicon™ Support Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Spectranetics Quick-Cross Support Catheter (K033678).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV - 9 2011
Boston Scientific Corporation c/o Glenn Jacques One Scimed Place Maple Grove, MN 55311
Re: K112303
Trade/Device Name: Rubicon™ Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 10, 2011 Received: August 11, 2011
Dear Mr. Jacques:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Glenn Jacques
Please be advised.that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
14112303 510(k) Number (if known):
Rubicon™ Support Catheter Device Name:
Indications for Use:
The Rubicon Support Catheter is intended to facilitate placement and support of guidewires and other interventional devices within the peripheral vasculature and to allow for exchange of guidewires, and provide a conduit for the delivery of saline or contrast solutions.
Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division/Sian-Off) Division of Cardiovascular Devices
510(k) Number 2303
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.