K112289 · Kang ZE Industrial Co. Limited · FLL · Dec 16, 2011 · General Hospital
Device Facts
Record ID
K112289
Device Name
DISPOSABLE THERMOMETER COVERS AND SHEATHS
Applicant
Kang ZE Industrial Co. Limited
Product Code
FLL · General Hospital
Decision Date
Dec 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
The Disposable Thermometer Covers and Sheaths are intended for use as barriers between digital or mercury thermometers and users' rectum or oral cavities to avoid the possible contamination and infection during temperature measuring. These covers and sheaths are non-sterile and intended for single use only.
Device Story
Disposable thermometer covers and sheaths; made of PE and EVA; non-sterile; latex-free. Used as physical barrier between digital or mercury thermometers and patient's oral or rectal cavities. Prevents cross-contamination and infection during temperature measurement. Single-use only. Used in clinical or home settings. Healthcare providers or patients place cover over thermometer probe before insertion; discard after use. Benefits patient by reducing risk of pathogen transmission during routine temperature monitoring.
Clinical Evidence
Bench testing only. Device met ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation and sensitization), and ASTM E1104-98 (clinical thermometer probe covers). Compliance with IEC 62321 for regulated substances and EPA Method 3546 for extraction confirmed safety.
Technological Characteristics
Materials: PE and EVA. Form factor: Disposable sheaths/covers sized for digital or mercury thermometers. Non-sterile. Latex-free. Standards: ISO 10993-5, ISO 10993-10, ASTM E1104-98(2003), IEC 62321.
Indications for Use
Indicated for use as a protective barrier during oral or rectal temperature measurement with digital or mercury thermometers to prevent contamination and infection. Intended for single use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
SaniTherm Thermometer Sheaths, Oral and Rectal, Digital and Mercury (K983406)
GOOD MEDY Disposable Thermometer Sheaths (K061007)
Submission Summary (Full Text)
{0}------------------------------------------------
#### KANG ZE INDUSTRIAL CO., LTD.
No. 9 Venture Industry Park, Jane Sandbar, Wanjiang Area, Dongguan, Guangdong, China 523063
# . Section 5
510(k) Summary
[As required by 21 CFR 807.92]
| 1. Submission Information: | |
|------------------------------|-----------------------------------------------------------------------------------|
| 510(k) Number: | K112289 |
| Date: | Aug 5th, 2011 |
| Type of 510(k) Submission: | Traditional |
| Basis for 510(k) Submission: | New device |
| Submitter/Manufacturer: | KANG ZE INDUSTRIAL CO., LTD. |
| | Office Add.: Room 1701, Henan Building, No. 90 Jaffe Road, Wan Chai, HK |
| | Factory Add.: No. 9 Venture Industry Park, Jane Sandbar, Wanjiang Area, Dongguan, |
| | Guangdong, China 523063 |
| Contactor: | Doris Dong |
| | [Consultant, from Shanghai CV Technology Co., Ltd.] |
| | Add.: Room 1706 Yuesha, No. 128 Songle Rd., Songjiang, Shanghai, China 201600 |
| | E-mail: doris_d@126.com |
| | Tel: 86 21-31261348 |
| | Fax: 86 21-37824346 |
| | |
| 2. Device Description: | |
| Proprietary Name: | Disposable Thermometer Covers and Sheaths |
| Common Name: | Disposable Thermometer Covers and Sheaths |
| Classification Name: | Clinical electronic thermometer / Clinical mercury thermometer |
| Regulation Number: | 880.2910 / 880.2920 |
| Product Code: | FLL / FLK |
| Device Class: | Class II |
| Submission Type: | 510(k) |
| Review Panel: | General Hospital |
| Indications for use: | The Disposable Thermometer Covers and Sheaths are intended for use as |
| | barriers between digital or mercury thermometers and users' rectum or oral |
| | cavities to avoid the possible contamination and infection during temperature |
| | measuring. These covers and sheaths are non-sterile and intended for single use |
| | only. |
| Device Description: | The Disposable Thermometer Covers and Sheaths, made of PE and EVA, are |
| | used for either oral or rectal measurements for digital or mercury thermometers. |
| | The products and packaging are non-sterile and are not made with natural rubber |
| | latex. |
| | The size may vary to accommodate differences in digital and mercury |
| | thermometers. |
| | The inner packing material is paper, while the outer packing material is carton. |
| | Each outer packing contains 100 pieces of covers and sheaths. |
| | |
| 3. Substantial Equivalence: | |
Detailed comparison data is included in "Section 9 - Substantial Equivalence Discussion" of this 510(k)
{1}------------------------------------------------
#### KANG ZE INDUSTRIAL CO., LTD.
KANG ZE INDUSTNTAL CO., LTD.
No. 9 Venture Industry Park, Jane Sandbar, Wanjiang Area, Dongguan, Guangdong, China 523063 submission
| Predicate 510(k) Numbers: | K983406 | K061007 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Marketing clearance date: | October 19th, 1998 | June 14th, 2006 |
| Product name: | SaniTherm Thermometer Sheaths, Oral and Rectal, Digital and Mercury | GOOD MEDY Disposable Thermometer Sheaths |
| Manufacturer: | Banta Healthcare Products | GOOD MEDY ENTERPRISES LTD |
| Differences: | The material and packing specification differ from the Predicate Devices of K983406 and K061007. | |
| Similarities: | Same intended use, non-sterile, and similar testing standards. | |
| Equivalence conclusion: | The new devices, Disposable Thermometer Covers and Sheaths, are substantially equivalent to the predicate devices: SaniTherm Thermometer Sheaths (K983406) and GOOD MEDY Disposable Thermometer Sheaths (K061007). The devices have same intended use, are non-sterile and latex-free, and passed biocompatibility tests by ISO 10993. Thus the new devices are substantially equivalent to the predicate devices. | |
#### 4. Safety and Effectiveness of the device:
The Disposable Thermometer Covers and Sheaths were tested and found to meet the standards of
ISO 10993-5: 1999: Tests for cytotoxicity: In vitro methods,
ISO 10993-10: 2002: Tests for Irritation and Sensitization,
ASTM E1104-98(2003): Standard Specification for Clinical Thermometer Probe Covers and Sheaths, IEC 62321 Ed. 1: Electrotechnical products - Determination of levels of six regulated substances (lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, polybrominated diphenyl ethers), as well as EPA Method 3546 Microwave extraction.
The conclusion drawn from the bench testing and safety testing is that the device is as safe and effective as the predicate device. Furthermore, the device complies with the recognized standards and performs its intended tasks as well as or better than the legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 1 6 2011
Kang ZE Industrial Company, Limited C/O Ms. Doris Dong Manager Shanghai CV Technology Company Limited RM 1706 Yuesha Building No 128 Songle Road Songjiang Area. Shanghai
Re: K112289
Trade/Device Name: Disposable Thermometer Covers and Sheaths Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electric Thermometer Regulatory Class: II Product Code: FLL、FLK Dated: November 18, 2011 Received: November 23, 2011
Dear Ms. Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2- Ms. Dong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /yom 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A.
Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Section 4 Indications for Use Statement
510(k) Number (if known); K 112289
Device Name:
Disposable Thermometer Covers and Sheaths
Indications for Use:
The Disposable Thermometer Covers and Sheaths are intended for use as barriers between digital or mercury thermometers and users' rectum or oral cavities to avoid the possible contamination and infection during temperature measuring. These covers and sheaths are non-sterile and intended for single use only.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __ √ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Weihny Crn for Richard Chapman
Division Sign Off
Division Sign-Off Division of Surgical, Orthopedic, And Restorative Devices
510(k) Number 112289
Page I of l
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.