← Product Code [DZE](/productcode/DZE) · K112279

# QUANTUM DENTAL IMPLANT SYSTEM (K112279)

_Quantum Bioengineering, Ltd. · DZE · May 4, 2012 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K112279

## Device Facts

- **Applicant:** Quantum Bioengineering, Ltd.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** May 4, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Quantum™ Dental Implant System implants are intended for immediate or delayed placement in the bone of the maxillary or mandibular arch. Quantum Dental Implant System abutments are intended for use as support for crowns, bridges or overdentures. When a one-stage surgical approach is applied, threaded implant may be immediately loaded when good primary stability is achieved and the functional load is appropriate. When using the push-in technique for fin-type or threaded implants, delayed loading is required.

## Device Story

Root form endosseous dental implants; internal Morse taper and external hex abutment interfaces. Titanium alloy construction; RBM, acid-etched, or HA-coated surfaces. Threaded or fin-type designs. Used by dental professionals in clinical settings for tooth replacement. Implants placed in maxillary or mandibular bone; abutments support crowns, bridges, or overdentures. Immediate loading possible for threaded implants with sufficient primary stability; delayed loading required for push-in techniques or short (5-6mm) implants. Provides stable foundation for prosthetic restoration; improves patient masticatory function and aesthetics.

## Clinical Evidence

Bench testing only. Evidence includes engineering analysis, dimensional analysis, surface area and bone-to-implant contact area analysis, and static/dynamic compression-bending testing per ISO 14801.

## Technological Characteristics

Titanium alloy implants and abutments. Internal Morse taper and external hex interfaces. Surface options: RBM, acid-etched, HA-coated. Threaded and fin-type designs. Diameters 4.0-8.0mm; lengths 5-14mm. Mechanical testing per ISO 14801.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- Quantum™ Versatility Dental Implant System ([K011223](/device/K011223.md))
- Quantum™ Versatility Dental Implant System ([K002241](/device/K002241.md))
- Quantum Versatility™ Implant System ([K991250](/device/K991250.md))
- Bicon Implants with a 2.5mm Internal Connection ([K092035](/device/K092035.md))
- Bicon, Inc., 4.5 x 6.0mm and 6.0 x 6.0mm Dental Implant ([K050712](/device/K050712.md))
- Bicon, Inc., 6.0 x 5.7mm Dental Implant ([K010185](/device/K010185.md))
- Astra Tech Implant System ([K101732](/device/K101732.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Quantum™ Dental Implant System

510(k) Summary K112279

MAY - 4 2012

# 510(k) Summary Quantum BioEngineering, Ltd. Quantum™ Dental Implant System K112279

### · May 1, 2012

#### ADMINISTRATIVE INFORMATION

Manufacturer Name:

Quantum BioEngineering, Ltd. 201 N. University Drive, Suite 101 Plantation, FL 33324 Telephone: +1 (954) 474-2556 +1 (954) 474-2533 Fax:

---

Official Contact:

Representative/Consultant:

Raul R. Mena, D.M.D., President

Linda K. Schulz Floyd G. Larson PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 Telephone: +1 (858) 792-1235 +1 (858) 792-1236 Fax: Email: lschulz@paxmed.com flarson@paxmed.com .

## DEVICE NAME AND CLASSIFICATION

| Trade/Proprietary Name:     | Quantum™ Dental Implant System     |
|-----------------------------|------------------------------------|
| Common Name:                | Dental Implant and abutment        |
| Classification Name:        | Implant, endosseous, root form     |
|                             | Endosseous dental implant abutment |
| Classification Regulations: | Class II, 21 CFR 872.3640          |
| Product Code:               | DZE, NHA                           |
| Classification Panel:       | Dental Products Panel              |
| Reviewing Branch:           | Dental Devices Branch              |

{1}------------------------------------------------

510(k) Summary K112279

#### INTENDED USE

Quantum™ Dental Implant System implants are intended for immediate or delayed placement in the bone of the maxillary or mandibular arch. Quantum Dental Implant System abutments are intended for use as support for crowns, bridges or overdentures. When a one-stage surgical approach is applied, threaded implant may be immediately loaded when good primary stability is achieved and the functional load is appropriate. When using the push-in technique for fin-type or threaded implants, delayed loading is required.

#### DEVICE DESCRIPTION

The Quantum Dental Implant System includes root form, endosseous dental implants with internal Morse taper and external hex abutment interfaces. The implants are made of titanium alloy with three surface options (RBM, acid etch, HA coated), and are provided in both threaded and a grooved (fin-type) designs. There are four sizes of the internal Morse taper (implant well), designated as the 2.0, 2.5, 2.75, and 3.0 Series. For all series, the threaded and fin-type implant designs are each provided in the following sizes: diameters of 4.5, 5.0, 5.5, and 6.0 mm, with each diameter in lengths of 5, 6, 9, 11, and 14 mm. The 2.0 Series of the threaded and fin-tvpe implant designs also include: 4.0 mm diameter in 5 and 6 mm lengths, 8.0 mm diameter implants in 5 and 14 mm lengths, and 5.0 mm diameter available only in lengths of 5 and 6 mm. The 2.5 Series of the threaded and fin-type implant designs also include 4.0 mm diameter in lengths of 5, 6, 9, 11, and 14 mm.

Abutments for cement-retained prostheses are provided for each diameter implant. All abutments are made of titanium alloy. Abutments with a Morse taper interface are provided for all implant series in straight, 15° and 25° angled designs. The 2.0 Series Morse taper abutments are provided in 5 mm and 7 mm platforms; all other Series are provided in 3.5, 4, and 5 mm platforms. External hex interface abutments are provided for the 2.5. 2.75. and 3.0 Series implants in straight, 15° and 25° angled designs. The 2.5 Series hex abutments are 4 mm platform; the 2.75 and 3.0 Series hex abutments are 4.5 mm platform. Healing plugs are provided for each implant series in titanium alloy and polyethylene.

#### EQUIVALENCE TO MARKETED DEVICE

The Quantum Dental Implant System is substantially equivalent in indications and design principles to the following predicate devices:

Quantum BioEngineering, Ltd., Components of the Quantum™ Versatility Dental Implant System, cleared under K011223.

Quantum BioEngineering, Ltd., Quantum™ Versatility Dental Implant System, cleared under K002241.

Quantum BioEngineering, Ltd., Quantum Versatility™ Implant System, cleared under K991250,

Bicon. LLC., Bicon Implants with a 2.5mm Internal Connection, cleared under K092035,

Bicon, Inc., 4.5 x 6.0mm and 6.0 x 6.0mm Dental Implant, cleared under K050712,

Bicon. Inc., 6.0 x 5.7mm Dental Implant, cleared under K010185, and

Astra Tech AB. Astra Tech Implant System, cleared under K101732.

{2}------------------------------------------------

# 510(k) Summary K112279

The subject device and the predicate devices have the same intended use and have the same technological characteristics. The subject device and the predicate devices also encompass the same range of physical dimensions for both the implants (diameter and length) and abutments (diameter, height, and angulation). The subject device implant and abutment designs, materials, and surfaces are the same as the predicates K011223, K002241 and K991250. The subject device implant largest diameter and shortest lengths are the same as the predicates K092035, K050712, K010185 and K101732.

The subject and predicate devices are packaged in similar materials and sterilized using similar methods, or provided nonsterile to be sterilized using similar methods.

Non-clinical testing data that was provided or referenced to demonstrate substantial equivalence included detailed engineering analysis, dimensional analysis, surface area and bone-to-implant contact area analysis, and static and dynamic compression-bending testing according to ISO 14801.

Anv differences in the technological characteristics between the subject device and the predicate devices do not raise new issues of safety or efficacy.

Overall, the Quantum Dental Implant System has the following similarities to the predicate devices:

- . has the same intended use,
- uses the same operating principle, .
- incorporates the same basic design, .
- incorporates the same or very similar materials, and .
- has similar packaging and is sterilized using the same materials and processes .

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized eagle-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Quantum BioEngineering, Ltd. C/O Ms. Linda K. Schulz Consultant PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, California 92130

Re: K112279

Trade/Device Name: Quantum™ Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: May 1, 2012 Received: May 2, 2012

Dear Ms. Schulz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

MAY - 4 2012

{4}------------------------------------------------

#### Page 2 - Ms. Schulz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Wh foc

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Summary K112279

# Indications for Use

K112279 510(k) Number:

Quantum™ Dental Implant System Device Name:

Indications for Use:

Quantum™ Dental Implant System implants are intended for immediate or delayed placement in the bone of the maxillary or mandibular arch. Quantum Dental Implant System abutments are intended for use as support for crowns, bridges or overdentures. When a one-stage surgical approach is applied, threaded implant may be immediately loaded when good primary stability is achieved and the functional load is appropriate. Delayed loading is required when using the push-in technique for fin-type or threaded implants, or when using any 5 mm or 6 mm length implant.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Siva Ramne

(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Division Control, Dental Devices

510(k) Number: K112279

---

**Source:** [https://fda-staging.innolitics.com/device/K112279](https://fda-staging.innolitics.com/device/K112279)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com