K112253 · Biolitec, Inc. · GEX · Sep 20, 2011 · General, Plastic Surgery
Device Facts
Record ID
K112253
Device Name
CERALAS 147NM DIODE LASER SYSTEM
Applicant
Biolitec, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ceralas 1470nm Fiber-Coupled Diode Laser family (and its delivery accessories used to deliver optical energy) is indicated for use in general surgical applications for incision, excision, ablation, cutting, vaporization, hemostasis, and coagulation of soft tissue contact or non-contact, open or closed endoscopic applications where incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors and lesions, tissue vaporization, hemostasis and/or coagulation may be indicated. Endovenous Occlusion of the Saphenous Veins in Patients with Superficial Vein Reflux Associated with Varicose Veins and Varicosities. Laser Assisted Lipolysis.
Device Story
Ceralas 1470nm Diode Laser system delivers optical energy via fiber-coupled delivery accessories to soft tissue. Used in general surgical, endoscopic, and dermatological applications; operated by physicians in clinical settings. Device performs incision, excision, ablation, cutting, vaporization, hemostasis, and coagulation. Also indicated for endovenous occlusion of saphenous veins and laser-assisted lipolysis. System provides controlled thermal energy to target tissues; clinical benefit includes precise tissue management and minimally invasive treatment of venous conditions and adipose tissue.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical technological characteristics and performance to predicate devices.
Technological Characteristics
1470nm diode laser system; fiber-coupled delivery. Complies with 21 C.F.R. §§ 1040.10 & 1040.11, ANSI/AAMI ES1, IEC 601-1, IEC 601-2-22, EN 60825-1, and ANSI/AAMI/ISO 10993-7. Identical to predicate devices.
Indications for Use
Indicated for patients with superficial vein reflux associated with varicose veins and varicosities requiring endovenous occlusion, and patients requiring soft tissue incision, excision, ablation, cutting, vaporization, hemostasis, or coagulation, including laser-assisted lipolysis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
SEP 2 0 2011
# 510(k) Summary Ceralas 1470nm Diode Laser Family
KI12253
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Biolitec Medical Devices, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028 Phone: (413) 525-0600 Facsimile: (413) 525-0611
Contact Person: Harry Hayes, Ph.D. - Regulatory Consultant Date prepared: August 3, 2011
### Name of Device and Name/Address of Sponsor
Ceralas 1470nm Diode Laser System Biolitec, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028
#### Classification Name
Surgical laser
#### Predicate Devices
Ceralas 1470nm Diode Laser System, (K073063, K082225 and K102755).
#### Intended Use/Indication for Use
The Ceralas 1470nm Fiber-Coupled Diode Laser family (and its delivery accessories used to deliver optical energy) is indicated for use in general surgical applications for incision, excision, ablation, cutting, vaporization, hemostasis, and coagulation of soft tissue contact or non-contact, open or closed endoscopic applications where incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors and lesions, tissue vaporization, hemostasis and/or coagulation may be indicated.
## Technological Characteristics
The Ceralas 1470nm family for Biolitec Medical Devices, Inc. is identical (contains the same components, technology and principles of operation and assembled by the same manufacturer) as the cleared Ceralas 1470nm family for Biolitec, Inc..
{1}------------------------------------------------
#### Performance Data
The device complies with the following voluntary consensus standards: 21 C.F.R. §§ 1040.10 & 1040.11; ANSI/AAMI ES1; IEC 601-1; IEC 601-2-22; EN 60825-1, and ANSI/AAMI/ISO 10993-7.
#### Substantial Equivalence
The Biolitec Medical Devices Inc Ceralas 1470nm family is as safe and effective as the Biolitec Inc. Ceralas 1470nm family as the products are identical in all aspects except labeling realating to the manufacturer/ distributor.
The Ceralas 1470nm family has the same intended uses, indications, technological characteristics, and principles of operation as its predicate devices. Thus, the Ceralas 1470nm family is substantially equivalent to its predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the USA. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hamoshire Avenue Document Control Room -WO66-G609 Silver Spring, MI) 20995-0002
Biolitec Medical Devices, Inc. % Genmarhav BDA Mr. Harry Haves 1349 Main Road Granville. Massachusetts 01034
SEP 20 2011
Re: K112253
Trade/Device Name: Ceralas 1470mm Diode Laser Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 3. 2011 Received: August 5, 2011
Dear Mr. Haves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract hiability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Mr. Harry Hayes
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
112253 510(k) Number (if known): ----------------------
Device Name: Ceralas 1470nm Diode Laser Family
Indications for Use:
The Ceralas Fiber-Coupled 1470nm Diode Laser family (and their delivery accessories used to deliver optical energy) are indicated for use in general surgical applications for incision, excision, ablation, cutting, vaporization, hemostasis, and coagulation of soft tissue contact or non contact, open or closed endoscopic applications where incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors and lesions, tissue vaporization, hemostasis and/or coagulation may be indicated. Endovenous Occlusion of the Saphenous Veins in
Patients with Superficial Vein Reflux Associated with Varicose Veins and Varicosities. Laser Assisted Lipolysis.
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
of CDRH, Office of Device Evaluation (ODE) Concurrenc MX m (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112253
Prescription Use __ V (Per 21 C.F.R. 801.109) OR
Over-The-Counter Use_ (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.