K112247 · Natus Medical Incorporated · EWO · Apr 2, 2012 · Ear, Nose, Throat
Device Facts
Record ID
K112247
Device Name
A BAER WITH ABAER I/O FUNCTION
Applicant
Natus Medical Incorporated
Product Code
EWO · Ear, Nose, Throat
Decision Date
Apr 2, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The 'ABaer System with ABaer I/O Function' is indicated for use when it is necessary for a trained health care professional to measure or determine cochlear function. The device can be used for patients of all ages, from newborn infants through adults, to include geriatric patients. The otoacoustic emissions test is especially indicated for use in testing individuals for whom behavioral results are deemed unreliable, such as infants, young children, and cognitively impaired adults.
Device Story
ABaer system performs automated auditory brainstem response (ABR) and otoacoustic emissions (OAE) screening. ABR input: electrical activity from three scalp electrodes following click stimuli. OAE input: acoustic signals in ear canal recorded via miniature microphone. ABaer I/O software option adds DPOAE Input/Output testing; measures DPOAE response across varying stimulus intensities/frequencies. System processes data via Point Optimized Variance Ratio (POVR) algorithm for ABR; calculates reproducibility, noise floor, and amplitude for OAE. Output: graphical representation of stimulus level vs. DPOAE level, plus Pass/Refer recommendations. Used by trained healthcare professionals in clinical settings. Output assists clinicians in evaluating cochlear outer hair cell function; aids in identifying hearing status in patients unable to provide reliable behavioral responses.
Clinical Evidence
Bench testing only. Design verification and validation were performed to confirm the ABaer with ABaer I/O Function meets performance specifications and demonstrates equivalence to predicate devices. No clinical study data provided.
Technological Characteristics
Windows-based software application running on PC hardware. Components include data collection box, insert earphones, headphones, OAE probe, and electrodes. Sensing: acoustic transducers for stimuli, electrodes for ABR, miniature microphone for OAE. Connectivity: USB. Algorithm: Point Optimized Variance Ratio (POVR) for ABR signal optimization.
Indications for Use
Indicated for patients of all ages (newborns to geriatric) requiring cochlear function assessment. Especially indicated for individuals where behavioral testing is unreliable, including infants, young children, and cognitively impaired adults.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
{0}------------------------------------------------
K112247
APR - 2 2012
## 510(k) SUMMARY
Manufacturer's Name:
Natus Medical Incorporated One Bio-logic Plaza Mundelein, IL 60060
Corresponding Official:
Telephone Number: Fax Number:
Summary Date:
Trade Name:
**Common or Usual Name:**
Classification Name and Number:
Predicate Devices:
Device Description:
Don Williams Vice President and General Manager Natus Medical Incorporated One Bio-logic Plaza Mundelein, IL 60060 800.323.8326 ext. 5424 847.949.8615
March 28, 2012
ABaer with ABaer I/O Function
Common or Usual Name: Audiometer and Evoked response auditory stimulator
Audiometer 21 CFR 874.1050, Product Code: EWO
Evoked response auditory stimulator 21 CFR 882.1900 Product Code: GWJ
K021801 ABaer Cub with Automated OAE and ABR
K964132 Bio-logic Scout and Scout Sport Otoacoustic Emissions (OAE) Test Instruments with TEOAE and DPOAE Software, incorporating the modifications of Automated Input / Output Software Functions
K072033 Otodynamics Otoport
The ABaer I/O Function is a Windows® based software application for use with the ABaer Hearing Screening System. The ABaer I/O software option enables the ABaer device user to perform DPOAE Input/Output (I/O) testing at different test frequencies, frequency ratios and intensity levels in addition to the ABR and OAE based hearing screening functions. The graphical representation of the test results in the form of stimulus level presented versus measured DPOAE level provides an effective way for the
{1}------------------------------------------------
Intended Use:
Technological Characteristics: user to view and evaluate stimulus level-sensitive information about DPOAE responses.
The 'ABaer System with ABaer I/O Function' is indicated for use when it is necessary for a trained health care professional to measure or determine cochlear function. The device can be used for patients of all ages, from newborn infants through adults, to include geriatric patients. The otoacoustic emissions test is especially indicated for use in testing individuals for whom behavioral results are deemed unreliable, such as infants, young children, and cognitively impaired adults.
The ABaer Hearing Screening System (HSS) performs an automated auditory evoked response (ABaer) screening and/or an automated otoacoustic emissions (AOAE) screening. The automated auditory brainstem response test (ABaer) involves placement of three recording electrodes on an infant's head. The electrodes record electrical activity generated by the auditory nervous system that results from the presentation of a click stimulus into the patient's ear via acoustic transducers (i.e. insert earphones, headphones, OAE probe, acoustic ear couplers). The system collects and averages evoked potential data in order to perform ABR based screening, recording and analysis functions, provides one channel of data recording, and includes the Point Optimized Variance Ratio (POVR) algorithm for optimizing signal quality, implementing the automated screening function and enhancing speed of test completion in the same manner as in the predicate device (K021801). The device presents the resulting POVR score and a Pass/Refer recommendation to the user.
With respect to the ABR based testing, the ABaer with ABaer I/O Function is equivalent to the predicate device cleared under K021801.
The automated otoacoustic emissions (AOAE) screening functionality of the ABaer system involves producing controlled acoustic signals in the ear canal and measures the resulting evoked otoacoustic emissions that are generated by the inner ear as a result of normal hearing processes. The ABaer device performs both distortion product otoacoustic emissions (DPOAE) tests and transient evoked otoacoustic emissions (TEOAE) tests.
{2}------------------------------------------------
The OAE stimuli are generated via miniature receivers and the sounds in the external ear canal are recorded via a miniature microphone, all embedded in the Bio-logic OAE probe. The system collects, averages and analyzes data samples until specified measurement and test parameters are achieved. For transient evoked otoacoustic emissions (TEOAEs), the reproducibility and the difference value between the TEOAE and the noise floor amplitudes are calculated and presented to the user. For distortion product otoacoustic emissions (DPOAEs), the DP and noise floor amplitudes are calculated and presented to the user. A pass or refer recommendation is assigned at the end of the test automatically based on the test protocol parameters and measured OAE parameters.
With respect to TEOAE and DPOAE testing, the ABaer with ABaer I/O Function is equivalent to the devices cleared under K021801. K964132. and K072033.
The ABaer I/O is a software option to be used in conjunction with the ABaer system. The standard DPOAE test measures otoacoustic response to a series of frequency-pairs of tones, varving the frequency while keeping the level or intensity of the stimulus tones at a constant level. The ABaer I/O software option enables the ABaer device user to perform DPOAE testing at different stimulus intensities in order to obtain the 'DPOAE Input / Output (I/O) function' for user defined test frequencies, frequency ratios and intensity levels. The graphical representation of the test results in the form of stimulus level vs. DPOAE level provides an effective way for the user to view and evaluate stimulus levelsensitive information about DPOAE response.
With respect to DPOAE I/O function, the ABaer with ABaer I/O Function is equivalent to the automated Input / Output Software functions present in the Scout and Scout Sport Otoacoustic Emissions (OAE) Test Instruments.
Nonclinical Tests:
Design verification and validation were performed to assure that the ABaer with ABaer I/O Function meets its performance specifications and demonstrates equivalence to the functionalities present in the respective predicate devices.
{3}------------------------------------------------
The verification and validation summary report and risk analysis documentation provided in this 510(k) support the conclusion that the ABaer System with ABaer I/O Function is safe and effective.
# Components:
| Hardware | |
|----------------|-----------------------------------------------------|
| 580-ABAER2 | ABaer Data Collection Box |
| 580-SINABR-008 | Insert Earphones |
| 206920 | Halo Adaptors & tubes |
| 580-MEPTDH-125 | Headphones |
| 301663 | Alligator Clips (3) |
| 580-PROAE3 | OAE Style Probe |
| 541-ABRC10-008 | Patient Cable - 3 inputs |
| 541-USB001 | USB Cable (ABaer & Printer) |
| 520-PSVDC | Power Supply for ABaer module |
| 540-LINECD-012 | Power Cord - 12 inch |
| 001308 | Seiko Smart Label Printer |
| 541-SEKBLE | Printer Serial Cable |
| 520-SEK120 | Printer Power Supply |
| 520-AMPS01 | Isolation Transformer - 3 outlet (laptop only) |
| 520-SWBXFS | Isolation Transformer - 6 outlet (touchscreen only) |
| 541-CORD12 | Power Cord for Isolation Transformer |
| 541-TSTCBL | Loop Test Cable |
| 001771 | Panel PC |
| 001316 | Notebook PC |
| Software | |
| 008634 | ABaer Software |
| 008164 | ABaer I/O Function Software |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo features a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Natus Medical Inc. c/o Mr. Don Williams Vice President and General Manager One Bio-Logic Plaza Mundelein, IL 60060
Re: K112247
Trade/Device Name: ABaer with ABaer I/O Function Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO, GWJ Dated: February 28, 2012 Received: February 29, 2012
Dear Mr. Williams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
APR - 2 2012
{5}------------------------------------------------
## Page 2 - Mr. Don Williams
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Makin R.E. Aldous, MD
Malvina B. Eydelman, M.L. Director Division of Ophthalmic. Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
i 12 247
### Indications for Use
# 510(k) Number (if known): KIL2247
### Device Name: ABaer with ABaer I/O Function
#### Indications for Use:
The Hearing Screening System (HSS) ABaer performs an automated brainstem auditory evoked response (ABaer) screening and/or an automated otoacoustic emissions (AOAE) screening. The automated auditory brainstem response test (AABR) involves placement of three recording electrodes on an infant's head. The electrodes record the electrical activity generated by the auditory nervous system that result from presentation of a click stimulus into the ear. The automated otoacoustic emissions (AOAE) screening involves producing controlled acoustic signals in the ear canal and measuring the resulting evoked otoacoustic emissions that are generated by the inner ear as a result of normal hearing processes. The ABaer device performs both distortion product otoacoustic emissions (DPOAE) tests and transient evoked otoacoustic emissions (TEOAE) tests.
The ABaer I/O is a software option to be used in conjunction with the ABaer System. The standard DPOAE test measures otoacoustic response to a series of frequency-pairs of tones, varying the frequency while keeping the level or intensity of the stimulus tones at a constant level. The ABaer I/O software option enables the ABaer device user to perform DPOAE testing at different stimulus intensities in order to obtain the 'DPOAE Input / Output (!/O) function' for user defined test frequencies, frequency. ratios and intensity levels. The graphical representation of the test results in the form of stimulus level presented versus measured DPOAE level provides an effective way for the user to view and evaluate stimulus level-sensitive information about DPOAE responses. DPOAE threshold is the lowest stimulus intensity level which produces DPOAE response amplitude that is distinguishable from the level of system distortion and from the noise floor. DPOAE threshold is expressed in much of the published literature as the lowest intensity of the F2 stimulus that generates a clear DPOAE response. DPOAE threshold does not equate to and should not be confused with audiometric and auditory threshold. Whereas auditory threshold measures assess the integrity of the entire auditory system from outer ear to cortex, DPOAE threshold measures, which the user can interpret from the DPOAE I/O functions, are limited to assessment of outer hair cell function in the cochlea only.
The 'ABaer System with ABaer I/O Function' is indicated for use when it is necessary for a trained health care professional to measure or determine cochlear function. The device can be used for patients of all ages, from newborn infants through adults, to include geriatric patients. The otoacoustic emissions test is especially indicated for use in testing individuals for whom behavioral results are deemed unreliable, such as infants, young children, and cognitively impaired adults.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)T. Neik
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
510(k) Number K112247
Prescription Use . (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.