K112182 · Angioscore, Inc. · PNO · Aug 26, 2011 · Cardiovascular
Device Facts
Record ID
K112182
Device Name
ANGIOSCULPT PTA SCORING BALLOON CATHETER
Applicant
Angioscore, Inc.
Product Code
PNO · Cardiovascular
Decision Date
Aug 26, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
Device Story
Two-lumen catheter with distal nylon-blend balloon and integrated scoring element; scoring element creates focal dilating force to minimize balloon slippage and assist luminal expansion of stenotic arteries; radiopaque markers aid positioning; used by physicians in clinical settings for peripheral artery and dialysis fistula interventions; device introduced via standard guidewires and sheaths; provides controlled expansion at specific pressures; benefits include improved lesion dilatation and reduced slippage compared to conventional balloons.
Clinical Evidence
Bench testing verified design outputs, including balloon profile, burst strength, compliance, inflation/deflation time, fatigue, and bond strength. Biocompatibility assessed per ISO 10993-1:2009 (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility). Acute GLP animal study demonstrated safety, deliverability, and device integrity in peripheral arteries with no dissection, embolization, or adverse events.
Technological Characteristics
Two-lumen catheter; nylon-blend balloon with integrated scoring element; radiopaque markers; compatible with standard guidewires and 6F/7F sheaths. Biocompatibility per ISO 10993-1. Sterilized finished device.
Indications for Use
Indicated for patients requiring dilatation of lesions in iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Contraindicated for coronary or neuro-vasculature use.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 31, 2016
AngioScore, Inc. c/o Mr. Jonathan S. Kahan Hogan Lovells US, LLP 555 Thirteenth Street NW Washington, DC 20004
Re: K112182
Trade/Device Name: AngioSculpt® PTA Scoring Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: PNO Dated: July 28, 2011 Received: July 28, 2011
Dear Mr. Kahan:
This letter corrects our substantially equivalent letter of August 26, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
Page 2 - Mr. Jonathan S. Kahan
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Misti L. Malone -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for AngioScore. The logo consists of a triangular shape on the left, with the word "AngioScore" to the right of the shape. The triangular shape is made up of three curved shapes that are arranged in a circular pattern. There is a small circle above the end of the word "AngioScore".
# INDICATIONS FOR USE STATEMENT
K112182 510(k) Number (if known): __
Device Name: AngioSculpt® PTA Scoring Balloon Catheter
Indications for Use:
The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|-----------------|
| Division Sign-Off | Page _1_ of _1_ |
| Division of Cardiovascular | |
| 510(k) Number | K112182 |
5055 Brandin Court Fremont CA 94538 www.angioscore.com Toll Free: 1.877.264.4692 Phone: 1.510.933.7900 Fax: 1.510.933.7901
{3}------------------------------------------------
K112182
AUG 26 2011
# 510(k) Summary for the AngioSculpt Scoring Balloon Catheter
# 1. Submitter's Name / Contact Person
| Submitter: | AngioScore, Inc.<br>5055 Brandin Court<br>Fremont, CA 94538 |
|-----------------|-------------------------------------------------------------------------------------------|
| Contact Person: | Kimberley Kline<br>Regulatory Affairs Manager<br>Phone: 510-933-7989<br>Fax: 510-933-7994 |
Summary Preparation Date: July 27, 2011
# 2. General Information
| Trade Name: | AngioSculpt® PTA Scoring Balloon Catheter |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Common / Usual Name: | Angioplasty catheter |
| Classification Name: | Percutaneous catheter |
| Product Codes: | DQY and LIT |
| Predicate Devices: | AngioSculpt® Scoring Balloon Catheter<br>(K100303/K110767)<br>VascuTrak® PTA Dilatation Catheter<br>(K103459) |
# 3. Intended Use / Indications
The AngioSculpt PTA Scoring Balloon Catheter is intended for the dilatation of lesions in the illiac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
#### 4. Device Description
The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.
{4}------------------------------------------------
# 5. Technological Characteristics .
There are no technological changes to the subject AngioSculpt catheters as compared to the 510(k) cleared AngioSculpt catheter family (K100303/K110767).
# 6. Summary of Bench Testing
Mechanical testing of the AngioSculpt catheter was conducted in accordance with AngioScore's Risk Analysis and all applicable FDA guidance documents and relevant standards.
The following bench tests were conducted to verify that design outputs met design requirements and to confirm proper function and durability. Test articles consisted of finished sterilized catheters.
- Catheter Diameter and Balloon Profile
- . Minimum Burst Strength (RBP)
- . Balloon Compliance (Diameter vs. Pressure)
- . Balloon Inflation and Deflation Time
- Device Fatigue, Bond (Tensile) Strength ●
- Tip Pull Strength
- Catheter Body Integrity
- . Catheter Diameter and Balloon Profile (with Scoring Element)
- . Flexibility and Kink
- Torque Strength
- . Pushability, Trackability, and Secure Edges
- Balloon Preparation
- Deployment and Retraction
- Freedom from Stent Interference
- . Focal Force
# 7. Summary of Biocompatibility Testing
The AngioSculpt catheter is categorized as an "External communicating device in contact with circulating blood with limited exposure time". The biocompatibility of the device was assessed in accordance with ISO 10993-1:2009 - Biological evoluation of medical devices, Part 1 -Evaluation and tests within a risk management process. In addition, the product complies with the requirements identified in FDA Guidance for the Submission of Research and Marketing Applications for Interventional Cardiology Devices: PTCA Catheters, Atherectomy Catheters, Lasers and Intravascular Stents(2008) – Biocompatibility testing for all interventional cardiology devices.
The blocompatibility tests listed below, except Thrombosis (in-vivo), were conducted in accordance with the provisions of the FDA GLP regulations 21 CFR Part 58. Thrombogenicity was evaluated as part of a GLP animal study.
- Cytotoxicity - MEM Elution Test using L-929 Mouse Fibroblast cells (ISO)
- Sensitization - Guinea Pig Maximization Sensitization Test
- Irritation Intracutaneous Reactivity (Irritation) Test ●
{5}------------------------------------------------
- Systemic Toxicity Acute Systemic Injection Test (ISO) .
- Hemocompatibility Partial Thromboplastin Time (PTT) ●
- Hemocompatibility - Thrombosis (in vivo)
- Hemocompatibility Hemolysis Direct Contact Method (NIH) .
The blocompatibility test results confirmed that the AngioSculpt catheter is non-cytotoxic, nonsensitizing, non-irritating, not systemically toxic, and non-hemolytic when evaluated under the respective test conditions. As shown in the GLP animal study, no thrombo-embolism was observed when the AngioSculpt catheter was evaluated under simulated use conditions.
#### 8. Summary of Animal Testing
An acute GLP study was conducted to determine the safety and deliverability of the AngioSculpt catheter.
The study results demonstrated that the AngloSculpt catheter was successfully introduced using standard guidewires in combination with a 6F sheath or 7F guide catheter and expanded in the targeted tissue while demonstrating no evidence of dissection, or embolization. All devices maintained integrity with no loss of components during the procedure. All treatment procedures were performed with ease and no adverse events occurred in any of the animals.
The objectives of the study were met. The catheters evaluated in the peripheral (femoral) arteries were found to be clinically acceptable.
### 9. Substantial Equivalence Comparison
The subject catheters share intended use, principles of operation, overall technical and functional capabilities, packaging and sterilization process, and similar design and materials as the predicate AngioSculpt catheters and are therefore substantially equivalent.
The subject catheters also have similar intended use and device operation, materials, design, dimensions/size configuration, guidewire platform and overall technical and functional capabilities as the BARD/YMed, Inc. VascuTraK® PTA Dilatation Catheters.
Although there are minor differences between the subject catheters and its predicate devices those differences do not raise new questions of safety or efficacy. Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised. The changes to the subject catheters do not affect the intended use of the device, alter the fundamental scientific technology of the device, or raise new issues of safety and effectiveness. The subject AngioSculpt PTA Scoring Balloon Catheters are therefore, substantially equivalent to the predicate catheters.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.