← Product Code [LON](/productcode/LON) · K112129

# VITEK(R) 2 STREPTOCOCCUS CLINDAMYCIN (K112129)

_bioMerieux, Inc. · LON · Sep 27, 2011 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K112129

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [LON](/productcode/LON.md)
- **Decision Date:** Sep 27, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

VITEK® 2 Streptococcus Clindamycin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 Streptococcus Clindamycin is a qualitative test intended for use with the VITEK® 2 and the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Clindamycin has been shown to be active in vitro against most strains of the microorganisms listed below, as describe in the FDA-approved package insert for this antimicrobial agent. Active in vitro and in clinical infections against: Streptococcus pneumoniae (penicillin-susceptible strains) Streptococcus pyogenes Active in vitro against: Streptococcus agalactiae Streptococcus anginosus Streptococcus mitis Streptococcus oralis The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

## Device Story

VITEK 2 Streptococcus Clindamycin is an AST card for use with VITEK 2 and VITEK 2 Compact systems; performs qualitative antimicrobial susceptibility testing. Input: bacterial isolate diluted in saline; rehydrates antimicrobial medium in card. Operation: system automatically fills/seals card (VITEK 2) or manual fill/seal (VITEK 2 Compact); incubates and monitors growth in wells over time. Output: MIC value and interpretive category result. Used in clinical laboratories by technicians/microbiologists. Output assists clinicians in selecting appropriate antibiotic therapy for Streptococcus infections. Benefits: standardized, automated susceptibility determination compared to manual broth microdilution.

## Clinical Evidence

Performance evaluated via clinical and challenge studies (n=1509 combined). Comparison to CLSI broth microdilution reference method showed 97.2% Category Agreement (CA). Discrepancy analysis: 26 minor, 15 major, 2 very major. Reproducibility >95% by category. Bench testing confirmed comparability between auto-dilution and manual dilution methods.

## Technological Characteristics

Miniaturized card-based broth microdilution system. Automated incubation and optical growth monitoring. Connectivity via VITEK 2/VITEK 2 Compact instrument platforms. Software-controlled automated filling, sealing, and reading. No specific material standards or algorithm architecture details provided.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK® 2 Gram Positive Amoxicillin for Streptococcus pneumoniae (k063597)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k112129

B. Purpose for Submission:
To obtain clearance for the addition of Clindamycin to the VITEK® 2 and VITEK®2 Compact Systems Antimicrobial Susceptibility Test (AST) Systems

C. Measurand
Clindamycin concentrations (≤ 0.25- ≥ 1μg/mL)

D. Type of Test:
Quantitative growth based detection algorithm using optics light detection

E. Applicant:
bioMérieux, Inc.

F. Proprietary and Established Names:
Vitek®2 Streptococcus Clindamycin (≤ 0.25- ≥ 1μg/mL)

G. Regulatory Information:
1. Regulation section:
866.1645 Short-Term Antimicrobial Susceptibility Test System
2. Classification:
II
3. Product Code:
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
4. Panel:
83 Microbiology

{1}

H. Intended Use:

1. Intended use(s):

VITEK® 2 Streptococcus Clindamycin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 Streptococcus Clindamycin is a qualitative test intended for use with the VITEK® 2 and the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Clindamycin has been shown to be active in vitro against most strains of the microorganisms listed below, as describe in the FDA-approved package insert for this antimicrobial agent.

Active in vitro and in clinical infections against:
Streptococcus pneumoniae (penicillin-susceptible strains)
Streptococcus pyogenes

Active in vitro against:
Streptococcus agalactiae
Streptococcus anginosus
Streptococcus mitis
Streptococcus oralis

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

2. Indication(s) for use:

VITEK® 2 Streptococcus Clindamycin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 Streptococcus Clindamycin is a qualitative test intended for use with the VITEK® 2 and the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Clindamycin has been shown to be active in vitro against most strains of the microorganisms listed below, as describe in the FDA-approved package insert for this antimicrobial agent.

Active in vitro and in clinical infections against:
Streptococcus pneumoniae (penicillin-susceptible strains)
Streptococcus pyogenes

Active in vitro against:
Streptococcus agalactiae
Streptococcus anginosus
Streptococcus mitis

Page 2 of 8

{2}

Page 3 of 8

# Streptococcus oralis

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, *Staphylococcus spp.*, *Enterococcus spp.*, *Streptococcus spp.* and clinically significant yeast.

3. Special condition for use statement(s)

Prescription Use Only

4. Special instrument Requirements:

VITEK® 2 and the VITEK® 2 Compact Systems

## I. Device Description:

Each VITEK® 2 test card contains 64 micro-wells. A control well which contains only microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 – 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

The VITEK® 2 AST-ST Clindamycin has the following concentrations in the card: 0.125, 0.25, 0.5 and 0.5/0.1 Clindamycin/Erythromycin µg/mL (equivalent standard method concentration by efficacy in µg/mL). The test provides only a clindamycin result; the clindamycin MIC result range for the VITEK® 2 card is ≤0.25 to ≥ 1 µg/mL.

## J. Substantial Equivalence Information:

1. Predicate device name(s)

VITEK® 2 Gram Positive Amoxicillin for *Streptococcus pneumoniae*

2. Predicate K number(s):

k063597

{3}

# 3. Comparison with predicate

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | Determine antimicrobial susceptibility to antimicrobial agents | Same  |
|  Test Card | VITEK® 2 card format with base broth | same  |
|  Instrument | VITEK® 2 and VITEK®2 Compact System | same  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Antibiotic | Clindamycin | Amoxicillin  |
|  Reading algorithm | Unique for Clindamycin | Unique for amoxicillin  |
|  Test organism | S. pneumoniae and other Streptococcus spp. | S. pneumoniae  |

# K. Standard/Guidance Document Referenced (if applicable):

Class II Special Controls Guidance Document: "Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA"

CLSI M7-A8 "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard"

CLSI M100-S19 "Performance Standards for Antimicrobial Susceptibility; Twenty-First Information Supplement"

# L. Test Principle:

Each VITEK® 2 test card contains 64 micro-wells. A control well which contains only microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with  $0.45 - 0.5\%$  saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

The VITEK® 2 AST-ST Clindamycin has the following concentrations in the card: 0.125, 0.25, 0.5 and 0.5/0.1 Clindamycin/Erythromycin  $\mu \mathrm{g} / \mathrm{mL}$  (equivalent standard method concentration by efficacy in  $\mu \mathrm{g} / \mathrm{mL}$ ). The test provides only a clindamycin result; the clindamycin MIC result range for the VITEK® 2 card is  $\leq 0.25$  to  $\geq 1\mu \mathrm{g} / \mathrm{mL}$ .

{4}

In addition to the automatic dilution, there is also a manual inoculation dilution procedure described in the packager insert.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

Reproducibility was demonstrated using 10 isolates at three sites on three separate days in triplicates. Results were &gt;95% reproducible by Category Agreement only. Of the ten isolates tested, nine had off-scale results, preventing an evaluation of reproducibility in MIC.

The study included the Auto-dilution and the Manual dilution with the VITEK®2, and the Manual dilution with the VITEK®2 Compact.

#### b. Linearity/assay reportable range:

Not Applicable

#### c. Traceability (controls, calibrators, or method):

The recommended QC isolates were tested on every test occasion with the reference method and the VITEK®2. The reference method QC results were in range for every day tested. The VITEK®2 was tested a sufficient number of times to demonstrate that the system can produce QC results in the recommended range.

Quality Control was performed during the studies using both the auto-dilution and the manual method of diluting the organisms. Results demonstrated that methods were comparable with the same mode.

Quality Control Summary (VITEK®2, Auto and Manual dilution)

|  Organism | Conc in μg/ml | Auto-dilution |   | Manual dilution  |   |
| --- | --- | --- | --- | --- | --- |
|  S. pneumoniae ATCC 49619
Expected Range 0.06- 0.25 μg/ml |  | Ref. | Test | Ref. | Test  |
|   |  ≤0.03 | 5 |  | 6 |   |
|   |  0.06 | 160 |  | 160 |   |
|   |  0.12 | 4 |  | 1 |   |
|   |  0.25 |  | 168 |  | 169  |
|   |  0.5 |  | 1 |  | 1  |

An additional QC study was performed with the VITEK®2 Compact, the secondary option, at three sites, with the following results.

Page 5 of 8

{5}

Quality Control Summary (VITEK®2 Compact, Manual dilution)

|  Organism | Conc. in μg/ml | Manual-dilution  |   |
| --- | --- | --- | --- |
|  S. pneumoniae |  | Ref. | Test  |
|  ATCC 49619 | ≤0.03 |  |   |
|  Expected Range | 0.06 | 48 |   |
|  0.06- 0.25 μg/ml | 0.12 | 12 |   |
|   | 0.25 |  | 60  |
|   | 0.5 |  |   |

Inoculum density control was monitored using the DensiChek2 instrument. This was standardized weekly with all results recorded and in the expected range. Verification was performed during internal testing.

d. Detection limit:
Not Applicable

e. Analytical specificity:
Not Applicable

f. Assay cut-off:
Not Applicable

2. Comparison studies:

The reference method follows the CLSI approved broth microdilution testing conditions:

- Medium: Mueller-Hinton broth supplemented with lysed blood
- Inoculum: Direct colony suspension
- Incubation: 35°C, ambient air, 20-24 hours

a. Method comparison with predicate device:

Clinical study was performed at four external sites using the VITEK2® Streptococcus Clindamycin and broth microdilution panels containing Clindamycin. A total of 1318 clinical isolates were tested at four external sites by auto inoculation. The no growth rate was 0.08% (1/1318) in the VITEK®2 AST-ST Clindamycin Test. Two hundred and nine (15.9%: 209/1318) were from stock isolates.

Page 6 of 8

{6}

Essential agreement was not calculated because the VITEK®2 AST-ST card contained &lt;5 dilutions of Clindamycin. The table below is the comparison to the reference method.

Performance Summary Table (VITEK 2, Auto Dilution)

|   | CA total | CA# | %CA | #R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Clinical | 1317 | 1276 | 96.9 | 163 | 26 | 13 | 2  |
|  Challenge | 192 | 190 | 99.0 | 24 | 0 | 2 | 0  |
|  Combined | 1509 | 1466 | 97.2 | 187 | 26 | 15 | 2  |

CA-Category Agreement
R-resistant isolates

maj-major discrepancies
vmj-very major discrepancies
min- minor discrepancies

Manual Dilution:

The challenge set of 192 organisms was also tested at one site using the manual method of inoculation for VITEK®2 and VITEK®2 Compact with the following performance.

Comparison Challenge Data - VITEK ®2, and VITEK®2 Compact (Manual dilution)

|   | CA total | CA# | %CA | #R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  VITEK2 | 192 | 189 | 98.4 | 24 | 1 | 2 | 0  |
|  VITEK2 Compact | 192 | 189 | 98.4 | 24 | 1 | 2 | 0  |

The performance of the optional VITEK®2 Compact was evaluated in the QC, challenge, and reproducibility studies.

b. Matrix comparison:

Not Applicable

3. Clinical studies:

a. Clinical sensitivity:

Not Applicable

b. Clinical specificity:

Not Applicable

c. Other clinical supportive data (when a and b are not applicable):

Page 7 of 8

{7}

4. Clinical cut-off:
Not Applicable

5. Expected values/Reference range
S. pneumoniae and other Streptococcus spp.
≤ 0.25 (S), 0.5 (I), ≥ 1 (R)

N. Proposed Labeling:
The labeling is sufficient and satisfies the requirement of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

Page 8 of 8

---

**Source:** [https://fda-staging.innolitics.com/device/K112129](https://fda-staging.innolitics.com/device/K112129)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com