The Contour Healer Temporary Abutment is intended for use with a root-form endosseous dental abutment to aid in prosthetic rehabilitation. The abutment is a provisional restoration that aids in creating an esthetic emergence through the gingiva during the healing period. The device is for use by dental professionals for single restorations in adults. The device is for single-use only and may not be re-processed. These abutments are designed to work with the following implant systems: Zimmer [Screw-Vent Dental Implant System] Nobel Biocare [Replace HA Coated Implant, Replace TiUnite Endosseous Implant, Nobel Biocare Endosseous Implants, and Groovy Implants] BioHorizons [BioHorizons Tapered Internal Implant System].
Device Story
Pre-manufactured dental abutment; PEEK Classix construction; stainless steel retaining screw. Used by dentists, periodontists, or oral surgeons in clinical settings. Device connects to endosseous dental implants to serve as temporary prosthesis during healing (max 90 days). Biconvex dome design mimics natural tooth morphology; shapes peri-implant soft tissue; facilitates easier restorative phase by preparing tissue for final impression. Clinician removes device after healing period for permanent crown installation. Benefits include improved esthetic gingival contour and simplified restorative workflow.
Clinical Evidence
No clinical studies conducted. Substantial equivalence supported by bench testing, design validation studies by dental professionals, and adherence to ISO 10993 (biocompatibility) and ISO 14971 (risk management) standards. Device design and material formulation are substantially equivalent to previously cleared devices.
Technological Characteristics
Material: PEEK Classix (polyetheretherketone); stainless steel retaining screw. Design: Biconvex dome shape; internal hex interface with friction fit. Dimensions: Various platform sizes (5.0, 4.3, 4.5, 5.7mm). Sterilization: Non-sterile; intended for autoclave by end-user. Standards: ISO 10993, ISO 14971.
Indications for Use
Indicated for adult patients requiring prosthetic rehabilitation with root-form endosseous dental implants. Used as a provisional restoration to facilitate esthetic gingival emergence during the healing period. Contraindicated for use as a final abutment, in occlusion, or as a 'cast-to' abutment.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
K//2099
FEB - 3 2012
510(k) Summary
This summary of information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.
Submitter Information
510(k) owner's name:Contour Healer, LLC. Address:1117 Ditchley Road, Virginia Beach, VA 23451 USA Phone number: 757 288 6671 Fax number:757-428-3470 Name of contact person:Allen Hardin White, III, D.D.S. Date of Summary:September 2011
{1}------------------------------------------------
Device Trade Name: Contour Healer Temporary Abutment
Device Classification Name: Endosseous Dental Abutment
Classification: Class II, 21 CFR 872.3630; Product Code NHA
Legally marketed device to which equivalence is claimed:Zimmer Dental Plastic Temporary Abutment [K092377]
Device Description
The Contour Healer. LLC product line includes pre-manufactured dental abutments intended to serve as a temporary dental prosthesis during the healing process until a permanent crown is fabricated. The abutment is made of PEEK Classix and is directly connected to the endosseous dental abutment by a stainless steel screw.
These abutments have biconvex domes to mimic emergence of the original tooth morphology. and therefore provide for a more esthetic gingival contour during the restorative phase. The abutments come in several abutment- to-implant diameters. The abutments are intended for single-use not to exceedninety days. At the end of the treatment period, the temporary abutment is to be removed by clinician and the permanent crown is installed.
Indications for Use
The Contour Healer Temporary Abutment is intended for use with a root-form endosseous dental abutment to aid in prosthetic rehabilitation. The abutment is a provisional restoration that aids in creating an esthetic emergence through the healing period. The device is for use by dental professionals for single restorations in adults. The device is for single-use only and may not be re-processed.
These abutments are designed to work with the following implant systems:
Zimmer [Screw-Vent Dental Implant System]
Nobel Biocare (Replace HA Coated Implant, Replace TiUnite Endosseous Implant, Nobel Biocare Endosseous Implants, and Groovy Implants]
BioHorizons [BioHorizons Tapered Internal Implant System].
## Testing
First article inspections are done on each molding run to ensure that the devices meet all dimensional specifications. Routine inspections are done on 100% of the Contour Healer devices to ensure that each temporary abutment meets design specifications. This finishing and inspection process includes hand polishing of each piece, visual inspection of each piece for flash and debris, a check for correct fit [passive] of temporary abutment into the corresponding implant analog fto ensure that each piece will fit into the corresponding implant], and a check for correct screw fit into the temporary abutment, before final packaging of each piece.
The stainless screw used with each model of the temporary dental abutment are purchased to specification from the manufacturer and are received by Contour Healer with certification testing results for chemical analysis and 11 different performance specifications per standards.
{2}------------------------------------------------
Biocompatibility of the finished, sterilized temporary abutments was demonstrated per ISO-Dioompanishility of the liner of otoxicity, irritation, delayed contact hypersensitivity, and acute systemic toxicity.
Formal clinical studies were not conducted on the device since [per the FDA Guidance r orniture from Endosseous Dental Implants and Endosseous Dental Implant Abutments" issued on May 12, 2004]: 1] the material formulation is identical to a device Anouthously cleared for this intended use, 2] the design is similar to the design of the predicate proviously diourou for the Contour Healer temporary dental abutment models are devices, of the imgth's or caphant diameters for each of the models are all greater than 3.25 greater than 7 mm, 17 the accompanying or recommended implant abutment is less than min, of the ungalation of the abogy required for the manufacture or the deployment of the 50 , of the now tow township, require identical to those of the previously cleared devices of the same type.
However, formal design validation studies have been done on the Contour Healer devices by roweral dental professionals to confirm the safety and efficacy of the device design for the intended use, and to ensure that the product labeling is clear and appropriate for theindications intended use, and to onears that the propected or device-related adverse events or complications were found in these studies.
| Parameter | Device | Predicate Device |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Contour Healer | Plastic Temporary Abutment |
| Company Name | Contour Healer, LLC | Zimmer Dental, Inc. |
| 510(k) # | TBD | K092377. |
| Class | II | Same |
| 21 CFR number | 872.3630 | Same |
| 21 CFR category | Endosseous dental abutment [temporary] | Same |
| Code | NHA | Same |
| Parameter | Device | Predicate Device |
| Description | Abutment is a plastic post with a predefined junction and abutment core. The plastic stem of the abutment engages the internal aspects of the endosseous implant and is secured with a separate retaining screw. The abutment is packaged with a stainless steel screw for retaining the temporary healing abutment to the endosseous implant. The abutments are provided in straight versions only. | Abutment is a plastic post with a predefined margin and abutment cone; available in multiple cuff heights and straight and angled cone variations. The plastic hex of the abutment engages the internal hex of the <i>Tapered Screw-Vent</i> or <i>Screw-Vent</i> Implant and is secured with a separate retaining screw.<br>Abutment is packaged with two screws, a long processing screw for fabricating a screw-retained temporary restoration and a short retaining screw for fabricating a cement-retained restoration |
| Use Features | Pre-manufactured prosthetic component; screw-retained | Same; cement-retained or screw-retained |
| Ancillary components | Not applicable; for attachment to endosseous permanent implant | Same |
| Intended Use | To be used in conjunction with an endosseous dental implant fixture to aid in prosthetic rehabilitation | Same |
| Indications for use | A provisional restoration that aids in creating an esthetic emergence through the gingiva during the healing period<br>The device is for use by dental professionals for single restorations in adults. The device is for single-use only and may not be re-processed. | To fabricate and support provisional restorations that aid in creating and esthetic emergence through the gingiva during the healing period, before final restoration<br>Single- or multi-unit restorations |
| End User | Dentist, periodontist, oral surgeon | Same |
| Frequency of Use | Single use | Same |
| Method of Use | Temporary prosthesis; 90 days maximum | Same; 180 days maximum |
| Parameter | Device | Predicate Device |
| Contraindications | Not to be used as a final abutment; not to be<br>used "in occlusion"; not to be used as a<br>"cast-to" abutment | Same |
| Standards met | ISO 10993; ISO 14971 | ISO 10993 |
| Abutment<br>Material | PEEK Classix; polyetheretherketone | Same |
| Screw material | Stainless steel | Titanium alloy |
| Design [style,<br>size] | Contoured shaped with interfaces to match<br>endosseous implants with various platform<br>size [abutment to implant diameters of 5.0,<br>4.3, 4.5, and 5.7mm], in straight versions<br>only | Straight and angled versions,<br>with interfaces to match three<br>diameters of tapered screw-<br>vent implants [3.5, 4.5, and 5.7<br>mm], 1 mm and 4 mm cuff<br>height options |
| Screw geometry | Internal hex with friction fit | Internal hex [anti-rotational]<br>with friction fit |
| Sterility | Non-sterile; intended for autoclave<br>sterilization by end user | Sterilized by gamma irradiation,<br>but instructions for autoclaving<br>by the end user are provided |
| Packaging | Round plastic vial with screw cap | Sealed Tyvek blister packs |
Summary Comparison Table of New Device to Predicate Device
{3}------------------------------------------------
. .
.
.
. . . .
.
:
{4}------------------------------------------------
## Device Comparison Statement
The new device and the predicate devices are temporary dental abutments consisting of a plastic post with a predefined junction and abutment core. In both devices the plastic stem of the abutment engages the internal aspects of the endosseous implant and is secured with a separate stainless steel retaining screw.
The Contour Healer differs from the predicate device in that the shape more closely mimics the shape of the permanent crown as opposed to a straight cylinder. The most difficult stage of implant surgery is to manage the peri-implant soft tissue contour. Shape and form of the periimplant tissues are equally as important as other important factors such as bone quantity/ quality, primary stability, and bone to implant contact. The Contour Healer abutment serves as a template to easily shape the soft tissue to proper form and allows the natural contours to be maintained without compromising the stability of the newly placed implant. The restorative phase is easier as the tissues are ready to be impressed along with the impression coping on the implant.
The Contour Healerproduct for the management of the peri-implant soft tissue contour issubstantially equivalent to the previously-cleared device, with the same indications for use, abutment material, and similar design and technological characteristics and therefore does not raise any new questions for safety. Contour Healer has further supported the safety and the efficacy of this product by user design validation.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in black and is in all caps. There is also a logo on the left side of the text.
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three overlapping, curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Allen Hardin White, III, D.D.S. CEO Contour Healer, LLC 1117 Ditchley Road Virginia Beach, Virginia 23451
FEB - 3 2012
Re: K112099
Trade/Device Name: Contour Healer Temporary Dental Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: January 9, 2012 Received: January 19, 2012
Dear Dr. White:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
## Page 2 - Dr. White
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours ..
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K112099
Device Name: Contour Healer Temporary Dental Abutment
Indications for Use:
The Contour Healer Temporary Abutment is intended for use with a root-form endosseous dental abutment to aid in prosthetic rehabilitation. The abutment is a provisional restoration that aids in creating an esthetic emergence through the gingiva during the healing period. The device is for use by dental professionals for single restorations in adults. The device is for single-use only and may not be re-processed.
These abutments are designed to work with the following implant systems:
Zimmer [Screw-Vent Dental Implant System]
Nobel Biocare [Replace HA Coated Implant, Replace TiUnite Endosseous Impiant, Nobel Biocare Endosseous Implants, and Groovy Implants]
BioHorizons [BioHorizons·Tapered Internal Implant System].
Susan Rose
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112099
Prescription Use _ × (Part 21 CFR 801 Subpart D)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.