K112089 · Molnlycke Health Care Us, LLC · KGO · Aug 24, 2011 · General, Plastic Surgery
Device Facts
Record ID
K112089
Device Name
BIOGEL ORTHOPOR BROWN SURGICAL GLOVE
Applicant
Molnlycke Health Care Us, LLC
Product Code
KGO · General, Plastic Surgery
Decision Date
Aug 24, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The Biogel® OrthoPro® Brown Surgical Glove is a disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.
Device Story
Biogel® OrthoPro® Brown Surgical Glove is a sterile, powder-free, natural rubber latex glove. It features a slightly curved former design with an independent/displaced thumb to improve ergonomics. The device is colored with brown pigmentation. It is intended for use by healthcare professionals in surgical settings to serve as a barrier against infectious materials and contaminants. The glove is manufactured using the same material and coating technology as previously cleared Mölnlycke surgical gloves. It is provided as a disposable, single-use device. The primary benefit is the provision of a protective barrier for the wearer during surgical procedures.
Clinical Evidence
No clinical data was required. Substantial equivalence is supported by bench testing, including primary skin irritation, ISO closed patch sensitization, physical properties (ASTM D3577), freedom from holes (21 CFR 800.20), protein extractables (ASTM D5712), and powder residual (ASTM D6142).
Technological Characteristics
Natural rubber latex surgical glove; powder-free; brown colorant. Sterilization via radiation (10^-6 SAL). Design features a slightly curved former with an independent/displaced thumb. Complies with ASTM D3577 (physical properties/dimensions), ASTM D5712 (protein extractables), and ASTM D6142 (residual powder). Biocompatibility assessed per ISO 10993.
Indications for Use
Indicated for use as a protective hand barrier in surgical settings for individuals requiring protection against infectious materials and contaminants. No specific age or gender restrictions; contraindications not specified.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Molnlycke Health Care. The logo consists of five black circles of varying sizes on the left side. To the right of the circles, the text "MÖLNLYCKE HEALTH CARE" is written in black, with "MÖLNLYCKE" on the top line and "HEALTH CARE" on the bottom line.
# MARY OF SAFETY AND EFFECTIVENESS
| Biogel® OrthoPro® Brown Surgical Glove | |
|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| This 510(k) summary of safety and effectiveness information is being submitted in<br>accordance with the requirements of 21 CFR 807.92(c). | |
| Date Prepared: | July 20, 2011 |
| Applicant: | Mölnlycke Health Care US, LLC<br>5550 Peachtree Parkway, Suite 500<br>Norcross, GA 30092<br>Registration number: 3004763499<br>Owner/Operator Number: 8030877 |
| Official Correspondent: | Angela L. Bunn, RAC<br>Director, Regulatory Affairs for the Americas<br>Tel: 678-250-7930<br>Fax: 678-245-7746<br>e-mail: angela.bunn@molnlycke.com |
| Trade/Proprietary Name: | Biogel® OrthoPro® Brown Surgical Glove |
| Common Name: | Surgeon's Glove |
| Classification Name: | Surgeon's Glove |
| Device Class: | Class 1 |
| Regulation Number: | 21 CFR 878.4460 |
| Product Code: | KGO |
| Predicate Device Name(s): | Biogel® Sensor Surgeons Glove (K071700)<br>Biogel® Orthopedic Surgical Glove (K071465) |
## Description of Device:
The proposed device, the Biogel® OrthoPro® Brown Surgical Glove is manufactured of
natural rubber latex colored with brown nigmentation. The Dieself on and Cate natured one, Inc. Diogel Ortholfrom pigmentation. The Bioget® Orthological of the Brown
natured Glove is manufactured of the exact same metarial Surging which colored of the exact same material and coated with the Biogel® Brown of Brown of Brown of Brown of Brown of Brown of Brown of - 1 - 2019 15 manuacuted of the exact same material and coated with the Bioght
Coating which is used on the currently cleared device(s) that have been legally marketed by Mology and in ased on the currently cleared device(s) that have been legally for
The closes for many years with the addition of the brown colorant.
The glove former design used in the manufacture of this glove allows the Biogel® OrthoPro® and good onliner design used in the manufacture of this glove allows
Brown Surgical Glove to provide the user with this additional feature:
- Slightly Curved Former with Independent/Displaced thumb .
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Molnlycke Health Care. The logo consists of five black circles of varying sizes arranged in a cluster on the left side. To the right of the circles, the text "MÖLNLYCKE HEALTH CARE" is displayed in a simple, sans-serif font, with "MÖLNLYCKE" on the top line and "HEALTH CARE" on the bottom line.
# Intended Use/Indication for Use:
The Bioge!® OrthoPro® Brown Surgical Glove is a disposable device made of natural rubber
latex that is intended to be worn on the hands, usually in a surgical catting at wor 1 a 1 - 2011 - Officil 10 - Brown Surgical Glove is a disposalsetting, to annural n
latex that is intended to be worn on the hands, usually in a surgical setting, to provide the world is intonded to be worn on the hands, usually in a surgicol seen
barrier against potentially infectious material and other contaminants.
# Technological Characteristics:
The Biogel® OrthoPro® Brown Surgical Glove is substantially equivalent to the Biogel®
Sensor Surgeons Glove (K071700) and Biogel® Orthopedic Surgical Change States of States - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 (100 ourgons Clove (1071700) and Biogel® Orthopedic Surgical Glove (K071465). All Allere (K071465).
of the assessed devices have similar indications for use, materials, pr claims and method of operation.
The difference in the proposed device, Biogel® OrthoPro® Brown Surgical Glove is the l
thickness, a slight change to the former and the addition of the broun ask the caller one in the proposed device, Biogel® OrthoPro® Brown Surgical Glor
thickness, a slight change to the former and the addition of the brown colorant.
> Mölnlycke Health Care 5550 Peachtree Parkway, Suite 500 Norcross. GA 30092 USA
Tel 678 250 7900 Fax 678 250 7994 www.molnlycke.com
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Molnlycke Health Care. The logo consists of five black circles clustered together on the left side. To the right of the circles is the text "MÖLNLYCKE" in all caps on the first line and "HEALTH CARE" in all caps on the second line.
Table 1-A below provides a clear detailed comparison of the proposed device (Biogel®
OrthoPro® Brown Surgical Glove) when compared to the predicate devices.
| Feature | Biogel® OrthoPro®<br>Brown<br>Surgical Glove<br>Proposed Device | Biogel® Sensor<br>Surgeons Glove<br>Predicate Device | Biogel® Orthopedic<br>Surgical Glove<br>Predicate Device |
|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance | TBD | K071700 | K071465 |
| Manufacturer | Mölnlycke | Mölnlycke | Mölnlycke |
| Common Name | Surgeon's Glove | Surgeon's Glove | Surgeon's Glove |
| Classification # | Class I | Class I | Class I |
| Classification<br>Name | 21 CFR 878.4460 | 21 CFR 878.4460 | 21 CFR 878.4460 |
| Product Code | KGO | KGO | KGO |
| Indication For<br>Use/Intended Use | The Biogel®<br>OrthoPro® Brown<br>Surgical Glove is a<br>disposable device<br>made of natural<br>rubber latex that is<br>intended to be worn<br>on the hands, usually<br>in a surgical setting,<br>to provide a barrier<br>against potentially<br>infectious material<br>and other<br>contaminants. | The Biogel® Sensor<br>Surgeons Glove<br>(K071700) is a<br>disposable device<br>made of natural<br>rubber latex that is<br>intended to be worn<br>on the hands, usually<br>in a surgical setting,<br>to provide a barrier<br>against potentially<br>infectious material<br>and other<br>contaminants. | The Biogel®<br>Orthopedic Surgical<br>Glove (K071465) is a<br>disposable device<br>made of natural<br>rubber latex that is<br>intended to be worn<br>on the hands, usually<br>in a surgical setting,<br>to provide a barrier<br>against potentially<br>infectious material<br>and other<br>contaminants. |
| Materials<br>Coated | Latex | Latex | Latex |
| Powder Free | Yes | Yes | Yes |
| Colored | Brown | Yes | Yes |
| Sterilization<br>Method and<br>Sterility Assurance<br>Level (SAL) | Radiation<br>10-6 SAL | Natural<br>Radiation<br>10-6 SAL | Natural<br>Radiation<br>10-6 SAL |
| Biocompatibility | Materials have been<br>assessed based to<br>ISO 10993 | Materials have been<br>assessed based to<br>ISO 10993 | Materials have been<br>assessed based to<br>ISO 10993 |
Mölnlycke Health Care
5550 D-5550 Peachtree Parkway, Suite 500
5550 Peachtree Parkway, Suite 500
Norcross, GA 30092 UC A Book Cochiree Parkway.
Norcross, GA 30092 USA
·
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Mölnlycke Health Care. The logo consists of five black circles of varying sizes arranged in a cluster on the left. To the right of the circles, the words "MÖLNLYCKE" and "HEALTH CARE" are stacked on top of each other. The text is in a simple, sans-serif font.
## Performance Testing:
Test
### Result
Primary Skin Irritation
ISO Closed Patch Sensitization
Dimensions
Physical Properties
Freedom from Holes
Protein Extractables
Powder Residual
Gloves are non-irritating.
Gloves do not display any potential for sensitization.
Gloves meet requirements of ASTM D3577.
Gloves meet requirements for rubber surgical gloves
per ASTM D3577 per ASTM D3577.
Gloves exceed the requirements of 21 CFR 800.20 and ASTM D3577.
Gloves meet requirements of ASTM D5712.
Gloves meet powder level requirements for "Powder-Free" designation per ASTM D3577 testing using ASTM standard D6142, Standard Test Method for Residual Powder on Medical Gloves. Results generated values below 2mg of residual powder per glove.
### Clinical Testing:
No clinical data was required.
#### Conclusion:
The Biogel® OrthoPro® Brown Surgical Glove met the technological and performance in characteristics of ASTM D3577 and are substantially equivalent to predicate devices
characteristics of ASTM D3577 and are substantially equivalent to predicate devices indentified in this summary with respect to intended use, materials, and design,
> Mölnlycke Health Care 5550 Peachtree Parkway, Suite 500 Norcross, GA 30092 USA
Tei 678 250 7900 Fax 678 250 7984 www.molnlycke.com
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Angela L. Bunn, RAC Director, Regulatory Affairs for the Americas Molnlycke Healthcare, US, LLC 5550 Peachtree Parkway, Suite 500 Norcross, Georgia 30092
AUG 2 4 2011
Re: K112089
Trade/Device Name: Biogel® OrthoPro® Brown Surgical Glove Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: August 9, 2011 Received: August 10, 2011
Dear Ms. Bunn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{5}------------------------------------------------
#### Page 2 - Ms. Bunn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthony, V, Water
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Molnlycke Healthcare. The logo consists of a cluster of five circles on the left side, resembling cells or molecules. To the right of the circles, the text "MÖLNLYCKE HEALTHCARE" is displayed in a simple, sans-serif font. The overall design is clean and professional, conveying a sense of scientific expertise and healthcare focus.
K11 2089
#### INDICATIONS FOR USE
510(k) Number (if known): ____________ K112089
Device Name:
Biogel® OrthoPro® Brown Surgical Glove with ≤50μg/g total water extinctable protein statement, Powder-Free, Sterile
Reorder Code -- 32160, 32165, 32170, 32175, 32180, 32185, and 32190
Indications For Use:
The Bioge!® OrthoPro® Brown Surgical Glove is a disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Concurrence of CERH, et al.
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112059
Page I of
Mölnlycke Health Care 5550 Peachtree Parkiray, Suite 500 Norcross, GA 30092 USA
Tol 228 250 7900 Fax 338 250-7964 Page 74 of 941; 40.00m
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.