K112074 · Everyway Medical Instruments Co.,Ltd · ITH · Feb 1, 2012 · Physical Medicine
Device Facts
Record ID
K112074
Device Name
EVERYWAY TRACTION UNIT, MODEL EVER-TRAC ET-800
Applicant
Everyway Medical Instruments Co.,Ltd
Product Code
ITH · Physical Medicine
Decision Date
Feb 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Everyway Traction System, model EVER-TRAC ET-800 is intended for medical purpose for use in conjunction with traction accessories, such as belts and harnesses , to exert therapeutic pulling forces on the patient's body.
Device Story
The Everyway Traction Unit (ET-800) is a microprocessor-controlled powered traction system used in clinical settings. It utilizes an induction motor, speed reduction gear, pulley assembly, and springs to generate and apply therapeutic pulling forces to a patient's body via belts and harnesses. The device is operated by healthcare professionals who set parameters for force, operation cycle, and duration. The internal microprocessor controls the exertion and release of force based on these settings. Safety features include manual release, auto-shutdown, and a 'safe protect' mechanism, supplemented by warning alarms for dangerous situations. The device operates on standard single-phase AC wall power. By providing controlled, repeatable traction, the device assists in therapeutic applications prescribed by a physician.
Clinical Evidence
No clinical data was conducted for this device. Safety and performance were established through bench testing, specifically electrical compliance (IEC 60601-1) and EMC compliance (IEC 60601-1-2).
Technological Characteristics
Powered traction equipment; induction motor and speed reduction gear drive; microprocessor-controlled; AC power (120V/60Hz); includes manual release, auto-shutdown, and alarm safety features; intended for use with belts/harnesses. Compliance with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for patients requiring therapeutic pulling forces applied to the body via traction accessories (belts/harnesses) as prescribed by a physician.
Regulatory Classification
Identification
Powered traction equipment consists of powered devices intended for medical purposes for use in conjunction with traction accessories, such as belts and harnesses, to exert therapeutic pulling forces on the patient's body.
Predicate Devices
Ever Prosperous Traction System model DIGIT-TRAC 930 (K052453)
Submission Summary (Full Text)
{0}------------------------------------------------
K112074
PAGE 1 OF 2
## EVERYWAY MEDICAL INSTRUMENTS CO.,LTD.
3F1. No. 5. Lane 155. Sec. 3.Peishen Rd. Shen Keng Hsiang, Taipei Hsien, Taiwan,
FEB = 1 2012
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being prepared in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 at June 15, 2011.
The assigned 510(k) number is:
- 1. Submitter's Identifications:
Establishment: EVERYWAY MEDICAL INSTRUMENT CO., LTD. Address: 3Fl., No. 5, Lane 155, Sec. 3, Peishen Rd., Shenkeng Hsiang, Taipei Hsien 222, Taiwan Registration Number: 9616877 Operations: Manufacturer
Owner/Operator:EVERYWAY MEDICAL INSTRUMENT CO., LTD. Address : 3Fl., No. 5, Lane 155, Sec. 3, Peishen Rd., Shenkeng Hsiang, Taipei Hsien 222, Taiwan Contact Person: Robert Tu Phone : 886-2-2662-0038 Fax No: 886-2-2664-5566 e-mail : tu922@ms35.hinet.net
- 2. Name of the Device: Everyway Traction Unit, model EVER-TRAC ET-800.
- 3. Information of the 510(k) Cleared Device (Predicate Device): Ever Prosperous Traction System model DIGIT-TRAC 930(K052453)
4. Classification Information: Trade/Device Name: Everyway Traction Unit, model EVER-TRAC ET-800. Regulation Number: 21 CFR Part 890.5900 Regulation Name: Powered Traction Equipment Regulatory Class: II Product Code: ITH
- 5. Device Description:
Everyway Traction Unit, model EVER-TRAC ET-800 is a medical device constructed for use in conjunction with traction accessories, such as belts and harnesses, to exert therapeutic pulling forces on the patient's body.
For this intended operation function, ET-800 is the combination of driving moter, speed reduction gear, pulley assembly, spring, force and speed control unit... etc. All of the operational parts as above mentioned are mounted on the frame of device housing.
{1}------------------------------------------------
## EVERYWAY MEDICAL INSTRUMENTS CO.,LTD.
3F1. No. 5, Lane 155, Sec. 3,Peishen Rd, Shen Keng Hsiang, Taipei Hsien, Taiwan,
The pulling force for ET-800 is generated by the combination of induction motor and speed reduction gear. For the operation of ET-800, the single phase AC power shall be supplied from the general wall socket of house (Note: In USA, the power supply specification is single phase, 60Hz, 120V).
Basically ET-800 is a microprocessor-controlled traction system in which the operational software is built inside the control unit to properly control the exertion and release of pulling force according to the setting for applying force, operation cycle, and operation time prescribed by the physician.
To ensure the safety of operation, the device was designed and constructed to provide the following safety protection: <1> Manual release, <2>Auto shutdown, and <3>Safe protect. In addition, the device also provides warning alarm function for same near dangerous situations as mentioned in user's manual.
- 6. Intended Use:
The Everyway Traction System, model EVER-TRAC ET-800 is intended for medical purpose for use in conjunction with traction accessories, such as belts and harnesses , to exert therapeutic pulling forces on the patient's body
- 7. Comparison to the 510(k) Cleared Device (Predicate Device): The Everyway Traction Unit, model EVER-TRAC ET-800 is substantially equivalent to the Ever Prosperous Traction System model DIGIT-TRAC 930(K052453) without any significant difference in main technological and operational feature.
- 8. Discussion of Non-Clinical Tests Verification Activities Performed to Determine the Safety and Performance of EVER-TRAC ET-800 are as the followings: 1> Electrical Compliance Test according to IEC 60601-1 by accredited laboratory. 2> EMC Compliance Test according to IEC 60601-1-2 by accredited laboratory.
- 9. Discussion of Clinical Test Validation Activities Performed to Determine the Effectiveness of Device are as the followings: No particular Clinical Test was conducted for Everyway Traction Unit, model EVER-TRAC ET-800.
- 10. Conclusions
The Everyway Traction Unit, model EVER-TRAC ET-800, has the same intended use and technological characteristics as the cleared device of Ever Prosperous Traction System model DIGIT-TRAC 930(K052453. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted model could maintain the same safety and effectiveness as that of cleared device.
In the other words, Everyway Traction Unit, model EVER-TRAC ET-800 is substantial equivalent with the Ever Prosperous Traction System model DIGIT-TRAC 930(K052453) .
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB - 1 2012
Everyway Medical Instruments Co., Ltd. % Mansour Consulting LLC Mr. Jay Mansour 845 Aronson Lake Court Roswell, Georgia 30075
Re: K112074
Trade/Device Name: Everyway Traction Unit, Model EVER-TRAC ET-800 Regulation Number: 21 CFR 890.5900 Regulation Name: Power traction equipment Regulatory Class: Class II Product Code: ITH Dated: December 30, 2011 Received: January 03, 2012
Dear Mr. Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Mr. Jay Mansour
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. .
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Millerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### EVERYWAY EVERYWAY MEDICAL INSTRUMENTS CO.,LTD.
3F1., No. 5, Lane 155, Sec. 3,Peishen Rd, Shen Keng Hsiang, Taipei Hsien, Taiwan,
# Indications For Use
510(k) Number (if known): _
Device Name: Everyway Traction Unit, model EVER-TRAC ET-800.
Indications For Use:
The Everyway Traction System, model EVER-TRAC ET-800 is intended for medical purpose for use in conjunction with traction accessories, such as belts and harnesses , to exert therapeutic pulling forces on the patient's body.
Prescription Use ブ (Part 21 CFR 801 Subpart D) OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
///
(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112074
Page 1 of _ 1
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.