OBC BONE ANCHORED HEARING SYSTEM

K112053 · Oticon Medical AB · MAH · Nov 18, 2011 · Ear, Nose, Throat

Device Facts

Record IDK112053
Device NameOBC BONE ANCHORED HEARING SYSTEM
ApplicantOticon Medical AB
Product CodeMAH · Ear, Nose, Throat
Decision DateNov 18, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3302
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Ponto bone anchored hearing implant system is intended for the following patients and indications; . Patient with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) of the indicated ear should be better than or equal to 45 dB HL for use with the Ponto and Ponto Pro sound processors, 55 dB HL for use with the Ponto Pro Power sound processor. . Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies. . Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e. single sided deafness or "SSD"). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz). Also indicated for any patient who is indicated for an air-conduction contralateral routing of . signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS. The placement of a bone anchored implant is contraindicated for patient below the age of 5.

Device Story

Ponto Bone Anchored Hearing Implant System; surgically implanted device for conductive/mixed hearing loss or single-sided deafness. System consists of sound processor, abutment, and titanium implant. Sound processor captures ambient sound; converts to mechanical vibrations; transmits via abutment to titanium implant integrated into skull bone; stimulates cochlea via bone conduction. Used in clinical settings by otolaryngologists/surgeons for implantation; patient wears processor daily. Improves hearing by bypassing damaged outer/middle ear. Provides amplification for patients unable to use traditional air-conduction hearing aids.

Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and technological comparison to predicate device.

Technological Characteristics

System includes titanium bone-anchored implant, abutment, and external sound processor. Operates via bone conduction principle. Sound processors (Ponto, Ponto Pro, Ponto Pro Power) provide varying levels of amplification based on patient hearing thresholds. Connectivity via mechanical coupling to abutment. Class II device.

Indications for Use

Indicated for patients aged 5+ with conductive or mixed hearing loss (PTA BC threshold ≤45-55 dB HL depending on processor) or single-sided deafness (profound sensorineural loss in one ear, normal hearing in other, AC threshold ≤20 dB HL). Also indicated for patients requiring AC CROS hearing aids who cannot/will not use them. Contraindicated for patients under age 5.

Regulatory Classification

Identification

A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Oticon Medical AB c/o Ms. Karolin Isberg Jernby Quality Assurance & Regulatory Affairs Manager Ekonomiv. 2 SE-436 33 Askim Sweden NOV 1 8 2011 Re: K112053 Trade/Device Name: OBC bone anchored hearing aid system Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid, Bone Conduction, Implanted Regulatory Class: Class II Product Code: MAH Dated: November 7, 2011 Received: November 10, 2011 Dear Ms. Jernby: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Zach Roher AM Malvina B. Eydelman, M. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for use ## ん112053 510(k) Number: _ Device Name: Ponto Bone Anchored Hearing Implant System ## Indications for Use: The Ponto bone anchored hearing implant system is intended for the following patients and indications; - . Patient with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) of the indicated ear should be better than or equal to 45 dB HL for use with the Ponto and Ponto Pro sound processors, 55 dB HL for use with the Ponto Pro Power sound processor. - . Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies. - . Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e. single sided deafness or "SSD"). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz). - Also indicated for any patient who is indicated for an air-conduction contralateral routing of . signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS. The placement of a bone anchored implant is contraindicated for patient below the age of 5. OR Prescription Use __ X_ (Part 21 CFR 801.109) · Over-The-Counter Use _ (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Vasut Dásku (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K112053
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...