K112050 · Biosense Webster, Inc. · OBJ · Nov 18, 2011 · Cardiovascular
Device Facts
Record ID
K112050
Device Name
SOUNDSTAR ECO DIAGNOSTIC ULTRASOUND CATHETER
Applicant
Biosense Webster, Inc.
Product Code
OBJ · Cardiovascular
Decision Date
Nov 18, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The Biosense Webster SOUNDSTAR® eco Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO® 3 EP Navigation Systems, the SOUNDSTAR® eco Catheter provides location information. Please refer to the Compatibility Matrix Insert for Compatible CARTO® 3 Systems as each catheter is compatible with a specific version of CARTO® 3 and is not backwards compatible with previous versions of CARTO® 3 EP Navigation Systems.
Device Story
The SOUNDSTAR® eco is a 10F intracardiac echo (ICE) catheter (90 cm) featuring an integrated acoustic array and magnetic location sensor. It functions by acquiring real-time ultrasound images while providing location data to the CARTO® 3 Navigation System (Version 2.2). The device connects to compatible ultrasound systems (Siemens Acuson or GE Vivid series) via a bifurcated tail and SwiftLink cable. It is used by physicians in clinical settings to visualize cardiac anatomy and other devices during electrophysiology procedures. The integration of ultrasound images with electromagnetic maps assists clinicians in navigation and decision-making, potentially improving procedural accuracy and patient outcomes. Modifications from predicate devices involve relocating catheter-specific electronics (calibration memory) to the connector and non-specific electronics (amplifiers) to the extension cable.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
10F ICE catheter with acoustic array and magnetic location sensor. Compatible with CARTO® 3 (v2.2) and specific Siemens/GE ultrasound systems. Complies with IEC 60601-2-37 for transducer surface temperature. Sterilization and materials are identical to predicate devices.
Indications for Use
Indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, and visualization of other devices in the heart. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
SOUNDSTAR 3D Ultrasound Catheter (SNDSTR10) (K070242)
SOUNDSTAR 3D Ultrasound Catheter (SNDSTR10G) (K092064)
Submission Summary (Full Text)
{0}------------------------------------------------
# K112050
# 15 510(K) SUMMARY
# SPONSOR'S NAME & ADDRESS
Biosense Webster. Inc. 3333 Diamond Canyon Road Diamond Bar, CA 91765
# OFFICIAL CORRESPONDENT
Marina Guevrekian Senior Specialist, Regulatory Affairs Phone: 909-839-8714 Fax: 909-839-8804 Email: mguevrek@its.jnj.com
# SUBMISSION DATE
July 13, 2011
# TRADE NAME
# SOUNDSTAR® eco DIAGNOSTIC ULTRASOUND CATHETER
(Part Number: M-5723-15 / Catalogue Number: 10438577, Compatible with Siemens Ultrasound systems)
# SOUNDSTAR® eco DIAGNOSTIC ULTRASOUND CATHETER
(Part Number: M-5723-16 / Catalogue Number: 10439072, Compatible with GE Ultrasound systems)
# COMMON NAME
Electrophysiology Mapping/Ultrasound Catheter
# CLASSIFICATION NAME/PRODUCT CODE
Intravascular Ultrasound Catheter/OBJ
# CLASSIFICATION
Class II, 21 CFR 870.1200
# PREDICATE DEVICE
The modified SOUNDSTAR eco10F and SOUNDSTAR eco10FG Catheters are substantially equivalent to the Biosense Webster SOUNDSTAR 3D Ultrasound Catheter (SNDSTR10) cleared on May 15, 2007 via 510(k) K070242 and also SOUNDSTAR 3D Ultrasound Catheter (SNDSTR10G) cleared on August 7, 2009 via 510(k) K092064 respectively.
{1}------------------------------------------------
# DESCRIPTION OF MODIFIED DEVICE
The modified SOUNDSTAR eco10F and SOUNDSTAR eco10FG Catheters are both 90 cm 10F IntraCardiac Echo (ICE) Catheters with an acoustic arrav identical to the currently cleared SNDSTR10 and SNDSTR10G Catheters. The catheters have a magnetic location sensor (providing location information to the CARTO 3 Navigation System Version 2.2) and an ultrasound transducer (acquiring real time ultrasound images) embedded in the tip.
The modified SOUNDSTAR eco Catheters have a bifurcated 'tail' originating from their handle which is nearly identical to the bifurcated tail of the predicate devices. One leg terminates in the SOUNDSTAR Flex Tab connector, which connects via the appropriate SwiftLink cable to the corresponding Ultrasound system. For the SOUNDSTAR eco10F Catheter the SwiftLink cable connects to Acuson Cypress, Acuson Sequoia or Acuson X300 Ultrasound systems. For the SOUNDSTAR eco10FG Catheter the SwifLink cable connects to GE Vivid-i or Vivid-q Ultrasound systems.
Both versions of SOUNDSTAR eco10F and SOUNDSTAR eco10FG Catheters are based on the existing SOUNDSTAR 3D 10F (SNDSTR10) and SOUNDSTAR 3D 10FG (SNDSTR10G) catheters and have the same intended use and clinical applications. The predicate and modified catheters share the majority of components and manufacturing process as explained previously.
The modified SOUNDSTAR eco10F and SOUNDSTAR eco10FG Catheters, when connected to the corresponding Ultrasound Systems, will provide real-time integration of ultrasound images with CARTO 3 Version 2.2 electromagnetic acquired maps.
# INDICATIONS FOR USE
The Biosense Webster SOUNDSTAR® eco Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO® 3 EP Navigation Systems, the SOUNDSTAR® eco Catheter provides location information. Please refer to the Compatibility Matrix Insert for Compatible CARTO® 3 Systems as each catheter is compatible with a specific version of CARTO® 3 and is not backwards compatible with previous versions of CARTO® 3 EP Navigation Systems.
# DESCRIPTION OF MODIFICATION
The modified SOUNDSTAR eco10F and SOUNDSTAR eco10FG Catheters are identical to the currently cleared SNDSTR10F and SNDSTR10FG Catheters in terms of: Indication for Use, Material, Manufacturing methods, Operating principles, Fundamental Scientific Technology, Performance, Array and Sensor specifications, Shelf Life and Sterilization methods. Modifications were only made in the handle of the catheter. Specifically, the catheter specific electronics (i.e. memory for sensor calibration data) were moved to the Hypertonics connector, and non catheter specific electronics (i.e. amplifiers) were moved to the Extension Cable.
{2}------------------------------------------------
1/6
# · SoundStar® eco ULTRASOUND CATHETER
COMPATIBILITY MATRIX
13 of 82
M-5276-702BC
·
-
{3}------------------------------------------------
English
# SoundStar®eco DIAGNOSTIC ULTRASOUND CATHETER COMPATIBILITY MATRIX INSERT
ADDENDUM TO SOUNDSTAR® eao DIAGNOSTIC ULTRASOUND CATHETER INSTRUCTIONS FOR USE (M-5276-701)
The SOUNDSTAR® eco Diagnostic Ultrasound Catheter has conjunction with the CARTO® EP Navigation Systems listed in the following table.
| Ultrasound Systems | CARTO® Systems |
|--------------------|------------------------|
| ACUSON Sequoia™ | CARTO® 3 version 2.2 * |
| ACUSON Cypress™ | CARTO® 3 version 2.2 * |
| ACUSON X300™ | CARTO® 3 version 2.2 * |
"The Soundstar® eco Diagnostic ULTrasound Catherer is not backwards compatible with previous versions of CARTO® 3 EP Navigation System,
TRANSDUCER SURFACE TEMPERATURE The following table provides the maximum surface temperature of the SoundSTAR® Catheter with the relevant ultrasound system. The tissue mirnicking material (TMM) temperature is displayed in accordance with IEC 60601-2-37 requirements.
| Ultrasound System | TMM (Max Temp) |
|-------------------|----------------|
| ACUSON Sequoia* | 42.1°C |
| ACUSON Cypress* | 42.1°C |
| ACUSON X300* | 41.8°C |
DISCLAIMER OF WARRANTY AND LIMITATION OF LIABILITY
THERE IS NO EXPRESS OR IMPLIED WARRANTY.
INCLUDING WITHOUT LIMITATION ANY IMPLIED INARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE, ON THE PRODUCT(S) FARTHOBEAR FOR COLLERE NE CRORUMS AFFILIATED
SHALL BIOSENSE WEBSTER NO. CRORUMS AFFILIATED
COMPANIES, BE LIASTER ANC. CORUNC AFFULACED
COMPANIES, BEE LIASLE FOR ANY SECURAL C OTHER THAN AS EXPRESSLY PROVIDED BY SPECIFIC
LAW. WITHOUT LIMITING THE FOREGOING, BIOSENSE WEBSTER, INC. OR ITS AFFILIATED COMPANIES, SHALL
NOT BE LIABLE FOR ANY SPECIAL, DIRECT. INCIDENTAL, CONSEQUENTIAL, OR OTHER DAMAGES, LABELED FOR SINGLE USE OR WHERE REUSE ISS TROSENSE WEBSTER, INC. PRINTED MATTER. INCLUDING THIS PUBLICATION, ARE INFORMATIONAL
ONLY AND MEANT SOLELY TO GENERALLY DESCRIBE THE PRODUCT AT THE TIME OF MANUFACTURE AND ARE NOT MADE OR GIVEN AS A WARRANTY OF THE PRESCRIBED PRODUCT IN ANY WAY.
7n catheter M-5723-15
eco 10F
| 2
M-5276-702B C
{4}------------------------------------------------
Biosense Webster. 0086
# SoundStar®eco
ULTRASOUND CATHETER
COMPATIBILITY MATRIX INSERT
Compatibility Matrix Inisert Engilish Page
M-5276-704XBXC
. . .
{5}------------------------------------------------
English
# SOUNDSTAR® eco DIAGNOSTIC ULTRASOUND CATHETER COMPATIBILITY MATRIX INSERT
ADDENDUM TO SOUNDSTAR® eco DIAGNOSTIC ULTRASOUND CATHETER INSTRUCTIONS FOR USE (M-5276-703)
The SoundSTAR® eco Diagnostic Ultrasound Catheter has been validated for use with the GE utrasound systems In conjunction with the CARTO® EP Navigation Systems listed in the following table.
| Ultrasound Systems | CARTO® Systems |
|--------------------|------------------------|
| Vivid-I* | CARTO® 3 version 2.2 * |
| Vivid-q* | CARTO® 3 version 2.2 * |
*The SoundSTAR® eoo Diegnostic Us ITRASOUND versions of CARTO 3 EP Navigation System.
TRANSDUCER SURFACE TEMPERATURE The following table provides the maximum surface
temperature of the SOUNDSTAR® Catheter with the relevant ultrasound system. The tissue mimicking material (TMM) temperature Is displayed in accordance with IEC 60601-2-37 requirements.
| Ultrasound System | TMM (Max Temp) |
|-------------------|----------------|
| Vivid-i* | 41.8°C |
| Vivid-q* | 41.8°C |
For catheter 175723-16
eco 10FG
| 2
LIABILITY
M-5276-704XBXC
DISCLAIMER OF WARRANTY AND LIMITATION OF
EADING IS NO EXPRESS OR IMPLIED WARRANTY,
INCLUDING WITHOUT LIMITATION ANY INPLIED
INCEDDING WITHOD COMPANIE ADA ANT INFORCED
PAARCAULA OF MERCEMANA ARTE OR FUTULOS FOR A
PARTACIAL OF MERCEMANA CARA AND CONSTITUTIOS FOR A
DESCRIBED HEBREWS UNDER INCOLORIES
INCIDENTAL, CONSEQUENTIAL, OR OTHER DAMAGES. LAW. WITHOUT LIMITING THE FOREGOING, BIOSENSE WEBSTER, INC. OR ITS AFFILIATED COMPANIES, SHALL NOT BE LIABLE FOR ANY SPECIAL, DREET, OR OTHER DAMAGES, ARISING OUT OF THE REUSE OF ANY PRODUCT(S) LABELED FOR SINGLE USE OR WHERE REUSE IS PROHIBITED BY APPLICABLE LAW.
DESCRIPTIONS AND SPECIFICATIONS APPEARING IN
BIOSENSE WEBSTER, INC. PRINTED MATTER, INCLUDING THIS PUBLICATION, ARE INFORMATIONAL
ONLY AND MEANT SOLELY TO GENERALLY DESCRIBE THE PRODUCT AT THE TIME OF MANUFACTURE AND
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 1 8 2011
Biosense Webster, Inc. c/o Marina Guevrekian, Ph.D. Senior Specialist, Regulatory Affairs 3333 Diamond Canyon Road Diamond Bar, CA 91765
Re: K112050
> Trade/Device Name: SOUNDSTAR® eco 10F and SOUNDSTAR® eco 10FG catheters Regulatory Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II (Two) Product Code: OBJ Dated: October 20, 2011 Received: October 21, 2011
Dear Dr. Guevrekian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Marina Guevrekian, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
W
Bram D. Zuckerman, M.D. Directør Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# 7 INDICATIONS FOR USE
# 510(k) No (if known): K112050
# Device Name:
SOUNDSTAR® eco 10F CATHETER SOUNDSTAR® eco 10FG CATHETER
# Indications for Use:
The Biosense Webster SOUNDSTAR® eco Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO® 3 EP Navigation Systems, the SOUNDSTAR® eco Catheter provides location information. Please refer to the Compatibility Matrix Insert for Compatible CARTO® 3 Systems as each catheter is compatible with a specific version of CARTO® 3 and is not backwards compatible with previous versions of CARTO® 3 EP Navigation Systems.
Prescription Use ----V — AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
1/5
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
| | 4 | |
|----------------|---|---------------|
| | | -<br>-4-4 |
| Strings<br>100 | | ------------- |
Division of Cardiovascular Devices 510(k) Number_K(2050
{9}------------------------------------------------
Image /page/9/Picture/5 description: The image shows the logo for Biosense Webster, a Johnson & Johnson company. The logo features the company name in a simple, sans-serif font, with the Johnson & Johnson tagline in a smaller font underneath. To the right of the company name is a stylized graphic that resembles a heart with a circle and lines inside.
# ULTRASOUND CATHETER
ULTRASOUND CATHETER
COMPATIBILITY MATRIX
INSERT
Exclusi
Companeraty Matrix Insert
Page
2/4
..
M-5276-702BC
{10}------------------------------------------------
# SOUNDSTAR® ECO DIAGNOSTIC ULTRASOUND CATHETER COMPATIBILITY MATRIX INSERT
ADDENDUM TO SOUNDSTAR® COO DIAGNOSTIC ULTRASOUND CATHETER INSTRUCTIONS FOR USE (M-5276-701)
The SoundSTAR® eoo Diagnostic Ultrasound Catheter has been validated for use with the Siemens ultrasound systems in conjunction with the CARTO® EP Navigation Systems listed in the following table.
| Ultrasound Systems | CARTO® Systems |
|--------------------|------------------------|
| ACUSON Sequola™ | CARTO® 3 version 2.2 * |
| ACUSON Cypress™ | CARTO® 3 version 2.2 * |
| ACUSON X300™ | CARTO® 3 version 2.2 * |
The SoundStar® eco Diagnostic Ul trasound Catherer is not backwards compatible with previous versions of CARTQ® 3 EP Navigation Syslem.
TRANSDUCER SURFACE TEMPERATURE The following table provides the maximum surface the relieving the SOUNDSTAR® Catheter with the relevant
ultrasound system. The tissue mimiciding material (TMM) temperature is displayed in accordance with IEC 60601-2-37 requirements.
| Ultrasound System | TMM (Max Temp) |
|-------------------|----------------|
| ACUSON Sequoia* | 42.1°C |
| ACUSON Cypress* | 42.1°C |
| ACUSON X300* | 41.8°C |
DISCLAIMER OF WARRANTY AND LIMITATION OF
THERE IS NO EXPRESS OR IMPLIED WARRANTY, THEREAKTY WITHORCLIANTABLITY OR FITHESS FOR A
WARRANTT WITHORCHANTABLITY OR FITNESS FOR A
WARRANTT WITHOSE, ON THE PRODUCT ISSUESTISS
PARTICULAR PURPOSE, ON THE PRODUCTION IS SHALL BIOSENSE WEBSTER, INC., OR ITS AFFILIATED
COMPANIES, BE LIABLE FOR ANY SPECIAL, DIRECT, OUTHER THAN AS EXPRESSSLY PROVIDED BY SPECIAL SECURITIES OF SPEENSE
OTHER THAN AS EXEREESSLY PROVIDED BY SPECIAL BE SPECIALS SE WEBSTER, INC. OR IT'S AFFILIATED COMPANIES, SHALL NOT BE LIABLE FOR ANY SPECIAL, DIRECT, ARISING OUT OF THE REUSE OF ANY PRODUCT(S) LABELED FOR SINGLE USE OR WHERE REUSE IS PROHIBITED BY APPLICABLE LAW.
DESCRIPTIONS AND SPECIFICATIONS APPEARING IN BIOSENSE WEBSTER, INC. PRINTED MATTER,
INCLUDING THIS PUBLICATION, ARE INFORMATIONAL ONLY AND MEANT SOLELY TO GENERALLY DESCRIBE THE PRODUCT AT THE TIME OF MANUFACTURE AND ARE NOT MADE OR GIVEN AS A WARRANTY OF THE PRESCRIBED PRODUCT IN ANY WAY.
ta catheter 11-5723-15
eco. 10F
2/1
LIABILITY
-2 M-5276-702B_G
{11}------------------------------------------------
Biosense Webster.
a *johnson* *johnson* company
CE
0086
# Sound SoundStak® eco
ULTRASOUND CATHETER
COMPATIBILITY MATRIX
INSERT
### Compatibility Matrix Insert English Page
.
M-5276-704XBXC · . .
{12}------------------------------------------------
# SOUNDSTAR® eco DIAGNOSTIC ULTRASOUND CATHETER COMPATIBILITY MATRIX INSERT
ADDENDUM TO SOUNDSTAR® ECO DIAGNOSTIC ULTRASOUND CATHETER INSTRUCTIONS FOR USE (M-5276-703)
The SOUNDSTAR® eco Diagnostic Ultrasound Catheter has been validated for use with the GE ultrasound systems in
conjunction with the CARTO® EP Navigation Systems listed in the following table.
| Ultrasound Systems | CARTO® Systems |
|--------------------|------------------------|
| Vivid-i* | CARTO® 3 version 2.2 * |
| Vivid-q* | CARTO® 3 version 2.2 * |
· The SoundStar® eco Diagnostic ULTRasoung Catherer is not backwards compatible with previous versions of CARTO
3 EP Navigation System.
TRANSDUCER SURFACE TEMPERATURE The following table provides the maximum surface
temperature of the SOUNDSTAR® Catheter with the relevant
ultrasound system. The tissue mimicking material (TMM) temperature is displayed in accordance with IEC 60601-2-37 requirements.
| Ultrasound System | TMM (Max Temp) |
|-------------------|----------------|
| Vivid-1* | 41.8°C |
| Vivid-g* | 41.8°C |
For catheter M5723-16
eco 10FG
DISCLAIMER OF WARRANTY AND LIKITATION OF LIABILITY
THERE IS NO EXPRESS OR IMPLIED WARRANTY, INPLIED WARRANTY OF MERCHANTABILITY OF ARRETHS
PARTICULAR PURPOSE, ON THE PRODUCTS OR FUTUS
PARTICULAR PURPOSE, ON THE PRODUCTS OF CORPORATIONS ANDER SHALL BIOSENSE WEBSTER, INC., OR ITS AFFILLATED
COMPANIES, BE LIABLE FOR ANY SPECIAL, DIRECT, OTHER THAN AS EXPRESSLY PROVIDED BY SPECIFIC WEBSTER INC. OR ITS AFFILIATED COMPANIES, SHALL NEBOTEN, IROFOR ANY SPECIAL, DIRECT,
INCIDENTAL, CONSEQUENTIAL, OR OTHER DAMAGES, ARISING OUT OF THE REUSE OF ANY PRODUCT(S) LABELED FOR SINGLE USE OR WHERE REUSE IS PROHIBITED BY APPLICABLE LAW. DESCRIPTIONS AND SPECIFICATIONS APPEARING IN BIOSENSE WEBSTER AND SPECIFICATION, ARE INFORMATIONS IN
INCLUDING THIS PUBLICATION, AREE MEDIRER
INCLUDING THIS PUBLICATION, AREE MEDIRER
INCLUDING THIS PUBLICATION, AREE MED ONLY AND MEANT SOLELY TO GENERALLY DESCRIBE THE PRODUCT AT THE TIME OF MANUFACTURE AND
| 2
M-5276-704XBXC
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.