The Koo SVN is a handheld, pneumatic nebulizer designed to aerosolize prescription drugs for inhalation by a patient. Its use is indicated whenever a healthcare professional administers or prescribes medical aerosol products to a patient using a small volume nebulizer. The Koo SVN is intended for use with pediatric (defined by the prescribed medication) and adult patients consistent with the indications for the aerosol medication. This includes hospital/institutional settings, home care use, schools and long term care facilities.
Device Story
Koo SVN is a handheld, pneumatic (gas-powered) jet nebulizer; utilizes external compressed air or oxygen source (6-8 lpm) to aerosolize liquid prescription drugs. Device consists of nebulizer chamber, mouthpiece/hose, face mask, and aerosol tee. Used in hospitals, clinics, schools, and home care settings; operated by healthcare professionals or patients. Continuous aerosolization occurs during inhalation and exhalation. Output is inhaled aerosolized medication; facilitates drug delivery to respiratory tract. Single-patient, multi-use design; accessories are DEHP-free PVC. Performance verified via cascade impactor testing for particle size (MMAD), respirable fraction, and total output. Benefits include effective delivery of prescribed aerosolized medications to patients.
Clinical Evidence
Bench testing only. Comparative performance testing via cascade impactor evaluated total output, particle size (MMAD), geometric standard deviation, respirable fraction, and respirable mass for Albuterol, Ipratropium, and Cromolyn. Results showed equivalent performance to the predicate. Additional testing included simulated life/cleaning validation (30 cycles), environmental (hot/cold cycles), and mechanical (drop/dimensional) testing. All tests met specifications at a 95% confidence level.
Technological Characteristics
Pneumatic jet nebulizer; gas-powered (6-8 lpm). Materials tested per ISO 10993 (cytotoxicity, sensitization, irritation). Accessories made of DEHP-free PVC. Single-patient, multi-use. No software or electronic components.
Indications for Use
Indicated for pediatric and adult patients requiring aerosolized prescription medication. No contraindications.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Koo SVN
K 112041
# 510(k) Summary Page 1 of 4 1 5-Jul-201 1
DEC 222 2011
| KOO (Shanghai) Industries Co., Ltd.<br>100 Zhongde Road, Dakun Industrial Park<br>Songjiang Shanghai 201614 China | T - 011 86 21 57858410<br>F - 011 86 21 57858410 x.107 |
|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Official Contact: | Chris Koo - President |
| Proprietary or Trade Name: | Koo Small Volume Nebulizer (SVN) |
| Common/Usual Name: | Small Volume Nebulizer |
| Classification Name: | Nebulizers (direct patient interface)<br>Procode - CAF - CFR 868.5630 |
| Predicate Devices: | K926055 - Miller (Vixone) Westmed |
### Device Description:
Device Deserription.
The Koo SVN is a simple handheld small volume nebulizer powered by compressed air to nebulize the I ne Noo S V It is a simple handler shall vith a mouthpiece or standard aerosol / oxygen face mask. The SVN can be packaged with optional accessories, i.e., oxygen tubing, mouthpiece and hose, and face mask. The nebulizer and its accessories are single patient, multi-use devices.
## Indications for Use:
The Koo SVN is a handheld, pneumatic nebulizer designed to aerosolize prescription drugs for inhe reo b r r louinent. Its use is indicated whenever a healthcare professional administers or prescribes medical aerosol products to a patient using a small volume nebulizer.
The Koo SVN is intended for use with pediatric (defined by the prescribed medication) and adult patients consistent with the indications for the aerosol medication. This includes hospital/institutional settings, home care use, schools and long term care facilities.
| Patient Population: | Pediatric (defined by the prescribed medication) and adult patients consistent with the indications for the aerosol medication |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Environment of Use: | Hospital/institutional settings, home care use, schools and long term care facilities |
| Contraindications: | None |
{1}------------------------------------------------
# Premarket Notification 510(k) Section 5 – 510(k) Summary
# 510(k) Summary
| Features | Proposed SVN | Predicate Miller (VixOne) K926055 |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The Koo SVN nebulizer is a handheld, pneumatic nebulizer designed to aerosolize prescription drugs for inhalation by a patient. Its use is indicated whenever a healthcare professional administers or prescribes medical aerosol products to a patient using a small volume nebulizer. | A handheld, pneumatic nebulizer designed to aerosolize prescription drugs for inhalation by a patient. Its use is indicated whenever a healthcare professional administers or prescribes medical aerosol products to a patient using a small volume nebulizer |
| Environment of Use | Hospital/institutional settings, home care use, schools and long term care facilities. | Hospital/institutional settings, home care use, schools and long term care facilities. |
| Patient Population | Pediatric (defined by the prescribed medication) and adult patients consistent with the indications for the aerosol medication | Pediatric and Adult<br>Pediatric population not defined |
| Contraindications | None | None |
| Principle of Operation | Pneumatic (gas powered) jet nebulizer | Pneumatic (gas powered) jet nebulizer |
| Aerosolization | Continuous during inhalation and exhalation | Continuous during inhalation and exhalation |
| Compressed gas source | Nebulizer compressor<br>Wall air / oxygen with flow rate control | Nebulizer compressor<br>Wall air / oxygen with flow rate control |
| Typical flow rate | 6-8 lpm | 6-8 lpm |
| Components available in kit with nebulizer | Mouthpiece / Hose<br>Face Mask<br>Oxygen / Delivery tubing<br>Aerosol tee | Mouthpiece / Hose<br>Face Mask<br>Oxygen / Delivery tubing<br>Acrosol tee |
| Component / Accessories intended use | All are single patient, multi-use | All are single patient, multi-use |
| Software driven | No | No |
| Performance | | |
| Materials tested per ISO 10993 or identical to another device | Cytotoxicity<br>Sensitization<br>Irritation | Identical tests to K091272<br>For the tests required |
| DEHP | PVC based components certified DEHP Free | Not labeled |
# Comparison to Predicates
{2}------------------------------------------------
. .
# 510(k) Summary Page 3 of 4 15-Jul-2011
| Features | Proposed<br>SVN | Predicate<br>Miller (VixOne)<br>K926055 |
|--------------------------------------------------|-----------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Particle Characterization per Cascade Impactor | | |
| Total Output (ug) | Albuterol -<br>$1005 \pm 21$ | Albuterol -<br>$1160 \pm 96$ |
| | Ipratropium -<br>$184 \pm 10$ | Ipratropium -<br>$196 \pm 7$ |
| | Cromolyn -<br>$4156 \pm 113$ | Cromolyn -<br>$4344 \pm 353$ |
| Particle size (MMAD)<br>(Microns) | Albuterol -<br>$1.80 \pm 0.2$ | Albuterol -<br>$2.10 \pm 0.17$ |
| | Ipratropium -<br>$1.90 \pm 0.1$ | Ipratropium -<br>$1.93 \pm 0.06$ |
| | Cromolyn -<br>$1.53 \pm 0.06$ | Cromolyn -<br>$1.57 \pm 0.21$ |
| Geometric Std. Dev.<br>(GSD) | Albuterol -<br>$2.84 \pm 0.14$ | Albuterol -<br>$3.00 \pm 0.25$ |
| | Ipratropium -<br>$2.71 \pm 0.1$ | Ipratropium -<br>$3.11 \pm 0.48$ |
| | Cromolyn -<br>$2.63 \pm 0.2$ | Cromolyn -<br>$2.83 \pm 0.05$ |
| Respirable Fraction<br>(% Mass 0.5-5<br>microns) | Albuterol -<br>$69.0\% \pm 2.0\%$ | Albuterol -<br>$67.3\% \pm 3.2\%$ |
| | Ipratropium -<br>$70.3\% \pm 0.6\%$ | Ipratropium -<br>$66.7\% \pm 5.8\%$ |
| | Cromolyn -<br>$71.0\% \pm 3.6\%$ | Cromolyn -<br>$68.0\% \pm 2.0\%$ |
| Respirable Mass (ug<br>0.5 -5.0 microns) | Albuterol -<br>$693 \pm 17$ | Albuterol -<br>$781 \pm 71$ |
| | Ipratropium -<br>$129 \pm 6$ | Ipratropium -<br>$131 \pm 13$ |
| | Cromolyn -<br>$2953 \pm 232$ | Cromolyn -<br>$2949 \pm 155$ |
| Treatment time (min) | Albuterol -<br>$4.00 \pm 0$ | Albuterol -<br>$4.33 \pm 0.38$ |
| | Ipratropium -<br>$2.67 \pm 0.29$ | Ipratropium -<br>$2.92 \pm 0.38$ |
| | Cromolyn -<br>$1.50 \pm 0$ | Cromolyn -<br>$1.58 \pm 0.38$ |
| Confidence level of<br>testing | The test and number of samples<br>tested provided a 95% confidence<br>level | The test and number of samples tested<br>provided a 95% confidence level |
| | Simulated Life / Mechanical and Environmental testing | |
| Nebulizer | Cleaned and tested after 30 cycles | Single patient, multi-use |
| Environmental testing<br>all components | Hot / Cold cycles | |
| Mechanical testing | Dimensional changes<br>Drop test | |
# Substantial Equivalence Discussion
The above table compares the key features of the proposed Koo SVN nebulizer with the identified predicate and demonstrates that the device can be found to be substantially equivalent.
In summary one can conclude that substantial equivalence is met based upon the following:
The SVN small volume, handheld nebulizer is viewed as substantially equivalent to the predicate device because:
{3}------------------------------------------------
## 510(k) Summary Page 3 of 4 15-Jul-2011
#### Indications -
The proposed indications for use are to aerosolize commonly prescribed medications is identical to predicate - K926055 - Miller - Vixone, now owned by Westmed.
## Technology -
The design as a jet (gas powered) nebulizer powered by an external compressed gas source is the identical principle of operation as the predicate - K926055 - Miller - Vixone.
#### Materials -
The materials in the gas and fluid pathway have been tested per ISO 10993 and found to be nonreactive.
In addition evidence to support DEHP free for accessories made of PVC is provided.
、
### Environment of Use -
The proposed environments of use are common and usual for handheld nebulizers and identical to predicate - K926055 - Miller - Vixone.
## Patient Population -
The patient population of adult and pediatric (defined by the prescribed medication) patients consistent with the indications for the aerosol medication. With the typical gas flow of 6-8 lpm, this is sufficient to satisfy the expected tidal volume of pediatric population as well as an adult. This is identical to predicate - K926055 - Miller - Vixone.
## Comparative Performance -
We performed comparative particle characterization testing via Cascade Impactor and the results demonstrated fequivalent performance to the predicate K926055.
In addition, ye performed testing related to intra- and inter-sample dose and particle variability. simulation life / cleaning validation, environmental and mechanical testing. The results demonstrated that the proposed device either passed or met its performance specifications after each test
All testing gemonstrated that the proposed device is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
KOO (Shanghai) Industries Company, Limited C/O Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
DEC 2 2 2011
Re: K112041
Trade/Device Name: Koo Small Volume Nebulizer (SVN) Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: December 14, 2011 Received: December 15, 2011
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{5}------------------------------------------------
Page 2 – Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely, yours,
hh for
Arthur D. Weston R.S.M.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
510(k) Number:
(To be assigned)
Device Name:
Koo Small Volume Nebulizer (SVN)
Indications for Use:
The Koo SVN is a handheld, pneumatic nebulizer designed to aerosolize prescription drugs for inhalation by a patient. Its use is indicated whenever a healthcare professional administers or prescribes medical aerosol products to a patient using a small volume nebulizer.
The Koo SVN is intended for use with pediatric (defined by the prescribed medication) and adult patients consistent with the indications for the aerosol medication. This includes hospital/institutional settings, home care use, schools and long term care facilities.
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K 11204 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.