# TI-MAX X450 (K112024)

_Nakanishi, Inc. · EFB · Jun 20, 2012 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K112024

## Device Facts

- **Applicant:** Nakanishi, Inc.
- **Product Code:** [EFB](/productcode/EFB.md)
- **Decision Date:** Jun 20, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Ti-Max X450 is an air-powered dental handpiece with intended use of being a surgical tool for impacted third molar removal and periodontal procedures for which a conventional handpiece would be used. The Ti-Max X450 is intended for use with a friction grip bur that conforms to ISO 1797-1 standard. Recommended supply air pressure is between 0.22 and 0.30 MPa, which results in high-speed bur rotation (approximately 380,000 to 450,000 RPM).

## Device Story

Air-powered high-speed dental handpiece; rotates friction grip burs at 380,000–450,000 RPM. Features 45-degree back angle; available in non-optic, fiber-optic, and LED models. Connects to dental units via quick-connect coupling or swivel adapter. Operated by dental professionals in clinical settings. Provides torque of at least 0.09Ncm; includes water jet for cooling/irrigation at bur point. Facilitates surgical access for third molar extraction and periodontal procedures.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Air-powered turbine; 380,000–450,000 RPM; 0.22–0.30 MPa supply pressure. 45-degree head angle. Connections comply with ISO 9168. Friction grip burs per ISO 1797-1. Features include water jet, fiber-optic, or LED illumination. Torque >= 0.09Ncm; noise <= 80dBA.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- Palisades Dental, Impact Air 45 Handpiece ([K952927](/device/K952927.md))
- Aerobine, Inc., Air Powered Dental Handpiece [Karam45, Dexor45] ([K101717](/device/K101717.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for NSK, which is a Japanese manufacturer of bearings. The logo is in white text on a black background. The letters are bold and sans-serif. The logo is simple and easy to read.

# 5. 510(k) SUMMARY

| Submitter:                                      | Nakanishi, Inc.<br>700 Shimohinata<br>Kanuma-Shi, Tochigi-Ken Japan 322-8666<br>JUN 20 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                 | Mr. Toshihiko Shinozaki<br>Assistant Manager, Technical Affairs Department<br>TEL: 0289(62)6011<br>FAX: 0289(62)6665<br>t-shinozaki@nsk-nakanishi.co.jp                                                                                                                                                                                                                                                                                                                                                                                                    |
| Date Prepared:                                  | July 7, 2011 (Revised June 18, 2012)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Trade Name:                                     | Ti-Max X450                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Common Name:                                    | High Speed Air Turbine Handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification Name:                            | EFB 872.4200 Handpiece, Air-Powered, Dental                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Predicate Device:                               | K952927 - Palisades Dental, Impact Air 45 Handpiece<br>K101717 - Aerobine, Inc., Air Powered Dental Handpiece [Karam45, Dexor45]                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Description:                             | The Ti-Max X450 air-powered high-speed dental handpiece capable of reaching<br>rotational speeds of 380,000 to 450,000 revolutions per minute. The device includes<br>non-optic, fiber-optic, and LED models. All models have a 45 degree back angle.<br>Models are available to connect directly to the couplings of specific brands.                                                                                                                                                                                                                     |
| Statement of<br>Intended Use:                   | The Ti-Max X450 is an air-powered dental handpiece with intended use of being a<br>surgical tool for impacted third molar removal and periodontal procedures for<br>which a conventional handpiece would be used.                                                                                                                                                                                                                                                                                                                                          |
|                                                 | The Ti-Max X450 is intended for use with a friction grip bur that conforms to ISO<br>1797-1 standard. Recommended supply air pressure is between 0.22 and 0.30 MPa,<br>which results in high-speed bur rotation (approximately 380,000 to 450,000 RPM).                                                                                                                                                                                                                                                                                                    |
| Summary of<br>Technological<br>Characteristics: | The Ti-Max X450 is capable of achieving high-speed rotation of 380,000 to<br>450,000 revolutions per minute when provided with a supply air pressure of 0.22 to<br>0.30 MPa (32 - 44 psi). This provides a torque of at least 0.09Ncm, while<br>generating a noise level of 80dBA or less. The Ti-Max X450 has a 45 degree back<br>angle. Also available are models with fiber optic or LED lighting. All models<br>include a water jet that is directed to the bur point.                                                                                 |
|                                                 | The Ti-Max X450 has a one-touch quick connect coupling system. The Ti-Max<br>X450 can connect directly to the dental unit, or a swivel adapter can be used. Direct<br>connection can be made with 4- and 5-hole connections that meet ISO 9168<br>specifications. A swivel adapter is connected between the handpiece and hose in the<br>case of 4-, 3-, and 2-hole connections, also in compliance with ISO 9168, type 2. In<br>addition, the Ti-Max X450 has models available that are designed to connect<br>directly to the couplings of other brands. |

NAKANISHI INC.
www.nsk-inc.com

ﺎ.

.

Section 5: Page 1 of 2 Headquarters : 700 Shimohihata Kanuma-shi Tochigi 322-8666, Japan Tokyo Office : MY Building 1-13-3 Ueno Taito-ku Tokyo 110-0005, Japan

·

Image /page/0/Picture/5 description: The image shows a signature with the words "Financial Partner of" written below it. The signature is illegible, but it appears to be written in black ink. The words "Financial Partner of" are written in a smaller font size than the signature. The image is a close-up of the signature and the words below it.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

11:33:33

KIJDO24

{1}------------------------------------------------

The Ti-Max X450 was developed and is produced under consideration of all Performance applicable technical standards, internal specifications, and FDA guidance Testing: documents. The product's conformance with applicable international and internal standards was verified in the course of bench testing. Nakanishi considers the Ti-Max X450 to be substantially equivalent to the predicate Conclusion:

devices listed above. This conclusion is based on the similarities in intended use, principles of operation, functional design, and established medical use, and demonstrates that the Ti-Max X450 is substantially equivalent to the declared predicate devices.

Image /page/1/Picture/2 description: The image shows a black and white drawing of a bird. The bird is in flight, with its wings spread and its head turned to the side. The bird's feathers are detailed, and its beak is slightly open. The background is plain, with some text at the bottom of the image.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of an eagle, with its wings spread and its head turned to the right.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

# JUN 2 0 2012

Nakanishi, Incorporated C/O Ms. Diane Rutherford Regulatory Engineer Ken Block Consulting 1201 Richardson Drive, Suite 280 Richardson, Texas 75080

Re: K112024

Trade/Device Name: Ti-Max X450 Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: June 4, 2012 Received: June 12, 2012

Dear Ms. Rutherford:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device _ Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2- Rutherford

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### INDICATIONS FOR USE

510(k) Number:

Device Name: Ti-Max X450

Indications for Use:

. .

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

.

The Ti-Max X450 is an air-powered dental handpiece with intended use of being a surgical tool for impacted third molar removal and periodontal procedures for which a conventional handpiece would be used.

The. Ti-Max X450 is intended for use with a friction grip bur that conforms to ISO 1797-1 standard. Recommended supply air pressure is between 0.22 and 0.30 MPa, which results in high-speed bur rotation (approximately 380,000 to 450,000 RPM).

Prescription Use X (21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDHR, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

K112024

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K112024](https://fda-staging.innolitics.com/device/K112024)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).