Published clinical literature; Clinical data from patients with bladder outlet obstruction
The sponsor utilized published clinical data and existing clinical experience with 1470nm lasers in patients with bladder outlet obstruction to support the substantial equivalence of the 200W power setting for prostate vaporization.
Ablation rates; Bladder outlet obstruction; Prostate vaporization; Literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published clinical data on 980nm vs 1470nm laser ablation; Literature review / Comparative analysis
—
Not applicable for this study
Ablation rates on a per Watt basis
Clinical use of 1470nm lasers in patients with bladder outlet obstruction; Retrospective clinical observation
Patients with bladder outlet obstruction
Not applicable for this study
Safety and effectiveness of 1470nm laser for prostate treatment
Indications for Use
The device is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures including via endoscopes. The Evolve HPD Multiwavelength 980/ 1470 Diode Laser (up to 150W) is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, dental applications, and endovenous occlusion of the saphenous veins in patients with superficial vein reflux. The Multiwavelength laser is further indicated for laser assisted lipolysis. Power from 151W to 200W is indicated in the vaporization of the prostate to treat Benign Prostatic Hyperplasia (BPH).
Device Story
Multiwavelength diode laser (980/1470nm) delivers laser light to soft tissue; contact or non-contact modes; endoscopic or open surgical use. Device increases power output to 200W for BPH vaporization. Operated by physicians in clinical/surgical settings. Laser energy interacts with tissue to achieve incision, excision, vaporization, ablation, hemostasis, or coagulation. Output controlled by surgeon via console; visual/audible indicators confirm operation. Benefits include precise tissue management, reduced bleeding, and minimally invasive access for various surgical procedures.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing and published literature. Bench testing verified compliance with 21 CFR 1040.10/1040.11, ANSI/AAMI ES1, IEC 601-2-22, EN 60825-1, and ISO 10993-7. Literature-based evidence quantified ablation rates per Watt for 980nm vs 1470nm lasers and referenced studies using 200W 980nm lasers on human cadaver prostates and 50W 1470nm lasers on patients with bladder outlet obstruction.
Technological Characteristics
Multiwavelength diode laser (980/1470nm); power output up to 200W. Complies with 21 CFR 1040.10/1040.11, ANSI/AAMI ES1, IEC 601-2-22, EN 60825-1, and ISO 10993-7. Operates in contact/non-contact modes for soft tissue surgery. Connectivity and software architecture details not specified.
Indications for Use
Indicated for incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue across various surgical specialties including ENT, arthroscopy, GI, general/plastic/podiatric surgery, urology, gynecology, neurosurgery, cardiac, pulmonary, and dental applications. Specific indications include endovenous occlusion of saphenous veins for superficial vein reflux, laser-assisted lipolysis, and vaporization of the prostate for BPH (151W-200W).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
JAN 1 3 2012
#### 510(k) Summary Evolve® HPD 980/ 1470nm Multiwavelength Diode Laser
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Biolitec, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028 Phone: (413) 525-0600 Facsimile: (413) 525-0611
Contact Person: Harry Hayes, Ph.D. - Regulatory Consultant Date prepared: July 5, 2011
#### Name of Device and Name/Address of Sponsor
Evolve® HPD 980/ 1470nm Multiwavelength Diode Laser; Evolve® HPD Dual Biolitec, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028
#### Classification Name
Surgical laser (21 CFR 878.4810, Product Code GEX)
#### Predicate Devices
Ceralas 150W 980/ 1470nm Multiwavelength Diode Laser, (K090164) Ceralas 180W 980nm Diode Laser, (K083682) Ceralas 980nm Diode Laser Family, (K100726 & K112324)
#### Intended Use/Indication for Use
The device is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures including via endoscopes. The Evolve HPD Multiwavelength 980/ 1470 Diode Laser (up to 150W) is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, dental applications, and endovenous occlusion of the saphenous veins in patients with superficial vein reflux. The Multiwavelength laser is further indicated for laser assisted lipolysis. Power from 151W to 200W is indicated in the vaporization of the prostate to treat Benign Prostatic Hyperplasia (BPH).
The device is specifically indicated for use as follows:
{1}------------------------------------------------
# K112013
Ear, Nose and Throat and Oral Surgery (Otolaryngology) Hemostasis, incision, excision, ablation, coagulation, and vaporization of tissue from the ear, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include: Removal of benign lesions from the ear, nose and throat Excision and vaporization of vocal cord nodules and polyps Incision and excision of carcinoma in situ Ablation and vaporization of hyperkeratosis Excision of carcinoma of the larynx Laryngeal papillomectomy Excision and vaporization of herpes simplex I and II Neck dissection
Arthroscopy
Hemostasis, incision, excision, coagulation, vaporization and ablation of joint tissues during arthroscopic surgery. Examples include: Menisectomy Synovectomy Chondromalacia
Gastroenterology
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts and also with endoscopic procedures. Examples include:
Hemostasis of upper and lower GI bleeding
Excision and vaporization of colorectal carcinoma
Excision of polyps
General Surgery, Dermatology, Plastic Surgery and Podiatry Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue and dermabrasion. Examples include: Matrixectomy Excision of neuromas Excision of periungual and subungual warts Excision of plantar warts Excision of keloids Liver resection
{2}------------------------------------------------
K 112013
pg 3 of 5
Excision of cutaneous lesions
Hemorrhoidectomy
Appendectomy
Debridement of decubitus ulcers
Hepatobiliary tumors
Mastectomy
Dermabrasion
Laser Assisted Lipolysis
Vaporization and hemostasis of capillary hemangioma
Excision, vaporization and hemostasis of abdominal tumors
Excision, vaporization and hemostasis of rectal pathology
Pilonidal cystectomy
Herniorapphy
Adhesiolysis
Parathyroidectomy
Laparoscopic cholecystectomy
Thyroidectomy
Resection of organs
Debridement of wounds
Photocoagulation of teleangectasia of the legs and face
Photocoagulation of vascular lesions of the face and extremities
Endovenous Occlusion of the Saphenous Veins in Patients with Superficial Vein
Reflux Associated with Varicose Veins and Varicosities
Treatment of reticular veins and branch varicosities
Urology
Excision, vaporization, incision, coagulation, ablation and hemostasis of urological tissues.
Examples include:
Vaporization of urethral tumors
Release of urethral stricture
Removal of bladder neck obstruction
Excision and vaporization of condyloma
Lesions of external genitalia
Vaporization of the prostate to treat benign prostatic (BPH). Note: powers from 151W to 200W should only be applied in the vaporization of the prostate to treat Benign Prostatic Hyperplasia (BPH).
Gynecology
Ablation, excision, incision, coagulation, hemostasis and vaporization of gynecological tissue.
Examples include:
Endometrial ablation
Excision or vaporization of condylomata acurninate
Vaporization of cervical intraepithelial neoplasia
{3}------------------------------------------------
K112013 874 af's
#### Cervical conization Menorrhagia
Neurosurgery Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue. Examples include: hemostasis in conjunction with menigiomas
Cardiac Surgery
Hemostasis and coagulation of soft tissue, including cardiac tissue.
Pulmonary Surgery
Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include: Tracheobronchial malignancy or stricture
Benign and malignant pulmonary obstruction
Endoscopic pulmonary applications
Dental Applications
Indicated for the following applications on intraoral and extraoral soft tissue (including marginal and interdental gingival and epithelial lining of free gingival); frenectomy, frenotomy, biopsy, operculectomy, implant recovery, gingivectomy, gingivoplasty, gingival troughing, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery,
debridement of diseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions, papillectomy, vestibuloplasy, excision of lesions, exposure of unerupted partially erupted teeth, leukoplakia, removal of hyperplastic tissues, treatment of aphthous ulcers and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket), pulpotomy, pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
Powers from 151W to 200W Powers from 151W to 200W should only be applied in the vaporization of the prostate to treat Benign Prostatic Hyperplasia (BPH).
{4}------------------------------------------------
K112013
pg 5 of 5
#### Technological Characteristics
The Evolve HPD 980/ 1470nm Multiwavelength Diode Laser contains the same basic components as the cleared Ceralas 150W 980/ 1470mm Diode Laser (K090164). The purpose of this submission is to increase the energy from 150W to 200W for application in the vaporization of the prostate to treat Benign Prostatic Hyperplasia. (BPH)
#### Performance Data
The device complies with the following voluntary consensus standards: 21 C.F.R. §§ 1040.10 & 1040.11; ANSI/AAMI ES1; IEC 601-2-22; EN 60825-1, and ANSI/AAMI/ISO 10993-7. Published clinical data supplied quantifies ablation rates on a per Watt basis for a 980nm laser versus a 1470nm laser, which together with studies using 200W 980nm lasers on human cadaver prostates and 50W 1470nm lasers on prostates of patients with bladder outlet obstruction along with its predicate devices support the conclusion that the device is safe and effective and substantially equivalent to its predicate devices.
#### Substantial Equivalence
The Evolve HPD 980/ 1470mm Multiwavelength Diode Laser is as safe and effective for these Indication for Use as the cleared Ceralas 150W 980/ 1470mm Diode Laser. The Evolve HPD 980/ 1470mm Multiwavelength Diode Laser has the same intended uses, indications, technological characteristics, and principles of operation as its predicate device. Thus, the Evolve HPD 980/ 1470nm Multiwavelength Diode Laser is substantially equivalent to its predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 1 3 2012
Biolitec, Incorporated % Genmarhay BDA Harry Hayes, Ph.D. 515 Shaker Road East Longmeadow, Massachusetts 01028
Re: K112013
Trade/Device Name: Evolve HPD 980/1470nm Multiwavelength Diode Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: January 5, 2012 Received: January 10, 2010
Dear Dr. Hayes:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assessment to May 28, 1976, the enactment date of the Medical Device Amendments, or to conniner of prior of ring 2011 - 12:00 accordance with the provisions of the Federal Food, Drug, de rices that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, therefore, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adultier. Thease noter Obras set, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elaborition (000 aron). Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r load be action of a determination that your device complies with other requirements of the Act that I Drivias intatutes and regulations administered by other Federal agencies. You must or any I oderar statute and securements, including, but not limited to: registration and listing (21 compy with an all love (21 CFR Part 801); medical device reporting (reporting of medical CI K Far 607), adoming (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Harry Hayes, Ph.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
1cgallanon (21 CFA Part 1007) process Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/defaulf.htm.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
pq 1 of 4
Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
# Device Name: Evolve® HPD 980/ 1470nm Multiwavelength Diode Laser
#### Indications for Use:
The device is intended for delivery of laser light to soft tissue in the contact and non The active is included a recedures including via endoscopes. The Evolve HPD contact article attring of 1470 Diode Laser (up to 150W) is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear. nose and throat and oral surgery (otolaryngology), arthroscopy, m car, nooo and the surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, dental applications, and endovenous occlusion of the saphenous veins in patients with surger); action approacted Multiwavelength laser is further indicated for laser assisted lipolysis. Bapernelar 151W to 200W is indicated in the vaporization of the prostate to treat Benign Prostatic Hyperplasia (BPH).
The device is specifically indicated for use as follows:
## Ear, Nose and Throat and Oral Surgery (Otolaryngology)
Har, I woo una rizion, ablation, ablation, coagulation, and vaporization of tissue from the ear, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include:
Removal of benign lesions from the ear, nose and throat
Excision and vaporization of vocal cord nodules and polyps
Incision and excision of carcinoma in situ
Ablation and vaporization of hyperkeratosis
Excision of carcinoma of the larynx
Laryngeal papillomectomy
Excision and vaporization of herpes simplex I and II
Neck dissection
#### Arthroscopy
Hemostasis, incision, excision, coagulation, vaporization and ablation of joint tissues during arthroscopic surgery. Examples include
Menisectomy
Synovectomy
Chondromalacia
Neil R. Ogden for mkn
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112013
{8}------------------------------------------------
K.11 2013
1 2 of 4
#### Gastroenterology
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts and also with endoscopic procedures. Examples include: Hemostasis of upper and lower GI bleeding
Excision and vaporization of colorectal carcinoma
Excision of polyps
# General Surgery, Dermatology, Plastic Surgery and Podiatry
Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue and dermabrasion. Examples include: Matrixectomy Excision of neuromas Excision of periungual and subungual warts Excision of plantar warts Excision of keloids Liver resection Excision of cutaneous lesions Hemorrhoidectomy Appendectomy Debridement of decubitus ulcers Hepatobiliary tumors Mastectomy Dermabrasion Laser Assisted Lipolysis Vaporization and hemostasis of capillary hemangioma Excision, vaporization and hemostasis of abdominal tumors Excision, vaporization and hemostasis of rectal pathology Pilonidal cystectomy (Division Sign-C Herniorapphy Division of Surgical, Orthopedic, Adhesiolysis and Restorative Devices Parathyroidectomy Laparoscopic cholecystectomy 112013 Thyroidectomy 510(k) Number. Resection of organs Debridement of wounds Photocoagulation of teleangectasia of the legs and face Photocoagulation of vascular lesions of the face and extremities I novenous Occlusion of the Saphenous Veins in Patients with Superficial Vein Reflux Associated with Varicose Veins and Varicosities Treatment of reticular veins and branch varicosities
{9}------------------------------------------------
1/2013 pg 3 of 4
#### Urology
Excision, vaporization, incision, coagulation, ablation and hemostasis of urological tissues.
Examples include: Vaporization of urethral tumors Release of urethral stricture Removal of bladder neck obstruction Excision and vaporization of condyloma Lesions of external genitalia Vaporization of the prostate to treat benign prostatic hyperplasia (BPH). Note: powers from Vaporization of the procuse to applied in the vaporization of the prostate to treat Benign Prostatic Hyperplasia (BPH).
#### Gynecology
Ablation, excision, incision, coagulation, hemostasis and vaporization of gynecological tissue. Examples include: Endometrial ablation Excision or vaporization of condylomata acurninate Vaporization of cervical intraepithelial neoplasia Cervical conization Menorrhagia
### Neurosurgery
Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue. Examples include:
hemostasis in conjunction with menigiomas
#### Cardiac Surgery
Hemostasis and coagulation of soft tissue, including cardiac tissue.
#### Pulmonary Surgery
Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue nitred. in the pulmonary system.
Examples include:
Tracheobronchial malignancy or stricture Benign and malignant pulmonary obstruction Endoscopic pulmonary applications
cision and ablation of soft tissue
Nilk for mam
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
**Dental Applications**
510(k) Number K112013
Indicated for the following applications on intraoral and extraoral soft tissue thatcated or the tonewal gingival and epithelial lining of free gingival). frenectomy, frenotomy, biopsy, operculectomy, implant recovery, gingivectomy,
{10}------------------------------------------------
K 11 2013 Pg 4 of 4
gingivoplasty, gingival troughing, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery,
debridement of diseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions, papillectomy, vestibuloplasy, excision of lesions, exposure of unerupted/ partially erupted teeth, leukoplakia, removal of hyperplastic tissues, treatment of aphthous ulcers and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket), pulpotomy, pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
#### Powers from 151W to 200W
x owers from 151 W to 200W should only be applied in the vaporization of the prostate to treat Benign Prostatic Hyperplasia (BPH).
(PLEASE DO NOT WRITE BELOW THIS LINE … CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _ V (Per 21 C.F.R. 801.109) OR
Over-The-Counter Use_ (Optional Format 1-2-96)
Neil RPOschelen for mxc
Division of Surnical, Orthopedic, and Restorative Devices
510(k) Number K112013.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.