K112008 · Touch Medical Solutions, Inc. · LLZ · Jan 6, 2012 · Radiology
Device Facts
Record ID
K112008
Device Name
TOUCHPACS SUITE
Applicant
Touch Medical Solutions, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 6, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The TouchPACS Suite software system is a picture archiving and communications system (PACS) intended to be used as a networked Digital Imaging and Communications in Medicine (DICOM) and non- DICOM information and data management system. The TouchPACS Suite PACS & DICOM Viewer Software is comprised of modular software programs that run on standard "off-the-shelf" personal computers, business computers, and servers running standard operating systems. TouchPACS TouchMGR Suite DICOM Viewer Software system is an image, data storage and display software that accepts DICOM data from any OEM modality which support DICOM standard imaging data. The system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulation, create graphical representations of anatomical areas, and perform quantitative measurements. The TouchPACS Suite software is not for mammographic use.
Device Story
TouchPACS Suite is a software-based PACS for diagnostic imaging workstations; runs on off-the-shelf Windows PCs/servers. Inputs: DICOM/non-DICOM imaging data from OEM modalities. Processing: Uses MyDICOM .NET C# libraries for DICOM v3 compliance; relational database management via Microsoft SQL Server 2008; DirectX for image rendering. Outputs: Organized image studies, graphical anatomical representations, and quantitative measurements displayed on multi-monitor workstations. Used by radiologists and clinicians in clinical environments. Workflow: DICOM SCP server receives/queues studies; TouchMGR module manages security/users/workflow; TouchPACS client provides diagnostic viewing. Benefits: Streamlines office efficiency; provides diagnostic environment designed for long-duration use; enables centralized storage and disaster recovery. Healthcare providers use the system to review, manipulate, and measure radiological images to support clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence established through bench testing, verification, and validation of software design specifications, hazard analysis, and compliance with DICOM v3.0 and hardware safety standards (IEC60950, IEC/CISPR 22).
Technological Characteristics
Software-only PACS; runs on standard Windows-based PCs/servers. Uses Microsoft SQL Server 2008 RDBMS. Connectivity: TCP/IP, DICOM v3.0 compliant via MyDICOM .NET libraries. Interface: WPF-based, multi-monitor support, DirectX-accelerated. Software architecture: Modular, C# language. Cybersecurity: DSN-less database connections, firewall port 1433, user-based security/access controls.
Indications for Use
Indicated for use as a networked DICOM and non-DICOM information and data management system for the storage, organization, display, digital manipulation, and quantitative measurement of medical images generated by OEM modality equipment. Not for mammographic use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# 5. 510(k) Summary
In accordance with the requirements of the Safe Medical Device Act, Touch Medical Solutions Inc. herewith submits a 510(K) summary for the following device.
KI12008
age lof H
| Submitter Name/Address: | Touch Medical Solutions, Inc.<br>3455 University Parkway<br>Winston-Salem NC 27106-4109 | |
|--------------------------------|-----------------------------------------------------------------------------------------|---------|
| Contact Person/Tel. #: | James Owen<br>(336) 413-6967 | |
| Date Summary Prepared: | December 2, 2010 | |
| Establishment No: | | |
| Trade/Proprietary Name: | TouchPACS Suite | |
| Common/Usual Name: | Medical Imaging Workstation | |
| Classification Name: | Picture Archiving and Communications System | |
| Class/Panel: | Class II, 90-LLZ, 21CFR 892.2050 | |
| Predicate Devices: | | |
| Manufacturer | Device | 510(k) |
| STAIR Systems Inc. | Stair Systems PACS & DICOM<br>Viewer | K100236 |
| Ramsoft Inc. | Ramsoft PACS | K031562 |
| Advance Imaging Solutions, LLC | EZPACS | K062878 |
| Fuji Medical System USA | Fuji Synapse Workstation<br>Software | K051553 |
| INTENDED USE STATEMENT | | |
The TouchPACS Suite software system is a picture archiving and communications system (PACS) intended to be used as a networked Digital Imaging and Communications in Medicine (DICOM) and non- DICOM information and data management system. The TouchPACS Suite PACS & DICOM Viewer Software is comprised of modular software programs that run on standard "off-the-shelf" personal computers, business computers, and servers running standard operating systems. TouchPACS TouchMGR Suite DICOM Viewer Software system is an image, data storage and display software that accepts DICOM data from any OEM modality which support DICOM standard imaging data. The system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulation, create graphical representations of anatomical areas, and perform quantitative measurements.
The TouchPACS Suite software is not for mammographic use.
{1}------------------------------------------------
KII 2008 2x 4 Page
## DEVICE DESCRIPTION
The TMS TouchPACS Suite is a collection of software applications coded in the Microsoft C# application language, using the Windows Presentation Foundation (WPF) libraries for user interface, that run on "off-the-shelf" Microsoft Windows-based personal or business computers to provide diagnostic imaging workstation capabilities. The system is designed to provide an environment for radiological diagnosticians which is "easy on the eyes"; so that the user can use the system for 10-12 hours per day without undue strain.
#### Note: the TouchPACS Suite imaging software uses the MyDICOM .NET C# Source Code Libraries, from myDICOM LLC, compiled directly into the imaging applications. to provide full DICOM v3 compliance.
The layout and design of the 3-screen user interface evolved from the legacy concept of 3 separate computers, each with a single monitor displaying a single element of the case under review, into a modern, multi-monitor native design which incorporates large performance gains through the use of Microsoft's DirectX imaging technologies.
The core of the system is the database, which is a relational design, using Microsoft SQL Server 2008 Enterprise edition as the Relational Database Management System (RDBMS). The database server provides a dedicated, central place to provide disaster recovery servicing (via standard SQL Server replication), as well as the large multiterabyte storage required for a PACS system.
Attached to the RDBMS are TouchPACS client workstations, TouchMGR client workstations, and servers required for system operations: TMS DICOM SCP Server and Microsoft SQL Server. All of the aforementioned software modules, except for Microsoft SQL Server, are TMS software products.
The TouchPACS Suite PACS client workstation is intended to be a desktop computer replacement product, with the interface dominating the screen space on a workstation computer. This design decision was made to accommodate non- technical doctors who, we found, typically prefer to have a simplified and unobtrusive environment to work in. Color-based exam status listings were evolutionary, growing from the legacy 'field of green' display listing into the more advanced dynamic, multicolored tree view seen today in the client. TouchPACS is a multi-monitor capable client, currently configurable in a 1, 2, or 3 monitor footprint.
The TouchMGR client workstation represents the nerve center of the TouchPACS Suite, providing services for securing the TMS network, maintaining users, allowing paperless workflow, and other such critical day-today system maintenance tasks. TouchMGR users will include hospital or practice administrators, security administrators, and functional personnel who may need to make modifications to data contained in the TMS database. TMS uses DSN-less connections to the database, usually requiring port 1433 to be set as an exception in the workstation's firewall rule set. The client performs many
{2}------------------------------------------------
KII2008 Page 30
tasks, some of which are not relevant to every customer, though we encourage adoption of TMS electronic processes by our clients in order to help them streamline their office efficiency. Users only see the functions of the Touch MGR modules which they are entitled to view or use, based on security settings captured at login time. The The TMS DICOM SCP Server provides multi-threaded DICOM v3 server capabilities of the TouchPACS system. It was built to be run unattended, but a user interface, available from either the desktop or the TouchMGR client, may be used to both help monitor and control the various DICOM storage processes necessary to import each radiological study from an external source (modality, another PACS, etc.). The server incorporates a DICOM SCP service grouping that allows for DICOM ping response, client protocol negotiation, transfer syntax negotiation, and several other system-level DICOM negotiations. As each case is received, it is stored via a Service host to the local hard drive, then it is queued for storage, and then is stored as data to the main database in one of 8 available simultaneous, threaded "put away" processes.
| Product Name | TouchPACS | STAIR<br>Systems<br>Constellation<br>PACS | RamSoft<br>PACS | Fuji<br>Synapse<br>Workstation<br>Software | EZPACS |
|--------------------------------------------|-----------------------------------|-------------------------------------------|-----------------|--------------------------------------------|---------|
| 510(k) Number | | K100236 | K031562 | K051553 | K062878 |
| Communications | TCP/IP | TCP/IP | TCP/IP | TCP/IP | TCP/IP |
| Image Archive | Yes | Yes | Yes | Yes | Yes |
| Image<br>Processing | Yes | Yes | Yes | Yes | Yes |
| Image Edit | Yes | Yes | Yes | Yes | Yes |
| Edit Patient<br>Demographics | Yes | Yes | Yes | Yes | Yes |
| Add and remove<br>images | Yes | Yes | Yes | Yes | Yes |
| Combine studies | Yes | Yes | Yes | Yes | Yes |
| Edit Patient<br>Orientation<br>information | Yes | Yes | Yes | Yes | Yes |
| Set and Edit<br>Routing<br>Information | Yes | Yes | Yes | Yes | Yes |
| JPEG<br>Lossy/Lossless<br>Compression | Display = Yes<br>Apply = LOSSLESS | Yes | Yes | Yes | Yes |
| JPEG 2000<br>Lossy/Lossless<br>Compression | Display = Yes<br>Apply = LOSSLESS | Yes | Yes | Yes | Yes |
| Image | Yes | Yes | Yes | Yes | Yes |
#### PREDICATE DEVICE COMPARISON
{3}------------------------------------------------
| Processing<br>(image scaling) | | | | | |
|----------------------------------------------------------------------|-----|-----|-----|-----|-----|
| Image<br>Processing<br>(Window Level,<br>Pan, Zoom, Cine<br>Display) | Yes | Yes | Yes | Yes | Yes |
| DICOM Print | Yes | Yes | Yes | Yes | Yes |
The TMS TouchPACS Suite has similar intended use and technical features of the predicate devices listed above. The TouchPACS Suite is not for mammographic use.
### SAFETY INFORMATION
The TouchPACS Workstation software introduces no new safety and efficacy issues other than those already identified with the predicate devices. The results of a hazard analysis, combined with the appropriate preventive measures taken indicate that the device is of minor level of concern as per the May 29, 1998 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The Instructions for Use contains precautions and warnings for safe and effective use of the device.
Potential hazards are identified through risk analysis and managed through the software development process and verification/validation testing.
The TouchPACS Workstation Software complies with the following mandatory and voluntary standards:
- ACR/NEMA DICOM Version 3.0 (Digital Imaging and Communications in o Medicine) - developed by the American College of Radiology and the National Electrical Manufacturers Association. The Synapse DICOM Conformance Statement has been included in the Labeling section as Appendix F3.
- Note: the TouchPACS Suite imaging software uses the MyDICOM .NET C# o Source Code Libraries, from myDICOM LLC, compiled directly into the imaging applications, to provide full DICOM v3 compliance.
Recommended Specifications for Workstation Hardware include compliance with the following standards:
- General Safety standards for all hardware: IEC60950 or equivalent. o
- Information Technology Equipment, Radio Disturbance (Emissions) . Characteristics-Limits and Methods of Measurements, IEC/CISPR 22 (EN55022)
## CONCLUSIONS
..
A comparison of the labeling, substantial equivalence table, and verification and validation testing has established that the device meets its intended use and design specifications.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion, surrounding the emblem.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
JAN - 6 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. James Owen Director of Development Touch Medical Solutions Inc. 3455 University Parkway WINSTON-SALEM NC 27106-4109
Re: K112008
Trade/Device Name: TouchPACS Suite Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 6, 2011 Received: December 14, 2011
Dear Mr. Owen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary S Patil
Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
KII2008
#### Indications for Use 510(k) Number (if known): K (12008 Device Name: TouchPACS Suite
Indications for Use:
The TouchPACS Suite software system is a picture archiving and communications system (PACS) intended to be used as a networked Digital Imaging and Communications in Medicine (DICOM) and non- DICOM information and data management system. The TouchPACS Suite PACS & DICOM Viewer Software is comprised of modular software programs that run on standard "off-the-shell" personal computers, business computers, and servers running standard operating systems. TouchPACS TouchMGR Suite DICOM Viewer Software system is an image, data storage and display software that accepts DICOM data from any OEM modality which support DICOM standard imaging data. The system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulation, create graphical representations of anatomical areas, and perform quantitative measurements.
The TouchPACS Suite software is not for mammographic use.
Prescription Use X (Part 21CFRS80ISubpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Mary S Postel
OK K 11.2008
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.