K111992 · Ivera Medical, Inc. · QBP · Jan 12, 2012 · General Hospital
Device Facts
Record ID
K111992
Device Name
CUROS PORT PROTECTOR
Applicant
Ivera Medical, Inc.
Product Code
QBP · General Hospital
Decision Date
Jan 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Curos is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. Curos ™ will disinfect the valve three (3) minutes after application and act as a physical barrier to contamination for up to seven (7) days (168 hours) if not removed. The effectiveness of Curos Protectors were tested in vitro against Staphylococcus aureus. Staphylococcus epidermidis, Escherichia coli and Pseudomonas aeruginosa, Candida glibrata, Candida albicans and was found to have >4 log reduction. The Curos Port Protector may be used in the home or healthcare facility.
Device Story
Curos Port Protector is a single-use, sterile device containing 70% isopropyl alcohol; designed to disinfect needleless luer-activated valves and provide a physical barrier against contamination. Device is applied to the valve; disinfection occurs within 3 minutes; barrier remains effective for up to 7 days if left in place. Used in home or healthcare settings by clinicians or patients. Output is a disinfected valve surface; facilitates safe line access; reduces risk of catheter-related bloodstream infections. Device is a passive mechanical cap; no electronic components or software.
Clinical Evidence
Bench testing only. In vitro efficacy testing performed against Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli, Pseudomonas aeruginosa, Candida albicans, and Candida glabrata. Results demonstrated >4 log reduction at 3 minutes and 168 hours (7 days), meeting pre-defined acceptance criteria.
Technological Characteristics
Single-use, sterile, passive cap containing 70% isopropyl alcohol. Materials meet ISO 10993 biocompatibility standards. Sterilized via Gamma radiation per ISO 11137-1, ISO 11137-2, and ISO 11137-3. Dimensions: 0.40 inches length, 0.54 inches diameter. Translucent green molded plastic housing.
Indications for Use
Indicated for use on swab-able luer access valves as a disinfecting cleaner and physical barrier to contamination. Suitable for use in home or healthcare facility settings.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized eagle. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
Ivera Medical, Inc. Don Canal Regulatory Affairs, Quality Manager 3525 Del Mar Heights Road, Suite #430 San Diego, California 92130
March 11, 2022
Re: K111992
Trade/Device Name: Curos Port Protector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: QBP
Dear Don Canal:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 12, 2012 and the correction letter dated December 14, 2018. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation 880.5440.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Payal Patel, OHT3: Office of GastroRenal, Ob-Gyn, General Hospital and Urology Devices, 240-402-6029, Payal.Patel(@fda.hhs.gov.
Sincerely,
Pror R
Payal Patel Assistant Director for General Hospital Devices DHT3C: Division of Drug Delivery and General Hospital Devices and Human Factors OHT3: Office of GastroRenal, Ob-Gyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol is a circular emblem with a stylized depiction of a caduceus, a symbol associated with medicine and healthcare. The text is arranged in three lines, with "FDA" in a bold, sans-serif font, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in a smaller, sans-serif font.
December 14, 2018
Ivera Medical, Inc. Don Canal RA/QA 2731 Loker Avenue West Carlsbad, California 92010
Re: K111992
Trade/Device Name: Curos Port Protector Regulatory Class: Unclassified Product Code: OBP Dated: July 13, 2011 Received: July 13, 2011
#### Dear Don Canal:
This letter corrects our substantially equivalent letter of January 12, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{2}------------------------------------------------
### K111992 - Don Canal
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FDA Tina Kiang -
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
#### Indications for Use Statement 4.
The Curos is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. Curos ™ will disinfect the valve three (3) minutes after application and act as a physical barrier to contamination for up to seven (7) days (168 hours) if not removed. The effectiveness of Curos Protectors were tested in vitro against Staphylococcus aureus. Staphylococcus epidermidis, Escherichia coli and Pseudomonas aeruginosa, Candida glibrata, Candida albicans and was found to have >4 log reduction. The Curos Port Protector may be used in the home or healthcare facility.
Prescription Use
AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS UNE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Khel C. Chapa 1/9/12
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k111992
Page 14 of 139
Title: <
Ivera Medical Curos 510(k) Notification
Confidential
{4}------------------------------------------------
JAN 1 2 2012
### 510(k) Summary
### General Company Information
| Name: | Ivera Medical Corporation |
|------------|---------------------------------------------------------------------------------------------|
| Contact: | Don Canal<br>Consultant RAQA |
| Address: | Ivera Medical Corporation<br>3525 Del Mar Heights Road<br>Suite #430<br>San Diego, CA 92130 |
| Telephone: | 760-612-6090 |
| Fax: | 858-228-1770 |
#### Date Prepared: October 23, 2011
### General Device Description
The Curos™ Port Protector device is a single use, sterile device that contains 70% lsopropyl Alcohol and is intended to be used as a disinfectant for needleless luer activated valves
| Common Name: | Pad, Alcohol |
|-----------------|---------------------------------------|
| Trade Name: | CurosTM Port Protector |
| Classification: | Unclassified Device, product Code LKE |
#### Predicate Devices
K080466 Curos Port Protector, Ivera Medical Corporation K083508 SwabCap, Excelsior Medical Corporation
#### Intended Use (Indications)
The Curos is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. Curos ™ will disinfect the valve three (3) minutes after application and act as a physical barrier to contamination for up to seven (7) days (168 hours) if not removed. The effectiveness of Curos Protectors were tested in vitro against Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli and Pseudomonas aeruginosa, Candida glibrata, Candida albicans and was found to have >4 log reduction. The Curos Port Protector may be used in the home or healthcare facility.
- -
{5}------------------------------------------------
# Comparison with Predicate Device
There is no change to the device for this 510(k) notification, the only change is to the Intended Use (indications) to reflect the test data for disinfection which includes 3 minutes to 7 days (168 hours), as described below. The materials of construction and technological characteristics are equivalent to the predicate device.
| Characteristic | Subject Device | Predicate Device | Predicate Device |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Curos Port Protector | Curos Port Protector | SwabCap |
| Common Name | Alcohol, disinfecting pad | Alcohol, disinfecting pad | Alcohol, disinfecting pad |
| Manufacturer | Ivera Medical | Ivera Medical | Excelsior Medical Corporation |
| 510(k) number | K110826 | K110826 | K083508 |
| Regulation<br>number,<br>product code | Unclassified,<br>Preamendment device,<br>product code: LKB | Unclassified,<br>Preamendment device,<br>product code: LKB | Unclassified,<br>Preamendment device, product code:<br>LKB |
| Indications for<br>use | The Curos is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. Curos ™ will disinfect the valve three (3) minutes after application and act as a physical barrier to contamination for up to seven (7) days (168 hours) if not removed. The effectiveness of Curos Protectors were tested in vitro against Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli and Pseudomonas aeruginosa, Candida glibrata. Candida albicans | The Curos Port Protector is a device containing 70% Isopropyl Alcohol when left in place for 5 to15 minutes, the Curos Port Protector decontaminates the injection Port; thereafter the Curos Port Protector provides a physical barrier during the intended use. | SwabCap is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between line accesses. SwabCap will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed. |
| Characteristic | Subject Device | Predicate Device | Predicate Device |
| | and was found to have >4<br>log reduction. The Curos<br>Port Protector may be used<br>in the home or healthcare<br>facility. | | |
| Disinfectant -<br>active<br>ingredient | 70% Isopropyl Alcohol | 70% Isopropyl Alcohol | 70% Isopropyl<br>Alcohol |
| Length | 0.40 inches | 0.40 inches | 0.50 inches |
| Diameter | 0.54 inches | 0.54 inches | 0.60 inches |
| User Population | Home and hospital use | Home and hospital use | Home and hospital<br>use |
| Colorants Used<br>(type, amount,<br>concentration) | Translucent green, molded<br>plastic, 3% concentration | Translucent green,<br>molded plastic, 3%<br>concentration | Orange pigment, %<br>concentration<br>unknown |
| Provided Sterile | Yes | Yes | Yes |
| Single Use<br>Device | Yes | Yes | Yes |
| Plastic Housing<br>to remain in<br>place | Yes | Yes | Yes |
## Subject Device to Predicate Technological Comparison Table
{6}------------------------------------------------
# Substantial Equivalence Performance Testing
Ivera medical has provided non-clinical performance test data that demonstrates the pre-defined acceptance criteria for a disinfecting device has been met. This acceptance criteria is defined as a bacteria count reduction of > 4 log reduction of 2 selected gram positive bacteria, 2 selected gram negative bacteria, and two selected fungus/yeast micro-organisms for a period of time from 3 minutes up to 168 hours (7 days). The efficacy testing was completed using a total of 4 bacteria, 2-gram negative and 2 gram positive as recommended in Draft Guidance for Industry and FDA Staff Premarket Notification [510(k)] Submissions for Medical Devices that Include Antimicrobial Agents DRAFT GUIDANCE. This guidance document is being distributed for comment purposes only. Document issued on: July 19, 2007. The test results are summarized in Table 1.
{7}------------------------------------------------
| Organism | Acceptance<br>Criteria<br>(bacterial count<br>reduction ( $\Delta$ Log)) | 3 minute exposure<br>(bacterial count<br>reduction ( $\Delta$ Log)) | 7 day (168 hours)<br>exposure<br>(bacterial count<br>reduction ( $\Delta$ Log)) |
|--------------------------|--------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Staphylococcus aureus | $\geq$ 4.0 | 6.0 | 6.9 |
| Staphylococcus epidermis | $\geq$ 4.0 | 6.8 | 7.3 |
| Escherichia coli | $\geq$ 4.0 | 5.2 | 5.2 |
| Pseudomonas aeruginosa | $\geq$ 4.0 | 5.1 | 5.1 |
| Candida Albicans | $\geq$ 4.0 | 5.6 | >4.8* |
| Candida Glabrata | $\geq$ 4.0 | 5.4 | >5.3* |
### Table 1 - Efficacy Test Results
* The entire population was killed, the reduction in bacteria count was limited to the population quantified by the positive controls used in the testing.
The Ivera Curos Port Protector is sterilized using a validated Gamma sterilization process which complies with ISO11137-1:2006/(R) Sterilization of health care products -Radiation - Part 2: Establishing the sterilization dose. Recognition number 14-225.
ISO11137-2:2006 Sterilization of health care products – Radiation – Part 1: requirements for development of validation and routine control of sterilization process for medical devices. Recognition number 14-297.
11137-3:2006/(R) 2010 10/04/2010 AAMI ANSI ISO 14-298 - Radiation - Part 3: Guidance on Dosimetric Aspects. Recognition number 14-298.FDA recognized standard ISO11137 Sterilization Standard.
Ivera Medical has completed testing to demonstrate the Curos Port Protector materials of construction meet FDA recognized standard ISO10993 for biocompatibility.
### Conclusion
The analysis arguments and test results demonstrate the Curos™ device is safe for its intended use and is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.