← Product Code [LCM](/productcode/LCM) · K111928

# PSYCHEMEDICS PHENCYCLIDINE EIA (K111928)

_Psychemedics Corp. · LCM · May 1, 2012 · CH · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K111928

## Device Facts

- **Applicant:** Psychemedics Corp.
- **Product Code:** [LCM](/productcode/LCM.md)
- **Decision Date:** May 1, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** CH
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K111928 · May 1, 2012 | PSYCHEMEDICS PHENCYCLIDINE EIA | Psychemedics Corp. | Clinical hair samples (head and body hair) from routine drug-of-abuse testing | Clinical hair samples were used to evaluate the performance of the Psychemedics Microplate EIA for Phencyclidine by comparing results against GC/MS confirmation and a predicate device. | Clinical performance; Drug-of-abuse testing; Hair samples; GC/MS confirmation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective clinical performance evaluation | Human head and body hair samples; Sample Size: Approximately 225 samples (140 negative, 24 below cutoff, 15 between cutoff and +100%, 46 >100% of cutoff) plus 224 additional negative samples | GC/MS and Predicate Device | Qualitative detection of phencyclidine (PCP) in hair |

## Indications for Use

The Psychemedics Microplate EIA for Phencyclidine is an enzyme immunoassay (EIA) for the preliminary qualitative detection of phencyclidine in human head and body hair samples using a phencyclidine calibrator at 3 ng /10 mg hair cutoff for the purpose of identifying phencyclidine use. This is an in vitro diagnostic device intended exclusively for Psychemedics use only and is not intended for sale to anyone. The Psychemedics EIA Phencyclidine Assay provides only a preliminary analytical test result. To obtain a quantitative analytical result or to confirm positive results, a more specific alternate chemical method (e.g. GC/MS) must be used. Clinical consideration and professional judgment should be applied to the interpretation of any drug-of-abuse test result.

## Device Story

Device performs preliminary qualitative screening for phencyclidine (PCP) in human hair. Process involves two stages: pre-analytical hair digestion to convert solid hair matrix to liquid, followed by competitive enzyme immunoassay (EIA). Input: hair sample extract. Mechanism: drug in sample competes with phencyclidine-BSA conjugate for binding sites on antibody-coated microplate wells; secondary antibody-HRP and TMB substrate added; absorbance measured at 450/630 nm via automated microplate reader. Output: absorbance value inversely proportional to drug concentration. Used in laboratory settings by trained personnel. Results normalized as B/B0 x 100. Samples exceeding cutoff are presumptively positive; require GC/MS confirmation. Benefits: identifies potential drug use; cosmetic treatments (bleach, dye, perm, relaxer, shampoo) do not cause false negatives in positive samples.

## Clinical Evidence

No clinical studies performed. Bench testing included precision (intra-assay and inter-assay) and method comparison against GC/MS. Comparison study of 217 donor hair samples (156 negative, 61 positive by GC/MS) showed high concordance. Two discordant results were observed near the cutoff (2.80 ng/10 mg and 1.97 ng/10 mg). Cosmetic treatment studies (bleach, dye, perm, relaxer, shampoo) confirmed no false negatives in positive samples.

## Technological Characteristics

Enzyme immunoassay (EIA) using microplate reader. Components: microplate wells coated with PCP-BSA conjugate, polyclonal rabbit anti-PCP, goat anti-rabbit HRP-conjugated secondary antibody, TMB substrate, HCl, and wash buffer. Extraction: nonproteolytic digestion. Cutoff: 3 ng/10 mg hair. Connectivity: standalone laboratory equipment.

## Predicate Devices

- Psychemedics RIA Phencyclidine Assay (k011275)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k111928

B. Purpose for Submission:
New submission

C. Measurand:
Phencyclidine (PCP) in Hair

D. Type of Test:
Qualitative Enzyme Immunoassay (EIA)

E. Applicant:
Psychemedics Corp.

F. Proprietary and Established Names:
Psychemedics Microplate EIA for Phencyclidine in Hair

G. Regulatory Information:
|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LCM | Unclassified | Enzyme Immunoassay
Phencyclidine | Toxicology (91)  |

H. Intended Use:

1. Intended use(s):
See indication for use below

2. Indication(s) for use:
The Psychemedics Microplate EIA for Phencyclidine is an enzyme immunoassay (EIA) for the preliminary qualitative detection of phencyclidine in human head and body hair samples using a 3 ng/10 mg hair cutoff for the purpose of identifying phencyclidine use. This is an in vitro diagnostic device intended exclusively for Psychemedics use only and is not intended for sale to anyone.

{1}

The Psychemedics Phencyclidine Enzyme Immunoassay provides only a preliminary analytical result. To confirm presumptive positive results, a more specific alternate chemical method (e.g., GC/MS) must be used. Clinical consideration and professional judgment should be applied to the interpretation of any drug abuse test results.

3. Special conditions for use statement(s):

This assay is for over the counter use.

4. Special instrument requirements:

The device is for use with an automated microplate reader capable of measuring at 450 and 630 nm. Plate washing also requires an instrument specifically designed to effectively and reproducibly wash all wells uniformly.

For confirmation testing, the sponsor uses GC/MS.

I. Device Description:

The test consists of two parts; a pre-analytical hair treatment procedure (to convert the solid matrix of hair to a measurable liquid matrix) and the screening assay, the Psychemedics Microplate EIA for Phencyclidine.

The drug is recovered from the hair using a patented method. The screening portion of the test system consists of (1) microplate wells coated with multiple drugs including phencyclidine conjugated to bovine serum albumin (BSA), (2) polyclonal rabbit anti-phencyclidine, (3) goat anti-rabbit secondary antibody conjugated to HRP (horseradish peroxidase), (4) substrate [3, 3', 5, 5' tetramethylbenzidine (TMB)], (5) HCl to acidify the final reaction, and (6) wash buffer for washing the plates (7) controls and calibrators.

Absorbance in the wells is read with a microplate reader.

Standard and control stock solutions are purchased from multiple vendors, prepared in the laboratory, and validated by GC/MS confirmation.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Psychemedics RIA Phencyclidine Assay

{2}

2. Predicate 510(k) number(s):

k011275

3. Comparison with predicate:

|  Similarities and Differences  |   |   |
| --- | --- | --- |
|  Items | Psychemedics Phencyclidine EIA k111928 (Candidate Device) | Psychemedics RIA Phencyclidine Assay k011275 (Predicate Device)  |
|  Intended Use | Intended to be used for the qualitative detection of PCP in hair | Same  |
|  Cutoff | 3 ng PCP/10mg hair | Same  |
|  Confirmation method | GC-MS | Same  |
|  Test Principle | Competitive EIA | Radioimmunoassay  |
|  Method of measurement | Microplate reader | Gamma counter  |
|  Extraction method | Non-enzymatic digestion | Enzymatic digestion  |
|  Antigens used in the reagent | BSA-benzoylecgonine, BSA-morphine, BSA-phencyclidine, and BSA-methamphetamine | \( ^{125} \)I-readiolabeled PCP  |

# K. Standard/Guidance Document Referenced (if applicable):

Premarket Submissions and Labeling Recommendations for Drugs of Abuse Screening Tests - Draft Guidance for Industry and FDA Staff, issued on December 2, 2003

# L. Test Principle:

PCP is first extracted from the hair sample with a patented extraction method. In the screening assay, the hair extracts and the primary antibody are combined in the wells and incubated. During this phase the enzyme-labeled drug conjugate competes with drug in the sample for a limited number of binding sites on the antibody-coated microwells. Subsequently, secondary antibody-HRP and the substrate are added to the reaction sequentially, the excess reagent are removed by washing solution, and after a final incubation, the wells are acidified and read with the microplate reader. Results are normalized by expression as  $\mathrm{B / B_0}\times 100$ . If phencyclidine is present in the sample, less primary antibody will be bound to the solid-phase antigen, thereby resulting in less binding of HRP-labeled secondary antibody; the absorbance produced is inversely proportional to the amount of phencyclidine in the sample (specimen, calibrator or control).

{3}

Samples with an absorbance value higher than the Cutoff Calibrator (3 ng PCP/10mg hair) are interpreted as negative. Either the sample does not contain PCP or PCP is present at a concentration below the cutoff level of the assay. Samples with an absorbance value equal to or lower than the Cutoff Calibrator are presumptively positive.

For samples that are presumptive positive by the screening assay, a new aliquot of the hair sample is weighed, washed extensively to remove and evaluate externally-derived methamphetamine contamination on the hair, digested by a different procedure, and confirmed by GC/MS.

## M. Performance Characteristics:

### 1. Analytical performance

#### a. Precision and Reproducibility

Hair samples known not to contain phencyclidine or related compounds were spiked in multiples of 15 with phencyclidine at zero, 0.75, 1.5, 2.25, 3, 3.75, 4.5, 5.25, and 6 ng per 10 mg hair and assayed by the EIA procedure. The phencyclidine spike solutions were prepared in methanol from a certified standard solution and confirmed by GC/MS. Intra-assay precision was performed in one run, and inter-assay precision was performed over 5 days. The results are presented in the tables below.

Table 1. Intra-assay Precision

|  PCP Concentration (ng PCP/10 mg hair) | Number Tested | Number Negative | Number Positive  |
| --- | --- | --- | --- |
|  Zero | 15 | 15 | 0  |
|  0.75 (-75%) | 15 | 15 | 0  |
|  1.5 (-50%) | 15 | 15 | 0  |
|  2.25 (-25%) | 15 | 15 | 0  |
|  3.75 (+25%) | 15 | 0 | 15  |
|  4.5 (+50%) | 15 | 0 | 15  |
|  5.25 (+75%) | 15 | 0 | 15  |
|  6.0 (+100%) | 15 | 0 | 15  |

Table 2. Inter-assay Precision

|  PCP Concentration (ng PCP/10 mg hair) | Number Tested | Number Negative | Number Positive  |
| --- | --- | --- | --- |
|  Zero | 75 | 75 | 0  |
|  0.75 (-75%) | 75 | 75 | 0  |
|  1.5 (-50%) | 75 | 75 | 0  |
|  2.25 (-25%) | 75 | 75 | 0  |
|  3.75 (+25%) | 75 | 0 | 75  |
|  4.5 (+50%) | 75 | 0 | 75  |
|  5.25 (+75%) | 75 | 0 | 75  |
|  6.0 (+100%) | 75 | 0 | 75  |

{4}

# b. Linearity

Not applicable. This assay is intended only for qualitative screening determinations.

# c. Traceability and Expected values for controls, calibrators, or methods

Psychemedics manufactures calibrators and control materials using drug stocks purchased the drugs from a commercial vendor. Each lot of drug is received with its specific certificate of analysis. The commercially obtained stock is made into the calibrators and controls to the desired concentrations. The concentrations are confirmed by GC/MS.

Stability studies for both controls and calibrators have been conducted. Protocols and acceptance criteria were described and found to be acceptable. The manufacturer claims the following expiration date for both controls and calibrators:

The calibrator and control solutions are stable for 1 year when stored frozen  $(&lt; -10^{\circ}\mathrm{C})$ .

# d. Detection Limit

Not required since this is a qualitative test.

# e. Analytical specificity:

Cross-reactivity was evaluated by spiking each compound to the negative hair samples in the extraction solution and evaluating the result against the cutoff control.

Results are expressed as a concentration of compound required to produce a response approximately equivalent to the cutoff concentration of the assay. The results are presented in the table below:

Table 3. Cross-reactivity of related Compounds in Phencyclidine EIA

|  Compound | Amount of Compound required to Produce a positive test at the cutoff of 3 ng phencyclidine/10 mg hair | Percent Cross-reactivity  |
| --- | --- | --- |
|  1-(4-Hydroxypiperidino) phenylcyclohexane | 1000 | 0.3  |
|  1-(1-Phenylcyclohexyl) morpholine (PCM) | 60 | 5.0  |
|  Nicotine | >1000 | < 0.5  |
|  Cotinine | >1000 | < 0.5  |
|  Metaphit | 10 | 30  |
|  Venlafaxine | >1000 | < 0.5  |
|  O-Desmethylvenlafaxine | >1000 | <0.5  |
|  Thioridazine | >1000 | < 0.5  |

{5}

Dextromethorphan 5000 &lt; 0.1

## Structually unrelated:

Negative hair samples were spiked with phencyclidine to -50%, and +50% of the cutoff. Structurally related and unrelated compounds were added to methanol to a concentration of 100 ng/10 mg hair then added to the negative hair sample. The following compounds do not cause interference at +/- 50% of the cutoff; oxazepam, glutethimide, amitryptyline, trimipramine, doxepin, imipramine, nordoxepin, nortriptyline, desipramine, protryptyline, barbital, phenobarbital, amobarbital, butabarbital, hexobarbital, secobarbital, medazepam, lorazepam, diazepam, flurazepam, medazepam, nordiazepam, temazepam, bromazepam, ethosuximide, methsuximide, a-methyl-a-propylsuccimide, metharbital, phensuximide, normethsuximide, mephenytoin, ethotoin, mephobarbital, PEMA, Methyl-PEMA, 10,11-dihydrocarbamazepine, carbamazepine, primidone, 5,5-diphenylhydantoin, 4-methylprimidone, acetaminophen, caffeine, dyphylline, methaqualone, theophylline, phenmetrazine, phenylpropanolamine, amitriptyline, dextromethorphan, lidocaine, methocarbamol, nordoxepin, pentazocine, phenylephrine, triamterene, meprobamate, methylprylon, Anhydroecgonine methyl ester, Atropine, Bupropion, Cotinine, Cannabinol, Chlorpheniramine maleate, O-Desmethyvenlafaxine, Desipramine, Doxylamine succinate, 1S, 2R Ephedrine, Ethosuximide, Ibuprofen, LSD, Haloperidol, Meperidine, Methadone, Methaqualone, Methyl phenidate, Naloxone, Naltrexone, Nicotine, Naproxen, Nortriptyline, Propoxyphene, R,R Pseudoephedrine, Thioridazine, Cis-Tramadol, Venlafaxine hydrochloride, 8(-)-11-nor-9-Carboxy-delta-9 THC, 11-nor-9-Carboxy-delta-9-THC, Delta 8-THC, Streptomycin, Procaine, Benzocaine, Erythromycin, Penicillin G, Mepivacaine, Phendimetrazine bitartrate, Diazepam, Despropionyl fentanyl, Ethylmorphine, Nalorphine, Codeine, Morphine, Hydromorphone, Oxycodone, Glutethimide, Meprobamate, Methyprylon, Flurazepam, Lorazepam, Medazepam, Temazepam, Carbamazepine, Diazepam, Nordiazepam, Oxazepam, Amitriptyline, Dextromethorphan, Lidocaine, Methocarbamol, Nordoxepin, Pentazocine, Phenylephrine, Triamterene, Ethosuximide, a-methyl-a-propylsuccimide, metharbital, barbital, methsuximide, phensuximide, N-Normethsuximide, Mephenytoin, Ethotoin, Mephobarbital, PEMA, Phenobarbital, Methyl PEMA, 10,11-Dihydrocarbamazepine, Primidone, Carbamazepine, 5,5-Diphenylhydantoin, 6,4-Methylprimidone, Butabarbital, Amobarbital, Secobarbital, Hexobarbital, Phenobarbital, Medazepam, Oxazepam, Lorazepam, Diazepam, Temazepam, Bromazepam, Amitriptyline, Desipramine, Doxepin, Imipramine, Nordoxepin, Nortriptyline, Protriptyline, Trimipramine, Chlorpromazine, Flurazepam, Amoxillin, Propanolol, Promethazine, Phenmetrazinbe, Phendimetrazine, Benzocaine, Ecgonine, Metanephrin.

## Cosmetic treatment of hair:

Twenty PCP-negative hair samples were treated with bleach, 20 with permanent wave, 20 with dye, 20 with relaxer, and 20 with shampoo, and the results

6

{6}

compared to the same samples without the treatments. In each case of the 20 samples treated with a type of cosmetic treatment, 10 samples were treated with one brand of a particular product and 10 other samples with a second brand. No significant differences were observed for the negative hair samples before and after the treatments; all samples remained negative after the treatments.

Twelve PCP-positive hair samples were treated with bleach, 12 with permanent wave, 12 with dye, 12 with relaxer, and 12 with shampoo, and the results compared to the same samples without the treatments. In each case of the 12 samples treated with a type of cosmetic treatment, 6 samples were treated with one brand of a particular product and 6 other samples with a second brand. The positive samples showed changes in the B/Bo x 100 (% of B₀) values between -5.2% to +5.4% for bleach, -5.4% to +3.8% for dye, -2.7% to +5.6% for perm, -7.6% to +9.1% for relaxer, and ---7.0% to +6.3% for shampoo. None of the originally positive samples tested negative by the EIA phencyclidine assay after any of the cosmetic treatments.

## Environmental Contamination of hair:

Preliminary positive hair sample results by the screening method could be due to environmental contamination. All positive should be sent for confirmation testing on a reference method to distinguish between true positive and those samples that were positive due to external exposure.

## f. Assay cut-off

Analytical performance of the device around the claimed cutoff is described in precision section M.1 above. The identified cutoff concentration of the device is consistent with other cleared DOA assays.

## 2. Comparison studies:

### a. Method comparison with predicate device:

The study was performed by comparing ELSIA results against the GC/MS results on the same head or body hair samples. A total of 217 donor hair samples were tested (156 negative and 61 positive by GC/MS). The results are shown in the tables below:

Table. Comparison of Positive Samples and Samples around the cutoff by GC/MS

|  GC/MS: | 0 | > 0 & <-50% of Cutoff | > -50% of Cutoff & < Cutoff | > Cutoff, & < +50% of Cutoff | > +50% of Cutoff  |
| --- | --- | --- | --- | --- | --- |
|  EIA Positive | 0 | 0 | 2 | 8 | 53  |
|  EIA Negative | 140 | 1 | 13 | 0 | 0  |

Discordant results: There are two discrepant results around the cutoff, the results was shown below:

{7}

|  Cutoff Value (ng/10 mg hair) | Candidate Device (+/-) | Phencyclidine GC/MS value (ng/10 mg hair)  |
| --- | --- | --- |
|  3 | pos | 2.80  |
|  3 | pos | 1.97  |

b. Matrix comparison:

Not applicable

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

No amount of PCP should be found in hair from those who are abstained from PCP use.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

At this stage, the submitted information in this premarket notification is complete to supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K111928](https://fda-staging.innolitics.com/device/K111928)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com