PSYCHEMEDICS OPIATES EIA

K111926 · Psychemedics Corp. · DJG · May 1, 2012 · Clinical Toxicology

Device Facts

Record IDK111926
Device NamePSYCHEMEDICS OPIATES EIA
ApplicantPsychemedics Corp.
Product CodeDJG · Clinical Toxicology
Decision DateMay 1, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3650
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K111926 · May 1, 2012PSYCHEMEDICS OPIATES EIAPsychemedics Corp.Clinical hair samples (head and body hair)Clinical hair samples were used to evaluate the performance of the assay, including agreement testing against a predicate device and confirmation of results via LC/MS/MS.Clinical hair samples; Agreement testing; LC/MS/MS confirmation

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective comparison of clinical hair samplesIndividuals providing head and body hair samples; Sample Size: 473 hair samples (383 head hair, 90 body hair)Psychemedics RIA Assay (Predicate)Qualitative detection of opiates; agreement with predicate device and LC/MS/MS confirmation

Indications for Use

The Psychemedics Microplate EIA for Opiates is an enzyme immunoassay (EIA) for the preliminary qualitative detection of opiates in human head and body hair using a morphine calibrator at 2 ng /10 mg hair cutoff for the purpose of identifying opiate use. This is an in vitro diagnostic device intended exclusively for Psychemedics use only and is not intended for sale to anyone. The Psychemedics Microplate EIA opiate assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Device Story

Device performs qualitative screening for opiates in human hair; utilizes enzyme immunoassay (EIA) technology. Process involves pre-analytical hair treatment to convert solid matrix to liquid; subsequent incubation with primary antibody in microplate wells coated with morphine-BSA conjugate; addition of HRP-labeled secondary antibody; substrate reaction; acidification. Absorbance measured via microplate reader at 450/650 nm; results normalized as B/B0 x 100. Inverse relationship between absorbance and opiate concentration. Used in laboratory settings by trained personnel. Output provides preliminary screening result; requires confirmation by GC/MS or LC/MS/MS for clinical decision-making. Benefits include rapid screening of hair samples for drug use history.

Clinical Evidence

No clinical studies performed. Evidence consists of analytical performance data including precision (intra-assay and inter-assay) and method comparison. Method comparison study tested 243 donor hair samples (97 negative, 146 positive) against LC/MS/MS reference method. Results showed high concordance. Cross-reactivity and interference studies were conducted with various structurally related and unrelated compounds. Cosmetic treatment studies (bleaching, dyeing, etc.) confirmed no false negatives in positive samples.

Technological Characteristics

Enzyme immunoassay (EIA) system. Components: microplate wells coated with morphine-BSA, polyclonal sheep anti-morphine, rabbit anti-goat secondary antibody-HRP, TMB substrate, HCl acidifier. Requires microplate reader (450/650 nm) and plate washer. Qualitative test; results normalized as B/B0 x 100. Not automated; manual/semi-automated processing.

Indications for Use

Indicated for preliminary qualitative detection of opiates in human head and body hair for individuals suspected of opiate use. Requires confirmatory testing via GC/MS or LC/MS/MS. For professional use only.

Regulatory Classification

Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k111926 B. Purpose for Submission: New device C. Measurand: Opiates D. Type of Test: Qualitative enzyme immunoassay (EIA) E. Applicant: Psychemedics Corp. F. Proprietary and Established Names: Psychemedics Microplate EIA for Opiates in Hair G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | DJG | Class II | 862.3650 – Opiate test system | 91-Toxicology | H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: The Psychemedics Microplate EIA for Opiates is an enzyme immunoassay (EIA) for the preliminary qualitative detection of opiates in human head and body hair using a morphine calibrator at 2 ng /10 mg hair cutoff for the purpose of {1} identifying opiate use. This is an in vitro diagnostic device intended exclusively for Psychemedics use only and is not intended for sale to anyone. The Psychemedics Microplate EIA for Opiates assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS or LC/MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. 3. Special conditions for use statement(s): This assay is for over the counter use. 4. Special instrument requirements: The device is for use with a microplate reader capable of measuring at 450 and 650 nm. Plate washing also requires an instrument specifically designed to effectively and reproducibly wash all wells uniformly. I. Device Description: The test consists of two parts; a pre-analytical hair treatment procedure (to convert the solid matrix of hair to a measurable liquid matrix) and the screening assay, the Psychemedics Microplate EIA for Opiates. The drug is recovered from the hair using a patented method. The screening portion of the test system consists of (1) microplate wells coated with multiple drugs including morphine conjugated to bovine serum albumin (BSA), polyclonal sheep anti-morphine, rabbit anti-goat secondary antibody conjugated to HRP (horseradish peroxidase), substrate [3, 3', 5, 5' tetramethylbenzidine (TMB)], HCl to acidify the final reaction, and wash buffer for washing the plates. Absorbance in the wells is read with a microplate reader. J. Substantial Equivalence Information: 1. Predicate device name(s): Psychemedics Morphine RIA Assay 2. Predicate K number(s): k000851 {2} 3 3. Comparison with predicate: | Item | Device | Predicate | | --- | --- | --- | | Indications/Intended Use | The Psychemedics Microplate EIA for Opiates is an enzyme immunoassay (EIA) for the preliminary qualitative detection of opiates in human head and body hair using a morphine calibrator at 2 ng /10 mg hair cutoff for the purpose of identifying opiate use. | Same | | Sample matrix | Hair | Same | | Method of measurement | Microplate reader | Gamma counter | | Cutoff | 2 ng morphine/10 mg hair | Same | | Test Principle | EIA | RIA | K. Standard/Guidance Document Referenced (if applicable): None were referenced. L. Test Principle: Hair sample extracts and primary antibody are combined in the wells and incubated. After washing, secondary antibody-HRP is added and incubated. After washing, substrate is added, and, after a final incubation, the wells are acidified and read with the microplate reader. Results are normalized by expression as B/B0 x 100. If morphine or related opiates are present in the sample, less primary antibody will be bound to the solid-phase antigen, thereby resulting in less binding of HRP-labeled secondary antibody; the absorbance produced is inversely proportional to the amount of opiates in the sample (specimen, calibrator or control). M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision studies were performed by taking commercially available materials consisting of morphine in methanol, to prepare spiking solutions at the following concentrations; negative, $\pm 75\%$, $\pm 50\%$, $\pm 25\%$ and $100\%$ of the cutoff. The concentration of each sample was confirmed by LC/MS/MS. The morphine solutions were then used to spike 15 replicates of negative hair samples. Intra-assay precision was performed in one run and inter-assay precision was performed over 5 non-consecutive days. The results are presented in the tables below: {3} 4 Intra-assay | Opiate (ng/10mg hair) | Percent of Cut-off | Replicate Number | Pos/Neg | | --- | --- | --- | --- | | 0 | -100% | 15 | 0/15 | | 0.5 | -75% | 15 | 0/15 | | 1.0 | -50% | 15 | 0/15 | | 1.5 | -25% | 15 | 0/15 | | 2.5 | +25% | 15 | 15/0 | | 3.0 | +50% | 15 | 15/0 | | 3.5 | +75% | 15 | 15/0 | | 4.0 | +100% | 15 | 15/0 | Inter-assay | Opiate (ng/10 mg hair) | Percent of Cut-off | Replicate Number | Pos/Neg | | --- | --- | --- | --- | | 0 | -100% | 75 | 0/75 | | 0.5 | -75% | 75 | 0/75 | | 1.0 | -50% | 75 | 0/75 | | 1.5 | -25% | 75 | 0/75 | | 2.5 | +25% | 75 | 75/0 | | 3.0 | +50% | 75 | 75/0 | | 3.5 | +75% | 75 | 75/0 | | 4.0 | +100% | 75 | 75/0 | b. Linearity/assay reportable range: Not applicable. This is a qualitative assay. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Psychemedics manufactures calibrators and control materials using drug stocks purchased from a commercial vendor. Each lot of drug is received with its specific certificate of analysis. The commercially obtained stock is made into the calibrators and controls to the desired concentrations. The concentrations are confirmed by GC/MS. {4} Stability studies for both controls and calibrators have been conducted. Protocols and acceptance criteria were described and found to be acceptable. The manufacturer claims the following expiration date for both controls and calibrators: When stored at less than or equal to 10 °C product is stable for 12 months. d. Detection limit: Not required since this is a qualitative test e. Analytical specificity: Cross-reactivity was established by spiking various concentrations of each substance into drug-free sample and evaluating the result against the cutoff control. Results are expressed as the concentration of compound required to produce a response approximately equivalent to the cutoff concentration of the assay. The results are presented in the table below: | Compound | Approximate concentration of compound (pg/mg) Equivalent to 2 ng/10 mg hair Opiate Cutoff | % Cross reactivity | | --- | --- | --- | | Codeine | 1.8 | 111 | | Hydromorphone | 38 | 5.2 | | Hydrocodone | 4.8 | 41.7 | | Acetylcodeine | 3.5 | 57.1 | | 6-Acetylmorphine | 4.6 | 43.5 | | Oxycodone | 46 | 4.3 | | Oxymorphone | 250 | <1 | | Methadone | >5000 | <1 | | Morphine Glucuronide | 13.3 | 15 | | Naloxone | >1000 | < 0.2% | | Naltrexone | >1000 | < 0.2% | | Propoxyphene | >1000 | < 0.2% | | Meperidine | >1000 | < 0.2% | Structurally unrelated: Negative hair samples were spiked with morphine to -50%, and +50% of the cutoff. Structurally related and unrelated compounds were added to methanol to a concentration of 100 ng/10 mg hair then added to the negative hair {5} sample. The following compounds do not cause interference at +/- 50% of the cutoff; Anhydroecgonine methyl ester, Atropine, Bupropion, Cotinine, Cannabinol, Chlorpheniramine maleate, O-Desmethyvenlafaxine, Desipramine, Doxylamine succinate, 1S, 2R Ephedrine, Ethosuximide, Ibuprofen, LSD, Haloperidol, Meperidine, Methadone, Methaqualone, Methyl phenidate, Naloxone, Naltrexone, Nicotine, Naproxen, Nortriptyline, Propoxyphene, R,R Pseudoephedrine, Thioridazine, Cis-Tramadol, Venlafaxine hydrochloride, 8(-)-11-nor-9-Carboxy-delta-9 THC, 11-nor-9-Carboxy-delta-9-THC, Delta 8-THC, Streptomycin, Procaine, Benzocaine, Erythromycin, Penicillin G, Mepivacaine, Phendimetrazine bitartrate, Diazepam, Despropionyl fentanyl, Ethylmorphine, Nalorphine, Codeine, Morphine, Hydromorphone, Oxycodone, Glutethimide, Meprobamate, Methyprylon, Flurazepam, Lorazepam, Medazepam, Temazepam, Carbamazepine, Diazepam, Nordiazepam, Oxazepam, Acetaminophen, Caffeine, Dyphylline, Methaqualone, Theophylline, Amitriptyline, Dextromethorphan, Lidocaine, Methocarbamol, Nordoxepin, Pentazocine, Phenylephrine, Triamterene, Ethosuximide, a-methyl-a-propylsuccimide, metharbital, barbital, methsuximide, phensuximide, phensuximide, N-Normethsuximide, Mephenytoin, Ethotoin, Mephobarbital, PEMA, Phenobarbital, Methyl PEMA, 10,11-Dihydrocarbamazepine, Primidone, Carbamazepine, 5,5-Diphenylhydantoin, 4-Methylprimidone, Butabarbital, Amobarbital, Secobarbital, Hexobarbital, Phenobarbital, Medazepam, Oxazepam, Lorazepam, Diazepam, Temazepam, Bromazepam, Amitriptyline, Desipramine, Doxepin, Imipramine, Nordoxepin, Nortiptyline, Protriptyline, Trimipramine, Glutethimide, Chlorpromazine, Flurazepam, Amoxillin, Propanolol, Promethazine, Phenmetrazinbe, Phendimetrazine, Benzocaine, Ecgonine, Metanephrin, amphetamine, caffeine, Imipramine, Methamphetamine, Phencyclidine, Phenmetrazine, Phenylpropanolamine, Cocaine, Cocaethylene, amitriptyline, Dextromethorphan, Lidocaine, Methocarbamol, Nordoxepin, Pentazocine, Phenylephrine, Triamterene, Pheniramine, Mephenytoin, Orphenadrine, Methylphenidate, Ibuprofen. There is the possibility that other substances and/or factors not listed above may interfere with the test and cause false results. ## Cosmetic Treatment Tests were performed to determine the effects of various hair treatments (i.e. bleaching, dyeing, relaxer, shampoo, permanent) on samples tested using the Psychemedics Microplate EIA for Opiates. The ethnic origin, hair color and curvature were documented. ## Effects on Positive Samples: 60 specimens determined to be positive for opiate were used in the study. The study was conducted with two different hair treatments. 6 {6} Six different hair samples were used for each hair treatment. ELSIA absorbance readings before and after treatment were compared. Average changes in absorbance values after were -2.25% for bleach, -4.65% for dye, -4.5% for perm, -3.15% for relaxer and 1.1% for shampoo, where a negative sign indicates a sample becoming “more negative” due to treatment and a positive sign indicates a sample becoming “more positive.” None of the positive samples became negative after treatment. Effect on Negative Samples: One hundred specimens previously determined to be negative were used in the study. The study was conducted with two different hair treatments. Ten different hair samples were used for each hair treatment. ELSIA absorbance readings before and after treatment were compared. All samples determined to be negative prior to treatment remained negative post treatment Environmental Study: Preliminary positive hair sample results by the screening method could be due to environmental contamination. All positive should be sent for confirmation testing on a reference method to distinguish between true positive and those samples that were positive due to external exposure. f. Assay cut-off: Analytical performance of the device around the claimed cutoff is described in precision section M.1a above 2. Comparison studies: a. Method comparison with predicate device: The study was performed by comparing ELSIA results against the LC/MS/MS results on the same head or body hair samples. A total of 243 donor hair samples were tested (97 negative and 146 positive). The results are presented in the table below: {7} | Optiate Test Result | Negative by GC/MS | Less than half the cutoff concentration by GC/MS | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (Greater than 50% above the cutoff concentration) | | --- | --- | --- | --- | --- | --- | | Positive | 0 | 2 | 8 | 10 | 136 | | Negative | 86 | 11 | 2 | 0 | 0 | | Screening Cutoff (ng/10 mg hair) | ELSIA Opiate Test Results (POS/NEG) | LC/MS/MS Drug Result (pg/10 mg hair) | | --- | --- | --- | | 2 | POS | 0.9 oxycodone | | 2 | POS | 0.9 hydrocodone | | 2 | POS | 1.2 hydrocodone | | 2 | POS | 1.4 hydrocodone | | 2 | POS | 1.4 hydrocodone | | 2 | POS | 1.5 hydrocodone | | 2 | POS | 1.5 6-MAM | | 2 | POS | 1.5 hydrocodone | | 2 | POS | 1.7 6-MAM and oxycodone | b. Matrix comparison: Not applicable. The assay is intended for only one sample matrix. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: {8} Not applicable 5. Expected values/Reference range: Not applicable **N. Proposed Labeling:** They have not provided adequate labeling. The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 9
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...