K111917 · Corventis, Inc. · QYX · Aug 24, 2011 · Cardiovascular
Device Facts
Record ID
K111917
Device Name
NUVANT, MOBILE CARDIAC TELEMETRY
Applicant
Corventis, Inc.
Product Code
QYX · Cardiovascular
Decision Date
Aug 24, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The NUVANT Mobile Cardiac Telemetry (MCT) System is intended to continuously measure, record, and periodically transmit physiological data. The System is indicated for those patients who require monitoring for the detection of nonlethal cardiac arrhythmias such as, but not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias and conduction disorders. The NUVANT system model monitors, derives and displays: - ECG - Heart rate
Device Story
Wearable, wireless mobile cardiac telemetry system; monitors ECG and heart rate for up to 7.5 days. Components: PiiX (adherent torso-worn device with ECG electrodes); Patient Trigger Magnet (manual symptom-triggered recording); zLink (hand-held gateway); Corventis Server (remote processing). PiiX transmits data to zLink via Bluetooth; zLink transmits to Server via cellular technology. Corventis cardiographic technicians (learned intermediaries) review data on Server; prepare/deliver information to prescribing physicians. Physicians review data online to identify suspected arrhythmias. Enables clinical decision-making for cardiac rhythm management.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, AAMI/ANSI EC38, and AAMI/ANSI EC57 standards provided to demonstrate safety and effectiveness.
Indicated for patients requiring monitoring for detection of non-lethal cardiac arrhythmias, including supraventricular tachycardias (atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias, and conduction disorders.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
CardioNet ECG Monitor with Arrhythmia Detection Model CN1005 (K072558)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
November 15, 2023
Corventis, Inc. Michele Chin-Purcell Senior Director 1410 Energy Park Drive, Suite 1 St. Paul, Minnesota 55108
Re: K111917
Trade/Device Name: Nuvant, Mobile Cardiac Telemetry Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: QYX, MHX
Dear Michele Chin-Purcell:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 24, 2011. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, jennifer.kozen@fda.hhs.gov.
Sincerely.
Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and the pursuit of well-being. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Corventis, Inc. c/o Michele Chin-Purcell, Ph.D. Senior Director, Quality and Regulatory Affairs 1410 Energy Park Drive, Suite 1 St. Paul, MN 55108
AUG 2 4 2011
ﻟﻠﺒﻨﺎﻧﻴﺔ
ﻷﺷﺘ
Re: K111917
Trade/Device Name: Nuvant Mobile Cardiac Telemetry Regulatory Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms)
Regulatory Class: II (two) Regulatory Class: II (two) Product Code: 74 MHX Dated: July 5, 2011 Received: July 6, 2011
Dear Dr. Chin-Purcell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications)
for use stated in the enclosure) to legally marked prodicated in the indicat for use stated in the enclosure) to legally marketed predicate devices marketed in interstated
for use stated in the enclosure) to legally marketed in interstate comments on the enclosin's to tegally marketed predical Devices marketed in interested
comments to May 28, 1976, the enactment date of the Medical Device Amendments, or o devices that have been reclassified in accordance with the provisions of the Federals, or to
and Cosmetic Act (Act) that do not require with the provisions of the Federal Foo and Cosmetic Act (Act) that do not require approval of the Federal Food, Drug,
and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market he device, subject to the general controls provisions of PMA),
You may, therefore, market the device, subject to the general controls of the Act. general controls provisions of the Act include requirements for annual registration, fisting of devices, good manufacturing practice, labeling, and prohibitions against misting
adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affective your device controls) or class il! (PMA),
found in the Code of Federal Repulations, Title 21 R found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Michele Chin-Purcell, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdaggov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
3
2
1
{3}------------------------------------------------
િન
111
Indications for Use
510(k) Number (if known): N/A
Device Name: NUVANT® Mobile Cardiac Telemetry
Indications for Use:
The NUVANT Mobile Cardiac Telemetry (MCT) System is intended to continuously measure, record and periodically transmit physiological data. The System is indicated for those patients who require monitoring for the detection of non-lethal cardiac arrhythmias such as, but not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias and conduction disorders.
The NUVANT system model monitors, derives and displays:
- ECG
- Heart Rate
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
(Division Sign Off)
Page 1 of 1
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number *K111912*
{4}------------------------------------------------
AUG 2 4 2011
## 2017 13:30 Presi
| Submitter Name & Address: | Corventis, Inc.<br>1410 Energy Park Drive, Suite 1<br>St. Paul, MN 55108 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Michele Chin-Purcell, Ph.D.<br>651-925-3803 (phone)<br>651-389-3251 (fax) |
| Trade/Proprietary Name: | NUVANT® Mobile Cardiac Telemetry System |
| Common/Usual Name: | Mobile Cardiac Telemetry System |
| Classification Name: | Arrhythmia Detector and Alarm<br>(21 CFR 870.1025, Product Code DSI)<br>Patient Physiological Monitor (with arrhythmia detection)<br>(21 CFR 870.1025, Product Code MHX) |
| Class: | Class II, Special Controls |
| 510(k): | Abbreviated 510(k) |
| Date Prepared: | July 5, 2011 |
### Predicate Devices:
- 1. NUVANT Mobile Cardiac Telemetry System, Corventis, Inc. cleared by FDA under 510(k) number K091971; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm", and 21 CFR 870.1025, MHX "Patient Physiological Monitor (with arrhythmia detection)"
- 2. CardioNet ECG Monitor with Arrhythmia Detection Model CN1005, CardioNet Inc., cleared by FDA under K072558; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm"
{5}------------------------------------------------
The following is the subject device of this Abbreviated 510(k):
- The NUVANT Mobile Cardiac Telemetry System (hereafter referred to as . "subject NUVANT")
This submission proposes to add clarifying language (e.g., listing the type of non-lethal arrhythmias) to the predicate NUVANT indications for use. This clarifying language is similar to that used in the predicate device from CardioNet. This change is for clarification and does not change to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, and does not affect the safety and effectiveness of the device when used as labeled [807.92(a)(5)].
The intent of the subject devices remains the same as that of the predicate devices, which is to monitor ECG and heart rate.
**INDICATION FOR USE:**
FOR USE ONLY WITH THE ALIGNER REMOVAL INSTRUMENT
The following are the proposed indications for use in this submission:
The NUVANT Mobile Cardiac Telemetry (MCT) System is intended to continuously measure, record, and periodically transmit physiological data. The System is indicated for those patients who require monitoring for the detection of nonlethal cardiac arrhythmias such as, but not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias and conduction disorders.
The NUVANT system model monitors, derives and displays:
- ECG .
- Heart rate .
### Section 7: Summary of Safety and Effectiveness
{6}------------------------------------------------
The NUVANT MCT System is wearable, wireless arrhythmia detection system that is used by clinicians to identify suspected cardiac arrhythmias. In combination with interpretation services provided by learned intermediaries in the Corventis Monitoring Center as well as online review of data (for prescribing physicians only), NUVANT MCT enables arrhythmia detection for up to 7.5 days for each PiiX application.
The NUVANT system components are:
- . PiiX® (aka: Adherent Device) - a patient-worn device which is applied to the patient's torso. It contains the ECG electrodes for recording ECG and heart rate data.
- Patient Trigger Magnet used by the patient to manually trigger the ECG . collection when he/she experiences symptoms.
- . zLink® (aka: Gateway) – hand-held device that receives information from the PiiX and transmits it to the Corventis Server via cellular technology.
- . Server - The Server receives sensor data from the PiiX via zLink. ECG and heart rate are presented to learned intermediaries, Corventis cardiographic technicians, who prepare and deliver the information to prescribing physicians
The communication between the PiiX and the zLink is enabled via BlueTooth Technology. Sensor data and ECGs collected by the PiiX are transmitted to the Server via zLink.
The system components of the subject NUVANT are identical to the system components of the predicate NUVANT.
{7}------------------------------------------------
Two (2) predicate devices have been identified:
- 1. NUVANT Mobile Cardiac Telemetry System, Corventis, Inc. cleared by FDA under 510(k) number K091971; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm", and 21 CFR 870.1025, MHX "Patient Physiological Monitor (with arrhythmia detection)"
- 2. CardioNet ECG Monitor with Arrhythmia Detection Model CN1005, CardioNet Inc., cleared by FDA under K072558; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm"
The following table summarizes the compared features and the corresponding predicate devices.
| Features being<br>compared | Subject Devices | Predicate Devices |
|----------------------------|-----------------|----------------------------------------------------------|
| All technical<br>features | Subject NUVANT | Predicate NUVANT (K091971)<br>CardioNet CN1005 (K072558) |
| Indications for Use | Subject NUVANT | Predicate NUVANT (K091971)<br>CardioNet CN1005 (K072558) |
Features comparison and corresponding predicate devices
# THE OF BEREORING FOR THE
Both the subject and predicate NUVANT MCT meet the requirements of following performance standards in accordance with FDA Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm.
IEC 60601-1
{8}------------------------------------------------
Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995
### IEC 60601-1-2
Medical Electrical Equipment - Part 1-2: General Requirements for Safety -Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004))
### AAMI/ANSI EC38
Medical electrical equipment - Part 2-47: Particular requirements for the safety, including essential performance, of ambulatory electrocardiographic systems, 2007.
### AAMI/ANSI EC57
Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms, 1998/(R) 2008
The modifications proposed in the subject NUVANT System make clarifications to the indications for use and do not change the fundamental scientific technology or use of the devices. As supported by the descriptive information and the bench tests, it is concluded that the NUVANT Mobile Cardiac Telemetry (NUVANT) Systems are as safe and as effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.