EZS05 CLEANING AND DISINFECTING SOLUTION, BAUSCH + LOMB EZS05 CLEANING AND DISINFECTING SOLUTION
K111877 · Bausch & Lomb, Inc. · LPN · Jan 24, 2012 · Ophthalmic
Device Facts
Record ID
K111877
Device Name
EZS05 CLEANING AND DISINFECTING SOLUTION, BAUSCH + LOMB EZS05 CLEANING AND DISINFECTING SOLUTION
Applicant
Bausch & Lomb, Inc.
Product Code
LPN · Ophthalmic
Decision Date
Jan 24, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
BAUSCH + LOMB EZS05 Disinfecting Solution is indicated for disinfecting. protein removal, and storage of soft (hydrophilic) contact lenses, as recommended by your eye care practitioner.
Device Story
BAUSCH + LOMB EZS05 Disinfecting Solution is a sterile, buffered 3% hydrogen peroxide solution used for cleaning, disinfecting, and storing soft (hydrophilic) contact lenses. The system includes a specialized lens case equipped with a platinum-coated neutralizing disc. The platinum disc catalyzes the decomposition of the 3% hydrogen peroxide into water and oxygen, neutralizing the solution during the disinfection process. The device is intended for consumer use at home. Users place lenses in the provided case with the solution for a minimum of 6 hours. The process removes proteins and microorganisms from the lens surface. The solution is compatible with various soft contact lens materials, including silicone hydrogels. By ensuring proper disinfection and protein removal, the device helps maintain lens hygiene and comfort, reducing the risk of ocular irritation or infection associated with contaminated lenses.
Clinical Evidence
No clinical data. Safety and efficacy were established through non-clinical bench testing, including biocompatibility (cytotoxicity, ocular irritation, sensitization), microbiological testing per EN ISO 14729 and EN ISO 14730, and in vitro protein removal studies using HPLC assay. Lens compatibility was confirmed for various soft contact lens materials.
Technological Characteristics
Chemical solution containing 3% hydrogen peroxide, phosphonic acid stabilizer, sodium chloride, and phosphate buffers. Includes a specialized lens case with a platinum-coated neutralizing disc. Disinfection requires a 6-hour minimum contact time. Compatible with soft (hydrophilic) and silicone hydrogel contact lenses.
Indications for Use
Indicated for disinfecting, protein removal, and storage of soft (hydrophilic) contact lenses for users as recommended by an eye care practitioner.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Ciba Clear Care Cleaning and Disinfecting Solution (K022687)
Submission Summary (Full Text)
{0}------------------------------------------------
KI11877
JAN 2 4 2012
## 510(k) SUMMARY
# BAUSCH + LOMB EZS05 3% Hydrogen Peroxide Disinfecting Solution
#### 1. Submitter Information
Bausch & Lomb Incorporated 1400 North Goodman Street Rochester, NY 14609
Contact Person Heather Michaels Specialist Global Regulatory Affairs
#### Telephone Number (585) 338-8493
## 2. Device Name
| Common Name: | Contact Lens Disinfection Solution | |
|------------------------|-----------------------------------------------|--|
| Trade Name: | TBD | |
| Classification: | Soft (hydrophilic) contact lens care products | |
| Device classification: | Class II (21 CFR §886.5928) | |
| Product Code: | LPN | |
# 3. Predicate Devices
Ciba AOSEPT Disinfecting Solution (K003345, K013512) Ciba Clear Care Cleaning and Disinfecting Solution (K022687)
#### 4. Description of the Device
BAUSCH + LOMB EZS05 Disinfecting Solution is a sterile, buffered 3% hydrogen peroxide solution with a phosphonic acid stabilizer, sodium chloride and phosphate buffers. A special lens case containing a platinum coated neutralizing disc is provided with the EZS05 Disinfecting Solution. The sterile solution is packaged in a plastic bottle with a tamper evident seal and labelled with a lot number and expiration date
{1}------------------------------------------------
#### 5. Intended Use
BAUSCH + LOMB EZS05 Disinfecting Solution is indicated for disinfecting. protein removal, and storage of soft (hydrophilic) contact lenses, as recommended by your eye care practitioner.
#### 6. Description of Safety and Substantial Equivalence
A series of preclinical testing was performed to demonstrate the safety and effectiveness of BAUSCH + LOMB EZS05 Disinfecting Solution. A summary of the test results is provided below:
#### Biocompatibility
EZS05 Disinfecting Solution was evaluated for non-clinical safety in accordance with FDA Guidance for Contact Lens Solutions and Accessory Products, May 1997, as well as referencing several recognized testing Standards which were performed under Good Laboratory Practice requiations. Cytotoxicity, ocular irritation, and sensitization studies were completed for BAUSCH + LOMB EZS05 Disinfecting Solution. The test results demonstrated that BAUSCH + LOMB EZS05 Disinfecting Solution is non-cytotoxic, and not an ocular irritant or sensitizing agent.
#### Microbiology
A series of studies were conducted according to EN ISO 14729:2001 Ophthalmic optics - Contact lens care products - Microbiological requirements and test methods for products and regimens for hygienic management of contact lenses and EN ISO 14730:2000 Ophthalmic optics -Contact lens care products - Antimicrobial preservative efficacy testing and guidance on determining discard date. The testing demonstrated BAUSCH + LOMB EZS05 Disinfecting Solution meets the criteria for disinfection and preservative efficacy.
#### Lens Compatibility
The results of lens compatibility studies demonstrate BAUSCH + LOMB EZS05 Disinfecting Solution is compatible with soft contact lenses including silicone hydrogel contact lenses.
#### Protein Removal
The protein removal ability of EZS05 and predicate device Ciba Clear Care was evaluated with Polymacon (38.6% water), Etafilcon A (55% water), Galyfilcon A (47% water), and Balafilcon A (36% water) Soft Contact Lenses. The study evaluated the efficacy of the test products' on lenses deposited in vitro with lysozyme and subjected to a single cleaning / disinfection regimen. The amount of lysozyme remaining on the lenses was quantitated by extraction and HPLC assay for total protein. The results demonstrated that EZS05 was statistically equivalent to the predicate device, Ciba Clear Care Cleaning and Disinfecting Solution.
{2}------------------------------------------------
# Clinical Study
EZS05 Disinfecting Solution contains 3% hydrogen peroxide as the active ingredient, which is the same active ingredient within marketed concentrations of peroxide disinfecting solutions currently marketed in the Bausch & Lomb assessed the market history of peroxides, the U.S. ingredients compared to marketed product and, the applicability of the US FDA's "Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products" and determined that further Clinical Studies are not warranted.
# Substantial Equivalence
The cumulative results of laboratory, in vitro and in vivo testing sponsored by Bausch & Lomb demonstrate that the safety, efficacy and performance of BAUSCH + LOMB EZS05 Disinfecting Solution are substantially equivalent to Ciba AOSEPT Disinfecting Solution, and Ciba Clear Care Cleaning and Disinfecting Solution for soft contact lenses.
| Feature | BAUSCH + LOMB<br>EZS05 Disinfecting Solution | Ciba Clear Care<br>Cleaning and<br>Disinfecting Solution K022687 | Ciba AOSEPT<br>Disinfecting Solution K003345,<br>K013512 |
|----------------------------------------------------|----------------------------------------------|------------------------------------------------------------------|----------------------------------------------------------|
| Indicated for soft<br>lenses | Yes | Yes | Yes |
| Indicated for the<br>removal of<br>Protein | Yes | Yes | No |
| Rub Regimen | No | Yes, only for Rigid<br>Gas Permeable<br>Lenses | No |
| Lens Case | Designed for<br>system | Designed for<br>system | Designed for<br>system |
| Minimum<br>Disinfection Time | 6 hours | 6 hours | 6 hours |
| Post disinfection<br>saline rinse prior to<br>wear | Optional | Optional | Yes |
| Hydrogen Peroxide<br>Content | 3% | 3% | 3% |
| Buffer System | Phosphate Buffer | Phosphate Buffer | Phosphate Buffer |
| Stabilizer | Phosphonic Acid | Phosphonic Acid | Phosphonic Acid |
| Lens Storage<br>Period | 7 days | 7 days | 24 hours |
## SUBSTANTIAL EQUIVALENCE SUMMARY TABLE
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird figure. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 2 4 2012
Bausch & Lomb, Inc. c/o Ms. Heather Michaels Specialist Global Regulatory Affairs 1400 North Goodman Street Rochester, NY 14609
Re: K111877
Trade/Device Name: Bausch & Lomb EZS05 Disinfecting Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) contact lens care products Regulatory Class: Class II Product Code: LPN Dated: January 17, 2012 Received: January 18, 2012
Dear Ms. Michaels:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to commerce prior to May 20, 1976, the excordance with the provisions of the Federal Food, Drug, uevices that have been recuire approval of a premarket approval application (PMA). alla Cosmetic Act (Fiel) that to her required to the general controls provisions of the Act. The T ou may, therefore, market the act include requirements for annual registration, listing of general controls provide of tractice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adultcrainon. Treasonote. OD142 801 at 11:15 am the truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (bee acove). Existing major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must Or any 1 occurated and survivements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of it I ar o 7), ideese events) (21 CFR 803); good manufacturing practice requirements as set de rice-related daverse overley (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you actual open fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Kesia Alexander
vina B. Evdelman. M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
BAUSCH + LOMB EZS05 Disinfecting Solution Device Name:
Indications for Use:
BAUSCH + LOMB EZS05 Disinfecting Solution is indicated for disinfecting, protein removal, and storage of soft (hydrophilic) contact lenses, as recommended by your eye care practitioner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use × (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K111877
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.