SYNGO.MR ONCO, SYNGO.MR 3D LESION SEGMENTATION, SYNGO.MR ONCO ENGINE
K111861 · Siemens Medical Systems USA · LLZ · Jan 6, 2012 · Radiology
Device Facts
Record ID
K111861
Device Name
SYNGO.MR ONCO, SYNGO.MR 3D LESION SEGMENTATION, SYNGO.MR ONCO ENGINE
Applicant
Siemens Medical Systems USA
Product Code
LLZ · Radiology
Decision Date
Jan 6, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
syngo.MR Oncology is a syngo.via-based image viewing, processing and reading software. This software allows for oncological MR image evaluation in a structured way. It is a reading application primarily for convenient reading of MR scans of single or multiple body regions up to the entire body which have been acquired for clinical purposes such as oncological screening, staging and grading. The reading application does not block the display of data from other modalities that syngo.via supports. syngo.MR Oncology is designed to support the oncological workflow including interpretation and evaluation of examinations, and follow-up documentation of findings. The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. syngo.MR Oncology is a tool to support the standard practices and visual comparisons. The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
Device Story
syngo.MR Oncology is a syngo.via-based software application for oncological MR image evaluation. It processes MR scans of single or multiple body regions to support clinical workflows, including screening, staging, and grading. The device includes specific workflow configurations (MR Onco Multi-Region, Liver, Brain, TimCT) and a 3D Lesion Segmentation tool. The segmentation tool utilizes Random walker or Level Set algorithms to provide volumetric evaluation of tumors, lymph nodes, and metastases. Clinicians use the software to interpret examinations and document findings. The device acts as a support tool for standard clinical practices; the clinician retains ultimate diagnostic responsibility. It is used in clinical settings where MR imaging is performed. The software facilitates structured reading and follow-up documentation, potentially improving the efficiency and consistency of oncological assessments.
Clinical Evidence
No clinical data provided. Safety and effectiveness are supported by software verification and validation testing, and adherence to recognized industry standards.
Technological Characteristics
Software-based image viewing and processing application. Operates on the syngo.via platform. Includes specific workflow modules and a 3D lesion segmentation tool. Segmentation algorithms include Random walker and Level Set. Software development follows ISO 14971:2007 risk management standards.
Indications for Use
Indicated for oncological MR image evaluation, including screening, staging, and grading of single or multiple body regions. Intended for use by clinicians to support interpretation, evaluation, and follow-up documentation of oncological examinations. Not intended for digital mammography diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
SIEMENS
KIII 861
# 510(k) Summary: syngo.MR Oncology
This summary of 510(k) safety and effectiveness information is being submitted in This burnmary of STO(1) Caroly at Safe Medical Device Act 1990 and 21 CFR § 807.92.
#### General Information l.
| Establishment | Siemens Medical Solutions USA, Inc.<br>51 Valley Stream Parkway<br>Mail Code D02<br>Malvern, PA 19355, USA |
|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number | 2240869 |
| Manufacturer | Siemens AG<br>Henkestrasse 127<br>D-91052 Erlangen, Germany |
| Registration Number | 8010024 |
| Contact Person | Ms. Nadia Sookdeo<br>Technical Specialist<br>Regulatory Affairs/Clinical Affairs<br>Phone: (610) 448-4918<br>Fax: (610) 448-1787<br>Email: nadia.sookdeo@siemens.com |
| Device Name<br>Classification Name:<br>Classification Panel:<br>Regulation number:<br>Device Class:<br>Product Code: | syngo.MR Oncology<br>Picture Archiving and Communications System<br>(PACS)<br>Radiology<br>21 CFR § 892.2050<br>II<br>LLZ, LNH |
{1}------------------------------------------------
Image /page/1/Picture/16 description: The image shows the word "SIEMENS" in all capital letters. The letters are outlined in black and filled with a textured pattern. The font is bold and sans-serif, giving the word a strong and clear appearance.
#### Safety and Effectiveness Information Supporting Substantial Equivalence II.
## Indications for Use:
manations for oncology is a syngo.via-based image viewing, processing and reading offware. This software allows for oncological MR image evaluation in a structured way. It is a reading application primarily for convenient reading of MR scans of way. It is a lually and regions up to the entire body which have been acquired for clinical purposes such as oncological screening, staging and grading. The reading application does not block the display of data from other modalities that syngo.via supports.
syngo.MR Oncology is designed to support the oncological workflow including interpretation and evaluation of examinations, and follow-up documentation of findings.
The clinician retains the ultimate responsibility for making the pertinent Note: diagnosis based on their standard practices and visual comparison of the separate unregistered images. syngo.MR Oncology is a tool to support the standard practices and visual comparisons.
The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
#### Device Description
In addition to the reading and reporting functions/tools that are available in the predicate syngo.via basic software configuration, Siemens Healthcare intends to offer MR specific workflow configurations that are adapted for MR Oncology (syngo.MR Oncology).
syngo.MR Oncology is a syngo-based image viewing, processing and reading software. This software allows for oncological MR image evaluation in a structured way. It is a reading application supporting convenient reading of MR scans of single or multiple body regions, up to the entire body which have been acquired for clinical purposes such as oncological screening, diagnosis and staging.
The medical device software syngo.MR Oncology consists of the following two applications: syngo.MR Onco Workflows and the syngo.MR 3D Lesion Segmentation Tool.
The syngo.MR Onco Workflow application consists of the following tools: MR Onco Multi-Region, MR Onco Liver, MR Onco Brain, MR Onco TimCT and an Onco report.
The syngo.MR 3D Lesion Segmentation Tool also known as MR Segmentation tool, is a tool using the Random walker algorithm (default algorithm) or the Level Set algorithm. 3D Lesion Segmentation provides convenient volumetric evaluation of lesions and/or other structure of interest as well as a particularly useful tool for oncology applications (for example volumetric evaluation of tumors, lymph nodes and metastases), and nononcology lesions or other structures of interest with sufficient contrast to surrounding tissue.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "SIEMENS" in a bold, outlined font. The letters are large and evenly spaced, filling the frame of the image. The font style gives the word a distinct and recognizable appearance.
syngo.MR Oncology 510(k)
## General Safety and Effectiveness Concerns:
General Oalery and Enternetions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product.
These potential hazards are controlled via software development, verification and validation testing. To minimize hazards, Siemens adheres to recognized and established industry practice and standards.
### Substantial Equivalence
The syngo.MR Oncology is substantially equivalent to the following devices:
| Predicate<br>Software Name | 510(K) Clearance<br>Number | 510(k) Clearance Date |
|-------------------------------------|----------------------------|-----------------------|
| Software<br>syngo MR B17 | K082427 | November 7, 2008 |
| Software<br>syngo.x (syngo.via) | K092519 | August 27, 2009 |
| Software<br>syngo PET&CT Oncology | K093621 | February 23, 2010 |
| Syngo Dosimetrist Workspace<br>v2.7 | K101119 | June 6, 2010 |
In summary, modifications made to the predicate device software do not affect the intended use of the device nor alter its fundamental scientific technology. Siemens believes that the subject software (syngo.MR Oncology) is substantially equivalent to the above listed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo includes the department's seal on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The seal features a stylized human figure and a circle of text around it.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
JAN - 6 2012
Ms. Nadia Sookdeo Technical Specialist Regulatory & Clinical Affairs Siemens Medical Solutions USA, Inc. Customer Solutions Group 51 Valley Stream Parkway, Mail Code D02 MALVERN PA 19355
Re: K111861
Trade/Device Name: syngo. MR Oncology Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 7, 2011 Received: December 8, 2011
Dear Ms. Sookdeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K111861
Device Name: syngo MR Oncology
Indications for Use:
syngo.MR Oncology is a syngo.via-based image viewing, processing and reading software. syngo.MR Oncology is a syngo.via Babea image evaluation in a structured way. It is a This soltware allows for orloological will image even of single or multiple body
reading application primarily for convenient reading of MR scans of single or reading application primanly for corrective en acquired for clinical purposes such as regions up to the entire body which have been application does not block the display of data from other modalities that syngo.via supports.
syngo.MR Oncology provides analytical tools to help the user assess and document syngo.MR Oneonogy provides andryther to the diagnostic and therapy follow-up examinations.
syngo.MR Oncology is designed to support the oncological workflow including interpretation Syngolian Shoology is examinations, and follow-up documentation of findings.
Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis Note. The Citilician relains the untinute respensibility to of the separate unregistered images. syngo.MR Oncology is a tool to support the standard practices and visual comparisons.
The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
Prescription Use _X_ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary Storts
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K111861/3
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.