SUPERA VERITAS INTERWOVEN SELF-EXPANDING NITINOL STENT TRANSHEPTIC BILIARY SYSTEMS
K111766 · Idev Technologies, Inc. · FGE · Oct 19, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K111766
Device Name
SUPERA VERITAS INTERWOVEN SELF-EXPANDING NITINOL STENT TRANSHEPTIC BILIARY SYSTEMS
Applicant
Idev Technologies, Inc.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Oct 19, 2011
Decision
SESU
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System is indicated for palliative treatment of biliary strictures produced by malignant neoplasms.
Device Story
The SUPERA VERITAS® is a 6Fr, multi-lumen, guidewire-compatible delivery system for a self-expanding Nitinol stent. The stent consists of 6 interwoven Nitinol wires joined by proprietary welded couplers. The delivery system features a thumb slide for stent advancement, a deployment lock to prevent premature deployment, and radiopaque markers for positioning. Used in clinical settings by physicians, the device is inserted over a 0.014" or 0.018" guidewire to the site of a biliary stricture. The physician actuates the thumb slide to deploy the stent, which expands to maintain patency of the biliary duct. The device is sterile, single-use, and intended to provide palliative relief for patients with malignant biliary obstructions.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated via bench performance testing, including radial force comparison, torque strength, deployment testing, and particulate matter evaluation. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, thrombogenicity, and complement activation) passed all requirements per ISO 10993 and 21 CFR Part 58.
Indicated for palliative treatment of biliary strictures caused by malignant neoplasms in patients requiring biliary stenting.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Cordis S.M.A.R.T. ®Nitinol Stent Transhepatic Biliary System (K062798)
Submission Summary (Full Text)
{0}------------------------------------------------
K 111766
Page 1 of 4
OCT 1 9 2011
Image /page/0/Picture/2 description: The image shows the logo for IDEV. The logo consists of the letters "IDEV" in a bold, sans-serif font. To the right of the letters, there are three concentric circles that appear to be incomplete, creating a sense of motion or dynamism. A registered trademark symbol is located above the circles.
inspiration · innovation · intervention
## 510(k) Summary
# SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System
Submitter:
IDEV Technologies, Inc. 253 Medical Center Boulevard Webster, Texas 77598 281/525-2000
Contact Person: Darlene Garner Regulatory Affairs Manager 281/525-2052 (phone) 281/525-2001 (fax) dgarner@idevmd.com
Date Prepared: June 22, 2011
SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Trade Name: Transhepatic Biliary System
Common Name: Stent Delivery Catheter
Classification Name: Catheter, Biliary, Diagnostic; Class II
Product Code: FGE
SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Predicate Devices: Transhepatic Biliary System (K093893)
Bard E-LUMINEXX® Biliary Stent (K063532)
Cordis S.M.A.R.T. ®Nitinol Stent Transhepatic Biliary System (K062798)
{1}------------------------------------------------
#### Device Description:
The SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System is a 0.014" or 0.018" guidewire compatible, multi-lumen sheath based delivery system and a SUPERA® Biliary Stent. The stent delivery catheter includes a radiopaque Stent Length Marker and Distal Sheath Marker embedded in the Outer Sheath to aid in proper stent positioning; a Thumb Slide connected internally for advancing the Stent out of the Outer Sheath while the Outer Sheath moves proximally in a de-coupled fashion; a Sheath Flush Port for flushing the central lumen of the device; a Deployment Lock that when actuated enables the final deployment stroke of the stent; a Guidewire Lumen with a radiopaque Catheter Tip located on the distal end of the Catheter Shaft; a Guidewire Flush Port used for flushing the Guidewire Lumen; a Stent Driver (ratchet) which moves the stent distally relative to the Outer Sheath; and the System Lock which eliminates the possibility of premature deployment. The working length of the delivery catheter is 80cm and 120cm.
The SUPERA® Stent is housed within the SUPERA VERITAS® stent delivery catheter and is a closed end interwoven self-expanding Nitinol stent. The SUPERA® stent is composed of 6 interwoven, closed loop Nitinol wires. The wire loops are closed via a proprietary welding process which utilizes small Nitinol tubes that act as a coupler to provide the mechanical means of joining the wire ends.
The table below includes the available sizes and model numbers for the 6Fr SUPERA VERITAS Stent Delivery System.
| Model No. | Stent Diameter | Stent Length | Catheter Length |
|-----------------|----------------|--------------|-----------------|
| S-04-040-80-6F | 4mm | 40mm | 80cm |
| S-04-060-80-6F | 4mm | 60mm | 80cm |
| S-04-080-80-6F | 4mm | 80mm | 80cm |
| S-04-100-80-6F | 4mm | 100mm | 80cm |
| S-04-120-80-6F | 4mm | 120mm | 80cm |
| S-05-040-80-6F | 5mm | 40mm | 80cm |
| S-05-060-80-6F | 5mm | 60mm | 80cm |
| S-05-080-80-6F | 5mm | 80mm | 80cm |
| S-05-100-80-6F | 5mm | 100mm | 80cm |
| S-05-120-80-6F | 5mm | 120mm | 80cm |
| S-06-040-80-6F | 6mm | 40mm | 80cm |
| S-06-060-80-6F | 6mm | 60mm | 80cm |
| S-06-080-80-6F | 6mm | 80mm | 80cm |
| S-06-100-80-6F | 6mm | 100mm | 80cm |
| S-06-120-80-6F | 6mm | 120mm | 80cm |
| S-07-040-80-6F | 7mm | 40mm | 80cm |
| S-07-060-80-6F | 7mm | 60mm | 80cm |
| S-07-080-80-6F | 7mm | 80mm | 80cm |
| S-07-100-80-6F | 7mm | 100mm | 80cm |
| S-04-040-120-6F | 4mm | 40mm | 120cm |
| S-04-060-120-6F | 4mm | 60mm | 120cm |
| S-04-080-120-6F | 4mm | 80mm | 120cm |
| S-04-100-120-6F | 4mm | 100mm | 120cm |
{2}------------------------------------------------
| 1<br>0<br>15<br>11<br>-<br>1 | t<br>1<br>16 |
|------------------------------|--------------|
|------------------------------|--------------|
| Model No. | Stent Diameter | Stent Length | Catheter Length |
|-----------------|----------------|--------------|-----------------|
| S-04-120-120-6F | 4mm | 120mm | 120cm |
| S-05-040-120-6F | 5mm | 40mm | 120cm |
| S-05-060-120-6F | 5mm | 60mm | 120cm |
| S-05-080-120-6F | 5mm | 80mm | 120cm |
| S-05-100-120-6F | 5mm | 100mm | 120cm |
| S-05-120-120-6F | 5mm | 120mm | 120cm |
| S-06-040-120-6F | 6mm | 40mm | 120cm |
| S-06-060-120-6F | 6mm | 60mm | 120cm |
| S-06-080-120-6F | 6mm | 80mm | 120cm |
| S-06-100-120-6F | 6mm | 100mm | 120cm |
| S-06-120-120-6F | 6mm | 120mm | 120cm |
| S-07-040-120-6F | 7mm | 40mm | 120cm |
| S-07-060-120-6F | 7mm | 60mm | 120cm |
| S-07-080-120-6F | 7mm | 80mm | 120cm |
| S-07-100-120-6F | 7mm | 100mm | 120cm |
The SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System is a sterile (via Ethylene Oxide sterilization) device and is intended for single use only.
## Intended Use:
The SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System is indicated for palliative treatment of biliary strictures produced by malignant neoplasms.
## Comparison to Predicate Devices:
The 6Fr SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System is substantially equivalent to the predicate devices: IDEV's 7Fr SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System (K093893); the Bard E-LUMINEXX® Biliary Stent (K063532) and the Cordis S.M.A.R.T® Nitinol Stent Transhepatic Biliary System (K. 062798).
A review of the product specifications concluded that there are no major differences in design, materials, performance, safety and product effectiveness. Substantial Equivalence to the predicate devices has been demonstrated via bench performance testing.
## Non-clinical Testing:
Non-clinical testing were performed per the FDA's "Guidance of Premarket notification 510(k) Submissions for Short Term and Long Term Intravascular Catheters', the "Guidance for the Content of Premarket Notifications for Metal Expandable Biliary Stents; ISO 10555 - Sterile, Single-use Intravascular Catheters; and ISO 1093 -International Standard for Biological Evaluation of Medical Devices". Clinical 1077yas not required in order to demonstrate safety and efficacy for the device modifications described in this 510(k).
{3}------------------------------------------------
Biocompatibility testing included:
| Test | Results |
|--------------------------------------------------|---------|
| Cytotoxicity | Pass |
| Klingman Maximization (NaCl and CSO extracts) | Pass |
| Intracutaneous Injection (NaCl and CSO extracts) | Pass |
| Systemic Injection (NaCl and CSO extracts) | Pass |
| Rabbit Pyrogen | Pass |
| Hemolysis | Pass |
| Thrombogenicity Study in Dogs | Pass |
| Complement Activation Assay (C3a and SC5b-9) | Pass |
| Unactivated Partial Thromboplastin Time (UPTT) | Pass |
All biocompatibility testing was performed under the 21 CFR, Part 58 Good Laboratory Practice for Non-Clinical Laboratory Studies. All tests passed the acceptance criteria identified in the applicable protocols.
Performance Bench Testing included:
| Test | Description of Test | Results |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Biliary Radial<br>Force | Comparison of the results of the SUPERA Stent deployed from a 6Fr<br>SUPERA VERITAS Stent Delivery System to the Bard E-<br>LUMINEXX in regards to compressive force (radial stiffness and<br>radial strength) and expansive force (radial stiffness and radial<br>strength. | Pass |
| Catheter Torque<br>Strength | Performed to characterize the torque strength of the SUPERA<br>VERITAS Stent Delivery system when the handle was rotated on its<br>central axis | Pass |
| Deployment<br>Testing | This following elements were assessed during deployment testing:<br>Working length, ratchet release, Tensile, Leak, Introducer<br>Compatibility, Guidewire Compatibility, Pushability/Flexibility,<br>Trackability, Torque, Distal stent deployment and Accuracy, Thumb<br>slide deployment force, Thumb slide retraction force and Deployed<br>stent length accuracy | Pass |
| Particulate | Performed using USP <788>, Particulate Matter in Injections and<br>AAMI TIR42, Evaluation of Particulates Associated with Vascular<br>Medical Devices. | Pass |
Performance bench testing demonstrated that the 6Fr SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System is substantially equivalent to its predicate devices. No additional safety risks were observed during testing.
## Conclusion:
The 6Fr SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System demonstrated to be substantially equivalent to the predicate devices based on design specifications and characteristics, principle of operation, indications for use and performance testing.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles a bird in flight or a human figure with outstretched arms, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Darlene Garner Regulatory Affairs Manager IDEV Technologies, Inc. 253 Medical Center Boulevard WEBSTER TX 77598
OCT 19 2011
Re: K111766
Trade/Device Name: SUPERA VERITAS® Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System Regulation Number: 21 CFR § 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: September 29, 2011 Received: September 30, 2011
Dear Ms. Garner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:
The safety and effectiveness of this device for use in the vascular system have not been established.
Furthermore, the indication for use must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
{5}------------------------------------------------
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (301) 796-5484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at (301) 796-6926. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Christy Foreman
Christy Foreman Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K11766
The SUPERA™ VERITAS™ Interwoven Self-Expanding Nitinol Stent Device Name: Transhepatic Biliary System
FDA's Statement of the Indications for Use for device:
The SUPERA™ VERITAS™ Interwoven Self-Expanding Nitinol Stent Transhepatic Biliary System is indicated for palliative treatment of biliary strictures produced by malignant neoplasms.
Prescription Use 1 (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
torgn m. Wh
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.