POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)

K111752 · Shandong Jianhong New Material Company · LYZ · Jul 27, 2011 · General Hospital

Device Facts

Record IDK111752
Device NamePOWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)
ApplicantShandong Jianhong New Material Company
Product CodeLYZ · General Hospital
Decision DateJul 27, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable vinyl patient examination gloves; worn on examiner's hand or finger; intended to prevent cross-contamination between patient and examiner. Used in clinical settings by healthcare providers. Device acts as physical barrier. No electronic, software, or algorithmic components.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including physical property testing, pinhole leak testing per 21 CFR 800.20, and biocompatibility testing (primary skin irritation and dermal sensitization) per ISO 10993-10.

Technological Characteristics

Vinyl material; powder-free; non-sterile. Meets ASTM D5250-06e1 (dimensions, physical properties, powder residual <2mg/glove). Biocompatibility per AAMI/ANSI/ISO 10993-10 (primary skin irritation, dermal sensitization). Pinhole AQL compliance per 21 CFR 800.20.

Indications for Use

Indicated for use as a non-sterile, disposable patient examination glove worn by examiners to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) Summary "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92." "The assigned 510(k) number is: Kill 75C " (applicant leave blank) Premarket Notification [510(k)] Summary | Submitter's name : | Shandong jianhong new material company | |--------------------------------|--------------------------------------------------------------------------------------------------| | Submitter's address : | Xisha, Guoli Town, Huantai County, Zibo City, Shandong Province, 255000,China | | Phone number : | 0086-533-8437998 | | Fax number : | 0086-533-8437998 | | Name of contact person: | Mr.Tian Jianguo | | Date the summary was prepared: | May 10, 2011 | | Device Name: | Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) | | Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Other clients private labeling | | Common Name: | Exam gloves | | Classification Name: | Patient examination glove | | Device Classification: | I | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital (80) | | Product Code: | LYZ | Class I* Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1. Predicate device: Powder-Free Vinyl Patient Examination Gloves, Clear (Non-colored), ZHANG JIA GANG FENGYUAN PLASTIC PRODUCTS CO LTD. K091663. Device Description: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1. {1}------------------------------------------------ Device Intended Use (indication for use): Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ## A summary of the technological characteristics of new device compared to the predicate device. The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard. | Characteristics | Standard | Device performance | |-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | Dimension | ASTM standard D 5250-06 el. | Meets | | Physical Properties | ASTM standard D 5250-06 el. | Meets | | Freedom from pinholes | 21 CFR 800.20 | Meets | | Powder Residual | ASTM standard D 5250-06 el<br>and D6124-06 | Meets<br><2mg/glove | | Biocompatability | Primary Skin Irritation in rabbits<br>AAMI / ANSI / ISO 10993-10<br><br>Dermal sensitization in the guinea pig<br>AAMI / ANSI / ISO 10993-10 | Passes<br>Not a Primary Skin Irritation<br><br>Passes<br>Not a Dermal sensitization | ## A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), meet requirements per ASTM D5250-06 e1, per ASTM D6124-06, per 21 CFR 800.20 and FDA recognition number 2-152: AAMI / ANSI / ISO 10993-10:2002/Amd. 1:2006. The performance test data of the non-clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above. ## A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Clinical data is not needed for gloves or for most devices cleared by the 510(k) process. ### Conclusions It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims. It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is as safe, as effective, and performs as well as the predicate device, Powder-Free Vinyl Patient Examination Gloves, Clear (Non-colored), ZHANG JIA GANG FENGYUAN PLASTIC PRODUCTS CO LTD.K091663. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around the left side of the emblem. To the right of the text is a stylized image of an eagle with three lines representing its wings and a wavy line representing its body. # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Shandong Jianhong New Material Company C/O Mr. Chu Xiaoan Beijing Easy-Link Company Room 1606 Building, 1 Jian Xiang Yuan No.209 Bei Si Huan Zhong Road, Beijing China 10083 JUL 2 7 2011 Re: K111752 Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: July 1, 2011 Received: July 8, 2011 Dear Mr. Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Xiaoan · Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of . the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOf fices/CDRH/C DRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProbl em/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Anthony D. Wilson Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Section B Indications for Use # INDICATIONS FOR USE Applicant: Shandong jianhong new material company 510(k) Number (if known): * Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Device Name: ## Indications For Use: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ × (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Rohit Ganguly's (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K111752
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...