VERSAPOINT II BIPOLAR ELECTROSURGERY GENERATOR, VERSAPOINT II 44MM ANGLED LOOP ELECTRODE, VERSAPOINT II FOOTSWITCH, VERS
K111751 · ETHICON, Inc. · HIN · Dec 2, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K111751
Device Name
VERSAPOINT II BIPOLAR ELECTROSURGERY GENERATOR, VERSAPOINT II 44MM ANGLED LOOP ELECTRODE, VERSAPOINT II FOOTSWITCH, VERS
Applicant
ETHICON, Inc.
Product Code
HIN · Obstetrics/Gynecology
Decision Date
Dec 2, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GYNECARE VERSAPOINT™ II Bipolar Electrosurgery System is intended for use in tissue cutting, removal and desiccation as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions, and septa, and benign conditions requiring endometrial ablation. Procedures include: Excision of intrauterine myomas Excision of intrauterine polyps Lysis of intrauterine adhesions Incision of uterine septa Endometrial ablation
Device Story
Electrosurgical system for gynecologic hysteroscopic procedures; utilizes bipolar technology for tissue vaporization, ablation, excision, and thermal coagulation/hemostasis. System components: generator, 4mm angled loop electrode, footswitch, and connector cable. Operated by physicians in clinical settings. Generator provides five output modes: Ablation (VC), Enhanced Ablation (VP), Blended Ablation (BL), Enhanced Blended Ablation (VBL), and Coagulation (DES). Footswitch allows clinician control of power and mode settings. Output enables precise tissue management, reducing surgical time and improving patient outcomes for intrauterine pathologies.
Clinical Evidence
Two bench studies and one animal study (conducted per 21 CFR Part 58) compared the subject device to the predicate. Endpoints included maximum thermal margin, depth of tissue necrosis, mean volume of gas generated, mean tissue removal rate, and hemostasis performance. Results demonstrated substantial equivalence.
Technological Characteristics
Bipolar electrosurgical system. Components: generator, 4mm angled loop electrode, footswitch, connector cable. Five output modes: VC, VP, BL, VBL, DES. Hardware and software updates accommodate the 4mm electrode. Bipolar technology maintains identical principle of operation to predicate.
Indications for Use
Indicated for gynecologic hysteroscopic electrosurgical procedures including excision of intrauterine myomas, polyps, lysis of adhesions, incision of septa, and endometrial ablation. Contraindicated in patients with acute cervicitis, pregnancy, cervical or uterine malignancy, active pelvic inflammatory disease, unaddressed adnexal pathology, or for tubal sterilization.
Regulatory Classification
Identification
A bipolar endoscopic coagulator-cutter is a device used to perform female sterilization and other operative procedures under endoscopic observation. It destroys tissue with high temperatures by directing a high frequency electrical current through tissue between two electrical contacts of a probe. This generic type of device may include the following accessories: an electrical generator, probes, and electrical cables.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance 2/12/90 (K-90),” and
(iii) “Guidance (‘Guidelines’) for Evaluation of Laproscopic Bipolar and Thermal Coagulators (and Accessories),”
(2) International Electrotechnical Commission's IEC 60601-1-AM2 (1995-03), Amendment 2, “Medical Electrical Equipment—Part 1: General Requirements for Safety,”
(3) American National Standards Institute/American Association for Medical Instrumentation's HF-18, 1993, “Electrosurgical Devices,”
(4) Labeling:
(i) Indication: For female tubal sterilization, and
(ii) Instructions for use:
(A) Destroy at least 2 centimeters of the fallopian tubes,
(B) Use a cut or undampened sinusoidal waveform,
(C) Use a minimum power of 25 watts, and
(D) For devices with ammeters: continue electrode activation for 5 seconds after the visual endpoint (tissue blanching) is reached or current flow ceases indicating adequate tissue destruction.
Predicate Devices
GYNECARE VERSAPOINT Bipolar Electrocautery System (K962482)
{0}------------------------------------------------
イ
# 510(k) Summary
ー
| Submitter<br>Information | Company: | Ethicon, Inc.<br>Route 22 West<br>Somerville, NJ 08876 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Brian A. Kanerviko<br>Director, WW Regulatory Affairs |
| | Email: | bkanervi@its.jnj.com |
| | Telephone: | 908-218-3392 |
| | FAX: | 908-218-2595 |
| | Date Prepared: | June 6, 2011 |
| Identification<br>of the Device | Common Name: | Electrosurgery System |
| | Device Name: | GYNECARE VERSAPOINT™ II Bipolar<br>Electrosurgery System |
| | Classification Name: | Bipolar endoscopic coagulator cutter and accessories<br>Hysteroscope and accessories |
| | Device<br>Classification: | Regulation Number: 21 CFR 884.4150<br>Regulatory Class: Class II<br>Product Code: HIN |
| | | Regulation Number: 21 CFR 884.1690<br>Regulatory Class: Class II<br>Product Code: HIH |
| Predicate<br>Devices | | GYNECARE VERSAPOINT Bipolar Electrocautery System (K962482)<br>VERSAPOINT VRS 2.5mm Angled Loop Electrode (K994418) |
| Description of<br>the Device | | The GYNECARE VERSAPOINT™ II Bipolar Electrosurgery System<br>(VERSAPOINT II System) is the next generation of the currently-marketed<br>GYNECARE VERSAPOINT™ Bipolar Electrosurgery System<br>(VERSAPOINT System). This electrosurgical system utilizes bipolar<br>technology specifically designed to provide a range of gynecological surgical<br>treatments, including vaporization to ablate and excise tissue, thermal<br>coagulation of tissue and hemostasis of blood vessels. The VERSAPOINT II<br>System is comprised of four main elements:<br>• the VERSAPOINT II Generator, |
| | | Continued on next page |
| Description of the Device, continued | the VERSAPOINT II 4mm Angled Loop Electrode, the VERSAPOINT II Footswitch, and the VERSAPOINT II Connector Cable. The VERSAPOINT II System offers five bipolar output modes: Ablation (VC [VaporCut]) – for tissue removal and cutting Enhanced Ablation (VP [VersaPulse]) - an enhanced version of the Ablation (VC) output mode attained by use of short-duration high power pulses Blended Ablation (BL) – simultaneous ablation and coagulation (the generator rapidly switches between ablation and coagulation output modes) Enhanced Blended Ablation (VBL) – a variant of the VP output mode that rapidly switches the output ON and OFF to deliver blended ablation/coagulation Coagulation (DES [Desiccate]) - for thermal coagulation and hemostasis | |
| Indications for Use | The GYNECARE VERSAPOINT™ II Bipolar Electrosurgery System is intended for use in tissue cutting, removal and desiccation as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions, and septa, and benign conditions requiring endometrial ablation. Procedures include: Excision of intrauterine myomas Excision of intrauterine polyps Lysis of intrauterine adhesions Incision of uterine septa Endometrial ablation | |
| Contraindications for Use | The GYNECARE VERSAPOINT™ II Bipolar Electrosurgery System is NOT intended for use in tubal sterilization procedures. The use of this device is contraindicated in patients with the following conditions: Acute cervicitis Pregnancy Cervical or uterine malignancy Active pelvic inflammatory disease Unaddressed adnexal pathology | |
{1}------------------------------------------------
# 510(k) Summary, Continued
ﺪ ---
K111751
page
{2}------------------------------------------------
ﺎ ﺗﺴﻤ
| Technological<br>Characteristics | This electrosurgical system utilizes bipolar technology specifically designed<br>to provide a range of gynecological surgical treatments, identical to the<br>predicate device.<br><br>This premarket notification is for modifications to the currently-marketed<br>GYNECARE VERSAPOINT™ Bipolar Electrosurgery System (generator,<br>connector cable, foot switch and loop electrode). Modifications to the<br>generator include hardware and software updates to accommodate a new 4mm<br>loop electrode. A new connector cable is provided to enable connection of<br>the new 4mm loop electrode and existing electrodes to the generator. In<br>addition, a new heavier weight footswitch has been added to the system which<br>allows the clinician to change power and mode settings. |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>Testing | Two bench studies and one animal study (conducted in compliance with the<br>Good Laboratory Practices regulation, 21 CFR Part 58) demonstrate the<br>substantial equivalence of the performance of the GYNECARE<br>VERSAPOINT™ II Bipolar Electrosurgery System to the predicate<br>GYNECARE VERSAPOINT™ Bipolar Electrosurgery System with regards<br>to:<br>maximum thermal margin and depth of tissue necrosis, mean volume of gas generated, mean tissue removal rate, and hemostasis performance. |
| Conclusions<br>Drawn from<br>Studies | Validity of Scientific Data:<br>The bench studies followed scientific protocols with pre-determined<br>acceptance criteria. The animal study was conducted in compliance with<br>Good Laboratory Practices (GLPs). The data were scientifically valid in<br>accordance with 21 CFR 860.7.<br><br>Substantial Equivalence:<br>Information presented in this premarket notification establishes that the<br>GYNECARE VERSAPOINT™ II Bipolar Electrosurgery System is as safe<br>and effective as the predicate device when used in accordance with the<br>labeled directions for use for the stated indication.<br><br>Risk and Benefits:<br>The risks of the subject device are the same as those of the predicate device.<br>The benefits to the patient are the same as those for the predicate device. |
KIII 15
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
DEC - 2 2011
Mr. Brian A. Kanerviko Director, WW Regulatory Affairs Ethicon, Inc. Route 22 West SOMERVILLE NJ 08876
K111751
Re:
Trade/Device Name: GYNECARE VERSAPOINT™ II Bipolar Electrosurgery System
Regulation Number: 21 CFR§ 884.4150
Regulation Name: Bipolar endoscopic coagulator-cutter and accessories Regulatory Class: II Product Code: HIN, HIH Dated: November 11, 2011 Received: November 17, 2011
· Dear Mr. Kanerviko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hubert Lenoir 19
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and · Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K | | 75 |
Device Name: GYNECARE VERSAPOINT™ II Bipolar Electrosurgery System
Indications for Use:
The GYNECARE VERSAPOINT™ II Bipolar Electrosurgery System is intended for use in tissue cutting, removal and desiccation as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions, and septa, and benign conditions requiring endometrial ablation. Procedures include:
- . Excision of intrauterine myomas
- ◆ Excision of intrauterine polyps
- . Lysis of intrauterine adhesions
- Incision of uterine septa
- . Endometrial ablation
Hebert Penn
ork) N
Prescription Use V (21 CFR Part 801 Subpart D)
and/or
Over-the-Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.