← Product Code [JJX](/productcode/JJX) · K111650

# IDS-ISYS CTX-1 (CROSSLAPS) CALIBRATION VERIFIERS, IDS-ISYS 25 HYDROXY VITAMIN D CALIBRATION VERIFIERS, AND IDS-ISYS... (K111650)

_Immunodiagnostic Systems , Ltd. · JJX · Sep 6, 2011 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K111650

## Device Facts

- **Applicant:** Immunodiagnostic Systems , Ltd.
- **Product Code:** [JJX](/productcode/JJX.md)
- **Decision Date:** Sep 6, 2011
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The IDS-iSYS CTX-I (CrossLaps) Control Set is intended for medical purposes for use in the IDS-iSYS CTX-I (Crosslaps) Assay on the IDS-iSYS Multi-Discipline Automated Analyser to monitor the accuracy and quality of the IDS-iSYS CTX-I (Crosslaps) Assay. The IDS-iSYS CTX-I (CrossLaps) Calibrator Verifier is a device intended for medical purposes for use in the quantitative verification of calibration and assay range of the IDS-iSYS CTX-I (Crosslaps®) Assay when performed on the IDS-iSYS Multi-Discipline Automated Analyzer. The IDS-iSYS 25-Hydroxy Vitamin D Calibration Verifiers are intended for use in the quantitative verification of calibration and assay range of the IDS-iSYS 25 Hydroxy Vitamin D Assay when performed on the IDS-iSYS Multi-Discipline Automated Analyzer.

## Device Story

IDS-iSYS CTX-I (Crosslaps) Control Set, IDS-iSYS 25-Hydroxy Vitamin D Calibration Verifiers, and IDS-iSYS CTX-I (Crosslaps) Calibration Verifiers are quality control and calibration verification materials. Used in clinical laboratory settings on the IDS-iSYS Multi-Discipline Automated Analyzer. Controls monitor assay accuracy and quality; verifiers confirm calibration and assay range. These materials ensure the performance of the automated analyzer system, allowing clinicians to rely on accurate diagnostic results for patient management.

## Clinical Evidence

No clinical data. Bench testing only. Stability studies (real-time closed vial, open vial, and on-board) confirmed performance within 10% of reference material concentrations. Value assignment performed via triplicate testing across multiple IDS-iSYS analyzers.

## Technological Characteristics

Liquid, bovine serum albumin or equine serum/buffer-based matrix. Contains <0.1% sodium azide. Multi-level packaging (2.5 mL or 1 mL vials). Designed for use on IDS-iSYS Multi-Discipline Automated Analyzer. Stability validated per CEN 13640.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k111650

B. Purpose for Submission:
New Device

C. Measurand:
Quality Control materials for CTX-I and 25 Hydroxy Vitamin D

D. Type of Test:
Quality Control Material

E. Applicant:
Immunodiagnostic Systems Ltd.

F. Proprietary and Established Names:
IDS-iSYS CTX-I (Crosslaps®) Calibration Verifiers, IDS-iSYS 25 Hydroxy Vitamin D Calibration Verifiers, and IDS-iSyS CTX-I (Crosslaps®) Control Set with model(s):IS-3035, IS-2735, and IS-3030

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JJX | Class I, reserved | 21 CFR 862.1660 | Chemistry (75)  |

H. Intended Use:

1. Intended use(s):
See indications for use.

2. Indication(s) for use:

{1}

2

# IDS-iSYS CTX-I (Crosslaps) Control Set

The IDS-iSYS CTX-I (CrossLaps) Control Set is intended for medical purposes for use in the IDS-iSYS CTX-I (Crosslaps) Assay on the IDS-iSYS Multi-Discipline Automated Analyser to monitor the accuracy and quality of the IDS-iSYS CTX-I (Crosslaps) Assay.

# IDS-iSYS CTX-I (Crosslaps) Calibrator Verifiers

The IDS-iSYS CTX-I (CrossLaps) Calibrator Verifier is a device intended for medical purposes for use in the quantitative verification of calibration and assay range of the IDS-iSYS CTX-I (CrossLaps®) Assay when performed on the IDS-iSYS Multi-Discipline Automated Analyzer.

# IDS-iSYS 25 Hydroxy Vitamin D Calibrator Verifiers

The IDS-iSYS 25-Hydroxy Vitamin D Calibration Verifiers are intended for use in the quantitative verification of calibration and assay range of the IDS-iSYS 25 Hydroxy Vitamin D Assay when performed on the IDS-iSYS Multi-Discipline Automated Analyzer.

3. Special conditions for use statement(s):

For prescription use

4. Special instrument requirements:

IDS iSYS Multi-Discipline analyzer (models IS-3035, IS-2735, IS-3030)

I. Device Description:

The IDS-iSYS CTX-I (CrossLaps®) Control Set is a single analyte, liquid, bovine serum albumin based product containing CTX-I and &lt;0.1% w/w sodium azide as a preservative. The packaging consists of three control levels (two vials per level containing 2.5 mL per vial).

The IDS-iSYS CTX-I (CrossLaps®) Calibration Verifiers are a single analyte, liquid, bovine serum albumin based product containing CTX-I and &lt;0.1% w/w sodium azide as a preservative. The packaging consists of four verifier levels (two vials per level, 2.5 mL for level 0, 1 mL for levels 1-3.

IDS-iSYS 25 Hydroxy Vitamin D Calibrator Verifiers are a single analyte, liquid equine serum/buffer based product containing 25-OH D and &lt;0.1% w/w sodium azide as a preservative. The packaging consists of four verifier levels (two vials per level

{2}

containing 2.5 mL per vial).

## J. Substantial Equivalence Information:

1.  **Predicate device name(s):**
Elecsys PreciControl Bone
Elecsys b Crosslaps® calcheck
IDS-iSYS 25-Hydroxy Vitamin D Control Set

2.  **Predicate 510(k) number(s):**
k051543
k000042
k091849

3.  **Comparison with predicate:**

Similarities and Differences

|  Characteristic | Candidate device:
IDS-iSYS CTX-I (Crosslaps) Control Set (k111650) | Predicate device:
Elecsys PreciControl Bone (k051543)  |
| --- | --- | --- |
|  Intended Use | Same | Elecsys PreciControl Bone is used for quality control of specified assays.  |
|  Levels | Same | 3 Levels  |
|  Matrix | Liquid, Phosphate bovine serum albumin | Lyophilized, based on equine serum  |
|  Stability and Storage | 2-8°C – unopened until expiration date
After opening: stable for 28 days at 4°C
On board the IDS-iSYS – 3 hours | 2-8°C – unopened until expiration date
Reconstituted/thawed: @ 20-25°C – 8 hours, @ 2-8°C – 5 days, @ -20°C (4 freeze thaw cycles possible) – 1 month  |
|  Analyte concentrations | CTX-I:
Low control: 0.2 ng/mL. | β-CTX (synthetic):
approximately 0.315, 0.75 and 3.0 ng/mL  |

{3}

|  Characteristic | Candidate device: IDS-iSYS CTX-I (Crosslaps) Calibrator Verifiers (k111650) | Predicate device: Elecsys b Crosslaps® calcheck (k000042)  |
| --- | --- | --- |
|  Intended Use | Same | For use in the verification of the calibration and assay range.  |
|  Levels | Levels 0, 1, 2,3 | Levels 1,2,3  |
|  Matrix | Liquid, phosphate bovine serum albumin | Crosslaps free human serum  |
|  Stability and Storage | 2-8°C – unopened until expiration date
On board stability: Single use-use then discard | Store unopened at 2-8°C.
Unopened stability at 2-8°C – until expiration date
Stability reconstituted: 4 hours at 20-25°C.  |
|  Analyte concentrations | CTX-I: 0.0 ng/mL, 0.6 ng/mL, 3.0 ng/mL, 5.0 ng/mL | β-CTX (synthetic)
0.5 ng/mL, 1.8 ng/mL, 3.8 ng/mL  |
|  Analyzer System | IDS-iSYS Multi-Discipline Automated Analyzer | Elecsys 1010 or 2010 immunoassay system  |
|  Characteristic | Candidate device: IDS-iSYS 25 Hydroxy Vitamin | Predicate device: IDS-iSYS 25-Hydroxy  |
| --- | --- | --- |

{4}

5

|   | D Calibrator Verifiers (k111650) | Vitamin D Control Set (k091849)  |
| --- | --- | --- |
|  Intended Use | Same | The IDS-iSYS 25-Hydroxy Vitamin D Control Set is intended for use as assayed quality control samples to monitor the accuracy and precision of the IDS-iSYS25-Hydroxy Vitamin D Assay.  |
|  Levels | Levels 0,1,2,3 | Levels 1, 2, 3  |
|  Matrix | Liquid, buffered equine serum | Buffered horse serum  |
|  Stability and Storage | 2-8°C – unopened until expiration date
On board stability: Single use-use then discard | 2-8°C – unopened until expiration date
On board the IDS-iSYS – 2.5 hours  |
|  Analyte concentrations | 25-Hydroxy Vitamin D: 0 ng/mL, 10 ng/mL, 70 ng/mL, and 120 ng/mL | 25-Hydroxy Vitamin D: 6.6ng/mL, 33.0ng/mL, and 72.5ng/mL  |
|  Analyzer System | Same | IDS-iSYS Multi-Discipline Automated Analyzer  |

K. Standard/Guidance Document Referenced (if applicable):

CEN 13640, Stability Testing of In Vitro Diagnostic Reagents.

L. Test Principle:

Not applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

{5}

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

## IDS-iSYS CTX-I (Crosslaps) Control Set

Traceability and Value Assignment:

Each control level is tested on three IDS-iSYS analyzers with a minimum of three runs for each cartridge batch tested in triplicate. Control solutions are prepared gravimetrically from an intermediate stock solution and concentrations are confirmed by spectrophotometric analysis. In the value assignment process, controls were calibrated using a master calibrator to generate a master curve. The assigned target value of each control level is defined as the mean of all the runs for the CTX-I assay and analyzer. The following are the expected values for each level of control: 0.2 ng/mL, 0.8 ng/mL, and 2 ng/mL.

Stability:

Closed vial:

For the real time closed vial testing, three lots of controls were stored at 4°C using cartridges with lids pierced by the IDS-iSYS then stored at 4°C. Each control material was tested in triplicate in 3 month intervals for up to a minimum of 15 months. Each control vial tested was compared back to reference control material stored at -20°C. Percent recoveries were within the sponsor's acceptance criteria of 10% of the reference material concentration and supports the sponsor's stability claim of 6 months when stored at 4°C.

Open Vial:

Open vial (in-use) stability of controls were performed at 4°C and tested against unopened vials of control material stored at 4°C. Controls were tested in duplicate at the timepoints stated in the stability protocol. Percent recoveries of each material were within the sponsor's acceptance criteria of 10% of the reference material concentration. Data supports the open vial stability claim of 28 days when stored at 4°C.

On-board stability studies were performed using two batches of controls using two IDS-iSYS instruments on the same day. Controls were tested at time 0, 3 hours, and 4.5 hours and compared to a reference material run at time 0. The on board stability data supports the sponsor's claimed on-board stability of 3 hours.

6

{6}

7

# IDS-iSYS CTX-I (Crosslaps) Calibrator Verifiers

Traceability and Value Assignment:

Four levels of calibrator verifiers were used to validate the calibration on the IDS-iSYS and validate the range of the analytical measurement. Each lot-specific value assignment was tested in three runs on at least three different IDS-iSYS analyzers in triplicate. The assigned target value of each calibrator verifier was defined as the mean of all the runs for each calibrator verifier. The following are the expected values for each calibrator verifier: 0.0 ng/mL, 0.6 ng/mL, 3.0 ng/mL, 5.0 ng/mL.

Stability:

For the real time closed vial testing, three lots of calibrators were stored at 4°C using cartridges with lids pierced by the IDS-iSYS then stored at 4°C. Each calibrator material was tested in triplicate in 3 month intervals for up to a minimum of 15 months. Each calibrator vial tested was compared back to reference control material stored at -20°C. Percent recoveries were within 10% of the reference material concentration and supports the sponsor's stability claim of 6 months when stored at 4°C.

The sponsor states that the IDS-iSYS CTX-I (Crosslaps) Calibrator Verifiers will be defined as "single use and then discard."

# IDS-iSYS 25 Hydroxy Vitamin D Calibrator Verifiers

Traceability and Value Assignment:

Four levels of calibrator verifiers were used to validate the calibration on the IDS-iSYS and validate the range of the analytical measurement. Master calibrators are prepared gravimetrically from stock solution and each concentration is traceable to spectrophotometric analysis. Each lot-specific value assignment was tested in three runs on at least three different IDS-iSYS analyzers in triplicate. The assigned target value of each calibrator verifier was defined as the mean of all the runs for each calibrator verifier. The following are the expected values for each calibrator verifier: 0, 10, 70, and 120 ng/mL.

Stability:

For real time closed vial testing, multiple sets of calibrators were stored at 4°C. Each calibrator was tested in quadruplicate using four separate 25 Hydroxy Vitamin D assays and the concentrations were compared back to reference material stored at -20°C. Percent recoveries were within the

{7}

sponsor's acceptance criteria of 10% of the reference material and support the sponsor's stability claim of 6 months.

The sponsor states that the IDS-iSYS Hydroxy Vitamin D Calibrator Verifiers will be defined as "single use and then discard."

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:

8

{8}

Not applicable

**N. Proposed Labeling:**

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

**O. Conclusion:**

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

9

---

**Source:** [https://fda-staging.innolitics.com/device/K111650](https://fda-staging.innolitics.com/device/K111650)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
