LANO-E

K111604 · Ug Global Resources Sdn. Bhd. · LYY · Jun 8, 2012 · General Hospital

Device Facts

Record IDK111604
Device NameLANO-E
ApplicantUg Global Resources Sdn. Bhd.
Product CodeLYY · General Hospital
Decision DateJun 8, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and the examiner.

Device Story

Disposable, non-sterile, powder-free latex examination glove; features inner coating of Lanolin and Vitamin E; green color; protein content labeled at <50ug/g. Used by healthcare professionals in clinical settings to provide a protective barrier between examiner and patient. Prevents cross-contamination during routine examinations. Device is donned manually by the user.

Clinical Evidence

Bench testing only. Evidence includes physical property testing (tensile strength, elongation), 1000ml water-tight testing (before and after aging), residual powder content testing, and biocompatibility testing. Protein content verified at <50ug/g.

Technological Characteristics

Material: Latex. Cuff: Beaded. Powder residue: Max 2mg/glove. Coating: Lanolin and Vitamin E. Compliance: ASTM D3578-10, ISO 2859-1:1999. Quality: ISO 9001:2000, ISO 13485. Dimensions: 240mm length, 0.08mm thickness. Tensile strength: 18.0 MPa (before aging), 14.0 MPa (after aging). Elongation: 650% (before aging), 500% (after aging).

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Suitable for general medical examination use.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## SUMMARY PREMARKET 510(k) NOTIFICATION Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim KIII604 JUN - 8 200 Submission Applicant - UG Global Resources Sdn Bhd 181 Floor No. 18 Jalan Dato' Abd Rahman 70000 Seremban Malaysia Official Correspondent - Kenneth J. Stanton 1565 Sunflower Avenue Costa Mesa, Ca 92626 Tel: (714)444-2248 Fax: (714)444-2271 Submitted: November 16, 2011 Description of the Device: Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim Trade Name: Lano-E Common Name: Examination Gloves A. Classification Name: Patient Examination Glove (per 21 CFR 880.6251) Class 1: Powder-Free Latex examination glove 80LYY that meets all of the requirements of ASTM D3578-10. Predicative Devices: Latex Powder-Free Examination Gloves Intended Use of the Device: A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and the examiner. Page - 9 {1}------------------------------------------------ SUMMARY PREMARKET 510(k) NOTIFICATION for - Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim 1011604 # Summary of Technological Characteristics: Material: Latex Cuff: Beaded Powder Residue: Maximum 2mg/glove In compliance with ASTM D3578-10.ISO 2859-1:1999. Quality Assurance: manufactured under ISO9001:2000 and ISO 13485 Inspection Parameters: | Criteria | Inspection Level | AQL | |-------------------------|------------------|-----| | Dimensions | S-2 | 4.0 | | Physical Properties | S-2 | 4.0 | | Water Tight Test 1000ml | G-1 | 1.5 | | Visual Major Defects | G-1 | 1.5 | | Visual Minor Defects | G-1 | 2.5 | # Physical Properties: | Dimensions: | | |-----------------|----------------------------------| | Overall Length: | 240 mm minimum | | Width: | 95 mm minimum (for medium glove) | | Thickness: | .08mm minimum | ## BEFORE AGING #### Tensile Strength: 18.0. Mpa minimum Ultimate Elongation:650% minimum # AFTER AGING 14.0 Mpa minimum 500% minimum # Special Properties: Lanolin and Vitamin E inner coating Packaging: 100 pcs per dispenser box, 10 boxes per case, 1,000 gloves per case Sizes: XS — XL Conclusion: The Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim meets the physical property requirements of ASTM D 3578-10, the FDA 1000 ml water test both before and after aging, and the Protein Labeling Claim Level at <50ug/g. This product is as safe, as effective, and performs as well or better than the legally marketed 510 #K013163. It has been supported by results of Biocompatibility Tests, Residual Powder Contents tests, Physical Property Tests and the 1000ml Water Tight Test. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized eagle or bird-like symbol with three curved lines representing its wings or body. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUN - 8 2012 UG Global Resources, SDN BHD % Mr. Kenneth J. Stanton President. UG Healthcare (USA) Incorporated 1565 Sunflower Ave. Costa Mesa, California 92626 Re: K11604 Trade/Device Name: Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: May 29, 2012 Received: May 29, 2012 Dear Mr. Stanton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. . . . . . . . . . {3}------------------------------------------------ ### Dear Mr. Stanton Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerelv vours. Fo Anthony D. Watson. B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {4}------------------------------------------------ #### 3.0 Indications for Use: 510K 111604 Device Name: Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim Indications for Use - The Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim (50 micrograms or less total water extractable protein per gram of glove), is a disposable device intended for Medical Purposes that is worn on the examiners hand to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) and/or Over-The-Counter Use (Part 21 CFR 801 Subpart C) (Please do not write below this line-continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) E.L.L. F. Currie Well Page - 5 (Division Sign-Off) (Division of Anesthesiology, General Hospital nfection Control, Dental Devices 510(k) Number: K 111 60.4
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...