K111557 · Seven Seas Distribtion and Manufacturing, LLC · NYN · Nov 10, 2011 · Neurology
Device Facts
Record ID
K111557
Device Name
X-FORCE
Applicant
Seven Seas Distribtion and Manufacturing, LLC
Product Code
NYN · Neurology
Decision Date
Nov 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The X-Force external, non-invasive, non-narcotic, electrotherapy system is indicated for use as an adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee. The X-Force is also indicated for use as an adjunctive therapy in reducing the level of pain, and stiffness associated with pain, from rheumatoid arthritis of the Hand. These devices are to be used or sold only under the direct supervision or order of a licensed practitioner. A prescription is required to obtain this product can be used in the home or clinic by all patients in need.
Device Story
X-Force is a microcomputer-controlled digital TENS device for electrotherapy. Input: user-selected treatment parameters via digital interface. Transformation: device generates pulsed monophasic (NT/HiVolt) electrical signals delivered through lead wires and electrodes to the treatment site. Output: electrical stimulation to reduce pain/stiffness. Used in clinics or homes under practitioner supervision. Features include dual-channel output, digital LCD, parameter locking, and automatic safety trips (overload, no-load, shut-off). Healthcare providers use the device to provide adjunctive therapy for osteoarthritis of the knee or rheumatoid arthritis of the hand. Patient benefits from non-narcotic, non-invasive pain management.
Clinical Evidence
No clinical data. Substantial equivalence is based on technological comparison and bench testing of electrical characteristics (leakage current, output parameters) against the predicate device.
Technological Characteristics
Microcomputer-controlled TENS device. ABS resin housing. Dual-channel output. Waveform: NT (HiVolt) pulsed monophasic. Power: 4x 1.2V rechargeable AA batteries. Connectivity: wired electrodes (21 CFR 898 compliant). Safety: automatic overload/no-load/shut-off, patient override, parameter locking. Dimensions: 142x62x26 mm. Weight: 365g with battery.
Indications for Use
Indicated for patients suffering from pain and symptoms of osteoarthritis of the knee or pain and stiffness associated with rheumatoid arthritis of the hand. Requires prescription; for use in home or clinic settings.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
NOV 10 2011 K///55 7
# ANNEXURE `IV′
# SUMMARY
·
# "Traditional 510(k) summary"
| SUBMITTER'S NAME | : | Seven Seas Distribution and Manufacturing LLC |
|----------------------------|---|-----------------------------------------------|
| ADDRESS | : | 2620 S Maryland Pkwy Ste 14 Unit 835 |
| | | Las Vegas, Nevada, 89109, United States |
| TELEPHONE | : | 213-6132356 |
| FAX | : | 213-6132344 |
| E-MAIL ADDRESSS | : | sales@sevenseasdm.com |
| CONTACT PERSON | : | Mark Ioele |
| TELEPHONE | : | 213-6132356 |
| FAX | : | 213-6132344 |
| Date of summary submission | : | 25th April 2011 |
| Resubmitting on | : | 12th October 2011 |
For Seven Seas Distribution and Manufacturing LLC Mark Ioele
MPL
{1}------------------------------------------------
#### ANNEXURE `IV′
## Traditional 510(k)
#### NEW DEVICE FOR WHICH SUBMITTING
| Trade Name | : X-Force |
|---------------------|-----------------------------------------------------------------------------------------------------|
| Common Name | : TENS (Transcutaneous Electrical Nerve Stimulator) |
| Classification Name | : Transcutaneous electrical nerve stimulator for pain<br>relief (21 CFR 882.5890, Product Code NYN) |
Device's Classification Panel : Neurology
## LEGALLY MARKETING DEVICE
J-Stim 1000 : Transcutaneous Electrical Nerve Stimulator
Manufacturer : Pain Management Technologies, Inc.
Address : 1340 Home Ave. , Bldg A, Akron, OH 44310
### DESCRIPTION OF NEW DEVICE X-Force
The X-FORCE device is a microcomputer controlled digital device featuring a NT Mode output. The device provides simple programming facility to customize the treatment. The panel and its keys provide easy selection for treatment. It is device that deploys the use of electrotherapy and compression to treat Osteoarthritis (OA) and rheumatoid arthritis of the knee or hand.
The X-Force consists of electrodes, the lead wires, and the signal generator. The device is portable and rechargeable battery operated. The lead wires connect the electrodes to the device. These electrodes complete an electrical circuit allowing current to flow. The stimulator produces a pulsed electrical signal through the lead wires and electrodes at the treatment site.
#### Device Features:
- l Digital LCD display
- . Dual channel
- . Soft fabric comfortable garment
- . Low battery indicator
- l ABS Portable handy device
- 1 Parameter locking
{2}------------------------------------------------
## ACCESSORIES:
X-Force comes complete with all the necessary accessories and below is a list of items that are included:
Units are supplied with:
- . Electrode Cable (2 Pin 2 Core); Length: 1.25 meter conforming to 21 CFR 898
- . Knee Belt listed in the 510(k) - K073386 of Jstim 1000
- . Elbow Belt and Conductive Patches listed in 510(k) - K943009
- . Conductive Gel
- . Battery Rechargeable AA size 1.2 V (4 in qty.)
- . External Battery Charging Adaptor Adaptor Specifications
- o INPUT: 100-240V AC-50/60Hz
- OUTPUT: 5.5V DC O
- . Instruction Manual
## INTENDED USE OF NEW DEVICE "X-FORCE"
The X-Force external, non-invasive, non-narcotic, electrotherapy system is indicated for use as an adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee.
The X-Force is also indicated for use as an adjunctive therapy in reducing the level of pain, and stiffness associated with pain, from rheumatoid arthritis of the Hand.
## INTENDED USE OF PREDICATE DEVICE "J-STIM 1000"
The J-Stim 1000 external, non-invasive, non-narcotic, electrotherapy system is indicated for use as an adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee.
The J-Stim 1000 is also indicated for use as an adjunctive therapy in reducing the level of pain, and stiffness associated with pain, from rheumatoid arthritis of the Hand.
{3}------------------------------------------------
# ANNEXURE 'IV'
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| S.<br>No | Description | New Device<br>X-Force | Predicate Device<br>JSTIM (K073386) | Remarks |
|----------|------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------|
| 1 | Energy Source | Rechargeable battery (4 X 1.2V)operated, more battery backup compare to Predicate device | 9V Alakaline battery operated | Similar |
| 2 | Device Housing | Material - ABS<br>Size - (142 X 62 X 26) mm<br>12.875 ounce (365 Grams) with battery,<br>8.924 ounce (253 Grams) without battery | Material - ABS<br>Size - (45 x 30 x 15) mm<br>2.8 ounce (65 Grams) with battery, 1.6 ounce (45.8 Grams) without battery | Similar |
| 3 | Number of Output<br>Channels | Two Channels | Two Channels | Same |
| 4 | Number of Modes | One | One | Same |
| 6 | Display | Customized LCD | Customized LCD | Similar |
| 7 | Waveform | NT (HiVolt) - Pulsed Monophasic | NT (HiVolt) - Pulsed Monophasic | Same |
| 7 | Treatment Timer | YES | YES | Same |
| 8 | Target of<br>populaces | Therapist in Neurology at health clinic and Chiropractor. | Therapist in Neurology at health clinic and Chiropractor. | Same |
:
.
{4}------------------------------------------------
# COMPARISON OF BASIC UNIT CHARACTERISTICS:
| S.N | FEATURES OF DEVICE | NEW DEVICE | PREDICATE DEVICE |
|-----|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 510(K) Number | K111557 | K073386 |
| 2 | Device Name, Model | X Force | JStim 1000 |
| 3 | Manufacturer | Seven Seas Distribution and<br>Manufacturing LLC | Pain Management<br>Technologies, Inc. |
| 4 | Power source | Rechargeable Battery(4 X 1.2V<br>- AA size) | 9V Li ion battery |
| 5 | Method of Line Current<br>Isolation | (a) Since Instruments<br>operates on battery, hence it<br>is completely isolated from<br>Mains Supply. | (a) Since Instruments<br>operates on battery, hence it<br>is completely isolated from<br>Mains Supply. |
| 6 | Patient Leakage Current<br>Normal condition<br>Single Fault condition | (b) Patient Leakage Current<br>Normal condition = less than<br>100μΑ*<br>Single fault condition = less<br>than 300µA*<br>* Reference: IEC 60601-1 | (b) Patient Leakage Current<br>Normal condition = less than<br>100μΑ*<br>Single fault condition = less<br>than 300µA*<br>* Reference: IEC 60601-1 |
| 7 | Number of Output Modes | 1 | 1 |
| 8 | Number of Output<br>Channels<br>Synchronous or Alternating? | Synchronous<br>Channel 1 and 2 are<br>completely isolated. Both have<br>designated Amplifier & control<br>circuit, Only power supply and<br>ground are common. | Synchronous<br>Channel 1 and 2 are<br>completely isolated. Both<br>have designated Amplifier &<br>control circuit Only power<br>supply and ground are<br>common |
| 9 | Regulated Current or<br>Regulated Voltage | Regulated Voltage | Regulated Voltage |
| 10 | Maximum Output Current | Iavg =0.2mA @ 500Ω | Iavg =0.2mA @ 500Ω |
| 11 | Maximum Output Voltage | Vavg. =0.108Vavg @ 500Ω | Vavg. =0.108Vavg @ 500Ω |
| 12 | Maximum Phase Charge | 24.3 μC @ 500Ω | 24.3 μC @ 500Ω |
| 13 | Maximum Current Density | 0.0021 mA/cm² @ 500Ω | 0.0021 mA/cm² @ 500Ω |
| 14 | Maximum Power Density | 0.00022 mW/cm² @ 500Ω | 0.00022 mW/cm² @ 500Ω |
| 15 | Software/Firmware/<br>Microprocessors Controls? | YES | YES |
| 16 | Automatic Overload Trip? | YES | YES |
| 17 | Automatic No-Load Trip | YES | YES |
| 18 | Automatic Shut off? | YES | YES |
| 19 | Patient Override Control? | YES | YES |
| 20 | Indicator Display:<br>On/Off Status?<br>Low Battery?<br>Voltage/Current Level? | YES<br>YES<br>YES | YES<br>YES<br>YES |
| 21 | Timer | YES | YES |
| 22 | Compliance with Voluntary<br>Standards | NO | NO |
| 23 | Compliance With 21 CFR<br>898 | YES, the electrode cable can<br>never be plugged in the AC | YES, the electrode cable can<br>never be plugged in the AC |
| 24 | Weight | 12.875 ounce (365 Grams) with battery, 8.924 ounce (253 Grams) without battery | 2.8 ounce (65 Grams) with battery, 1.6 ounce (45.8 Grams) without battery |
| 25 | Dimension (L X B X H) | (142 X 62 X 26) mm | (45 x 30 x 15) mm |
| 26 | Housing Material | ABS resine | ABS resine |
{5}------------------------------------------------
#### ANNEXURE 'IV'
# SUMMARY:
XForce functions in the same way as the predicate device Jstim 1000. The intended
use, operational characteristics of both the devices are similar.
| Indication Comparison: | Indented use of X Force and Predicate Device are same i.e.<br>X Force is indicated for use as an adjective<br>Therapy in reducing the level of pain and symptoms<br>associated with Osteoarthritis of the knee and<br>Rheumatoid Arthritis of the Hand. |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Technological Comparison: The technological characteristics (Energy Source, Design,<br>Material and Physical Characteristics) of modified device<br>has been compared with Jstim 1000 and found to be<br>same as the predicate device Jstim 1000.<br>Below are the differences: |
| 1) | Power source: Jstim 1000 has power source of 9Volt<br>where as X-Force have Power source of 1.2 DC X 4<br>Battery) which are more power full compare to 9VDC<br>battery thus giving more battery backup. This change<br>only increase battery backup of Unit and there is no<br>effect on the safety or effectiveness of the device. |
| 2) | Program Lock feature: X-Force have Program Lock<br>feature which is not present in Jstim 1000. This feature<br>give user to lock the used parameter, this function<br>doesn't affect any parameter which can make New device<br>different from Predicate |
| 3) | Digital Control of Output Power: Jstim 1000 output<br>power is controlled through Analog pots where in X-Force<br>Output power is controlled digitally through<br>Microcontroller. |
| Labeling Comparison | : Labeling of the device compares to that of predicate<br>device. |
| Non-Clinical Testing | : Not Applicable as Jstim 1000 is already FDA Approved<br>and the New Device X-Force is same as Predicate Device |
| Clinical Testing | : Not Applicable as Jstim 1000 is already FDA Approved<br>and the New Device X-Force is same as Predicate Device |
| Conclusion | : Drawn from the comparison between the modified device<br>X-Force with predicate device J Stim 1000, it<br>demonstrates that X-Force is as safe, as effective and<br>performs as well as its predicate device. |
ﺎﺭ ﺗﻮ
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird-like figure with three wing-like shapes extending from its body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Seven Seas Distribution and Manufacturing, LLC c/o Mr. Mark Ioele President 2620 S. Maryland Pkwy, Ste. 14 Unit 835 Las Vegas, Nevada 89109
NOV 1 0 2011
Re: K111557
Trade/Device Name: X Force Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NYN Dated: October 12, 2011 Received: October 27, 2011
Dear Mr. Ioele:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act., The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Mark loele
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
![]()
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices 、 Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a black and white drawing of a person running. The person is in a sprinting position, with one leg forward and the other leg back. The arms are bent at the elbows, and the hands are clenched into fists. The person's head is turned to the side, and the face is not visible. The background is white.
SEVEN SEAS DISTRIBUTION ANI) MANUFACTURING LLC
# INDICATIONS FOR USE
510(K) No : K111557
Device Name : X Force
Indications For Use:
The X-Force external, non-invasive, non-narcotic, clectrotherapy system is indicated for use as an adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee.
The X-Force is also indicated for use as an adjunctive therapy in reducing the level of pain, and stiffess associated with pain, from rheumatoid arthritis of the Hand.
These devices are to be used or sold only under the direct supervision or order of a licensed practitioner. A prescription is required to obtain this product can be used in the home or clinic by all patients in need.
Prescription Use __ V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ERIC NELSON
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K111557
2620 S. Maryland Parkway suite 14, Unit 835 Las Vegas, Nevada 89109 800.571.1846 Tel 702.974.1651
Page 1 of __ 1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.