WIRELESS/WIRED FDR D-EVO FLAT PANCEL DETECTOR SYSTEM MODEL DR-ID600 W/DR-ID611SE
K111548 · Fujifilm Medical System U.S.A., Inc. · MQB · Aug 30, 2011 · Radiology
Device Facts
Record ID
K111548
Device Name
WIRELESS/WIRED FDR D-EVO FLAT PANCEL DETECTOR SYSTEM MODEL DR-ID600 W/DR-ID611SE
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
MQB · Radiology
Decision Date
Aug 30, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The Wireless/Wired FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications wherever conventional film/screen or CR systems may be used. The FDR D-EVO is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Device Story
Portable digital flat panel detector system; acquires and digitizes x-ray exposures from standard radiographic systems. Input: x-ray radiation; Output: digital radiographic images for display. Uses Irradiated Side Sampling (ISS) design; Cesium Iodide (CsI) scintillator. Used in clinical environments; placed in wall bucky, table bucky, or used for non-grid exams. Operated by radiology staff. Output viewed by radiologists for diagnostic assessment; facilitates clinical decision-making by providing high-quality digital images equivalent to or better than predicate GOS-based detectors.
Clinical Evidence
Bench testing completed per IEC requirements and FDA guidance. Clinical image quality reader study performed on 30 image pairs comparing proposed CsI detector to predicate GOS detector. Three board-certified radiologists reviewed images; 100% deemed diagnostic. Consensus results indicated the proposed device provided equivalent or better diagnostic capability in 93.3% of cases.
Technological Characteristics
Portable 14x17" flat panel detector; Cesium Iodide (CsI) scintillator; Irradiated Side Sampling (ISS) design. Wireless and wired connectivity. Complies with IEC testing standards. Solid state x-ray imager.
Indications for Use
Indicated for general projection radiographic imaging of human anatomy in clinical settings where conventional film/screen or CR systems are used. Not indicated for mammography, fluoroscopy, tomography, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Wireless/Wired FDR D-EVO flat panel detector system (DR-ID600 w/DR-ID601SE) (K103596)
{0}------------------------------------------------
# KII1548
Image /page/0/Picture/1 description: The image shows the Fujifilm logo. The logo is in black and white and features the word "FUJIFILM" in large, bold letters. Below the main logo, there is the text "Fujifilm Medical Systems, USA" in a smaller font. The logo appears to be a company logo.
AUG 300 2011
# 510(k) Summary Wireless/Wired FDR D-EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE)
### Date: June 1, 2011
### Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, CT, 06902, USA
## Contact Person:
| Name: | Katherine Y. Choi, RAC |
|------------|-------------------------------|
| Title: | Regulatory Affairs Specialist |
| Telephone: | (203) 602-3568 |
| Facsimile: | (203) 363-3950 |
## Identification of the Proposed Device:
Proprietary/Trade Name:
Classification Name: Regulations Number: Product Codes: Device Class: Review Panel: Common Name:
Wireless/Wired FDR D-EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE) Solid State X-ray Imager (Flat Panel/Digital Imager) 21 CFR 892.1650 90 MQB Class II Radiology Flat Panel Digital Detector
#### 1. INDICATIONS FOR USE
The Wireless/Wired FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications wherever conventional film/screen or CR systems may be used. The FDR D-EVO is not intended for mammography, fluoroscopy, tomography, and angiography applications.
#### ll. DEVICE DESCRIPTION
The proposed device, DR-ID600 w/DR-ID611SE is a modified version of our currentlycleared predicate device, Wired FDR D-EVO flat panel detector system (DR-ID600 w/DR-ID601SE), K103596. The predicate device includes the 14x17" fat panel detector model DR-ID601SE using GOS (Gadolinium Oxysulfide) scintillator whereas the proposed device introduces the 14x17" flat panel detector model DR-ID611SE using a different scintillator material, Csl (Cesium lodide).
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Fujifilm Medical Systems, USA. The word "FUJIFILM" is written in large, bold, black letters. Below the main logo, the words "Fujifilm Medical Systems, USA" are written in a smaller, sans-serif font, also in black.
FUJIFILM's unique Irradiated Side Sampling (ISS) design delivering high image quality, and wireless communication specifications remains unchanged in the proposed device.
The proposed DR-ID600 w/DR-ID611SE is a portable digital detector system that acquires and digitizes x-ray exposures from standard radiographic systems, which is the same as our legally marketed predicate device, DR-ID600 w/DR-ID601SE. The proposed device is designed to be used in any environment that would typically use a radiographic cassette. It can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid exams. All of which are also the same as our legally marketed predicate device.
#### 1111. SUMMARY OF STUDIES
The proposed device, DR-ID600 w/DR-ID611SE successfully completed internal and international IEC testing requirements. In addition, the FDA Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, issued on August 6, 1999 was followed to describe the detector characteristics.
FUJIFILM performed an image quality reader study on 30 image pairs comparing the images obtained between the proposed CsI detector (DR-ID611SE) and our legally marketed GOS detector (DR-ID601SE). Three board-certified radiologists reviewed the images and all images were deemed to be of diagnostic capability. In addition, the consensus results of the three radiologists shows that the proposed device provides images of equivalent or better diagnostic capability to those of our cleared predicate device (K103596) by 93.3%. Test protocol summary and analysis results are enclosed in the submission.
#### IV. SUBSTANTIAL EQUIVALENCE
The proposed device, DR-ID600 w/DR-ID611SE is substantially equivalent to the following legally marketed devices.
| Legally Marketed Device | 510(k)# |
|-----------------------------------------------------------------------------|---------|
| Wireless/Wired FDR D-EVO flat panel detector system (DR-ID600 w/DR-ID601SE) | K103596 |
| Canon CXDI-70C Wireless | K102012 |
The proposed DR-ID600 w/DR-ID611SE has the same Indications for Use and very similar functional and technical requirements as our currently-cleared predicate device, K103596. The proposed device's wireless communication specifications remain unchanged as K103596. In addition, the scintillator material used in the proposed device is the same material as the legally marketed device, K102012.
#### V. CONCLUSION
The proposed DR-ID600 w/DR-ID611SE is substantially equivalent to the legally marketed devices and conforms to applicable medical device safety standards.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings outstretched, symbolizing care and protection. The eagle is positioned within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Katherine Y. Choi, RAC Regulatory Affairs Specialist Fuji Medical Systems U.S.A., Inc. 419 West Avenue STAMFORD CT 06902
AUG 23 2013
Re: K111548
Trade/Device Name: Wireless/Wired FDR D-EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE)
Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: July 26, 2011 Received: July 27, 2011
Dear Ms. Choi:
This letter corrects our substantially equivalent letter of August 30, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I va may , and controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirents as set form in the quality of evice as described in your Section 510(k) premarket whit anow you to begin marketing your arrance of your device to a legally marketed nouncation. The FDA imania of successification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFF Parts 801 and If you desire specific advice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-007), product the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
under the MDK regulation (27 CFR Pat 003), products go the fault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tour may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): KIII548
## Device Name: Wired FDR D-EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE)
Indications for Use:
The Wireless:Wired FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human analomy. It is intended for use in general projection radiographic applications wherever conventional film /screen or CR systems may be used. The FDR D-EVO is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device (OIVD)
Michael O'Kane
7
(Division Sign-On)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K 111548
Page 39 of 711
r.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.