K111535 · Exact Medical Manufacturing, Inc. · FYA · Aug 26, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111535
Device Name
EMM SURGICAL GOWN - SPUNLACE WITH PE LAYER
Applicant
Exact Medical Manufacturing, Inc.
Product Code
FYA · General, Plastic Surgery
Decision Date
Aug 26, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Exact Medical Manufacturing Surgical Gown - Spunlace w/PE Layer are sterile or non-sterile single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The Exact Medical Manufacturing Surgical Gown - Spunlace w/PE Layer are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide sterilization.
Device Story
Single-use, disposable surgical gown; used in operating room by staff as protective barrier against microorganisms, body fluids, and particulates. Comprised of nonwoven spunlace fabric with polyethylene (PE) layer, polyester cuffs, and Velcro neck closure. Provides physical barrier to reduce cross-contamination between patient and surgical staff. Sold as sterile or non-sterile bulk items for further processing/sterilization by repackagers.
Clinical Evidence
Bench testing only. Testing included biocompatibility (cytotoxicity, irritation, sensitization per ISO 10993), barrier properties, tensile strength, tear strength, flammability, linting, and sterility validation per AAMI 11135-1:2007.
Indicated for operating room personnel during surgical procedures to protect patient and personnel from transfer of microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for EMM Exact Medical Manufacturing, Inc. The logo consists of a geometric design on the left and the text "EMM Exact Medical Manufacturing, Inc." on the right. The text is in a simple, sans-serif font, with "EMM" in a larger, bolder font than the rest of the text. The geometric design is made up of small squares.
### Sec. 6: 510(k) Summary – Surgical Gown-Spunlace w/PE Layer, rev. 072511
#### 510(k) Summary for Exact Medical Manufacturing Inc., Surgical Gowns – Spunlace w/PE Layer
T
100
1
| Date Summary was Prepared | June 1, 2011 |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter | David Nowicki, President<br>Exact Medical Manufacturing Inc.<br>4917 William Street<br>Lancaster, NY 14043<br>dnowicki@exactmm.com<br>(p)716-681-0866, (f) 716-681-4110 |
| Primary Contact for this 510(k)<br>Submission | David Nowicki, President<br>Exact Medical Manufacturing Inc.<br>4917 William Street<br>Lancaster, NY 14043<br>dnowicki@exactmm.com<br>(p)716-681-0866, (f) 716-681-4110 |
| Device Common Name | Surgical Gown |
| Trade Name | Surgical Gown - Spunlace w/PE Layer |
| Device Product Codes and<br>Classification Name | FYA, 21CFR878.4040, Surgical Apparel, Class II |
| Predicate Device | Medline (Proxima) Surgical Gowns 510(k)964142 |
| Device Description | Exact Medical Manufacturing Surgical Gown-SPL are single use, disposable surgical gowns<br>using in the OR as a protective covering, for the operating room staff, from the transfer of<br>microorganisms, body fluids and particulates. Exact Medical Manufacturing Surgical Gowns-<br>Spunlace w/PE Layer are comprised of disposable nonwoven (Spunlace), 100% polyester<br>cuffs, Velcro neck closure, reinforced areas fabric or polyethylene |
| Intended Use | Device Name: Exact Medical Manufacturing Surgical Gowns-Spunlace w/PE Layer Model Numbers 18-<br>201L, 18-201-XL, 18-201 XXL and 18-201 XLXL Indications for Use: Exact Medical Manufacturing<br>Surgical Gown - Spunlace w/PE Layer are sterile or non-sterile single use devices that are intended to be<br>worn by operating room personnel during surgical procedures to protect both the surgical patient and the<br>operating room personnel from transfer of microorganisms, body fluids, and particulate material. |
| | The Exact Medical Manufacturing Surgical Gown - Spunlace w/PE Layer are also sold as bulk non-<br>sterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide<br>sterilization. |
| | Sterilization method: ETO (ethylene oxide)<br>Standard: AAMI 11135-1:2007, Sterilization of Healthcare Products<br>Pre-Conditioning: Temperature C o = 42<br>RH% = 60<br>Dwell Time Hours = 6<br>EO Cycle: Temperature C o = 54<br>RH% =/> 30<br>EO Concentration = 730mg/m3<br>Dwell Time Hours = 9<br>Aeration: Temperature = ambient<br>Time =/> 72 hours<br>EO Residuals; EO=4mg max., ECH=9mg. max. |
| Technological Characteristics | Exact Medical Manufacturing Surgical Gowns-Spunlace w/PE Layer have the same design,<br>material and performance characteristics of the predicate device. |
| Summary of Testing | Exact Medical Manufacturing Surgical Gowns-Spunlace w/PE Layer is substantially equivalent<br>and meets the same acceptance criteria as the predicate device/gown in K964142. Non-<br>clinical performance testing includes:<br>Biocompatibility (cytotoxicity, irritation, sensitization) in compliance with the methods of ISO<br>10993, Barrier properties, tensile, tear strength, flammability, linting and sterility. All results of<br>the testing met acceptance criteria. |
| Substantial Equivalence<br>Conclusion | The surgical gowns described in this 510(k) submission are substantially equivalent in all<br>specifications and performance compared to the predicate device indentified in K964142 and<br>do not present any differences that could raise concerns related safety and effectiveness. |
4917 William Street - Suite C ・ Lancaster, NY 14086 ・ (716) 681-0866
··
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Exact Medical Manufacturing, Inc. % Mr. Robert O. Dean President Compliance Systems International, LLC. 1083 Delaware Avenue Buffalo, NY 14209
AUG 26 2011
Re: K111535
RT11955
Trade/Device Name: EMM Surgical Gowns-Spunlace w/PE Layer ade/Device Name: Linner 2011 XL, 18-201 XXL, and 18-201 XLXL Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA Dated: July 28, 2011 Received: August 1, 2011
Dear Mr. Dean:
We have reviewed your Section 510(k) premarket notification of intent to market the device We nave reviewed your Section 910(x) promised is substantially equivalent (for the referenced above and have determined the accession marketed predicate devices marketed in
indications for use stated in the enclosure) to legally marketed in of the Medical D indications for use stated in the cholosation to readers of the Medical Device interstate commerce prior to Niay 20, 1 70, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 1 Amendments, or to uevices that have been to the do not require approval of a prematice the Federal Food, Drug, and Cosment Act (710) that sket the device, subject to the general
approval application (PMA). You may, therefore, market the Aet include approval application (1 MA). Tou may, more of controls provisions of the Act include confors provisions of the Act. "The genting of devices, good manufacturing practice, requirements for annual registration, insting and adulteration. Please note: CDRH does labeling, and promotions against insoftanting warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III If your device is classifica (see above) into child in the regulations affecting your (PMA), It may be subject to adultions, Title 21, Parts 800 to 898. In the Feder device can be found in the Code of Poderal Regulations, wour device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA s Issuance of a suscises with other requirements
mean that FDA has made a determination that your devices with other Federal agencies. mean that FDA has made a delemination that your administered by other Federal agencies.
of the Act or any Federal statutes and regulations a byt not limited to: registration of the Act or any Federal statues and reguirements, including, but not limited to: registration
You must comply with all the Act's requirements, and not device reporting You must comply with an the Act s requirements, most 801); medical device reporting and listing (21 CFK Part 807), laoching (21 CFR 803); good manufacturing
(reporting of medical device-related adverse events) (21 CFR 803); goverlight (21 CFR Part 82 (reporting of medical device-lefated adverse verse) (D) regulation (21 CFR Part 820);
practice requirements as set forth in the quality systems (QS) regulations 531-542 of practice requirements as set form in the quality systems (@) . . .
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific advice for your actives of these Offices/CDRH/CDRHOffices
please go to http://www.fda.gov/AboutFDA/Centers@ffices/CDRH's) ( please go to http://www.ida.gov/A00an Drives and Radiological Health's (CDRH's) Office of
/ucm115809.htm for the Center for Devices and Radios by reference to /ucm113809.mill for the Center 101 DOYlots a mixtled, "Misbranding by reference to Compliance. "Also, please note the regardions regarding the reporting the reporting of
premarket notification" (21CFR Part 807.97). For questions regarding of premarket nothlication (21CFR Part 800), (21 CFR Part 803), please go to adverse events under the MDR regulation (21 Or KaProblem/default.htm for the CDRH's
http://www.fda.gov/MedicalDevices/Safety/Reportaliance http://www.lua.gov/Medical/Devision of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may obtain other general intonnational on 9 and Consumer Assistance at its toll-free
Division of Small Manufacturers, International and Consumer Assistance at its toll-fr DIVIsion of Sinan Manufacturers. 11, 2019 796-7100 or at its Internet address
number (800) 638-2041 or (301) 796-7100 or at its Internet start defentlation number (800) 038-2041 01 (2017) 790 1700 views for You/Industry/default.htm.
incerely yours,
F. Swart-Tuorinen
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Sec. 7: Intended Use, rev.072511
# Indications for Use Form
#### Indications for Use:
510(k) Number (if known): K111535
Device Name: Exact Medical Manufacturing Surgical Gowns-Spunlace w/PE Layer Model Numbers 18-201L, 18-201-XL, 18-201 XXL and 18-201 XLXL Indications for Use: Exact Medical Manufacturing Surgical Gown - Spunlace w/PE Layer are sterile single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
The Exact Medical Manufacturing Surgical Gown - Sounlace w/PE Laver are also sold as built nonsterile, single use items, to repackagerrelabeler establishments for further packaging and ethylene oxide sterilization.
Sterilization method: ETO (ethylene oxide) Standard: AAMI 11135-1:2007, Sterilization of Healthcare Products Pre-Conditioning: Temperature C = 42 . -RH% = 60 Dwell Time Hours = 6 EO Cycle: Temperature C° = 54 RH% =/> 30 EO Concentration = 730mg/m3 Dwell Time Hours = 9 Aeration: Temperature = ambient Time =/> 72 hours EO Residuals; EO=4mg max., ECH=9mg. max.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Susa Pease
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111535
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.