Multi-Ethnic Study of Atherosclerosis (MESA) cohort
The MESA cohort data was used to define the expected values and reference ranges for LDL-P concentrations in a diverse population.
Reference range; Population-based cohort; MESA
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Multi-Ethnic Study of Atherosclerosis (MESA); Prospective population-based cohort study
Multi-ethnic population; Sample Size: 5,362
Not applicable for this study
LDL-P concentration distribution
Indications for Use
The NMR LipoProfile® test, when used with the NMR Profiler, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease. This test is performed and provided as a service by LipoScience Laboratory.
Device Story
The NMR LipoProfile Test utilizes a 400 MHz proton NMR spectrometer to analyze human serum or plasma samples. The system measures the composite NMR signal at approximately 0.8 ppm, which corresponds to methyl group protons of lipids in LDL, HDL, and VLDL subclasses. An automated analysis server running Gen2.1 software performs spectral deconvolution to isolate unique frequencies and line shapes for each lipoprotein subclass. These signal amplitudes are converted into particle concentrations (LDL-P) and cholesterol/triglyceride concentrations. The device is operated as a centralized laboratory service by LipoScience Laboratory. Results are provided to clinicians to assist in the management of cardiovascular disease risk. The modular software architecture allows for independent updates to analyte-specific algorithms without impacting signal generation or other modules.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A2; LDL-P CVs ranged 3.7-6.9%. Linearity confirmed across 300-3500 nmol/L (LDL-P), 7-140 mg/dL (HDL-C), and 5-1100 mg/dL (Triglycerides). Method comparison against predicate (n=1555-1600) showed high correlation (r=0.97-1.00) and bias <5% at medical decision limits. Interference testing screened 27 substances. Reference ranges bridged from MESA study (n=5,362) using CLSI C28-A2.
Indicated for patients requiring management of lipoprotein disorders associated with cardiovascular disease. Measures LDL-P, HDL-C, and triglycerides in human serum and plasma.
Regulatory Classification
Identification
A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
Predicate Devices
NMR Profiler and NMR Lipoprofile Assay (k063841)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k111516
B. Purpose for Submission:
Modification of a previously cleared device to change software architecture for NMR LipoProfile Test (LDL-P, HDL-C and triglycerides) and NMR curve fit algorithm for LDL-P
C. Measurand:
LDL-P (low density lipoprotein particle number), HDL cholesterol, triglycerides
D. Type of Test:
Nuclear magnetic resonance (NMR)
E. Applicant:
LipoScience Inc.
F. Proprietary and Established Names:
NMR LipoProfile Test
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MRR | I, meets limitations per 21 CFR 862.9 (c)(4) | 21 CFR §862.1475 – Lipoprotein test system | Chemistry (75) |
| LBS | I, meets limitations per 21 CFR 862.9 (c)(4) | 21 CFR §862.1175 – Cholesterol test system | Chemistry (75) |
| CDT | I, meets limitations per 21 CFR 862.9 (c)(4) | 21 CFR §862.1705 – Triglyceride test system | Chemistry (75) |
H. Intended Use:
1. Intended use(s):
See Indications for use below.
{1}
2. Indication(s) for use:
The NMR LipoProfile® test, when used with the NMR Profiler, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease. This test is performed and provided as a service by LipoScience Laboratory.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
This device is cleared as a service (in-house). Samples are sent to Liposcience Laboratories to be run on the NMR Profiler, using software version Gen2.1.
I. Device Description:
The test system includes a 400 MHz proton NMR spectrometer interfaced with a commercial sample handler, an analysis server containing the software to analyze digitized spectral data, and the following reagents:
- NMR Diluent - aqueous solution containing Na₂EDTA (5.0 mM), CaCl₂ (1.0 mM), KCL (120 mM), Na₂HPO₄ (50 mM), NaN₃ (0.02%), pH 7.4
- NMR Wash - Triton X-100-0.1% v/v, Liqui Nox 0.1% v/v in deionized water, pH 10.0
- NMR Calibrator - aqueous solution of Trimethyl Acetate (TMA) disodium salt (15.0 mM) containing Na₂EDTA (5.0 mM), CaCl₂ (3.0 mM), KCl (120 nM), D₂O 10% v/v
- NMR LipoProfile Quality Control material - two levels of pooled human serum-based control material, (Control A and Control B), with pre-determined target ranges.
The control materials contain human source material. Each donor unit is tested by FDA – approved methods and found non-reactive for hepatitis B surface antigen (HBsAg), antibody to hepatitis C, and antibody to HIV-1/2, all products using human source material should be handled as potentially infectious, because no test method can offer complete assurance that infectious agents are absent.
J. Substantial Equivalence Information:
1. Predicate device name(s):
NMR Profiler and NMR Lipoprofile Assay
2. Predicate K number(s):
k063841
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Predicate device (k063841) | Proposed device |
| Indications for use | Measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease. | Same |
| Specimen type | Serum and plasma | Same |
| Reference range study | 5^{th} through 95^{th} percentile values for LDL-P obtained for samples from the Multi-Ethnic Study of Athlerosclerosis | Same
(Ranges from predicate transferred to new device via CLSI Guideline C28-A2; see section M5 below) |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Predicate device (k063841) | Proposed device |
| Software version that includes algorithm used to curve fit to the NMR peak and calculate the LDL-P results and software architecture | LP2 | Gen2.1 |
| Software architecture | Calculation of all three analytes (LDL-P, HDL concentration and triglycerides) performed by one algorithm in one software module. Changes to the algorithm result in changes in calculations for all analytes. | Modular (separate software modules for each analyte that allows for modules to change and new modules to be added with no impact to the signal generation or to other unmodified analytes) |
K. Standard/Guidance Document Referenced (if applicable):
- CLSI Guideline EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods
- CLSI Guideline EP 17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation
- CLSI Guideline EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach
- CLSI Guideline EP9-A2: Method Comparison and Bias Estimation Using Patient Samples
- CLSI Protocol EP7-A2: Interference Testing in Clinical Chemistry
- CLSI Guideline C28-A2: How to Define and Determine Reference Intervals in the Clinical Laboratory
L. Test Principle:
The NMR LipoProfile® test and NMR Profiler involves measurement of the 400 MHz proton NMR spectrum of a plasma/serum sample, deconvolution of the composite signal at approximately 0.8 ppm to produce signal amplitudes of the lipoprotein subclasses that contribute to the composite plasma/serum signal, and conversion of these subclass signal amplitudes to lipoprotein subclass concentrations. The ~0.8 ppm plasma NMR signal arises from the methyl group protons of the lipids
{4}
carried in the LDL, HDL and VLDL subclasses of varying diameters. The NMR signals from the various lipoprotein subclasses have unique and distinctive frequencies and line shapes, each of which is accounted for in the deconvolution analysis model. Each subclass signal amplitude is proportional to the number of subclass particles emitting the signal, which enables subclass particle concentrations to be calculated from the subclass signal amplitudes derived from the spectral deconvolution analysis. LDL subclass particle concentrations, in units of nanomoles of particles per liter (nmol/L), are summed to give the reported total LDL particle concentration (LDL-P). By employing conversion factors assuming that the various lipoprotein subclass particles have cholesterol and triglyceride contents characteristic of normolipidemic individuals, HDL cholesterol and triglyceride concentrations are also derived.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Within-run and within-laboratory precision were evaluated in accordance with the methods defined in the CLSI Guideline EP5-A, “Evaluation of Precision Performance of Clinical Chemistry Devices.”
Three serum pools with different analyte concentrations were prepared and tested in multiple runs over multiple days. For the Within-Lab, multiple instruments, operators and reagent lots were incorporated into the testing and a variance component analysis was conducted to estimate the individual sources of the total system variability.
**Within-run precision** - A single run of 20 replicates for the low and the medium pool was conducted on one instrument and a single run of 20 replicates for the high pool was conducted on a second NMR instrument. A single operator conducted all three runs.
LDL-P (nmol/L)
| Pool | N | Mean | SD | SD 95% CI | CV |
| --- | --- | --- | --- | --- | --- |
| Low | 20 | 908 | 45.4 | 34.5 – 66.3 | 5.0% |
| Medium | 20 | 1493 | 64.8 | 49.3 – 94.7 | 4.3% |
| High | 19* | 1967 | 72.8 | 55.0 – 107.6 | 3.7% |
*One replicate of the high pool produced no data due to instrument failure (sample delivery failure) and therefore only 19 replicates were analyzed.
{5}
6
HDL-C (mg/dL)
| Pool | N | Mean | SD | SD 95% CI | CV |
| --- | --- | --- | --- | --- | --- |
| Low | 20 | 23.7 | 0.5 | 0.4 - 0.7 | 2.0% |
| Medium | 19* | 54.9 | 1.0 | 0.8 - 1.5 | 1.9% |
| High | 20 | 95.1 | 0.9 | 0.6 - 1.2 | 0.9% |
Triglycerides (mg/dL)
| Pool | N | Mean | SD | SD 95% CI | CV |
| --- | --- | --- | --- | --- | --- |
| Low | 20 | 81.0 | 2.1 | 1.6 - 3.1 | 2.6% |
| Medium | 19* | 140.6 | 2.5 | 1.9 - 3.7 | 1.8% |
| High | 20 | 649.5 | 8.7 | 6.6 - 12.7 | 1.3% |
* One replicate of the medium pool produced no data due to an instrument failure (sample delivery failure) and therefore, only has 19 replicates analyzed.
Within-laboratory precision – Two runs per day with two replicates per run for three pools were tested on three instruments for a total of 20 testing days. Five operators conducted the study over the 20 testing days. Additionally, six lots of NMR Diluent 1 were used over the 20 days of testing.
LDL-P (nmol/L):
| Pool | N | Mean | SD | SD 95% CI | CV |
| --- | --- | --- | --- | --- | --- |
| Low | 240 | 932.6 | 64.6 | 58.8 - 71.7 | 6.9% |
| Medium | 240 | 1539.3 | 86.9 | 77.3 – 99.3 | 5.6% |
| High | 240 | 2046.1 | 111.8 | 99.3 – 127.9 | 5.5% |
{6}
7
HDL-C (mg/dL):
| Pool | N | Mean | SD | SD 95% CI | CV |
| --- | --- | --- | --- | --- | --- |
| Low | 240 | 23.5 | 0.9 | 0.8 – 1.0 | 3.9% |
| Medium | 240 | 56.5 | 1.3 | 1.2-1.6 | 2.4% |
| High | 240 | 96.2 | 2.5 | 2.1 – 2.9 | 2.6% |
Triglycerides (mg/dL):
| Pool | N | Mean | SD | SD 95% CI | CV |
| --- | --- | --- | --- | --- | --- |
| Low | 240 | 78.0 | 2.9 | 2.6 – 3.3 | 3.7% |
| Medium | 240 | 146.0 | 4.2 | 3.6 – 4.9 | 2.8% |
| High | 240 | 627.5 | 20.7 | 17.7 – 24.8 | 3.3% |
b. Linearity/assay reportable range:
The claimed measuring ranges of this device are:
LDL-P: 300 - 3500 nmol/L
HDL-C: 7 - 140 mg/dL
Triglycerides: 5 - 1100 mg/dL
LDL-P:
The dilution series consisted of eleven samples that were prepared by combining selected de-identified, residual specimens from LipoScience’s commercial laboratory. The samples covered the range from 184 nmol/L to 3868 nmol/L. All the pools were run on the same day in 4 replicates. The study was conducted in a single run on one instrument. Significance of the terms for both the linear and polynomial regression was examined.
{7}
| R=0.9987, R²=0.9974 | | | |
| --- | --- | --- | --- |
| Linear Term | Coefficient | SE | p |
| Intercept | -11.9944 | 19.4903 | 0.5424 |
| X | 0.9843 | 0.0086 | < 0.0001 |
| R=0.9990, R²=0.9980 | | | |
| Non-Linear Term | Coefficient | SE | p |
| Intercept | 39.4779 | 23.4440 | 0.1016 |
| X | 0.8929 | 0.0293 | < 0.0001 |
| X² | 0.0000 | 0.0000 | 0.0027 |
In the polynomial regression, the $\mathrm{X}^2$ term was significant (p=0.0027). Given the significance of the nonlinear term, acceptance criteria for allowable nonlinearity were evaluated. Samples 5-11 met the sponsor's pre-determined allowable nonlinearity criteria (10%). The low concentration samples 2-3, compared to the absolute nonlinearity criteria of 52.1 nmol/L, meet the sponsor's allowable nonlinearity criteria. Therefore the linearity observed for LDL-P in this study supports the claimed measuring range of 300 to 3500 nmol/L.
HDL-C and triglycerides: Twelve concentration levels were prepared across the range of the assays by serial dilutions of the source pools and tested. The study was conducted in a single run on one instrument and each level was tested in replicates of four. For HDL-C, the samples covered the range of 6.3 - 205 mg/dL; for triglycerides the samples covered the range 5.4 - 1880 mg/dL. Linear regression analysis yielded the following results:
HDL-C:
$\mathrm{R}^2 = 0.9998$
$\mathrm{y} = 1.004\mathrm{x} - 0.5956$
Triglycerides:
$\mathrm{R}^2 = 0.9998$
$\mathrm{y} = 1.006 + 2.057$
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Reagents, calibrators and controls are unchanged from the previously cleared
{8}
device (k063841).
d. Detection limit:
The lower ends of the measuring ranges for LDL-P, HDL-C and triglycerides were determined by linearity, above in 2b, and unchanged from the predicate (k063841):
LDL-P: 300 mmol/L
HDL-C: 7 mg/dL
Triglycerides: 5 mg/dL
e. Analytical specificity:
A new interference study was performed to evaluate whether substances could be potential interferents of the device and its modified software. In addition, two new therapeutics for the intended use population became available and were also evaluated in the new study.
In order to evaluate potential interferents with the new version of the LDL-P software Gen2.1, LipoScience employed the following study design: first, identify potentially interfering substances that show resonance(s) in the vicinity of lipoprotein resonances. Each potential interferent was diluted in the appropriate solvent and analyzed on the NMR Profiler to determine if it demonstrated peak(s) in the 0.7 – 1.0 ppm region. Five endogenous substances and 22 drugs were screened in this step using concentrations in accordance to CLSI EP7-A2.
Second, once a potentially interfering substance was identified, a spiking study was completed where the substance was added to sample pools containing two different levels of LDL-P, HDL-C and triglycerides for a paired difference test.
9
{9}
10
Endogenous
Exogenous (OTC drugs, etc.)
| Potential Interferent | Test Concentration | Potential Interferent | Test Concentration |
| --- | --- | --- | --- |
| Hemoglobin | 2 mg/mL | Simvastatin | 48 μg/mL |
| Bilirubin, unconj. | 200 μg/mL | Fenofibrate | 45 μg/mL |
| Creatinine | 50 μg/mL | Nicotinic Acid Sodium salt | 1.2 mg/mL |
| Urea | 2.6 mg/mL | Acetylsalicylic acid | 652 μg/mL |
| Uric acid | 235 μg/ml | Acetaminophen | 200 μg/mL |
| | | Naproxen Sodium | 547 μg/mL |
| | | Ibuprofen Sodium salt | 210^{1}, 70^{2} μg/mL |
| | | Piroxicam | 60 μg/mL |
| | | Hydrochlorothiazide | 6.0 μg/mL |
| | | Triamterene | 8.9 μg/mL |
| | | Pravastatin | 48 μg/mL |
| | | Furosemide | 60 μg/mL |
| | | Metoprolol tartrate | 6.4 μg/mL |
| | | Nifedipine | 400 μg/mL |
| | | Enalaprilat Dihydrate | 0.3 μg/mL |
| | | Hydralazine hydrochloride | 180 μg/mL |
| | | Isosorbide dinitrate (lactose mixture 2:3) | 375 μg/mL |
| | | Clopidogrel hydrogensulfate | 360 μg/mL |
| | | Glipizide | 2.0 μg/mL |
| | | Metformin Hydrochloride | 600 μg/mL |
| | | Pioglitazone hydrochloride | 27 μg/mL |
| | | Atorvastatin | 48 μg/mL |
1 For LDL-P and HDL concentration
2 test results for triglycerides were ~4 to 6% lower at the 140 and 210 μg/mL concentrations of ibuprofen
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
**LDL-P**: The comparison was conducted in agreement with CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples. The study involved testing singlicates followed by statistical analysis of 1600 freshly-collected clinical de-identified specimens across the range of the LDL-P test. Bias was calculated for LDL-P from the linear regression line at the medical decision limits of 1000, 1300 and 1600 nmol/L as well. 1555 samples are included in the final analysis (44 samples were outside of the claimed
{10}
measuring range; one sample failed QC check due to an instrument shim error and could not be repeated due to insufficient sample volume). The range of the samples was 315 to 3497 nmol/L.
## Linear regression/least squares analysis:
$$
r = 0.970
$$
Slope = 0.98 95% CI: 0.97 – 0.99
Intercept = 46.42 95% CI: 22.44 – 70.40
## Deming Regression analysis:
Slope = 1.01 95% CI: 1.00 - 1.03
Intercept = -7.97 95% CI: -31.45 – 15.50
In addition to conducting a method comparison for the overall data set, the data were classified into the three following segments based on LDL-P values; (1) results below the medical decision limits of 1000 nmol/L, (2) results between 1000 and 1600 nmol/L and (3) results above 1600 nmol/L. For each segment the bias was determined. The determined % Bias for each segment was less than the acceptable allowable bias (< 5% for each level).
Observed LDL-P Bias per linear regression at Medical Decision Limits
| | Linear regression | | Deming regression | |
| --- | --- | --- | --- | --- |
| LDL-P (nmol/L) | Absolute Bias | % Bias | Absolute Bias | % Bias |
| 1000 | 28.0 | 2.8% | 3.9 | 0.4% |
| 1300 | 22.6 | 1.7% | 7.5 | 0.6% |
| 1600 | 17.0 | 1.1% | 11.1 | 0.7% |
## HDL-C and triglycerides:
A total of 1600 freshly-collected clinical de-identified specimens from the LipoScience commercial lab were evaluated. The performance of these samples run on the proposed device was compared to results given by the predicate (k063841).
$$
\mathrm{HDL - C:}
$$
$$
y = 1.00x + 0.03
$$
$$
R = 0.999
$$
$$
\text{Range} = 12.0 - 140.0 \, \mathrm{mg/dL}
$$
Triglycerides:
n = 1597 (two samples outside the measuring range; one sample failed as indicated above under LDL-P method comparison)
$$
y = 1.00x + 1.17
$$
$$
r = 1.000
$$
$$
\text{Range} = 20 - 925 \, \mathrm{mg/dL}
$$
{11}
b. Matrix comparison:
The various specimen tubes recommended by the manufacturer for use with this assay are serum drawn in gel barrier NMR LipoTubes (Greiner, Inc. Part #456293), or red-top plain serum collection tubes; or plasma drawn into EDTA or heparin collection tubes. A study was conducted to validate the equivalence of NMR LipoProfile results using these serum and plasma specimens. Blood was collected in the specified collection tubes and immediately tested. The study comparing samples collected in serum tubes (NMR LipoTube, reference) to plain serum (red-top), EDTA plasma (purple top) or Na Heparin plasma (green top) tubes tested individual specimens from 30 volunteers. Samples tested spanned most of the assay range. Average biases observed ranged between 2.6 to 6.1% for LDL-P. No concentration-dependent trends were observed.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The sponsor used procedures in CLSI Guideline C28-A2 to bridge the transference of a reference range established with the predicate method in k063841 to the proposed device. For the predicate devices, in order to determine the distribution of LDL-P levels expected in a representative sampling of the general population, plasma samples (n=5,362) were analyzed from apparently healthy men and women (mean age 61, ranging from 44 to 84 years) enrolled in the Multi-Ethnic Study of Atherosclerosis (MESA), a large epidemiologic study sponsored by the National Heart, Lung, and Blood Institute. Results were presented as percentiles.
In the sponsor’s bridging study, the plasma samples from the MESA study were re-tested. Statistical analysis demonstrated that the percentiles determined for the predicate are also valid for the proposed device.
{12}
| | All (n=5,362) | Men (n=2,529) | Women (n=2,833) |
| --- | --- | --- | --- |
| Percentile | LDL-P (nmol/L) | LDL-P (nmol/L) | LDL-P (nmol/L) |
| 5 | 770 | 800 | 760 |
| 10 | 870 | 900 | 850 |
| 20 | 1000 | 1040 | 970 |
| 30 | 1100 | 1150 | 1060 |
| 40 | 1190 | 1250 | 1150 |
| 50 | 1280 | 1330 | 1230 |
| 60 | 1380 | 1430 | 1330 |
| 70 | 1480 | 1530 | 1440 |
| 80 | 1610 | 1640 | 1570 |
| 90 | 1790 | 1820 | 1760 |
| 95 | 1980 | 1990 | 1970 |
HDL concentration and triglycerides: The proposed device has similar reference values as the predicate. The reference values for patient classification have been recommended by the NCEP for HDL cholesterol and triglycerides for the assessment and management of CVD risk. Each laboratory should verify the validity of these reference values for the population it serves.
| HDL Cholesterol, mg/dL Classification | |
| --- | --- |
| Low | High |
| <40 | ≥60 |
| Triglycerides, mg/dL Classification | | | |
| --- | --- | --- | --- |
| Normal | Borderline-high | High | Very high |
| < 150 | 150-199 | 200-499 | ≥500 |
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.