SWEETSPOT DIABETES DATA MANAGEMENT SERVICE
Device Facts
| Record ID | K111509 |
|---|---|
| Device Name | SWEETSPOT DIABETES DATA MANAGEMENT SERVICE |
| Applicant | Sweetspot Diabetes Care, Inc. |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Nov 9, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Software as a Medical Device, Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K111509 · Nov 9, 2011 | SWEETSPOT DIABETES DATA MANAGEMENT SERVICE | Sweetspot Diabetes Care, Inc. | User evaluations in typical clinical settings; Patient surveys; Healthcare professional (HCP) feedback | The device was evaluated in real-world clinical settings to assess usability, workflow impact, and the accuracy of the data retrieval (Fetch) process when performed by lay users in a clinical environment. | Usability; Workflow; Clinical setting; Lay user evaluation |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Lay User Evaluation; Observational user evaluation in typical clinical settings | Typical outpatients; Sample Size: 88 respondents; Number of Sites: 2 | Not applicable for this study | Ease of use, likelihood to bring meter to appointments, meter download success rate |
| Back-Office Kiosk HCP Evaluation; Observational user evaluation in a clinical setting | Healthcare professionals (HCP); Sample Size: 11 respondents; Number of Sites: 1 | Not applicable for this study | Workflow efficiency, ease of use, report consistency |
Indications for Use
The SweetSpot Diabetes Data Management Service is intended for use in in clinical settings by both patients and healthcare professionals to assist people with diabetes and their healthcare professionals in the review, analysis and evaluation of historical blood glucose test results to support effective diabetes management. It is intended for use as an accessory to blood glucose meters with data interface capabilities.
Device Story
SweetSpot Diabetes Data Management Service is a web-based SaaS platform for diabetes management. It retrieves historical blood glucose data from various manufacturers' meters via a 'Fetch Utility' and device-specific cables. Input data is processed and stored on remote cloud servers. The system provides three access points: Front-Office Kiosk (patient-operated), Back-Office Kiosk (HCP-operated), and Back-Office Web Application (HCP-operated). The service consolidates data from multiple meters into a standardized report format. HCPs use these reports to monitor patient glucose levels, simplify clinical decision-making, and improve workflow efficiency. The system benefits patients by facilitating data sharing with clinicians and potentially increasing adherence to meter usage.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing to verify data transfer compatibility and accuracy between meters and the software, and a human factors study (n=21) to verify ease of use, label comprehension, and meter data transfer by lay users.
Technological Characteristics
Web-based SaaS platform utilizing cloud computing for data storage and processing. Interfaces with blood glucose meters via device-specific cables. Compatible with PC and Mac OS. Software-based data retrieval and report generation. Usability testing conducted per ISO 15197:2003.
Indications for Use
Indicated for patients with diabetes and healthcare professionals to assist in the review, analysis, and evaluation of historical blood glucose test results to support diabetes management. Used as an accessory to blood glucose meters with data interface capabilities.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- MyCareTeam-Diabetes (MCT-Diabetes) data management software (k073699)