K111483 · Sciton, Inc · PDZ · Dec 7, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111483
Device Name
JOULE CLEARSENSE LASER SYSTEM
Applicant
Sciton, Inc
Product Code
PDZ · General, Plastic Surgery
Decision Date
Dec 7, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The JOULE ClearSense Laser System is indicated for ablation, vaporization, incision, excision and coagulation of soft tissue, including: - Matrixectomy - Periungual and subengual warts - Planter warts - Radical nail excision - Neuromas It is also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
Device Story
JOULE ClearSense Laser System is a transportable Nd:YAG laser (1064 nm) for soft tissue surgery and onychomycosis treatment. System comprises control console, foot switch, articulated arm, and handpiece with stainless steel guide tip. Laser energy heats target tissue to destroy it while sparing surrounding skin. Operated by clinicians in a medical setting. Output is pulsed laser energy; clinician monitors via LCD touchscreen interface. Benefits include temporary increase of clear nail in onychomycosis patients and surgical precision for soft tissue procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and performance specifications compared to predicate devices.
Technological Characteristics
Nd:YAG laser, 1064 nm wavelength; pulsed output mode; 20-3500 mJ energy per pulse; 100-3000 µsec pulse duration; 1 mm spot size; 5-100 Hz repetition rate; Class IV laser; Class I, Type BF electrical safety; LCD touchscreen interface; stainless steel guide tip.
Indications for Use
Indicated for patients with onychomycosis (dermatophytes or yeasts) for temporary increase of clear nail; also indicated for soft tissue procedures including matrixectomy, warts (periungual, subungual, plantar), radical nail excision, and neuromas.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K11148/3
pg 1 of 3
DEC - 7 2011
## Attachment IV
## 510(k) Summary
| Submitter: | Sciton, Inc. | |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Address: | 925 Commercial Street, Palo Alto, CA 94303 | |
| Phone: | (650) 493-9155 | |
| Fax : | (650) 493-9146 | |
| Contact Person: | Jay M. Patel, VP of Regulatory Affairs | |
| Date Prepared: | September 26, 2011 | |
| Device Trade Name: | JOULE ClearSense Laser System | |
| Common Name: | Laser Powered Surgical Device (and Accessories) | |
| Classification: | Regulation Number: 21 CFR 878.4810<br>Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology<br>Regulatory Class: Class II<br>Product Code: GEX | |
| Legally Marketed<br>Predicate Device: | K103626: Cutera GenesisPlus Laser System<br>K093547: PinPointe FootLaser<br>K093545: PinPointe FootLaser | |
| Description of<br>JOULE ClearSense<br>Laser System: | The JOULE ClearSense Laser System is a transportable medical device used for the treatment of temporary increase in clear nail in patients with onychomycosis. It uses Nd:YAG laser with a wavelength of 1064 nm.<br><br>ClearSense uses the laser energy to heat the target to a temperature that is sufficient to destroy it, but not to the point that the heat damages skin and surrounding tissue.<br><br>The ClearSense system consists of a control console, a foot switch, articulated. arm and a handpiece with a stainless steel guide tip (spacer) attached at its end.<br><br>Control Console houses the power supply, control electronics, cooling system and optics to direct the laser beam to the input of the articulated arm.<br><br>The following specifications list significant physical and performance characteristics of this device: | |
| | | |
| | Laser Medium | Nd:YAG |
| | Laser Wavelength | 1064 nm |
| | Energy per Pulse | 20 - 3500 mJ |
| | Max Power | ≤ 100 W |
| | Spot Size | 1 mm |
| | Output Mode | Pulsed |
| | Aiming Beam Wavelength | 630 - 680 nm |
| | Aiming Beam Power | ≤ 2.5 mW |
| Rated Voltage | 200-240 V~ | |
| Rated Frequency | 50/60 Hz | |
| Rated Current | 25 A | |
| Classification | Class I, Type BF | |
| Laser Output | Class IV | |
{1}------------------------------------------------
(111483 Pg 2 of 3
Intended Use:
The JOULE ClearSense Laser System is indicated for ablation, vaporization, incision, excision and coagulation of soft tissue, including:
- · Matrixectomy
- · Periungual and subengual warts
- · Planter warts
- · Radical nail excision
- · Neuromas
It is also indicated for use for the temporary increase of clear nail in patients with cnychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
Technological Characteristics: The JOULE ClearSense Laser System shares the same indications for use, and as noted below, shares similar design features (Including wavelength, laser medium and delivery systems, power supply, cooling and control system), functional features (including power output, repetition rate, energy, spot size and fluence), and is therefore substantially equivalent to the above legally marketed predicate devices.
| Specification | JOULE<br>ClearSense<br>Laser System | Cutera<br>GenesisPlus<br>Laser | PinPointe<br>FootLaser |
|------------------|-------------------------------------------------|-------------------------------------------------|----------------------------------------------|
| Ref. 510(k) | K111483 | K103626 | K093547,<br>K093545 |
| Wavelength | 1064 nm Nd:YAG | 1064 nm Nd:YAG | 1064 nm Nd:YAG |
| Aiming Beam | 630 - 680 nm<br>(≤ 2.5 mW) | 630 - 680 nm<br>(≤ 2.5 mW) | 630 - 680 nm<br>(≤ 2.5 mW) |
| Energy per Pulse | 20 - 3500 mJ | 20 - 3500 mJ | 20 - 3500 mJ |
| Fluence | 25.5 J/cm2 | 25.5 J/cm2 | 25.5 J/cm2 |
| Max Power | ≤ 100 W | ≤ 100 W | ≤ 100 W |
| Pulse Duration | 100 - 3000 µsec | 100 - 3000 µsec | 100 - 3000 µsec |
| Spot Size | 1 mm*<br>(up to 13 mm for<br>other indications) | 1 mm*<br>(up to 13 mm for<br>other indications) | 1 mm*<br>(other spot sizes<br>not published) |
| Output Mode | Pulsed, multimode | Pulsed, multimode | Pulsed, multimode |
| Repetition Rate | 5 - 100 Hz | 5 - 100 Hz | 5 - 100 Hz |
| Laser Media | Flashlamp<br>pumped solid<br>state rod | Flashlamp<br>pumped solid<br>state rod | Flashlamp<br>pumped solid<br>state rod |
| User Interface | LCD touchscreen | LCD touchscreen | LCD touchscreen |
* for increase in clear nail
Safety and Effectiveness: The indications for use are based upon the indications for use for predicate systems. Technologically, the JOULE ClearSense Laser System is substantially equivalent to the listed predicate devices. Therefore, the risks and benefits for the JOULE ClearSense Laser System are comparable to the predicate devices.
{2}------------------------------------------------
K111493
1
pg. 3 of 3
Conclusion:
The JOULE ClearSense Laser System shares similar inclications for use, design
features, and similar functional features as, and therefore is substantially equivalent to, the currently marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing and body. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the left and top portions of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Sciton, Inc. % Mr. Jay M. Patel Vice President of Regulatory Affairs 925 Commercial Street Palo Alto, California 94303
MAY 13 2013
Re: K111483
Trade/Device Name: JOULE ClearSense Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: PDZ, GEX Dated: November 21, 2011 Received: December 01, 2011
Dear Mr. Patel:
This letter corrects our substantially equivalent letter of December 07, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Jay M. Patel
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely yours. FOR
Peter D. Rúmm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Attachment III
## Statement of Indications for Use
| 510(k) Number (if known): | K111483 |
|---------------------------|---------|
|---------------------------|---------|
pg 1 of 1
| Device Name: | JOULE ClearSense Laser System |
|--------------|-------------------------------|
|--------------|-------------------------------|
Indications for Use:
The JOULE ClearSense Laser System is indicated for ablation, vaporization, incision, excision and coagulation of soft tissue, including:
- Matrixectomy .
- Periungual and subengual warts ●
- Planter warts .
- Radical nail excision ●
- . Neuromas
It is also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
Neil R. Doyle firma
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
111483 510(k) Number.
× Prescription Use (Per 21CFR801)
OR
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.