K111475 · Starmedtec GmbH · GEX · Sep 28, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111475
Device Name
AURIGA XL
Applicant
Starmedtec GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 28, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Auriga XL laser system including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialities including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Device Story
Auriga XL is a holmium laser system (approx. 2.1 µm wavelength) for surgical tissue management. System components include a laser unit with control panel, foot switch, and optical application fiber. Operated by physicians in OR, clinic, or endoscopic settings. Laser energy is delivered via fiber to target tissue for incision, excision, resection, ablation, vaporization, coagulation, or hemostasis. Clinical benefit includes precise tissue removal and lithotripsy across multiple specialties. Output is controlled by the surgeon via the user interface and foot switch; clinical decision-making relies on surgeon visualization and procedural requirements.
Clinical Evidence
Bench testing only. Laboratory testing verified that the Auriga XL met all design specifications and is substantially equivalent to the predicate device. No clinical information is required.
Indicated for patients requiring surgical incision, excision, resection, ablation, vaporization, coagulation, and hemostasis of soft, cartilaginous, and bony tissue across various specialties including urology (lithotripsy, BPH, tumors), gastroenterology, arthroscopy, spinal surgery, thoracic/pulmonary, gynecology, ENT, dermatology, and plastic surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K111475
1 of 4.
SEP 2 8 2011
Image /page/0/Picture/2 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol inside, followed by the word "starmedtec" in a sans-serif font. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font.
# 510(k) Auriga XL Summary (Section 5)
510(k) summary of safety and effectiveness information according 21 CFR Part 807.87(h)
- 1. General Information:
- a. Applicant:
StarMedTec GmbH Kreuzstrasse 22 82319 Starnberg Germany +49815126861-0 (phone) +49815126861-35 (fax)
- b. Contact:
Date Prepared: C.
- 2. Names:
- a. Device Name:
- b. Common Name:
- Classification Name: C.
- d. Product code:
Gregor Weidemann May 20, 2011
- Auriga XL
Auriga XL Laser instrument, surgical, powered GEX
- 3. Predicate Device:
Lisa Laser Products - Sphinx (510(k) Number: K033437)
- 4. Product Description:
The Auriga XL is a holmium laser system, which emits laser radiation with a wavelength of approximately 2.1 um. The laser power transfers an optical application fiber. The indications are lithotripsy, dissection, ablation, resection and coagulation of tissue. The laser system consists of:
- laser system including control panel (user interface) -
- foot switch
- application fiber -
- 5. Indications for use:
The Auriga XL laser system including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialities including:
Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectorny, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
## Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Strictures of urethra and ureter -
- Bladder Neck Incisions (BNI) -
{1}------------------------------------------------
K111475
Image /page/1/Picture/1 description: The image shows the text "510(k) Auriga XL Summary (Section 5)". The text is in bold and is black. The text is likely a title or heading for a document or section of a document.
Image /page/1/Picture/2 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol on the left, and the word "starmedtec" in a sans-serif font on the right. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font with a line extending from the right side of the phrase.
- Ablation and resecetion of Bladder Tumors, Uretheral Tumors and Ureteral -Tumors
- Ablation of Benign Prostatic Hyperplasia (BPH) -
- Transurethral incision of prostate (TUIP)
- Holmium Laser Resection of the prostate (HoLRP) -
- Holmium Laser Enucleation of the prostate (HoLEP)
- Holmium Laser Ablation of the prostate (HoLap) -
- Bladder/Renal calculi (Lithotripsy) 1
- Condylomas -
- Lesions or external genitalia -
- Lithotripsy
- Percutaneous Urinary Lithtripsy -
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragemnation of kidney calculi -
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
#### Gastroenterology
Open and endoscopic gastroenterological surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy
- Polyps
- Biopsv
- Gall Bladder Calculi (Lithotripsy) ー
- Biliary/Bile duct calculi (Lithotripsy)
- Ulcers
- Gastric ulcers
- Duodenal ulcers
- Non bleeding ulcers ﮯ
- Pancreatitis
- Hemorrhoids
- Cholecystectomy
- Benign and malignant neoplasm
- Angiodysplasia
- Colorectal cancer ー
- Telangiectasias
- Telangiectasias of Osler-Weber-Renu disease
- Vascular malformation
- Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric Erosions
i
{2}------------------------------------------------
KII1475
3 of 4
Image /page/2/Picture/1 description: The image shows the title "510(k) Auriga XL" in bold, followed by the subtitle "Summary (Section 5)" also in bold. The text is black and the background is white. The text appears to be part of a document or presentation.
Image /page/2/Picture/2 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol on the left, and the word "starmedtec" in black letters on the right. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font, with a line above and below the phrase.
### Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and crtilaginous tissue) in small and large joints of the body excluding the spine but including:
- Ligament and tendon Release -
- Countouring and sculpting of articular surfaces -
- Capsulectomie in the knee ﺖ
- Chondreoplasty in the knee l
- Debridement of inflamed synovial tissue -
- Chondromalacia ablation …
- Chondtromalcia and tears "
- Plica removal -
- Meniscectomy -
- Loose body debridement ー
- Lateral retineacular release -
Ablation of soft, cartilanginous and bony tissue in minimal invasive spinal surgery including
- Percutaneous laser disc decompression/discectomy of the L4-5 and L5-S1 । lumbar discs, including forminoplasty
- Percutaneous cervical disc decompression/discectomy -
- Percutaneous thoracic disc decompression/discectomy -
### Thoracic and Pulmonary
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
### Gynaecology
Open and endoscopic/laparoscopic gynaecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis).
### ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue and cartilage) including:
- Endonasal / sinus surgery -
- Partial turbinectomy -
- Polypectomy -
- Dacryocystorhinostomy ー
- Frontal sinustomy t
- Ethmoidectomy -
- Maxillary antrostomy 1
- Functional endoscopic sinus surgery -
{3}------------------------------------------------
K111475
4 of 4
# 510(k) Auriga XL Summary (Section 5)
Image /page/3/Picture/2 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol on the left, and the word "starmedtec" in black letters on the right. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font.
### Dermatology and Plastic Surgery
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue, muscosal tissue, fatty tissue and cartilaginous tissue in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- Basal Cell Carcinomas -
- Lesion of skin and subcutaneous tissue 1
- Skin tags -
- Plantar warts 1
- Lesions of skin and subcutaneous tissue -
- Port wine stains ー
- Papillomas -
### General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy -
- Biopsy -
- Skin incision ー
- ﺖ Excision of external tumours and lesions
- Complete or partial resection of internal organs, turnours, lesions -
- 6. Performance testing:
The Auriga XL is tested according to following standards:
ISO 14971:2006 IEC 60601-1:2005 (DIN EN 60601-1:2006) IEC 60601-1-2:2007 (DIN EN 60601-1-2:2007-12) IEC 60601-1-6:2006 (DIN EN 60601-1-6:2007) / IEC 62366:2007 (DIN EN 62366:2008) IEC 60601-2-22:2005 IEC 60825-1:2007 (DIN EN 60825-1:2007) IEC 62304:2006 (DIN EN 62366:2006) The device also complies with European Medical Device 93/42/EEC + Amendment 2007/47/EC
- 7. Performance data
Laboratory testing was conducted to verify and validate that the Auriga XL met all design specifications and is substantially equivalent to the predicate device. Clinical data: No clinical information is required.
- 8. Conclusion:
The Auriga XL is as effective and save as the predicate device. The Auriga XL is substantially equivalent to the cited legally marked predicate device.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle faces left and is enclosed in a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper part of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
StarMedTec GmbH % Mr. Gregor Weidemann Kreuzstrasse 22 Starnberg, Germany 82319
SEP 2 8 2011
Re: K111475
Trade/Device Name: Auriga XL Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX
Dated: September 21, 2011 Received: September 23, 2011
Dear Mr. Weidemann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 – Mr. Gregor Weidemann
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 331-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutliDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Sincerely, yours,
Eunice Keith
for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol on the left, and the word "starmedtec" in black letters on the right. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font. A horizontal line extends from the target symbol to the word "starmedtec".
# Indications for Use
510(k) Number (if known): K111475
Device Name: Auriga XL
- 1. Indications for use:
The Auriga XL laser system including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialities including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office o
the RP Usler for mxn
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111475
{7}------------------------------------------------
nnovative solutions
### Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Strictures of urethra and ureter
- Bladder Neck Incisions (BNI) -
- Ablation and resecetion of Bladder Tumors, Uretheral Tumors and i Ureteral Tumors
- Ablation of Benign Prostatic Hyperplasia (BPH) -
- Transurethral incision of prostate (TUIP) -
- Holmium Laser Resection of the prostate (HoLRP) -
- Holmium Laser Enucleation of the prostate (HoLEP) -
- Holmium Laser Ablation of the prostate (HoLap)
- Bladder/Renal calculi (Lithotripsy) ー
- Condylomas ー
- Lesions or external genitalia -
- Lithotripsy
- Percutaneous Urinary Lithtripsy
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragemnation of kidney calculi t
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ODE)
Neikpegh for mkn
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K114725
{8}------------------------------------------------
# 510(k) Auriga XL Indication for Use Statement (section 4)
starmedtec onovative solutions
### Gastroenterology
Open and endoscopic gastroenterological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy -
- Polyps -
- Biopsy -
- Gall Bladder Calculi (Lithotripsy) -
- Biliary/Bile duct calculi (Lithotripsy) -
- Ulcers -
- Gastric ulcers -
- Duodenal ulcers -
- Non bleeding ulcers ー
- Pancreatitis -
- Hemorrhoids "
- Cholecystectomy -
- Benign and malignant neoplasm -
- Angiodysplasia ー
- Colorectal cancer -
- Telangiectasias -
- Telangiectasias of Osler-Weber-Renu disease -
- Vascular malformation
- Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric Erosions
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE
Neil R.P. Dofe for man
vision Sign-Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111475
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for starmedtec. The logo consists of a black square with a white circle and crosshairs inside, followed by the word "starmedtec" in black letters. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font.
### Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilangenous tissue) in small and large joints of the body excluding the spine but including:
- Ligament and tendon Release -
- Countouring and sculpting of articular surfaces -
- Capsulectomie in the knee -
- Chondreoplasty in the knee ー
- Debridement of inflamed synovial tissue -
- Chondromalacia ablation ー
- Chondtromalcia and tears -
- Plica removal -
- Meniscectomy -
- Loose body debridement -
- Lateral retineacular release -
Ablation of soft, cartilanginous and bony tissue in minimal invasive spinal surgery including
- Percutaneous laser disc decompression/discectomy of the L4-5 and L5--S1 lumbar discs, including forminoplasty
- Percutaneous cervical disc decompression/discectomy -
- Percutaneous thoracic disc decompression/discectomy
## Thoracic and Pulmonary
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (OI
(ODE)
Neil R. Adler for mam
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111475
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol on the left, and the word "starmedtec" in black letters on the right. Below the word "starmedtec" is the phrase "innovative solutions".
### Gynaecology
Open and endoscopic/laparoscopic gynaecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis).
# ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue and cartilage) including:
- Endonasal / sinus surgery -
- Partial turbinectomy -
- Polypectomy ー
- Dacryocystorhinostomy -
- Frontal sinustomy -
- Ethmoidectomy -
- -Maxillary antrostomy
- Functional endoscopic sinus surgery -
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <i>Neil</i> RP <i>Ogden for mam</i> |
|---------------|-------------------------------------------------------------------------------------------------------|
| | (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices |
| 510(k) Number | K111475 |
{11}------------------------------------------------
## Dermatology and Plastic Surgery
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue, muscosal tissue, fatty tissue and cartilaginous tissue in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- Basal Cell Carcinomas -
- Lesion of skin and subcutaneous tissue -
- Skin tags -
- Plantar warts -
- Lesions of skin and subcutaneous tissue -
- Port wine stains -
- Papillomas -
### General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy -
- -Biopsy
- Skin incision =
- Excision of external tumours and lesions -
- Complete or partial resection of internal organs, tumours, lesions -
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) *Evaluation code for mxm*
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
| 510(k) Number | K111475 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.