GRI STERILE AND NON-STERILE SURGICAL DRAPES AND SURGICAL EQUIPMENT COVERS
K111458 · Gri Medical & Electronic Technology Co., Ltd. · KKX · Dec 1, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111458
Device Name
GRI STERILE AND NON-STERILE SURGICAL DRAPES AND SURGICAL EQUIPMENT COVERS
Applicant
Gri Medical & Electronic Technology Co., Ltd.
Product Code
KKX · General, Plastic Surgery
Decision Date
Dec 1, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4370
Device Class
Class 2
Indications for Use
GRI single use drapes (both sterile and non-sterile) are made of natural or synthetic material intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. Non-sterile GRI surgical drapes are sold, in bulk as single use items, to repackager/relabeler establishments for further packaging and ethylene oxide (EtO) sterilization. GRI single use surgical equipment covers (both sterile and non-sterile) are protective barrier covers intended to be used to cover surgical equipment and provide a protective barrier for that equipment. Non-sterile GRI surgical equipment covers are sold, in bulk as single use items, to repackager/relabeler establishments for further packaging and ethylene oxide (EtO) sterilization.
Device Story
Single-use surgical drapes and equipment covers; constructed from three-layer polypropylene laminate (spunbond/meltblown/spunbond) or polyester knit stockinette; thermally embossed; topically treated for water repellency and static dissipation. Provided in various sizes/shapes; may include fluid collection pouches or clear film panels. Used in clinical settings to isolate surgical sites or cover equipment. Non-sterile versions sold in bulk for repackaging and EtO sterilization. Acts as a physical barrier to microbial and fluid contamination. Benefits patient by reducing risk of surgical site infection.
Clinical Evidence
Bench testing only. Performance testing included biocompatibility (cytotoxicity, primary skin irritation, sensitization), hydrostatic pressure, impact penetration, tear strength, tensile strength, linting, and flammability. All results met acceptance criteria.
Technological Characteristics
Materials: Polypropylene (SMS laminate) and polyester knit. Barrier performance: AAMI Level 3 per PB70. Construction: Thermally embossed layers, topically treated for water repellency and static dissipation. Form factor: Various sizes/shapes including fenestrated, split, and equipment-specific covers. Sterilization: Ethylene oxide (EtO) for sterile versions.
Indications for Use
Indicated for use as protective patient coverings to isolate surgical incision sites from contamination and as protective barrier covers for surgical equipment in healthcare facilities.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for GRI, which stands for Global Reporting Initiative. The logo is in black and white and features the letters "gri" in a stylized font. There is a curved line above the "i" in the logo.
ﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴﻴ
## 510(k) Summary
510(k) Applicant/Owner: GRI MEDICAL & ELECTRONIC TECHNOLOGY CO., LTD.
> 1805 HongGao Road, XiuZhou Industry Zone, ZheJiang, China, 314031 86-573-83916501 86-573-83916520
Primary Contact Penny Northcutt Person for this 510(k): REGSolutions, LLC. 717 Lakeglen Drive Suwanee, GA 30024 Fax: (678) 513-0937 Phone: (678) 428-6978 Email: pennynorthcutt@theregsolutions.com
| Summary Date: | November 17, 2011 | | |
|------------------------|------------------------------------------------------------------------------|--|--|
| Device Trade Name: | GRI Sterile and Non-Sterile Surgical Drapes and<br>Surgical Equipment Covers | | |
| Common/Usual Name: | Surgical Drape and Drape Accessories | | |
| Device Classification: | General & Plastic Surgery<br>Class II - 21 CFR 878.4370 | | |
| Device Product Code: | KKX Surgical Drapes<br>MMD Surgical Equipment Covers | | |
#### Device Description:
GRI's single use drapes and surgical equipment covers with AAMI Liquid Barrier Level 3 claim are constructed with various base sheet fabric and reinforcement fabric types, provided in various sizes and shapes, and may contain fluid collection pouches or clear film panels as components. The base sheet fabric is a three layer laminate comprised of polypropylene .
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for GRI, which is a company that provides gas research and insights. The logo is in black and white and features the letters "gri" in a stylized font. The "i" has a curved extension above it. The logo is simple and modern, and it is likely used on the company's website and marketing materials.
spunbond/polypropylene mettblown/polypropylene spunbond. Layers are thermally embossed together producing a single layer with various basis weights. Fabric is topically treated to enhance water repellency and to assure static dissipation. The drape reinforcement fabrics are constructed of various configurations of spunbond, meltblown and film. The reinforcement materials are spunbond/film laminate. The primary component of these materials is polypropylene with surfactant and blue pigment.
#### Predicate Device:
GRI single use drapes and surgical equipment covers are substantially equivalent to the Kimberly Clark K083234 sterile and non-sterile surgical drapes and surgical equipment covers.
#### Indications for Use:
#### GRI Drapes
GRI single use drapes (both sterile and non-sterile) are made of natural or synthetic material intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.
Non-sterile GRI surgical drapes are sold, in bulk as single use items, to repackager/relabeler establishments for further packaging and ethylene oxide (EtO) sterilization.
#### GRI Surgical Equipment Covers
GRI single use surgical equipment covers (both sterile and non-sterile) are protective barrier covers intended to be used to cover surgical equipment and provide a protective barrier for that equipment.
Non-sterile GRI surgical equipment covers are sold, in bulk as single use items, to repackager/relabeler establishments for further packaging and ethylene oxide (EtO) sterilization.
#### Technological Characteristics and Substantial Equivalence:
GRI single use drapes and surgical equipment covers, that is subject of this premarket notification is substantially equivalent to the predicate KC100 Surgical Drapes and Equipment Covers (K083234) in intended use statements, equivalent device construction, materials, drape design, chemical characteristics and technology performance characteristics. The performance attributes of the GRI single use drapes and surgical equipment covers are substantially equivalent to the predicate drape.
{2}------------------------------------------------
## K-111458
Image /page/2/Picture/1 description: The image shows the logo for GRI, which stands for Global Reporting Initiative. The logo is in black and features the lowercase letters "gri" in a stylized font. The "i" has a curved accent mark above it, and there is a trademark symbol to the bottom right of the "i".
#### Summary of Testing:
GRI Surgical Drapes and Equipment Covers have met acceptance criteria for performance testing including biocompatibility (cytotoxicity, primary skin irritation testing and sensitization testing), hydrostatic pressure, impact penetration, tear strength, tensile strength, linting, and flammability. All tests results met acceptance criteria and were substantially equivalent to the predicate devices.
#### Conclusion:
Based on the performance testing and device attributes, it can be concluded that the GRI surgical drapes and surgical equipment covers are equivalent to the predicate devices with respect to the indications for use and technological characteristics and do not present any further safety or effectiveness questions.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular arrangement of the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" surrounding a stylized eagle-like symbol. The symbol is composed of three curved lines that resemble the wings and body of a bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
GRI Medical & Electronic Technology C/O Ms. Penny Northcutt President/CEO REGSolutions, LLC 717 Lakeglen Drive Suwanee, Georgia 30024
DEC - 1 2011
Re: K111458
·
Trade/Device Name: GRI Sterile and Non-sterile Surgical Drapes GRI Sterile and Non-sterile Surgical Equipment Covers Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape and Drape Accessories Regulatory Class: II Product Code: KKX Dated: November 17, 2011 Received: November 17, 2011
Dear Ms. Northcutt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent.(for the_ indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Northcutt
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
## Indications for Use Form
### Indications for Use
510(k) Number (if known): _ K 11458
Device Name: Surgical Drapes (Model Numbers List Attached)
### GRI Sterile and Non-sterile Surgical Drapes (various models and configurations as attached)
GRI single use drapes (both sterile and non-sterile) are made of natural or synthetic material intended to be used as a protective patient covering, such as to isolate a te of surgical incision from microbial and other contamination.
Non-sterile GRI surgical drapes are sold in bulk as single use items, to repackager/ relabeler establishments for further packaging and ethylene oxide (EtO) sterilization.
The drapes are classified as level 3 per AAMI Standard PB70 Liquid barrier performance and classification of protective apparel and drapes intended for use in health-care facilities.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (OD)
Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111458
{6}------------------------------------------------
# |<||||458
Image /page/6/Picture/1 description: The image shows the logo for GRI, which is a company that specializes in sustainability reporting. The logo is a stylized version of the letters "gri" in a bold, sans-serif font. The "i" in the logo has a curved line above it, which gives the logo a modern and dynamic look. The logo is simple, yet effective, and it is easily recognizable.
| Surgical Drapes Indications for Use Model Number List | | | | |
|-------------------------------------------------------|-----------------------|--------------------------------------------------------------------------------|----------------------------------|----------------------|
| Code<br>(Sterile) | Code<br>(Non-Sterile) | Description | Fabric | Product<br>Family |
| 80-01101-S | 80-01101 | Femoral Angiography Drape (no window) | | |
| 80-01102-S | 80-01102 | Femoral Angiography Drape (single window) | | |
| 80-01103-S | 80-01103 | Femoral Angiography Drape (dual windows) | | |
| 80-01104-S | 80-01104 | Femoral Angiography Drape | | |
| 80-01105-S | 80-01105 | Femoral Angiography Drape (dual windows) Tri laminate material, Extra Long | | |
| 80-01106-S | 80-01106 | Femoral Angiography Drape (dual windows) Extra-long | 50gsm SMS | Angiography |
| 80-01107-S | 80-01107 | Femoral Angiography Drape (single window) with Trough | | |
| 80-01108-S | 80-01108 | Femoral Angiography Drape (dual windows) Tri laminate Material with Side Panel | | |
| 80-01109-S | 80-01109 | Femoral Angiography Drape (no window) Extra-long | | |
| 80-01110-S | 80-01110 | Femoral Angiography Drape (single window) Extra-long | | |
| 80-01201-S | 80-01201 | Femoral Angiography Drape (no window) | | |
| 80-01202-S | 80-01202 | Femoral Angiography Drape (single window) | | |
| 80-01203-S | 80-01203 | Femoral Angiography Drape (dual windows) | | |
| 80-01204-S | 80-01204 | Femoral Angiography Drape | | |
| 80-01205-S | 80-01205 | Femoral Angiography Drape (dual windows) Tri laminate material, Extra Long | | |
| 80-01206-S | 80-01206 | Femoral Angiography Drape (dual windows) Extra-long | 35gsm SMS | |
| 80-01207-S | 80-01207 | Femoral Angiography Drape (single window) with Trough | | |
| 80-01208-S | 80-01208 | Femoral Angiography Drape (dual windows) Tri laminate Material with Side Panel | | |
| 80-01209-S | 80-01209 | Femoral Angiography Drape (no window) Extra-long | | |
| 80-01210-S | 80-01210 | Femoral Angiography Drape (single window) Extra-long | | |
| 80-02101-S | 80-02101 | Bilateral Under Leg/Perineal Drape | | Cardiovascular |
| 80-02102-S | 80-02102 | CV Slip Drape | | |
| 80-02103-S | 80-02103 | Cardiovascular Incise W/Armboard cover | 50gsm SMS | |
| 80-02104-S | 80-02104 | Cardiovascular Universal set | | |
| 80-02105-S | 80-02105 | Clearview Split Drape | | |
| 80-02203-S | 80-02203 | Cardiovascular Incise W/Armboard cover | 35gsm SMS | |
| 80-02204-S | 80-02204 | Cardiovascular Universal set | | |
| 80-02205-S | 80-02205 | Clearview Split Drape | | |
| 80-03101-S | 80-03101 | Cystoscopy Drape | 50gsm SMS | Cystoscopy |
| 80-03102-S | 80-03102 | Cystoscopy T Drape | | |
| 80-03201-S | 80-03201 | Cystoscopy Drape | 35gsm SMS | |
| 80-03202-S | 80-03202 | Cystoscopy T Drape | | |
| 80-04101-S | 80-04101 | Fenestrated EENT Drape | | EENT Drape |
| 80-04104-S | 80-04104 | Head Drape | | |
| 80-04105-S | 80-04105 | Head Drape with Detachable Turban | 50gsm SMS | |
| 80-08102-S | 80-08102 | Eye Drape W/Fluid Pouch | | |
| 80-08103-S | 80-08103 | Basic ophthalmic Drape | | |
| 80-08104-S | 80-08104 | Ophthalmic Drape | | |
| 80-04201-S | 80-04201 | Fenestrated EENT Drape | | |
| 80-04204-S | 80-04204 | Head Drape | | |
| 80-04205-S | 80-04205 | Head Drape with Detachable Turban | 35gsm SMS | |
| 80-08202-S | 80-08202 | Eye Drape W/Fluid Pouch | | |
| 80-08203-S | 80-08203 | Basic ophthalmic Drape | | |
| 80-08204-S | 80-08204 | Ophthalmic Drape | | |
| 80-05101-S | 80-05101 | Side Drape 29.5x39 in | | General<br>Surgery |
| 80-05102-S | 80-05102 | Bottom Drape 69x71 in | | |
| 80-05104-S | 80-05104 | Top Drape 59x94.5 in | | |
| 80-05107-S | 80-05107 | U-Drape | | |
| 80-05109-S | 80-05109 | Bottom drape 72x75 in double cord holders | | |
| 80-05110-S | 80-05110 | Top drape 50x108 in double cord holders | | |
| 80-05111-S | 80-05111 | Breast Chest U Drape | | |
| 80-05112-S | 80-05112 | Breast Drape | 50gsm SMS | |
| 80-05113-S | 80-05113 | Bar Drape 96x44 in w/ Arm Board Covers | | |
| 80-05114-S | 80-05114 | Bar Drape 106x62 in w/ Arm Board Covers | | |
| 80-05115-S | 80-05115 | Ortho Bar Drape | | |
| 80-05116-S | 80-05116 | Laparotomy Drape 100x124 in | | |
| 80-05117-S | 80-05117 | Laparotomy 104x120 in | | |
| 80-05118-S | 80-05118 | Transverse Laparotomy Drape | | |
| 80-05119-S | 80-05119 | Pediatric Laparotomy Drape | | |
| 80-05120-S | 80-05120 | Thyroid Drape | | |
| | | | | |
| 80-05122-S | 80-05122 | Chest Drape | | |
| 80-05123-S | 80-05123 | Laparoscopic Drape | | |
| 80-05125-S | 80-05125 | Laparoscopic Colecystectomy Drape | | |
| 80-05128-S | 80-05128 | Side drape 38.5x77 in | | |
| 80-05129-S | 80-05129 | Side Drape 90x90 in | | |
| 80-05201-S | 80-05201 | Side Drape 29.5x39 in | | |
| 80-05202-S | 80-05202 | Bottom Drape 69x71 in | | |
| 80-05204-S | 80-05204 | Top Drape 59x94.5 in | | |
| 80-05207-S | 80-05207 | U-Drape | | |
| 80-05209-S | 80-05209 | Bottom drape 72x75 in double cord<br>holders | | |
| 80-05210-S | 80-05210 | Top drape 50x108 in double cord<br>holders | | |
| 80-05211-S | 80-05211 | Breast Chest U Drape | | |
| 80-05212-S | 80-05212 | Breast Drape | | |
| 80-05213-S | 80-05213 | Bar Drape 96x44 in w/ Arm Board<br>Covers | | |
| 80-05214-S | 80-05214 | Bar Drape 106x62 in w/ Arm Board<br>Covers | 35gsm<br>SMS | |
| 80-05215-S | 80-05215 | Ortho Bar Drape · | | |
| 80-05216-S | 80-05216 | Laparotomy Drape 100x124 in | | |
| 80-05217-S | 80-05217 | Laparotomy 104x120 in | | |
| 80-05218-S | 80-05218 | Transverse Laparotomy Drape | | |
| 80-05219-S | 80-05219 | Pediatric Laparotomy Drape | | |
| 80-05220-S | 80-05220 | Thyroid Drape | | |
| 80-05222-S | 80-05222 | Chest Drape | | |
| 80-05223-S | 80-05223 | Laparoscopic Drape | | |
| 80-05225-S | 80-05225 | Laparoscopic Colecystectomy Drape | | |
| 80-05228-S | 80-05228 | Side drape 38.5x77 in | | |
| 80-05229-S | 80-05229 | Side Drape 90x90 in | | |
| 80-07101-S | 80-07101 | C-Section Fluid Collection Drape<br>Fenestrated Incise | | |
| 80-07102-S | 80-07102 | C-Section Drape with Fenestrated<br>Incise | 50gsm<br>SMS | Obstetrics |
| 80-07103-S | 80-07103 | C-Section Drape with full Incise | | |
| 80-07104-S | 80-07104 | C-Section Fluid Collection Drape full<br>Incise | | |
| 80-07201-S | 80-07201 | C-Section Fluid Collection Drape<br>Fenestrated Incise | | |
| 80-07202-S | 80-07202 | C-Section Drape with Fenestrated<br>Incise | 35gsm<br>SMS | |
| 80-07203-S | 80-07203 | C-Section Drape with full Incise | | |
| 80-07204-S | 80-07204 | C-Section Fluid Collection Drape full<br>Incise | | |
| 80-09101-S | 80-09101 | Lower Extremity Drape | 50gsm | Orthopedics |
| 80-09104-S | 80-09104 | Hip Drape W/Pockets | | |
| 80-09105-S | 80-09105 | Body Split Drape | | |
| 80-09106-S | 80-09106 | Split Drape | | |
| 80-09107-S | 80-09107 | U-Drape | | |
| 80-09108-S | 80-09108 | Hand Drape | | |
| 80-09109-S | 80-09109 | Shoulder Arthroscopy Drape | | |
| 80-09110-S | 80-09110 | Shoulder Split Drape | | |
| 80-09111-S | 80-09111 | Beach Chair Shoulder | | |
| 80-09201-S | 80-09201 | Lower Extremity Drape | | |
| 80-09202-S | 80-09202 | Knee Arthroscopy Drape | | |
| 80-09204-S | 80-09204 | Hip Drape W/Pockets | | |
| 80-09205-S | 80-09205 | Body Split Drape | 35gsm | |
| 80-09207-S | 80-09207 | U-Drape | SMS | |
| 80-09208-S | 80-09208 | Hand Drape | | |
| 80-09209-S | 80-09209 | Shoulder Arthroscopy Drape | | |
| 80-09210-S | 80-09210 | Shoulder Split Drape | | |
| 80-11101-S | 80-11101 | Lithotomy Drape | | |
| 80-11102-S | 80-11102 | Lithotomy T-Drape | 50gsm | |
| 80-11105-S | 80-11105 | Leggings Drape 54x31 in | SMS | |
| 80-11201-S | 80-11201 | Lithotomy Drape | | Peri-GYN |
| 80-11202-S | 80-11202 | Lithotomy T-Drape | 35gsm | |
| 80-11205-S | 80-11205 | Leggings Drape 54x31 in | SMS | |
| 80-12101-S | 80-12101 | Half Drape | | |
| 80-12102-S | 80-12102 | Medium Drape | 50gsm | |
| 80-12103-S | 80-12103 | Large Drape | SMS | |
| 80-12104-S | 80-12104 | XL Drape | | |
| 80-12201-S | 80-12201 | Half Drape | | Ancillary |
| 80-12202-S | 80-12202 | Medium Drape | 35gsm | |
| 80-12203-S | 80-12203 | Large Drape | SMS | |
| 80-12204-S | 80-12204 | XL Drape | | |
| 85-07341-S | 85-07341 | 7"<br>with pull tab Stockinette Drape | | |
| 85-09481-S | 85-09481 | 9" with pull tab Stockinette Drape | | |
| 85-12481-S | 85-12481 | 12" with pull tab Stockinette Drape | | |
| 85-16481-S | 85-16481 | 16" with pull tab Stockinette Drape | Polyester<br>knit<br>stockinette | Stockinette<br>Drape |
| 85-07340-S | 85-07340 | 7" without pull tab Stockinette Drape | | |
| 85-09480-S | 85-09480 | 9" without pull tab Stockinette Drape | | |
| 85-12480-S | 85-12480 | 12" without pull tab Stockinette Drape | | |
| 85-16480-S | 85-16480 | 16" without pull tab Stockinette Drape | | |
፡
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for GRI, which stands for Global Reporting Initiative. The logo is in black and white and features the lowercase letters "gri" in a stylized font. There is a curved line above the "i" that resembles a crescent moon. The letters are connected and have a modern, sleek appearance.
{8}------------------------------------------------
# KII1458
Image /page/8/Picture/1 description: The image shows the logo for GRI, which is written in a stylized lowercase font. The letters 'g', 'r', and 'i' are connected and written in black. The 'i' has a curved line above it, resembling a crescent moon. The letters are slightly slanted to the right.
{9}------------------------------------------------
K111458
Image /page/9/Picture/1 description: The image shows the logo for GRI, which is a company that specializes in sustainability reporting. The logo is a stylized version of the letters "gri" in a sans-serif font. The letters are black, and the "i" has a curved top. The logo is simple and modern.
### Key for Surgical Drape Codes:
80=Surgical drape set
Prefix
80=Surgical drape series
Suffix - 1&2
-
- First two positions represent product category
- 01 Angiography
- 02 Cardiovascular
- 03 Cystoscopy
04 EENT
-
{10}------------------------------------------------
- 05 General Surgery
- 07 Obstetrics
- 09 Orthopedic
- 11 Peri-GYN
- 12 Ancillary
- 14 Accessories
- Suffix 3
- Third position represents "fabric" 1 = SMS - 50gsm
- 2 = SMS 35gsm
- Suffix -- 4&5
- Fourth and Fifth positions represent Product identification in sequential order of development Sixth position represents "packaging"
- . Suffix 6
- S = Sterile
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for GRI, which is a company that specializes in sustainability reporting. The logo is in black and white and features the letters "gri" in a stylized font. The "i" in the logo has a curved line above it, which gives the logo a unique look. The logo is simple and modern, and it is easily recognizable.
## Indications for Use Form
Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Surgical Equipment Covers (Model Numbers List Attached)
GRI Sterile and Non-sterile Surgical Equipment Covers (various models and configurations as attached)
GRI single use surgical equipment covers (both sterile and non-sterile) are protective barrier covers intended to be used to cover surgical equipment and provide a protective barrier for that equipment.
Non-sterile GRI surgical equipment covers are sold in bulk as single use items, to repackager/ relabeler establishments for further packaging and ethylene oxide (EtO) sterilization.
The covers are classified as level 3 per AAMI Standard PB70 Liquid barrier performance and classification of protective apparel and drapes intended for use in health-care facilities.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K/11458
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for GRI, which is a company that specializes in sustainability reporting. The logo is a stylized lowercase "gri" in black. There is a curved line above the "i" that resembles a check mark.
| Surgical Equipment Covers | Model Number | Indications for Use | List | |
|---------------------------|--------------|-----------------------------|-------|----------------------|
| 80-13101-S | 80-13101 | Back Table Cover 44x75in | | |
| 80-13102-S | 80-13102 | Back Table Cover 44x90in | | |
| 80-13103-S | 80-13103 | Back Table Cover 50x90in | | |
| 80-13104-S | 80-13104 | Back Table Cover 80x90in | 43gsm | Back Table<br>Covers |
| 80-13107-S | 80-13107 | Back Table Cover 55x75in | SMS | |
| 80-13108-S | 80-13108 | Back Table Cover 60x75in | | |
| 80-13109-S | 80-13109 | Back Table Cover 60x90in | | |
| 80-13110-S | 80-13110 | Back Table Cover 44x44in | | |
| 80-13105-S | 80-13105 | Mayo Stand Cover 23x54in | 43gsm | Mayo Stand<br>Covers |
| 80-13106-S | 80-13106 | Mayo Stand Cover 29.5x57in | SMS | |
| 80-14101-S | 80-14101 | Underbuttocks Drape w/Pouch | | |
| 80-14102-S | 80-14102 | Underbuttocks Drape 48x36in | 50gsm | Accessories |
| 80-14103-S | 80-14103 | Drape Underbuttocks W/Pouch | SMS | |
| 80-14104-S | 80-14104 | Underbuttocks Drape 38x44in | | |
### Key for Surgical Drape Codes:
80=Surgical drape series
Suffix - 1&2
Prefix
bo-Surgical drape series
First two positions represent product category
- 01 Angiography
- 02 Cardiovascular
- 03 Cystoscopy
- 04 EENT
- 05 General Surgery
- 07 Obstetrics
- 09 Orthopedic
- 11 Peri-GYN
- 12 Ancillary
- 14 Accessories
Suffix - 3
- 1 = SMS 50gsm
- 2 = SMS 35gsm
- Suffix 4&5 Fourth and Fifth positions represent Product
Suffix — 6
- identification in sequential order of development Sixth position represents "packaging"
Third position represents "fabric"
- S = Sterile
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.