B1P Adhesive is an adhesive designed to be used for all direct and indirect applications including, but not limited to, the following: Direct Applications - Direct bonding to dentin and/ or enamel . - Bonding composite to composite, porcelain and/ or metal Opaquing metal (used . in conjunction with an opaque resin stain or cement) - Bonding with dual cure and self cure composite resins, such as cements and core . build up materials - Amalgam sealing . - Dentin sealing . Indirect applications - Indirect bonding of all composite, all ceramic, PFM and alloy crowns, bridges, . inlays, onlays and veneers (used in conjunction with resin luting agents) - Bonding of fiber and metal posts (in conjunction with resin luting agents) .
Device Story
B1P Adhesive is a fifth-generation dental bonding agent used by dentists for direct and indirect restorative procedures. It functions via a total-etch technique, combining primer and adhesive into a single component. The device is light-curable, with an optional dual-cure capability when mixed with a dual-cure activator. It incorporates glutaraldehyde as a desensitizer and utilizes filler technology to enhance bond strength. The adhesive is applied in a single coat following acid etching and rinsing. It is available in both bottle and single-dose configurations and is stored at ambient temperature. By facilitating the adhesion of restorative materials (composites, ceramics, metals) to tooth structure or other restorations, the device enables the repair and reconstruction of teeth, potentially improving clinical outcomes for patients requiring dental restorations.
Clinical Evidence
No clinical data. Evidence consists of biocompatibility testing per ISO 10993 and bench testing comparing direct and indirect bonding strengths to predicate devices.
Technological Characteristics
Fifth-generation dental bonding agent; total-etch technique; light-curable (optional dual-cure with activator); contains glutaraldehyde (desensitizer) and filler technology; acetone-based; single-coat application; ambient temperature storage; bottle or single-dose form factor. Biocompatibility per ISO 10993.
Indications for Use
Indicated for direct bonding to dentin/enamel; bonding composite to composite, porcelain, or metal; opaquing metal; bonding with dual/self-cure composite resins; amalgam and dentin sealing; and indirect bonding of crowns, bridges, inlays, onlays, veneers, and fiber/metal posts.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
Predicate Devices
Dentin Conditioning and Adhesive (Bond 1) (K973388)
{0}------------------------------------------------
K111431
# AUG - 1 2011
# Section III - 510(k) Summary of Safety and Effectiveness
# Submitter:
Sybron Dental Specialties, Inc. 1717 W. Collins Avenue Orange, California 92867 (714) 516-7602 - Phone (714) 516-7472 - Facsimile Wendy Garman - Contact Person
Date Summary Prepared: May 2011
#### Device Name:
- . Trade Name - B1P Adhesive
- . Common Name – Bonding Agent
- . Classification Name - Resin Tooth Bonding Agent, per 21 CFR § 872.3200
- . Product Code: KLE
## Devices for Which Substantial Equivalence is Claimed:
- . Pentron Clinical, Dentin Conditioning and Adhesive (Bond 1), K973388
- . Kerr Corporation, Optibond Solo Plus 2, K991808
- Heraeus Kulzer GmbH, IBond Total Etch, K083652
# Device Description:
B1P Adhesive is a fifth generation dental bonding agent that is intended to be used for direct composite bonding and indirect restoration in combination with a dental restorative material. The formula is used based on a total etch technique and combines the primer and adhesive into one component. The application procedure requires steps of acid etching and rinsing, followed by the application of the B1P Adhesive. It is a light curable adhesive with the option of being dual cured when mixed with the dual cure activator.
{1}------------------------------------------------
# Intended Use of the Device:
B1P Adhesive is an adhesive designed to be used for all direct and indirect applications including, but not limited to, the following:
# Direct Applications
- Direct bonding to dentin and/ or enamel .
- Bonding composite to composite, porcelain and/ or metal Opaquing metal (used . in conjunction with an opaque resin stain or cement)
- Bonding with dual cure and self cure composite resins, such as cements and core . build up materials
- Amalgam sealing .
- Dentin sealing .
# Indirect applications
- Indirect bonding of all composite, all ceramic, PFM and alloy crowns, bridges, . inlays, onlays and veneers (used in conjunction with resin luting agents)
- Bonding of fiber and metal posts (in conjunction with resin luting agents) .
# Summary of Technological Characteristics:
B1P Adhesive is substantially equivalent to three other legally marketed devices in the United States: Dentin Conditioning and Adhesive (Bond 1), Optibond Solo Plus 2 and lBond Total Etch.
B1P Adhesive functions in a manner similar to and is intended for the same use as Dentin Conditioning and Adhesive (Bond 1) that is currently marketed by Pentron Clinical. Both products are acetone based and utilize the same adhesive monomer. B1P Adhesive differs in that it incorporates glutaraldehyde as a desensitizer, can be stored at ambient temperature, has both a bottle and a single dose configuration, and requires the use of only one coat.
B1P Adhesive functions in a manner similar to and is intended for the same use as Optibond Solo Plus 2 that is currently marketed by Kerr Corporation. Both products can be stored at ambient temperature, are both applied in a single coat, utilize filler technology to increase bond strength, and are both offered in a bottle and a single dose configuration. B1P Adhesive differs in that it uses a different adhesive monomer and it contains glutaraldehyde as a desensitizer.
B1P Adhesive functions in a manner similar to and is intended for the same use as IBond Total Etch that is currently marketed by Heraeus Kulzer GmbH. Both products contain glutaraldehyde as a desensitizing agent, require the use of only one coat, bond to both dentin and enamel, employ filler technology to increase bond strength, require the use
{2}------------------------------------------------
of an etching agent and can both be stored at ambient temperature. B1P Adhesive differs in that it may be dual cured when mixed with the dual cure activator, and employs different main adhesive components.
# Non-Clinical Test Data
Biocompatibility studies have been completed according to ISO 10993, which demonstrates that B1P Adhesive is safe for its intended use.
This 510(k) submission also includes data from bench testing used to evaluate the performance characteristics of B1P Adhesive compared to the predicate devices. The characteristics evaluated include direct and indirect bonding strengths.
# Clinical Testing
Clinical testing has not been conducted on this product.
#### Conclusion:
Based upon biocompatibility tests, similar technological characteristics to the predicate devices and bench testing, the clinical performance of B1P Adhesive is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle-like bird with three curved lines representing its wings. The bird is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Pentron Clinical C/O Ms. Wendy Garman Director, Regulatory Affairs Sybron Dental Specialties, Incorporated 1717 West Collins Avenue Orange, California 92867
AUG - 1 2011
Re: K111431
Trade/Device Name: BIP Adhesive Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: May 20, 2011 Received: May 23, 2011
Dear Ms. Garman: ·
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 – Ms. Garman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
( 11 431
Device Name: B1P Adhesive
Indications For Use:
B1P Adhesive is an adhesive designed to be used for all direct and indirect applications including, but not limited to, the following:
# Direct Applications
- Direct bonding to dentin and/ or enamel .
- Bonding composite to composite, porcelain and/ or metal .
- . Opaquing metal (used in conjunction with an opaque resin stain or cement)
- Bonding with dual cure and self cure composite resins, such as cements and core build up . materials
- Amalgam sealing
- Dentin sealing ●
# Indirect applications
- Indirect bonding of all composite, all ceramic, PFM and alloy crowns, bridges, inlays, ● onlays and veneers (used in conjunction with resin luting agents)
- Bonding of fiber and metal posts (in conjunction with resin luting agents) ●
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
rrence of CDRH, Office of Device Evaluation (ODE)
Susan Fraser
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111431
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.