K111423 · Carestream Health, Inc. · NBL · Mar 2, 2012 · Dental
Device Facts
Record ID
K111423
Device Name
CS 1600 INTRAORAL CAMERA
Applicant
Carestream Health, Inc.
Product Code
NBL · Dental
Decision Date
Mar 2, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1745
Device Class
Class 2
Indications for Use
The CS 1600 Intraoral Camera (CS 1600) is intended for chairside use by health professionals during oral health examinations. It provides magnified digital color images of intraoral or extraoral anatomy via a video monitor or Personal Computer (PC). Both still and video images can be captured and stored. In Caries Detection mode, the CS 1600 is intended as an aid in the detection and diagnosis of dental caries. It provides both video mode and still image detection mode. In still image mode, it provides a color-coded indication of the degree of fluorescence loss for affected areas of the tooth. The digital video and still images may be digitally recorded for use with computerized patient records.
Device Story
CS 1600 is an intraoral camera system for dental clinical settings. It captures intraoral/extraoral digital color video and still images. In caries detection mode, it uses fluorescence imaging with reflectance enhancement to identify dental caries. The system includes a camera handpiece with a 0.25-inch CCD sensor, connected via USB to a PC. Accessories include hygienic barrier sheaths and a collar attachment for maintaining optimal working distance. Healthcare professionals use the device chairside to view images on a monitor. The system provides color-coded visual feedback regarding fluorescence loss in affected tooth areas. This aids clinicians in diagnosing caries and facilitates patient communication by showing treatment areas before and after procedures. Images are stored in computerized patient records.
Clinical Evidence
Clinical testing was conducted to verify the efficacy of the caries detection modes under clinical conditions. Results confirmed the device's performance for its intended use. No specific sensitivity, specificity, or AUC metrics were provided in the summary.
Technological Characteristics
Intraoral camera system featuring a 0.25-inch CCD sensor. Connectivity via USB to a PC. Includes hygienic barrier sheaths and a collar attachment for working distance maintenance. Operates via fluorescence imaging with reflectance enhancement. Complies with international standards for medical device safety, electrical, mechanical, and EMC/EMI requirements.
Indications for Use
Indicated for use by health professionals as an aid in the detection of dental caries and for viewing/capturing intraoral color video images to magnify oral cavity regions, assist patient communication, and document patient records.
Regulatory Classification
Identification
A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.
1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;
2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.
{0}------------------------------------------------
*Carestream Health, Inc.*
K///423
MAR - 2 2012 1 of 3
# 510(k) Summary CS 1600 Intraoral Camera
#### 1. Company Identification
Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 Establishment Registration: 1315356
#### 2. Contact Person
Daniel Hoefer Manager, Regulatory Affairs, Carestream Dental 1765 The Exchange Atlanta, GA 30339 Tel 770 226 3287 Fax 770 850 5011
#### 3. Device Name
Commercial name: CS 1600 Intraoral Camera Common name: Intraoral Camera with Fluorescence Caries Detection Classification name: Laser Fluorescence Caries Detection Device
#### 4. Device Classification
Class: II Product Code: NBL
#### 5. Intended Use
The CS 1600 Intraoral Camera (CS 1600) is intended for chairside use by health professionals during oral health examinations. It provides magnified digital color images of intraoral or extraoral anatomy via a video monitor or Personal Computer (PC). Both still and video images can be captured and stored.
In Caries Detection mode, the CS 1600 is intended as an aid in the detection and diagnosis of dental caries. It provides both video mode and still image detection mode. In still image mode, it provides a color-coded indication of the degree of fluorescence loss for affected areas of the tooth.
The digital video and still images may be digitally recorded for use with computerized patient records.
#### 6. Device Description
The CS 1600 Intraoral Camera (CS 1600) is an intraoral camera system that also includes an optical caries detection system based on fluorescence imaging with reflectance
{1}------------------------------------------------
K111423
Carestream Health, Inc.
2 of 3
enhancement. The system is intended for use by healthcare professionals in dental and dental sub-specialty clinical settings.
The system consists of an intraoral camera assembly which is connected via USB connection to a PC, and associated image acquisition software and accessories. Accessories include hygienic barrier sheaths and a collar attachment for the camera assembly that is used for maintaining an optimal working distance in either caries detection mode.
#### 7. Substantial Equivalence
The CS 1600 is substantially equivalent the Kodak 1500 Intraoral Camera (Carestream Health, Inc., K091186), to the Inspektor Pro (Inspektor Dental Care by, K040063) and to the SoproLife (K092583, Sopro).
- The Kodak 1500, SoproLife, and CS 1600 are each intended for use in . dental or dental sub-specialty intraoral video image acquisition, for the purposes of viewing the oral cavity and patient communication.
- . The CS 1600, Inspektor Pro, and SoproLife are each intended for use as an aid in the detection and/or diagnosis of dental caries. The principle of operation for each fluorescence caries detection device is substantially equivalent.
- The technological characteristics are equivalent. Each device consists of a ● digital video camera with 0.25 inch CCD sensor assembled within a camera hand piece. Each connects either to a docking station or directly to a computer via USB, and displays on a chairside monitor. Video formats are similar.
- The intended users of each device are the same or equivalent. Each is ● intended for use by dentists and other oral health specialists.
- All of the devices are substantially equivalent in terms of energy used or . delivered, labeling, materials, biocompatibility, workflow, and technology.
- All of the devices conform to equivalent international standards for safety of . medical devices.
#### 8. Non-Clinical testing
Bench top performance testing of the CS 1600 was performed on Hardware sub-systems, Software sub-systems, and on the complete assembled device. The system also has undergone laboratory testing for mechanical, electrical, biocompatibility and EMC/EMI compliance with standards.
Results of testing demonstrate that the device is safe and effective in meeting user requirements in accordance with its intended use.
{2}------------------------------------------------
K111423
3073
### 9. Clinical Testing
Clinical testing of the system in each caries detection mode has been completed. The conclusion of the study verified the efficacy for the intended use of the device under clinical conditions.
#### 10. Conclusion
The CS 1600 Intraoral Camera is substantially equivalent to the predicate devices listed above. Testing has demonstrated that it is safe and effective for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized eagle with three wing-like extensions, suggesting movement or flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Carestream Health, Inc. C/O Mr. Daniel Hoefer Manager, Regulatory Affairs, Carestream Dental Carestream Dental, LLC 1765 The Exchange Atlanta, Georgia 30339
MAR - 2 2012
Re: K111423
Trade/Device Name: CS 1600 Intraoral Camera Regulation Number: 21 CFR 872.1745 Regulation Name: Laser Fluorescence Caries Detection Device Regulatory Class: II Product Code: NBL Dated: February 23, 2012 Received: February 24, 2012
Dear Mr. Hoefer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Hoefer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for Anthony, DiNotson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital; Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
f
KIII423
## Indications for Use
The CS 1600 is indicated for use by health professionals as an aid in the detection of dental caries.
It is also indicated for use in viewing and capturing intraoral color video images for the purpose of:
- Allowing practitioners to view and magnify all regions of the oral cavity .
- Assisting in communication with the patient by providing a view of treatment areas before and after . a procedure
- Providing images for documentation in patient records.
#### Concurrence of CDRH, Office of Device Evaluation
Prescription Use √
(Per 21 CFR 801.109)
OR
Over-The-Counter
Susan Ruanes
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.