K111413 · Molnlycke Health Care Us, LLC · KGO · Jan 30, 2012 · General, Plastic Surgery
Device Facts
Record ID
K111413
Device Name
BIOGEL (R) INDICATOR (R) UNDERGLOVE
Applicant
Molnlycke Health Care Us, LLC
Product Code
KGO · General, Plastic Surgery
Decision Date
Jan 30, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The Biogel® PI Indicator® Underglove is a disposable device made of polyisoprene, blue in color, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.
Device Story
Disposable polyisoprene surgical glove; blue color; worn on hands in surgical settings. Provides barrier against infectious materials and contaminants. Manufactured with Biogel® coating and added surfactant. Used by surgeons and clinical staff to protect against cross-contamination. Benefit: barrier protection during surgical procedures.
Clinical Evidence
No clinical data was required; bench testing only.
Technological Characteristics
Polyisoprene material; blue pigmentation; Biogel® coating with added surfactant. Meets ASTM D3577 (dimensions, physical properties, freedom from holes), ISO 10993-1 (biocompatibility), and ASTM D7102 (LAL test).
Indications for Use
Indicated for use as a disposable surgical glove to provide a barrier against infectious materials and contaminants for individuals in a surgical setting.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Biogel® PI Indicator® Underglove (K081180 - originally cleared under Skinsense Polyisoprene Underglove)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Molnlycke Health Care. The logo consists of a cluster of five circular shapes on the left, resembling cells or bubbles. To the right of the shapes, the text "MOLNLYCKE HEALTH CARE" is displayed in two lines.
# (11413
JAN 3 0 2012
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
# Biogel® PI Indicator® Underglove
This SIO(k) summary of safety and effectiveness information'is being subnitted in accordance with the requirements of 21 CFR 807.92(c).
Date Prepared:
August 25, 2011
Applicant:
Mölnlycke Health Care US, LLC
Angela L. Bunn, RAC
Tel: 678-250-7930
5550 Peachtree Parkway, Suite 500 Norcross, GA 30092 Registration number: 3004763499 Owner/Operator Number: 9067000
Director, Regulatory Affairs for the Americas
Official Correspondent:
Fax: 678-245-7746 e-mail: angela.bunn@molnlycke.com Trade/Proprietary Name: Biogel® PI Indicator Underglove Common Name: Surgeon's Glove Classification Name: Surgeon's Glove Device Class: Class 1 Regulation Nümber: 21 CFR 878.4460 Product Code: KGO Predicate Device Name(s): Biogel® PI Indicator® Underglove (K081180 - originally cleared under Skinsense Polyisoprene Underglove)
Description of Device:
The proposed device, the Biogel PI Indicator Underglove is manufactured of polyisoprene colored writh blue pigmentation. The Blogel PI Indicator Underglove is namifactured of the exact same material and coaled with *the Biogel® Coating which is used on the currently cleared device that has been legally marketed by M8lhycke Health Care for many years with the addition of a surfactant.
Intended Use/Indication for Use:
The Bloged® PI Indicator 4 Underglove is a disposable device made of polyisoprene, blue in color, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other containinants.
> Holnlycke Health Care 5550 Peachired Parkway. Suite 500 Norcross: GA 30092 USA
011 678 2511 7940 Fax 673 250 7904 Page 18 BP20 % ke.con
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Molnlycke Healthcare. The logo consists of a cluster of five circles on the left side, arranged in a pattern resembling a flower or a cell cluster. To the right of the circles, the text "MOLNLYCKE HEALTHCARE" is displayed in a simple, sans-serif font. The overall design is clean and professional, suggesting a focus on health and medical solutions.
#### Technological Characteristics:
The Biogel® Pl Indicator Underglove is substantially equivalent to the Biogel® PI Indicator Underglove (KOST 180 originally cleared under Skinsense Polyisoprene Underglove). The assessed devices have the same indications for. use, materials, product design, labeling claims and method of operation.
The only difference in the proposed device, Biogel® PI Indicator® Underglove is the addition of a surfactant.
The Bioget® Plandicator Underglove characteristics are summarized below as compared to ASTW requirements.
Characteristic Dimensions Physical Properties Freedom from Holes Biocompatibility LAL Test Results
Standard Meets ASTM D3577 Meets ASTM D3577 Meets ASTM D3577 Meets ISO 10993-1 ASTM D7102
#### Performance Data!
The performance data are summarized above.
#### Clinical Testing:
No clinical data was required:
#### Conclusion:
Based on the performance testing, it can be concluded that the Biogel® Indicator® Underglove is equivalent to the Biogel® Pl. Indicator® Underglove. (K031180 - originally cleared under Skinsense Poly isoprene Underglove) predicate with respect to intended use, materials; design, and technological characteristics:
> Mointycke Health Care 101 676-290 7900 5550 Peachtree Parkway, Suile 500 Fax 678 250 7984 Norce oss: GA 30092 USA Page 44 88 8 8 x 8 con
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of three human figures, with their heads tilted back and arms raised.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Angela L. Bunn, RAC Director. Regulatory Affairs for the Americas Molnlycke Health Care, US LLC 5550 Peachtree Parkway, Suite 500 Norcross, Georgia 30092
JAN 3 0 2012
Re: K111413
Trade/Device Name: Biogel® PI Indicator® Underglove Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: January 13, 2012 Received: January 17, 2012
Dear Ms. Bunn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Ms. Bunn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542.of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for Anthony D. Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Molnlycke Healthcare. The logo consists of a cluster of five circles on the left side, arranged in a pattern resembling a bunch of grapes. To the right of the circles, the text "MOLNLYCKE HEALTHCARE" is written in a simple, sans-serif font, with "MOLNLYCKE" on the top line and "HEALTHCARE" on the bottom line.
### INDICATIONS FOR USE
510(k) Number (if known):
K 1.1/1413.
Biogel® PI Indicator® Underglove Device Name:
Indications For Use:
The Biogel® PI Indicator® Underglove is a disposable device made of polyisoprene, blue in color, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier. against potentially infectious material and other contaminants.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth D. Clauser-Williams
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
111412
Page.1 of i
678 250 7784
Page & Bryglycke.com
olnlycke Health Care
Nortross, GA 30092 USA
50 Peachtree Parkway, Suil
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.