The BEVER™ intermittent catheter is indicated for intermittent catheterization of the urethra for those individuals who are unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder -voiding episode. The catheter is inserted into urethra to reach the bladder allowing urine to drain.
Device Story
BEVER™ Intermittent Catheter is a sterile, single-use device for bladder drainage. Available in uncoated and hydrophilic-coated variants; features Nelaton (straight/rounded) or Tiemann (curved/tapered) tips; includes two drainage eyelets. Coated version includes a sterile water packet for activation, creating a slippery surface to reduce urethral friction. Used by patients or clinicians for intermittent self-catheterization to drain residual urine. Device is inserted through the urethra into the bladder. Benefits include reduced friction during insertion and potentially lower risk of long-term complications associated with repeated catheterization.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-5 (in vitro cytotoxicity), ISO 10993-10 (vaginal irritation and guinea pig maximization sensitization tests). Packaging integrity validated via ISTA Procedure 2A. Sterilization validated per ISO 11135-1 and ISO 11137 series. Shelf life validated via real-time and accelerated aging studies (3 years for coated, 4 years for uncoated).
Technological Characteristics
Materials: PVC shaft, PVP hydrophilic coating. Tip configurations: Nelaton (straight/rounded), Tiemann (curved/tapered). Sizes: 6Fr-22Fr. Sterilization: EO (catheter), radiation (water packet). Standards: EN 1616:1997, EN 1618:1997, ISO 10993, ISO 11135-1, ISO 11137. Connectivity: None.
Indications for Use
Indicated for intermittent urethral catheterization in men, women, and children unable to promote natural urine flow or with significant residual urine volume following bladder voiding.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Astra Tech AB LoFric® Primo™ Single Use Urinary Catheter (K050874)
Submission Summary (Full Text)
{0}------------------------------------------------
Hangzhou Bever N ical Devices Co., Ltd.
Add: No. 8-1, Longquan Rd., Cangqian Town, Yuhang District, Hangzhou, China Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515
# 510(k) Summary
Date of summary: D December 30, 2011
# 1. Submitter (Owner) of 510 (k):
Hangzhou Bever Medical Devices Co., Ltd. No. 8-1, Longquan Rd., Cangqian Town, Yuhang District 311121, Hangzhou, China Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515 Registration Number: 3008729910
# 2. Contact person:
Allyson Zhou Management Representative Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515 Email: zlshio1224@sina.com
### 3. Device Name:
Common Name: Intermittent Urethral Catheter Trade Name: BEVER™ Intermittent Catheter Classification name: Urological catheter and accessories (21 CFR 876.5130, Product Code GBM)
11:10:1
# 4. Predicate Device:
BEVER is claiming substantial equivalence to the following medical devices: Rüsch Intermittent Urethral Catheters - K033023 Astra Tech AB LoFric® Primo™ Single Use Urinary Catheter - K050874
# 5. Device Description
BEVERTM Intermittent Catheter is sterile, single use device to be designed as an intermittent pathway for drainage of the bladder. It is available for men, women and children, in uncoated and coated variants and in two different tip configurations of Nelaton (straight and rounded) and Tiemann (curved and tapered) respectively. There are two polished drainage eyelets on the catheter in various configurations and types.
The uncoated catheter consists of a tubular polyvinyl chloride catheter shaft with attached a drainage funnel. The catheter is available in sizes 6Fr ~ 22Fr in 2Fr increments for Nelaton-tip and sizes 8Fr ~ 22Fr in 2Fr increments for Tiemann-tip.
The coated catheter consists of a tubular polyvinyl chloride catheter shaft, coated with a hydrophilic low-friction coating, with attached a drainage funnel, and a sterile water packet is placed in the package. The surface of coated catheter is hydrophilic and when the coated catheter is activated with the sterile water in the attached water
{1}------------------------------------------------
KIII405
pg. 2 of 4
packet, it becomes slippery and thus reduces friction against the urethra. This allows the coated catheter to slide in and out of the urethra in the most comfortable way, And, just as importantly, there's less risk of long-term complications from the repeated friction of self-catheterizing. The coated catheter is available in sizes 6Fr ~ 22Fr in 2Fr increments for Nelaton-tip and sizes 8Fr ~ 22Fr in 2Fr increments for Tiemann-tip.
### 6. Indications for Use
The BEVERTM Intermittent Catheter is indicated for intermittent catheterization of the urethra for those individuals who are unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder - voiding episode. The catheter is inserted into urethra to reach the bladder allowing urine to drain.
#### 7. Substantial Equivalence
The BEVERTM Intermittent Catheter maintains the same intended use as the predicate devices. It is a device that is inserted through the urethra and used to drain urine from the bladder.
The BEVERTM Intermittent Catheter is composed essentially of the same materials as the predicate devices. The material for both devices is PVC and PVP for coating.
The BEVER™ Intermittent Catheter has the same design and performance characteristic as the predicate devices.
#### 8. Device Performance
The dimension, design, material, sterility, packaging and labeling of BEVERTM Intermittent Catheter are conformed with EN 1616:1997.
According to the device performance testing protocol, the device is compliance with EN 1616:1997, EN 1618:1997.
#### 9. Summary of Testing
# 9.1 Biocompatibility testing was performed based on ISO 10993 requirements
The indwell time of BEVERTM Intermittent Catheter is about 1-3 minutes. And it contact mucosal membrane. According to 510(k) Memorandum - #G95-1 Table 1 Initial Evaluation Tests for Consideration, the biocompatibility testing show consider the following 3 items:
- . Cytotoxicity
- . Sensitization
- Irritation .
{2}------------------------------------------------
KIII 405
pg. 3 of 4
# a) Test for in vitro cytotoxicity (MTT cytotoxicity test)
An in vitro cytotoxicity study was conducted to assess the potential for cytotoxicity of the test article, Intermittent Catheter, based on the International Organization for Standardization ISO 10993-5:2009: Biological evaluation of medical devices-Part 5:Tests for in vitro cytotoxicity; ISO 10993-12:2007; Biological evaluation of medical devices-Part 12: Sample preparation and reference materials.
Four concentrations (100%, 75%, and 25%) of the test article extracts, the blank control, 100% of the negative control and the positive control were prepared using Minimum Essential Medium (MEM) supplemented with 10% calf serum. The semi-confluent monolayers of L-929 mouse fibroblast cells were incubated with the test extract and other three controls, supplemented with 10% calf serum in a 96-well microplate respectively at 37℃ under the condition of 5% CO2.. At 24h, the MTT
colorimetric assay was employed and plate was read on a microplate reader at 570 and 650nm. The viability of the cells was calculated.
Under the conditions of this study, the viability of 100% extract of the test article was 86%, it had not a cytotoxic potential.
# b) Test for irritation (vaginal irritation test)
The purpose of the study was to evaluate the potential irritation response of the extract of the test article, Intermittent Catheter contact with vaginal tissue. This study was conducted based on the requirements for the International Organization for Standardization ISO 10993-10:2010: Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization: ISO 10993-12:2007: Biological evaluation of medical devices Part 12: Sample preparation and reference materials.
The test extract were contacted with vaginal tissue 1 time (once a day) directly for five consecutive days. At 24h after the final treatment, the tissues were examined macro- scopically and microscopically. The irritant effects on vaginal tissue were evaluated and scored.
Under the conditions of this study, the macroscopic reaction of test extract was relatively similar to that of the reagent control. Microscopically, the extract of the test article was classified as a non- irritant as compared to the control.
# c) Test for skin sensitization (Maximization test)
A guinea pig maximization test of the test article, Intermittent Catheter, was conducted to evaluate the skin sensitizing potential. This study was based on the International Organization for Standardization ISO 10993-10:2010: Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization; ISO 10993-12:2007: Biological evaluation of medical devices Part 12: Sample preparation and reference materials.
{3}------------------------------------------------
KIII405
Pg. Hoff
The test article was extracted in 0.9% sodium chloride solution (SC) and cotton seed oil (CSO). Each extract was injected intradermally and patched occlusively to ten test guinea pigs (per extract) in an attempt to induce sensitization. The vehicle was similarly injected and patched occlusively to five control guinea pigs (per vehicle). Following a recovery period, the test and control animals were received a challenge patch of the appropriated test article extract and the reagent control. All sites were scored at 24h and 48h after patch removal,
Under the conditions of this study, the SC and CSO extracts of the test article showed no evidence of causing sensitization in the guinea pig.
### 9.2 Packaging Shipment Testing was performed based on ISTA Procedure 2A
2 pieces of packaged-products were submitted for the packaging shipment testing in accordance with ISTA Procedure 2A. After all tests completed, no tearing was observed on the inner packaging and outer packaging.
#### 10. sterility and shelf life
The device is sterilized by EO. The water packet in coated Intermittent Catheter is sterilized by radiation. The sterilization process is compliance with ISO11135-1: 2007, ISO11137-1: 2006, ISO11137-2: 2006, ISO11137-3: 2006.
Regarding the accelerated aging testing result, after 137 days accelerated aging, BEVER " coated Intermittent Catheter is still compliance the requirements of device specification. The shelf life of BEVER™ coated Intermittent Catheter could be considered as 3 years. And, according to the real time stability study, the 3 years shelf life of BEVER"10 Coated Intermittent Catheter has been validated. Herewith we declare that the shelf life of BEVERTM Coated Intermittent Catheter is 3 years.
Regarding the accelerated aging testing result, after 183 days accelerated aging, BEVER™ uncoated Intermittent Catheter is still compliance the requirements of device specification. The shelf life of BEVER™ Uncoated Intermittent Catheter could be considered as 4 years. And, according to the real time stability study, the 4 years shelf life of BEVER™ Coated Intermittent Catheter has been validated. Herewith we declare that the shelf life of BEVER 110 Coated Intermittent Catheter is 4 years.
## 11. Conclusions
These proposed devices have the same intended use and technological characteristics to the currently-marketed predicate devices. No new issues of safety or effectiveness are introduced by using these devices. Therefore we believe the proposed devices are substantially equivalent to the currently-marketed predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features an abstract eagle design with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple, yet recognizable, representing the department's role in promoting the health and well-being of the nation.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Allyson Zhou Management Representative Hangzhou Bever Medical Devices Co., Ltd. No. 8-1, Longquan Rd., Cangqian Town, Yuhang District HANGZHOU ZHEJIANG 311121 CHINA
FEB 1 3 2012
Re: K111405
Trade/Device Name: BEVER™ Intermittent Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: GBM Dated: May 16, 2011 Received: January 11, 2012
Dear Ms. Zhou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Twichell
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): Device Name: BEVERTM Intermittent Catheter Indications For Use:
The BEVER™ intermittent catheter is indicated for intermittent catheterization of the urethra for those individuals who are unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder -voiding episode. The catheter is inserted into urethra to reach the bladder allowing urine to drain.
Over-The-Counter Use Prescription Use _ Yes AND/OR No (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| (Division Sign-Off) | |
|----------------------------------------------------------------|---------|
| Division of Reproductive, Gastro-Renal, and Urological Devices | |
| 510(k) Number | K111405 |
Section 1, Page 1 of 1 Page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.