K111303 · Sharplight Technologies, Ltd. · GEX · Dec 21, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111303
Device Name
DERMATOLOGICAL TREATMENT DEVICE FAMILY
Applicant
Sharplight Technologies, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 21, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dermatological Treatment Systems device family and optional Hand Piece family are intended for use in aesthetic and cosmetic applications and in selective treatments required in the medical specialties of dermatology.
Device Story
Dermatological Treatment Systems (DTS) platform supports multiple interchangeable handpieces using IPL, Infrared (IR), and Laser (Nd:YAG, Er:YAG) technologies. Device delivers energy to human skin for aesthetic/dermatological procedures. System comprises base unit with embedded control software and touch-screen interface; each handpiece features dedicated operating screen. Operator selects parameters via interface to control energy delivery. Device used in clinical settings by trained professionals. Output (light/laser energy) targets melanin, hemoglobin, or water in tissue to achieve therapeutic effects like hair reduction, lesion removal, or skin resurfacing. Benefits include non-invasive or minimally invasive treatment of dermatological conditions.
Clinical Evidence
Bench testing and clinical data provided. No specific quantitative metrics (e.g., sensitivity, AUC) reported in the summary. Evidence supports safety and effectiveness for the claimed dermatological indications.
Technological Characteristics
Platform system supporting IPL (415-1200 nm), IR (850-1750 nm), and Laser (Nd:YAG 1064nm, Nd:YAG 532nm, Er:YAG 2940nm) handpieces. Features contact cooling. Embedded software control system with touch-screen interface. Class II device.
Indications for Use
Indicated for hair removal/permanent hair reduction (Fitzpatrick I-VI), vascular lesions, inflammatory acne (I-V), benign pigmented epidermal/cutaneous lesions (warts, scars, striae), wrinkles/rhytids, skin resurfacing (acne scars), tattoo removal, and photocoagulation of soft tissue. Specific indications vary by handpiece (IPL, IR, Nd:YAG 1064nm, Nd:YAG 532nm, Er:YAG 2940nm) and include treatment of port wine stains, hemangiomas, telangiectasias, rosacea, psoriasis, and various pigmented lesions (lentigos, nevi, etc.) across specified skin types.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
SharpLight - BEAMAX/FORMAX Pulse Light device family (K082876)
Alma lasers - Harmony XLTM Multi-Application Platform (K07256-)
Cutera Optional Pulsed Light Hand Piece Family (K05004)
{0}------------------------------------------------
K111303
Image /page/0/Picture/1 description: The image contains a logo with the word "Sharplight" written in a stylized font. Above the word, there is a rectangular shape on the left and some text on the right. In the middle, there is a small square shape.
# DEC 2 1 2011
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92
#### Submitter Information
SharpLight Technologies Ltd. 33 Lazarov St. Rishon Le Zion 75658, Israel Tel: (972) 3 9611969, Fax: (972) 3 9528890 Submission contact person: Ilan Sharon P.O.B. 4262, Zichron Yaacov 30900, Israel TEL: 972-52-8704904
#### Device Classification
| Proprietary Device Name: | Dermatological Treatment Systems device family |
|----------------------------|--------------------------------------------------------------------------------------|
| Common name: | Multi-Application Dermatological Treatment device family |
| Product Code: | Primary GEX, Secondary NUV, ONG, ONE, ONF |
| Classification Name: | Laser Surgical Instrument, for use in General and Plastic Surgery<br>and Dermatology |
| Classification Regulation: | 21 CFR § 878.4810 |
| Regulatory Class: | II |
#### Identification of Legally Marketed Predicate Devices
| SharpLight - BEAMAX/FORMAX Pulse Light device family - | K082870 |
|---------------------------------------------------------|---------|
| Alma lasers - Harmony XLTM Multi-Application Platform - | K07256- |
| Cutera Optional Pulsed Light Hand Piece Family - | K05004 |
| Palomar LuxIR Fractional Hand piece - | K060065 |
#### Device Description
The Dermatologic Treatment Systems (DTS) device family and optional hand pieces, including Intensive Pulse Light (IPL) Infrared light (IR) and Laser light technologies for dermatological treatments by delivery of energy to the human skin. The device family is a modification to the BeaMax/ ForMax family (K082876), pulsed light energy device based on a filtered, Xenon flashlamp.
The DTS device family modification enhances capabilities of the legally cleared Beamax/ Formax device family by adding new hand pieces that are now supporting Laser and IR technologies as well as the IPL technology.
The Basic unit is a platform system, designed to support all treatment technologies hand pieces (i.e. IPL, IR and Laser).
The embedded software is the heart of the control system which provides a control of all of the parameters to the relevant hand piece, where the operator has the access to modify some parameters via the touch screen and every hand piece has its own operating screen.
{1}------------------------------------------------
#### Intended Use of Device
The DTS device family, a modification to the legally cleared device family (K082876), uses IPL, IR, Laser and RF energy sources, is intended for dermatological treatments.
The DTS device family is indicated for dermatological treatments such as, but not limited to:
- * Hair removal in all skin types to the Fitzpatrick scale. Permanent Hair Reduction.
- * Treatment of Vascular Lesions
- * Treatment of Inflammatory Acne (acne vulgaris)
- * Treatment of Benign Pigmented Epidermal and Cutaneous Lesions including warts, scars and striae.
- * Treatment of Wrinkles, rhytids and Periorbital Wrinkles
- * Skin resurfacing (Acne scars)
- * Treatment of Tattoos
- * Treatment of Leg and Facial Vein Removal
- * Treatment of photocoagulation of soft tissue in dermatologic applications
- * Resurfacing of the skin, and for the treatment of facial wrinkles.
## Safety & Effectiveness
The DTS device family was compared to the predicate devices in terms of intended use, indications for use, components, principles of operation, technological characteristics and safety features.
Based on the performance results provided (including test results and clinical data) and the analysis of similarities and differences presented above, SharpLight Technologies Ltd. believes that the proposed device family safe & effectiveness is substantially equivalent to the predicate devices without raising new safety and/or effectiveness issues.
## Rational for Substantial Equivalency
Similarities:
The intended use and indications of the DTS device familar to the legally cleared predicate devices.
The technology, performance and most of the specifications of the DTS device family are similar to the legally marketed predicate devices
Differences:
The DTS device family wider indications for use enhances their capability to perform dermatological treatments, by using the different technologies (IPL, Laser and IR), while predicate devices are capable of supporting partial treatments only.
## Substantial Equivalence Statement
Based on the above, it is SharpLight's opinion that the proposed DTS device family is substantially equivalent in terms design principles, performance features and of safety & effectiveness to the unmodified BeaMax/ ForMax family (K082876) legally cleared device and to the legally cleared predicate devices referred to in chapter 4 of this submission.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 2 1 2011
SharpLight Technologies, Ltd. . % Mr. Ilan Sharon P.O. Box 4262 Zichron Yaacov Israel 30900
Re: K111303
Trade/Device Name: Dermatological Treatment Device Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, GEI Dated: December 16, 2011 Received: December 20, 2011
Dear Mr. Sharon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Ilan Sharon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K111303
Device Name: Dermatological Treatment Systems Device Family
Indications for Use:
The Dermatological Treatment Systems device family and optional Hand Piece family are intended for use in aesthetic and cosmetic applications and in selective treatments required in the medical specialties of dermatology.
The Dermatological Treatment device family and optional Hand pieces with 415 -1200 nm wavelengths (with and without contact-cooling) are indicated for:
- * Hair removal and Permanent Hair Reduction in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelengths in the range of 635-950 nm, 730-950 nm or 580-950 nm
- * Treatment of Pigmented and Vascular Lesions in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelength in the range of 535-950 nm
- * Treatment of Inflammatory Acne (acne vulgaris) in skin types (I-V) to the Fitzpatrick scale - Recommended wavelength in the range of 415-950 nm
- * Treatment of Benign Pigmented Epidermal and Cutaneous Lesions including warts, scars and striae in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelength in the range of 580-950 nm
- *Treatment of Benign Pigmented Epidermal, Cutaneous and Vascular Lesions including warts; scars and striae in skin types (I-V) to the Fitzpatrick scale - Recommended wavelength in the range of 535-680 nm & 860-1200nm
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nil Re ach for mxm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111303
Page 1 of 5
{5}------------------------------------------------
510(k) Number (if known): K111303
Device Name: Dermatological Treatment device family
Indications for Use: Continued from last page.
The Dermatological Treatment device family and optional Infrared (IR) Hand Pieces with 850 - 1750 nm wavelengths (with contact cooling) is indicated for:
- * Dermatologic Treatment such as, but not limited to: photocoagulation of soft tissue (Scars, Wrinkles, Rhytids and Periorbital Wrinkles)
## The Dermatological Treatment device family and optional Nd: YAG Laser 1064 nm Long Pulse (LP) Hand Pieces is indicated for:
- * Removal of unwanted hair, for stable long-term, or permanent, hair reduction through selective targeting of melanin in hair follicles.
- * Removal or lightening of unwanted hair (with and without adjuvant preparation)
- * Treatment of pseudofolliculitis barbae (PFB)
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
- Neil RP agle for nxn
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 2 of 5_
510(k) Number.
{6}------------------------------------------------
510(k) Number (if known): K111303
Device Name: Dermatological Treatment device family
Indications for Use: Continued from last page.
## The Dermatological Treatment device family and optional Nd:YAG Laser 1064 nm Long Pulse (LP) and Q-Switched Hand Pieces is indicated for:
- * Benign vascular lesions such as, but not limited to treatment of: Port wine stains; Hemangiomas; Warts, Superficial and deep telangiectasias (venulectasias); Reticular veins (0.1-4.0 mm dia.) of the leg; Rosacea; Venus lake; Leg veins; Spider veins; Poikiloderma of Civatte; Angiomas
- * Benign cutaneous lesions, such as, but not limited to: Warts; Scars; Striae; Psoriasis
- * Benign pigmented lesions such as, but not limited to: Lentigos (age spots); Solar lentigos (sun spots); Cafe-au-lait macules; Seborrheic; keratoses; Nevi and nevus of Ota, Chloasma; Verrucae; Skin tags; Keratoses; The removal of black, blue or green tattoos (significant reduction in the intensity of black and /or blue/black tattoos); Plaques
- * Pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment and for patients with lesions that have not responded to other laser treatments.
- * The non-ablative treatment of facial wrinkles, such as, but not limited to: Periocular wrinkles; Perioral wrinkles
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil RP Oden for man
(Division Sign Division of Surgical, Orthopedic, and Restorative Devices
· 510(k) Number.
Page 3 of 5
{7}------------------------------------------------
510(k) Number (if known): K111303
Device Name: Dermatological Treatment device family
Indications for Use: Continued from last page.
## Continued : The Dermatological Treatment device family and optional Nd:YAG Laser 1064 nm Long Pulse (LP) and Q-Switched Hand Pieces is indicated for:
* Laser skin resurfacing procedures for the treatment of: Acne scars; Wrinkles
- * Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
* Indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.
The Dermatological Treatment device family and optional Nd:YAG Laser 532 mm Frequency Dabbler (FD) Long Pulse (LP) and Q-Switched Hand Pieces is indicated for:
* Tattoo removal: Light blue; Yellow; Red; Green
- * Vascular lesions: Hemangiomas (Port wine stains/birthmarks, cavernous, cherry, spider); hemangiomas; Angiomas (cherry, spider); Telangiectasia; Spider nevi
- * Benign pigmented lesions, such as, but not limited to: Cafe-au-lait (macules); Lentigines (senile and solar); Freckles (ephelides); Chloasma; Nevi; Nevus spillus, Nevus of Ota: Becker's Nevi;
- * Other pigmented cutaneous lesions, such as, but not limited to: Verrucae; Skin tags; Keratoses; Plaques
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil RP Dade formxn
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KIII 303 510(k) Number_
Page 4_ of 5_
{8}------------------------------------------------
510(k) Number (if known): K111303
Device Name: Dermatological Treatment device family
Indications for Use: Continued from last page.
The Dermatological Treatment device family and optional Er:YAG Laser 2940 nm Long Pulse (LP) Hand Pieces, with standard and scanner accessory tips is indicated for:
Use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs, and glands) such as, but not limited to:
* Dermatology and plastic Surgery: Skin resurfacing; Treatment of wrinkles; Epidermal nevi; Telangiectasia; Spider veins; Actinic chelitis; Keloids; Verrucae; Skin tags; Anal tags; Keratoses; Scar revision (including acne scars);
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Prescription Use X (Part 21 CFR 801 Subpart D)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Net RPD sale for mxm
(Division Sign bm)
Division Sign Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111303
Page 5_ of 5_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.