K111292 · Siemens Medical Solutions USA, Inc. · JAA · Jun 3, 2011 · Radiology
Device Facts
Record ID
K111292
Device Name
LUMINOS AGILE
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAA · Radiology
Decision Date
Jun 3, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Luminos Agile is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. Luminos Agile is applicable for emergency treatment on an outpatient basis, as well as for bedside examinations.
Device Story
Luminos Agile is a floor-mounted universal fluoroscopic/radiographic (R/F) X-ray system; utilizes X-ray tube and solid-state flat panel detector (Pixium 5100) to capture diagnostic images. System replaces traditional image intensifiers with digital imaging chain (Fluorospot COMPACT). Configurable as single or dual-tube system; supports under-table and over-table exposures. Operated by healthcare professionals in clinical settings (outpatient, emergency, bedside). Digital images are processed and displayed for clinician review to support diagnostic decision-making. System includes safety monitoring with automated error blocking and visual/audible alerts.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by adherence to industry practices, final performance testing, and substantial equivalence to previously cleared Siemens R/F systems.
Technological Characteristics
Floor-mounted R/F system; solid-state flat panel detector (Pixium 5100); Fluorospot COMPACT digital imaging system. Components include X-ray generator, X-ray tube/housing, beam-limiting device, Bucky cassette/detector holders. Connectivity via digital imaging chain. System is not stand-alone; functions as platform for integrated X-ray components. Software-based monitoring and error-handling features included.
Indications for Use
Indicated for universal radiographic and fluoroscopic diagnostic imaging, including general R/F, angiography, and pediatric examinations. Applicable for emergency outpatient treatment and bedside examinations.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
SIEMENS
# 510(k) Summary: Luminos Agile
JUN - 3 2011
Company: Siemens Medical Systems, Inc. 1 Valley Stream Parkway Malvern, PA 19355
Date Prepared: May 4, 2011
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
- 1. General Information:
Importer / Distributor: Siemens Medical Solutions, Inc. 51 Valley Stream Parkway, E-50 Malvern, PA 19355 Establishment Registration Number: 2240869
Manufacturing Site: SIEMENS AG Sector Healthcare Siemensstr. 1 91301 Forchheim
## 2. Contact Person:
Ms. Patricia D Jones Technical Specialist. Requlatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway · G-01 Malvern, PA 19355 Phone: (610) 448 -3536 Fax: (610) 448-1787 Email: patricia.d.jones@siemens.com
#### 3. Device Name and Classification:
Trade Name: Classification Name: Classification Panel: CFR Section: Device Class: Device Code:
Luminos Agile Solid state x-rav imager (flat panel/digital imager) Radiology 21 CFR § 892.1650 Class II 90 MQB
{1}------------------------------------------------
## 4. Device Description:
Luminos Agile is a floor-mounted universal fluoroscopic x-ray diagnostic system (R/F system), that was developed for tableside examinations in combination with an Explorator. The modification is to replace the Image Intensifier / TV Camera with a solid state detector (flat panel/digital imager). The Explorator houses the flat panel solid state detector.
This modified AXIOM SIRESKOP SD will be marketed under the trade name Luminos Agile. The modification does not affect the intended use of the device nor does it alter its fundamental scientific technology.
Luminos Agile may be configured as a single tube system, with an under table tube or a dual tube system which features an additional 3D overhead tube crane, that can be moved longitudinally and laterally as well as vertically. The dual x-ray tube configuration provides a quick change between under table and over table exposure modes. The under table tube is used for fluoroscopy and radiographic exposures taken with the Digital Imaging System. The over table tube is suspended from an overhead tube support for exposures with a table Bucky or Bucky wall stand, either with film cassettes or solid state detector. The digital images produced by the solid state detector are recorded and displayed by the Fluorospot COMPACT digital imaging system. The table design remains unchanged while the new imaging chain is based on the AXIOM Luminos dRF described in premarket notification K062623 which received FDA Clearance on August 22. 2007.
#### 5. Intended Use:
The Luminos Agile is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations.
Luminos Agile is applicable for emergency treatment on an outpatient basis, as well as for bedside examinations.
#### 6. Substantial Equivalence:
The Luminos Agile with flat panel solid state detector is substantially equivalent to the commercially available Siemens systems, the AXIOM Luminos TF (AXIOM Sireskop SD) and the AXIOM Luminos dRF. The AXIOM Sireskop SD was described in premarket notification K051602 which received FDA Clearance on 07/07/2005. The AXIOM Sireskop SD was renamed later to AXIOM Luminos TF for marketing reasons. The change was documented in an NFJ. The AXIOM Luminos dRF was described in premarket notification K062623 which received FDA Clearance on August 22. 2007.
{2}------------------------------------------------
# SIEMENS
Special 510(k) Submission: Luminos Agile
X-ray generation and control used with the Luminos Agile is identical to the AXIOM Luminos dRF. The Flat Detector Pixium 5100 and the Fluorospot Compact digital imaging system equipped with Luminos Agile are identical to the detector and imager used in the AXIOM Luminos dRF.
7. Summary of Technological Characteristics of the Principal Device as Compared with the Predicate Device:
Luminos Agile is not a stand-alone device, but functions as the platform for specific X-ray components (X-ray generator, X-ray tube and housing, Exploratory with flat panel solid state detector, mobile flat detector, digital imaging system, Bucky cassette holder, Bucky detector holder and Bucky wall stand etc.).
Luminos Agile is a modified AXIOM Sireskop SD. The principal device Luminos Agile features a solid state detector instead of an x-ray image intensifier like the predicate AXIOM Sireskop SD. The design of the Luminos Agile's imaging chain is based on the design of the second predicate the AXIOM Luminos dRF. Also, the Luminos Agile shares the same x-ray- and software components with the AXIOM Luminos dRF. (Table base, Generator, X-rav tube and housing, beam-limiting-devise flat detector, digital image processing device, etc.).
Many of these components used in Luminos Agile are either commercially available with current Siemens systems or include minor modifications to existing components.
#### 8. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Luminos Agile is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed. Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
## 9. Conclusion as to Substantial Equivalence:
The Luminos Agile is intended for the same indications for use as the predicate AXIOM Sireskop SD. The imaging chain has been modified to include a flat panel detector. The portfolio of accessories is the same as with the predicate AXIOM Sireskop SD to compliment the needs of the Fluoroskopy suite. It is Siemens opinion, that the Luminos Agile is substantially equivalent to the AXIOM Sireskop SD.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in blue font. The text is left-aligned and appears to be a header or title. A blue arrow is to the left of the text.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Patricia D. Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway, E-50 MALVERN PA 19355
AUG 2 0 2013
Re: K111292
Trade/Device Name: Luminos Agile Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA and IZI Dated: May 4, 2011 Received: May 6, 2011
Dear Ms. Jones:
This letter corrects our substantially equivalent letter of June 3, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "SIEMENS" in all capital letters. The letters are outlined in black, with a white fill. The font is bold and sans-serif.
# Indications for Use Form
510(k) Number (if known): K111292
Device Name: Luminos Agile
Indications for Use:
The Luminos Agile is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations.
Luminos Agile may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Patil.
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) C111292
Page 1 of
Siemens Medical Systems, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.