K111290 · Guilin Woodpecker Medical Instrument Co., Ltd. · DZI · Nov 18, 2011 · Dental
Device Facts
Record ID
K111290
Device Name
PIEZO BONE SURGERY
Applicant
Guilin Woodpecker Medical Instrument Co., Ltd.
Product Code
DZI · Dental
Decision Date
Nov 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4120
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Piezo Bone Surgery is intended for use in the following dental applications: - Bone cutting for use in oral surgery - Removing supra and sub-gingival calculus deposits and stains from teeth - Periodontal pocket lavage with simultaneous ultrasonic tip movement - Scaling and root planning - Retrograde preparation of root canals
Device Story
Handheld ultrasonic surgical device for dental applications; connects to control console via cord. Uses piezoelectric technology to generate mechanical micro-vibrations at 24-29.5 kHz. Practitioner selects from three modes (Bone, Root, Clean) via console; irrigation provided by peristaltic pump using purified water or saline. Used by dental professionals in oral surgery, implantology, and periodontal settings. Device output (ultrasonic vibration at tip) enables precise bone cutting and cleaning with minimal soft tissue trauma. Benefits include improved surgical precision and reduced tissue damage compared to traditional rotary instruments.
Clinical Evidence
Bench testing only. Performance validated through electrical, mechanical, and environmental safety testing (IEC 60601-1, IEC 60601-1-2) and biocompatibility testing (ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 7405). No clinical data provided.
Technological Characteristics
Handheld ultrasonic surgical device; piezoelectric wafer transducer. Materials: stainless steel tips. Frequency: 24-29.5 kHz. Power: 100-120VAC 50/60Hz. Peristaltic pump for irrigation. Intermittent operation (60s ON/10s OFF). Biocompatibility per ISO 10993-1. Sterilization: autoclave up to 135°C.
Indications for Use
Indicated for dental patients requiring oral surgery (bone cutting), periodontal treatment (calculus/stain removal, pocket lavage, scaling, root planning), or endodontic procedures (retrograde root canal preparation).
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
Predicate Devices
MECTRON, Piezo Bone Surgery, Piezosurgery (K091227)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for a company called "Woodpecker". The logo features a stylized drawing of a woodpecker's head and neck on the left side. To the right of the woodpecker image, the word "WOODPECKER" is written in a bold, sans-serif font. The logo appears to be simple and clean, with a focus on the company name and a recognizable image of a woodpecker.
幸林市 陈 木 鸟 医 疗 器 械 有 限 A WILLINE WOODDECKED Medical Instrument
NOV 1 8 2011
ﺴﺘﻨﺪ
#### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
# Date: February 4, 2011
- 1. Company:
Name - Guilin Woodpecker Medical Instrument Co., Ltd. Address - Information Industrial Park, Guilin National High-Tech Zone District, Guilin, Guangxi, 541004, P.R. China Telephone - +86-773-5855340 Fax – +86-733-5855351 Contact - Mr. Wu Xunxian Email – woodpeckera@mailgl.cn
Correspondent:
Name- IRC Address- 77325 Joyce Way, Echo, Oregon 97826 Telephone- 931-625-4938 Fax- 541-376-5063 Contact- Charlie Mack Email- charliemack@irc-us.com
2. Device :
Trade/proprietary name: Piezo Bone Surgery, Model Ultrasurgery Common Name : Drill, bone, powered Classification Name : Bone cutting instrument and accessories
#### 3. Predicate Devices :
MECTRON, Piezo Bone Surgery, Piezosurgery, K091227
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo for "WOODPECKER". To the left of the word "WOODPECKER" is a cartoon drawing of a woodpecker. The woodpecker is facing left and has a small circle with an R in it below it.
## 4. Classifications Names & Citations :
21CFR 872.4120, DZI, Drill, Bone, Powered, Class2
Description :
## 5.1 General
The Guilin Woodpecker Medical Instrument Co., Ltd. Piezo Bone Surgery device is a dental device used in oral surgery situations. In this submission, it is intended to be used for bone cutting in oral surgery, removing supra and sub-gingival calculus deposits, stains from teeth, periodontal pocket lavage with simultaneous ultrasonic tip movement, scaling, root planning, and retrograde preparation of root canals.
The device is a hand held ultrasonic surgical device, which is connected via a cord to the control console. The device operates at frequency range of 24 to 29.5 kHz. There are three modes of operation, which are selectable from the control console. The practitioner can select the Bone, Root or Clean modes of operation. Each mode has a different power mode, with the Bone mode giving the most power. Irrigation to the tip is provided and adjustable via the control console. Water flow for the irrigation is provided via a peristaltic pump.
A selection of tips is available for the dental professional to select and use for the specific dental procedure. The available tips are shown in the User's manual and also in the advertisement brochure.
This device is not delivered sterile, but must be sterilized after each use. Instructions for cleaning and sterilization are provided within the User's Manual.
5. Indication for use :
The Piezo Bone Surgery is intended for use in the following dental applications:
- Bone cutting for use in oral surgery
- Removing supra and sub-gingival calculus deposits and stains from teeth
- Periodontal pocket lavage with simultaneous ultrasonic tip movement
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with a stylized drawing of a woodpecker on the left and the word "WOODPECKER" on the right. The woodpecker is drawn with simple lines, and the word "WOODPECKER" is in a sans-serif font. The logo is black and white.
- Scaling and root planning
- Retrograde preparation of root canals
- 6. Comparison with predicate device :
Guilin Woodpecker Medical Instrument Co., Ltd. believes that the Piezo Bone Surgery device, model Ultrasurgery is substantially equivalent to the Mectron, Piezosurgery® (K091227).
Please see the next two pages for a comprehensive comparison with the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "WOODPECKER" in a vertical orientation. Below the word is a logo of a woodpecker. The woodpecker is facing left and has a long beak.
| Element of comparison | Subject Device | Claimed SE Device |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | WOODPECKER | MECTRON |
| Device name | Piezo Bone Surgery | Piezo Bone Surgery |
| Device model | Ultrasurgery | Piezosurgery ® |
| FDA510(K) No. | N/A | K091227 |
| Intended use(s) | The Piezo Bone Surgery is intended for use in the<br>following dental applications:<br>- Bone cutting for use in oral surgery<br>- Removing supra and subgingival calculus<br>deposits and stains from teeth<br>- Periodontal pocket lavage with simultaneous<br>ultrasonic tip movement<br>- Scaling and root planning<br>- Retrograde preparation of root canals | The Piezosurgery 3 is intended for use in the<br>following dental applications:<br>- Bone cutting for use in oral surgery<br>- Removing supra and subgingival calculus<br>deposits and stains from teeth<br>- Periodontal pocket lavage with simultaneous<br>ultrasonic tip movement<br>- Scaling and root planning<br>- Retrograde preparation of root canals |
| Operation | Using piezoelectric ultrasonic technology to<br>generate mechanical micro vibrations for bone<br>cutting and ultrasonic scaling, with minimal trauma<br>to soft tissue. | Using piezoelectric ultrasonic technology to<br>generate mechanical micro vibrations for bone<br>cutting and ultrasonic scaling, with minimal trauma<br>to soft tissue. |
| Medium used | Purified water or normal saline | Purified water or normal saline |
| Tip material | Stainless steel | Stainless steel |
| Ultrasonic vibration style | Piezoelectric Wafer | Piezoelectric Wafer |
| Device for intermittent operation | Intermittent Operation 60" ON 10" OFF | Intermittent Operation 60" ON 30" OFF |
| Working frequency | 24KHz~29.5 KHz | From 24 KHz to 36 KHz |
| Voltage supply | 100-120VAC 50/60Hz | 100-240 VAC 50/60 Hz |
| APC circuit protection systems | No hand piece connected<br>Cord interrupted<br>Insert broken or not correctly tightened | No hand piece connected<br>Cord interrupted<br>Insert broken or not correctly tightened |
| Power Modes | ROOT mode<br>BONE mode | ROOT mode<br>BONE mode<br>IMPL mode |
| IEC60601-1 Class | Type B<br>Class I | Type B<br>Class I |
| Device classification using Directive 93/42EEC | | Class II a |
| Peristaltic pump volume delivery | | From 0 to 90 ml / min approx |
| Fuses | 2×T1.0AL 250V | 230 VAC 2 X 2 A T |
| Environmental operating conditions | from +10°C to +40°C<br>Relative humidity from 30% to 70%<br>from -10°C to +50°C | from +10°C to +40°C<br>Relative humidity from 30% to 75%<br>from -10°C to +70°C |
| Transport and storage environmental conditions | Relative humidity from 10% to 90%.<br>Air pressure P: 500hPa/1060hPa | Relative humidity from 10% to 90%.<br>Air pressure P: 500hPa/1060hPa |
| Where used | Oral surgery<br>Implantology<br>Periodontal surgery<br>Surgical orthodontics | Oral surgery<br>Implantology<br>Periodontal surgery<br>Surgical orthodontics |
| Biocompatibility | Complying with ISO10993-1 | Complying with ISO10993-1 |
| Weight and Size | 3.8KG<br>L×W×H:333×255×167mm | 3.2 kg<br>L×W×H:340 X 210 X 150 mm |
| Clean and disinfection method | Clean and disinfect the surfaces of the casting, the cords and their connectors using a cloth moistened with a mild detergent or disinfectant solution with a neutral pH (pH 7). | Clean and disinfect the surfaces of the casing, the rod, the hand piece-holder, the cords and their connectors using a low fiber release cloth moistened with a detergent solution (pH 6-9) and/or a mild disinfectant with a neutral pH (pH7). |
| Sterilization method | Maximum temperature of 135°C for a maximum of 20 minutes. | Maximum temperature of 135°C for a maximum of 20 minutes. |
| Components can be sterilized | Hand piece, Tips, Tip holder, Torque wrench, Pump tube, Cord/peristaltic pump tube connection, Hand piece holder | Hand piece, Inserts, Wrench for tightening the inserts, Tube for the peristaltic pump, Connection for the cord / tube of the peristaltic pump, Rod for supporting the bag, Support for the hand piece |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "WOODPECKER" in a vertical orientation. Below the word is a logo of a woodpecker. The woodpecker is facing left and has a long beak.
桂 林 市 啄 木 鸟 医 疗 器 械 有 限 公 司
GUILIN WOODPECKER Medical Instrument Co.,LTD.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for a company called "WOODPECKER". To the left of the company name is a stylized drawing of a woodpecker. The woodpecker is facing left and has a long beak. Below the woodpecker is the registered trademark symbol.
# 7. Safety and Performance Data :
Electrical, mechanical, environmental safety and performance testing according to EN/ IEC 60601-1[1990] Medical electrical equipment Part 1: General Requirement for safety, IEC60601-1-2, EMC Compatibility, ISO10993-5 Cvtotoxicity. ISO10993-10 Cvtotoxicity. ISO 10993-1 Biological evaluation of Medical Devices Part-1; ISO 7405:2008 Dentistry --Evaluation of biocompatibility of Medical devices used for dentistry: ISO 13485- Risk Management; ISO 14971, Risk Management of Medical Devices. Performance testing was used to validate the effectiveness and accuracy of the device. All test results were satisfactory.
### 9. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Guilin Woodpecker Medical Instrument Co., Ltd. concludes that the Piezo Bone Surgery device, model Ultrasurgery is substantially equivalent to predicate devices as described herein.
END
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with its wings spread, and three wavy lines below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Guilin Woodpacker Medical Instrument Company, Limited C/O Mr. Charlie Mack Principal Engineer International Regulatory Consultants 77325 Joyce Way Echo, Oregon 97826
NOV 1 8 2011
Re: K111290
Trade/Device Name: Piezo Bone Surgery Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: II Product Code: DZI, ELC Dated: November 6, 2011 Received: November 9, 2011
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
#### Page 2- Mr. Mack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
. Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo for "WOODPECKER". To the left of the word "WOODPECKER" is a cartoon drawing of a woodpecker. The woodpecker is facing left and has a small circle with an R in it below it.
GUILIN WOODPECKER Medical Instrument Co
#### Indications for Use
# Indications for Use
510(k) Number (if known):
K11 290
Device Name: Piezo Bone Surgery
Indications For Use:
The Piezo Bone Surgery is intended for use in the following dental applications:
- -Bone cutting for use in oral surgery
- Removing supra and subgingival calculus deposits and stains from teeth -
- Periodontal pocket lavage with simultaneous ultrasonic tip movement -
- Scaling and root planning -
- Retrograde preparation of root canals -
AND/OR Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign Division of Anesthesiology, General Hospital Intection Control, Dental Devices
Page 1 of 1
510(k) Number
1 ર
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.