AQUAMANTYS 3 PUMP GENERATOR, BIPOLAR SEALER, BIPOLAR SEALER WITH CUTTING
K111285 · Salient Surgical Technologies, Inc. · GEI · Sep 9, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111285
Device Name
AQUAMANTYS 3 PUMP GENERATOR, BIPOLAR SEALER, BIPOLAR SEALER WITH CUTTING
Applicant
Salient Surgical Technologies, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 9, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aquamantys3 Generator is an electrosurgical generator with monopolar and bipolar RF outputs. It is intended to be used with specified disposables for delivery of RF energy for cutting of soft tissue and RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone at the operative site during surgical procedures. The system is not intended for contraceptive tubal coagulation (permanent female sterilization). The Aquamantys3 Bipolar Sealer is a bipolar, single use, sterile, disposable device intended for use with the Aquamantys3 Generator. The device delivers bipolar RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone. It is intended for, but not limited to, orthopaedic, spine, thoracic, and open abdominal surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization). The Aquamantys3 8.2L Bipolar Sealer with Cutting is a monopolar/bipolar, single use, sterile, disposable device intended for use with the Aquamantys3 Generator. The device delivers monopolar RF energy for cutting of soft tissue and bipolar RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone. It is intended for, but not limited to, orthopaedic, spine, thoracic, and open abdominal surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
Device Story
System comprises electrosurgical generator and sterile, single-use disposable handpieces (6.0 Bipolar Sealer; 8.2L Bipolar Sealer with Cutting). Generator provides monopolar/bipolar RF energy and peristaltic pump for saline delivery. Handpieces connect via unique cassette interface to ensure compatibility. 6.0 Sealer delivers bipolar RF with saline for haemostasis (Transcollation technology). 8.2L Sealer features dual electrodes: yellow button for monopolar RF cutting; blue button for bipolar RF with saline coagulation. Used in OR by surgeons; generator sits on shelf. Output provides haemostatic sealing/coagulation without charring; aids surgical site management; reduces blood loss. System monitors patient return electrode circuit for safety.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and in-vivo animal (porcine) studies. Verification and validation activities confirmed tissue effects comparable to predicate devices for both monopolar cutting and bipolar RF/saline coagulation.
Technological Characteristics
Shelf-top electrosurgical generator with internal rotary peristaltic pump. Sheet metal housing; LCD touchscreen interface. Disposable handpieces with unique cassette interface for electrical and saline connection. Monopolar and bipolar RF energy outputs. Split-pad patient return electrode monitoring. Sterile, single-use disposables.
Indications for Use
Indicated for patients undergoing orthopaedic, spine, thoracic, or open abdominal surgery requiring soft tissue cutting, haemostatic sealing, or coagulation of soft tissue and bone. Contraindicated for contraceptive tubal coagulation (permanent female sterilization).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Salient Surgical Aquamantys Pump Generator System (K052859)
{0}------------------------------------------------
K 111 285
SEP - 9 2011
Page 1 of 5
# 510(k) Summary for the Salient Surgical Technologies, Inc. Aquamantys 3 Pump Generator System, Aquamantys 3 6.0 Bipolar Sealer and Aquamantys 3 8.2L Bipolar Sealer with Cutting As Required by 21 CFR 807.87(k)
## 1. SPONSOR/510(K) HOLDER
Salient Surgical Technologies, Inc. 180 International Drive Portsmouth, NH 03801
| Phone: | 603-742-1515 |
|----------|----------------|
| Fax: | 603-742-5943 |
| Contact: | Rita M. Calnan |
May 5, 2011 Date Prepared:
## 2. DEVICE NAME
| Proprietary Name: | Aquamantys®3 Pump Generator System, which includes the<br>Aquamantys®3 Pump Generator, Aquamantys®3 6.0 Bipolar<br>Sealer and Aquamantys®3 8.2L Bipolar Sealer with Cutting |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Electrosurgical Generator and accessory |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
## 3. PREDICATE DEVICES
- Salient Surgical (formerly TissueLink) Aquamantys Pump Generator System . (includes Aquamantys Pump Generator, Aquamantys 6.0 Bipolar Sealer and Aquamantys 2.3 Bipolar Sealer), K052859
- ValleyLab ForceFX Electrosurgical Generator, K944602 .
- ValleyLab E2516 Electrosurgical Pencil, K813071 .
## 4. DEVICE DESCRIPTION
The Salient Surgical Aquamantys3 Pump Generator System is a general reference term used to describe use of the Salient Aquamantys3 Pump Generator together with specific, design compatible, Aquamantys3 electrosurgical instruments. The Aquamantys3 Pump Generator itself is an electrosurgical generator with, monopolar and bipolar, high
{1}------------------------------------------------
KIII 285
Page 2 of 5
frequency current outputs and an internal rotary peristaltic pump for delivery of saline. The Aquamantys3 Pump Generator is for use only with specific, design compatible Aquamantys3 electrosurgical instruments, for delivery of radio frequency (RF) energy for cutting of soft tissue and delivery of RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone at the operative site. Currently included as part of the system in this 510(k) submission are the Aquamantys3 Pump Generator, and two hand-held, sterile, single use disposable electrosurgical devices that are compatible with the Aquamantys3 Pump Generator system. These disposable devices are the Aquamantys3 6.0 Bipolar Sealer and the Aquamantys3 8.2L Bipolar Sealer with Cutting.
The Aquamantys3 Pump Generator is a shelf-top unit consisting of a sheet metal housing and a front LCD control panel. The generator has a peristaltic pump located within the housing, which is capable of transferring saline through the disposable accessory device concurrent with the generator's provision of RF energy. The LCD control panel is touchscreen and serves as the user interface for power and saline settings. The Aquamantys3 Pump Generator has a receptacle in its front panel that provides for the disposable device connection to the generator's RF power and the peristaltic pump. The Aquamantys3 disposable device connection to the Aquamantys3 Pump Generator is configured as a cassette; the configuration is unique to the Aquamantys3 system and ensures only Aquamantys3 disposable devices with the cassette can be used with the Aquamantys3 Pump Generator. The generator accepts designated, commercially available, split-pad patient return electrode pads (neutral electrodes) for monopolar applications, and provides monitoring of the patient return circuit for safety purposes.
The Aquamantys3 6.0 Bipolar Sealer device is a modified disposable, that when connected to the Aquamantys3 Pump Generator, uses bipolar RF energy applied concurrently with saline to the operative site. The effect of RF energy applied concurrently with saline results in haemostatic sealing and coagulation of soft tissue and bone without charring; this result has been trademarked as Transcollation® technology by Salient Surgical Technologies. The proposed Aquamantys3 6.0 Bipolar Sealer shares the same handpiece, shaft and tip configuration as its predicate device, the Aquamantys 6.0 Bipolar Sealer. The proposed and predicate devices differ only in how they connect to their designated generators. The proposed Aquamantys3 device has a cassette to connect it to the Aquamantys3 Pump Generator, allowing the user to connect the device's saline tubing to the peristaltic pump at the same time as the device is connected electrically to the generator. The predicate Aquamantys disposable device, comparatively, has a standard three-prong plug for electrical connection to the Aquamantys Pump Generator and the user is required to manually feed the predicate's saline tubing into the peristaltic pump. The unique cassette connection helps prevent the
{2}------------------------------------------------
K 111 285
connection of Aquamantys3 devices to other, unintended or unspecified equipment.
The Aquamantys3 8.2L Bipolar Sealer with Cutting device is a disposable that, when connected to the Aquamantys3 Pump Generator, uses monopolar RF energy for cutting, and bipolar RF energy concurrent with saline for haemostatic sealing and coagulation. The device is equipped with dual electrodes at the distal tip. Saline and electrical lines exit the opposite end (proximal end) of the device's handpiece from the distal electrodes. The device's handpiece is equipped with two activation buttons; the distal yellow button activates monopolar RF energy for cutting, and the proximal blue button simultaneously activates bipolar RF energy concurrent with saline flow for haemostatic sealing and coagulation. A saline fluid delivery line is provided with the device, which includes a spike for insertion into saline bags. The electrical connection of the device to the generator, as well as the saline tubing insertion into the generator's pump, is made via a cassette interface that is part of the disposable device. The cassette assures the unique connection of Aquamantys3 disposable devices with Aquamantys3 Pump Generators.
## 5. INDICATION FOR USE/INTENDED USE
## Aquamantys3 Generator:
The Aquamantys3 Generator is an electrosurgical generator with monopolar and bipolar RF outputs. It is intended to be used with specified disposables for delivery of RF energy for cutting of soft tissue and RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone at the operative site during surgical procedures. The system is not intended for contraceptive tubal coagulation (permanent female sterilization).
## Aquamantys3 6.0 Bipolar Sealer:
The Aquamantys3 Bipolar Sealer is a bipolar, single use, sterile, disposable device intended for use with the Aquamantys3 Generator. The device delivers bipolar RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone. It is intended for, but not limited to, orthopaedic, spine, thoracic, and open abdominal surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
## Aquamantys3 8.2L Bipolar Sealer with Cutting:
The Aquamantys3 8.2L Bipolar Sealer with Cutting is a monopolar/bipolar, single use, sterile, disposable device intended for use with the Aquamantys3 Generator. The device delivers monopolar RF energy for cutting of soft tissue and bipolar RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone. It is intended
{3}------------------------------------------------
K11 285
Page 4 of 5
for, but not limited to, orthopaedic, spine, thoracic, and open abdominal surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
## 6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICE/S
The Aquamantys3. Pump Generator applies the same fundamental technology as the predicate Aquamantys Pump Generator (previously cleared under K052859) in that it delivers bipolar RF energy with saline to provide haemostatic sealing and coagulation of soft tissue and bone at the operative site. The proposed generator is additionally similar to the predicate Valleylab ForceFX Generator (previously cleared under K944602) in that it delivers monopolar RF energy, in addition to bipolar RF energy, for cutting and coagulation during surgery.
The modified single-use disposable accessory device, the Aquamantys3 6.0 Bipolar Sealer, is similar to the Aquamantys 6.0 Bipolar Sealer such that both devices provide concurrent delivery of bipolar RF energy with saline. The two devices share identical handpiece, shaft and tip configurations and differ only in regard to how they connect to their respective, designated generators. The cord of the predicate Aquamantys 6.0 terminates in a standard three-prong electrical connector and has saline tubing that is manually loaded as an additional step into the peristaltic pump. The cord of the proposed Aquamantys3 6.0 terminates in a cassette that is designed to uniquely insert into the Aquamantys3 Pump Generator, providing simultaneous connection to both RF power and the peristaltic pump.
The Aquamantys3 8.2L Bipolar Sealer with Cutting applies the same fundamental scientific technology as the existing Salient Surgical Bipolar Sealers, with the addition of monopolar RF energy applied without saline, similar to a standard monopolar electrosurgical pencil, such as Valleylab's E2516 (K813071). The device's handpiece is equipped with two activation buttons; the distal yellow button activates monopolar RF energy for cutting, and the proximal blue button simultaneously activates bipolar RF energy concurrent with saline flow for haemostatic sealing and coagulation. The cord of the Aquamantys3 8.2L terminates in a cassette that is designed to uniquely insert into the Aquamantys3 Pump Generator, providing simultaneous connection to both RF power and the peristaltic pump.
{4}------------------------------------------------
## 7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
K 111 285
The tissue effect and general performance of the Aquamantys3 Pump Generator was evaluated with each of the proposed disposable devices, the Aquamantys3 6.0 Bipolar Sealer and the Aquamantys3 8.2L Bipolar Sealer with Cutting, in in-vivo animal (porcine) studies and related verification bench tests. These verification and validation activities demonstrated the tissue effect of the Aquamantys3 system was comparable to the predicate devices for both monopolar RF for cutting and bipolar RF concurrent with saline for haemostatic sealing and coagulation (Transcollation® technology) applications. In-vivo animal study participants included surgeons, circulating nurses and Salient Surgical Technologies' personnel.
# 8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Clinical testing was not required to establish substantial equivalence between the proposed and predicate devices.
## 9. SUMMARY OF OTHER INFORMATION
No other testing was required to establish substantial equivalence between the proposed and predicate devices.
## 10. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
Testing performed, including in-vivo animal, verification and validation testing, indicates the Aquamantys3 Pump Generator, the Aquamantys3 6.0 Bipolar Sealer and the Aquamantys3 8.2L Bipolar Sealer with Cutting are substantially equivalent to the predicate devices. The proposed devices do not introduce considerations for safety and efficacy different from the considerations of their predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three lines representing its wings. The eagle is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room WO66-G609 Silver Spring, MD 20993-0002
Salient Surgical Technologies, Inc. % Ms. Rita M. Calnan 180 International Drive Portsmouth, New Hampshire 03801
SEP - 9 2011
Re: K111285
Trade/Device Name: Aquamantys®3 Pump Generator System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: August 29, 2011 Received: August 31, 2011
Dear Ms. Calnan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
Page 2 - Ms. Rita M. Calnan
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 331-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# KIII285
# Indications for Use
510(k) Number (if known): 《《III 285
Device Name: Aquamantys®3 Pump Generator System
Indications for Use:
## The Aquamantys3 Pump Generator
The Aquamantys3 Generator is an electrosurgical generator with monopolar and bipolar RF outputs. It is intended to be used with specified disposables for delivery of RF energy for cutting of soft tissue and RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone at the operative site during surgical procedures. The system is not intended for contraceptive tubal coagulation (permanent female sterilization).
## Aquamantys3 6.0 Bipolar Sealer:
The Aquamantys3 Bipolar Sealer is a bipolar, single use, sterile, disposable device intended for use with the Aquamantys3 Generator. The device delivers bipolar RF energy concurrent with saline for haemostatic sealing and coagulation of soft tissue and bone. It is intended for, but not limited to, orthopaedic, spine, thoracic, and open abdominal surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
## Aquamantys3 8.2L. Bipolar Sealer with Cutting:
The Aquamantys3 8.2L Bipolar Sealer with Cutting is a monopolar/bipolar, single use, sterile, disposable device intended for use with the Aquamantys3 Generator. The device delivers monopolar RF energy for cutting of soft tissue and bipolar RF energy concurrent with saline for haemostatic sealing and coaqulation of soft tissue and bone. It is intended for, but not limited to, orthopaedic, spine, thoracic, and open abdominal surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Neil R. Podell (44c) for man
Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111285
Page 4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.