K111273 · Shenzhen Anpan Technology Co., Ltd. · IRT · Nov 8, 2011 · Physical Medicine
Device Facts
Record ID
K111273
Device Name
FIR HEAT THERAPHY SYSYEMS
Applicant
Shenzhen Anpan Technology Co., Ltd.
Product Code
IRT · Physical Medicine
Decision Date
Nov 8, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5740
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FIR Heat Therapy Systems are indicated for the temporary relief of minor muscle and joint pain and stiffness; the temporary relief of joint pain associated with arthritis, muscle spasms, minor strains and sprains and minor muscular back pain; muscular relaxation; and the temporary increase of local circulation where applied.
Device Story
FIR Heat Therapy Systems provide infrared heat to body areas for pain relief and muscle relaxation. Device consists of nylon-cotton blend application cover with adjustable pads containing carbon fiber heating elements; secured via Velcro fasteners. Powered by 6Vdc (4 AA batteries). Operates by emitting far-infrared electromagnetic spectrum (5-15 microns, center ~9 microns) to achieve therapeutic skin temperatures of 40-45°C. Reaches therapeutic range in ~5 minutes; maximum temperature in ~10 minutes. Intended for OTC use; patient applies device to affected area for 30-45 minute sessions. Heat application promotes local circulation and provides temporary relief of minor musculoskeletal pain and stiffness.
Clinical Evidence
Bench testing only. Testing included skin temperature tests, infrared energy/spectrum tests, and biocompatibility testing (ISO 10993-1, ISO 10993-5, ISO 10993-10). Spectrum test confirmed emission of 5-15 microns (center ~9 microns). Thermal distribution map confirmed core temperature range of ~44°C. Skin temperature testing demonstrated therapeutic range (40°C) reached in 5 minutes and maximum temperature (42°C) in 10 minutes.
Indicated for temporary relief of minor muscle/joint pain and stiffness, arthritis-associated joint pain, muscle spasms, minor strains/sprains, minor muscular back pain, muscular relaxation, and temporary increase of local circulation. No specific age or gender restrictions noted.
Regulatory Classification
Identification
A powered heating pad is an electrical device intended for medical purposes that provides dry heat therapy for body surfaces. It is capable of maintaining an elevated temperature during use.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E part 807 of this chapter subject to § 890.9.
{0}------------------------------------------------
K111273
NOV - 8 2011
Sponsor: Shenzhen Anpan Technology Co., Ltd. Subject Device: FIR Heat Therapy Systems Supplemental Information 3 for Response the 3d FDA Comment File No.:
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
#### 1. Submitter Information
Sponsor Name: Shenzhen Anpan Technology Co., Ltd. Address: No. 44, 2nd Road, XiaKeng, TongLe Community, LongGang Town, LongGang District, 518116 Shenzhen, China Contact Person: Mr. Wang Tao (Certification Engineer) Tel: +86-755-8964 0313-862 Fax: +86-755-2570 0826 Email: wangtao3801495@126.com
# Application Correspondent Information:
MEDLAB (Shenzhen) Information Service Co., Ltd. Address: Room 2706, Block A, ZhongFang JinYuan Buiding, Xinwen Road, Shenzhen, Guangdong, P.R. China, 518034 Contact Person: Ms. Sabrina Wei (Manager) Tel: +86-755-8308 9699 Fax: +86-755-8632 9134 Email: sabrinawei@hotmail.com
# 2. Subject Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Powered heating pad and infrared lamp |
| Trade Name: | FIR Heat Therapy Systems |
| Models: | EH-6601, EH-6603, EH-6604, EH-6605, EH-6606, EH-6607,<br>EH-6608, EH-6609, EH-6610, EH-6611, EH-6612, EH-6613,<br>EH-6614 |
| Classification Name: | Powered heating pad and infrared lamp |
| Review Panel: | Physical Medicine |
| Product Code: | IRT, ILY |
| Regulation Number: | CFR 890.5740, 890.5500 |
| Regulation Class: | 2 |
{1}------------------------------------------------
### 3. Predicate Device Information
| Sponsor: | EAS Consulting Group LLC. |
|--------------------|--------------------------------|
| Device Name: | Thermotex Heat Therapy Systems |
| 510(k) Number: | K092589 |
| Product Code: | IRT, ILY |
| Regulation Number: | CFR 890.5740, 890.5500 |
| Regulation Class: | 2 |
# 4. Device Description
The FIR Heat Therapy Systems provide infrared heat to different areas of the patient's body. The Systems consist of an outer application cover with adjustable pads that enclose the infrared heating elements to treat different areas of the body. The pad's covers are fabricated of nyloncotton blend. Velcro fasteners on the cover allow for adjusting the cover and pads for optimum contact to the patient's body areas.
# 5. Intended Use
The FIR Heat Therapy Systems are indicated for the temporary relief of minor muscle and joint pain and stiffness; the temporary relief of joint pain associated with arthritis, muscle spasms, minor strains and sprains and minor muscular back pain; muscular relaxation; and the temporary increase of local circulation where applied.
# 6. Conformity Standards
The manufacturer has evaluated the subject device by lab testing according to the following standards:
- IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, � 1988+A1:1991+A2:1995
- . IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. 2007
- � ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process, 2009
- . ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity, 2009
- � ISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity, 2002+A1:2006
{2}------------------------------------------------
### 7. Determination of Substantial Equivalence
Compare with predicate device, the subject device is very similar in design principle, intended use, indication for use, the applicable standards, infrared electromagnetic spectrum and therapeutic temperature range.
The difference for the key specifications listed in the following table will not affect the safety and effectiveness for the subject device.
| Item | Predicate Device | Subject Device | Note |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|--------|
| Device Name | Thermotex Heat Therapy<br>Systems | FIR Heat Therapy Systems | -- |
| Design Principle | FIR (far-infrared) therapy | FIR (far-infrared) therapy | Same |
| Patient Contacting<br>Material | Velours (nylon face cover,<br>50/50 nylon/cotton back) | Nylon-cotton blend pad<br>ABS plastic cover<br>Nylon Velcro fastener | Note 1 |
| Energy Source | 110Vac, 50W | 6Vdc (4 x "AA" alkaline<br>batteries), 8W (maximum) | Note 2 |
| Heating Element | Carbon black fiberglass<br>(glass-fabric impregnated with<br>electric conducting plastic<br>"PTFE-Carbon") | Carbon fibre | Note 3 |
| Infrared<br>Electromagnetic<br>Spectrum | 5~15 microns, with center<br>around 9 microns | 5~15 microns, with center<br>around 9 microns | Same |
| Therapeutic<br>Temperature Range | 40~45 °C | 40~45 °C | Same |
| Approximate Skin<br>Temperature | 44 °C (for Low setting) | 41~42 °C | -- |
| Time to Maximum<br>Temperature | -- | 10 minutes | Note 4 |
| Time to Reach<br>Therapeutic<br>Temperature Range | -- | 5 minutes | |
| Recommended<br>Treatment Time | 30-45 minutes | 30-45 minutes | Same |
### Note 1:
Although the patient contacting material of subject device is different from the predicate device, both of them are complied with ISO 10993-1. So this difference will not raise any safety or effectiveness issue.
# Note 2:
Although the power supply of subject device is different from the predicate device, both of them are complied with IEC 60601-1. So this difference will not raise any safety or effectiveness issue.
{3}------------------------------------------------
#### Note 3:
Carbon black fiberglass is an E-Glass that is woven and dyed black to look like a 3K Twill Weave carbon fiber. The carbon black fiberglass does not have the strength properties of a real carbon fiber. But their far infrared thermal specification is similar, so this difference will not raise any safety or effectiveness issue,
#### Note 4:
Although Approximate Skin Temperature and Time to Reach Temperature Range of predicate device are different from the predicate device, they are both in the FDA generally considering therapeutic temperature range 40-45°C, so this difference will not raise any safety or effectiveness issue.
#### 8. Performance Testing
We have provided the performance tests including the skin temperature test and the infrared energy and spectrum test to support the safety and effectiveness of the subject device.
The spectrum test shows that the subject device emits the main infrared electromagnetic spectrum 5~15 microns, with center around 9 microns.
The infrared energy test (Thermal Distribution Map) shows that the core temperature range of heating pad is about 44°C.
The skin temperature test shows that the time for skin surface to reach therapeutic temperature range 40°C is about 5 minutes; the time to reach maximum temperature 42°C is about 10 minutes.
#### 9. Conclusion
The subject devices have all features of the predicate devices. The few differences do not affect the safety and effectiveness of the subject devices.
Thus, the subject devices are substantially equivalent to the predicate devices.
### 10. Summary Prepared Date: 3 November 2011
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shenzhen Anpan Technology Co., Ltd. % MEDLAB (Shenzhen) Information Service Co., Ltd. Ms. Sabrina Wei 11900 Dulongpre Apt. 207 Street Montreal, Quebec Canada H4J 0A1
NOV - 8 2011
Re: K111273
Trade/Device Name: FIR Heat Therapy Systems, models: EH-6601, EH-6603, EH-6604, EH-6605, EH-6606, EH-6607, EH-6608, EH-6609, EH-6610, EH-6611, EH-6612, EH-6613, EH-6614 Regulation Number: 21 CFR 890.5740 Regulation Name: Powered heating pad Regulatory Class: Class II Product Code: IRT, ILY
Dated: October 19, 2011
Received: October 26, 2011
Dear Ms. Wei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
# Page 2 - Ms. Sabrina Wei
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Mark N. Melke Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Sponsor: Shenzhen Anpan Technology Co., Ltd. FIR Heat Therapy Systems Subject Device: 510(k) submission report (V1.0), Chapter 3 File No.:
Chapter 3. Indications for Use . .
# Indications for Use
-- •
510(k) Number (if known):
Device Name: FIR Heat Therapy Systems, models: EH-6603, EH-6604, EH-6604, EH-6606, EH-6607, EH-6608, EH-6609, EH-6610, EH-6611, EH-6612, EH-6613, EH-6614
Indications for Use:
The FIR Heat Therapy Systems are indicated for the temporary relief of minor muscle and joint pain and stiffness; the temporary relief of joint pain associated with arthritis, muscle spasms, minor strains and sprains and minor muscular back pain; muscular relaxation; and the temporary increase of local circulation where applied.
Prescription Use_ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-Off) (Division Sign-Old)
Division of Surgical, Orthopedic,
Division of Surgices Division of Surgers and Restorative Devices and Restorative Devices
510(k) Number K111273
Report by Shenzhen Anpan Technology Co., Ltd.
Page 3-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.