A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. This device is for over-the counter use.
Device Story
Disposable nitrile examination gloves; non-sterile; powder-free. Available in Original Blue and Cobalt Blue. Used by healthcare professionals or in over-the-counter settings to provide a protective barrier against contamination. Tested for resistance to permeation by chemotherapy drugs per ASTM D6978-05. Gloves act as a physical barrier; worn on hands/fingers. Protects both patient and examiner from cross-contamination. Performance verified via watertight testing and physical property assessment (tensile strength, elongation).
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ASTM D6319-05 (physical properties, dimensions, powder content) and ASTM D6978-05 (chemotherapy drug permeation). Watertight integrity confirmed via FDA 1000ml water leak test (AQL 2.5). Biocompatibility confirmed via primary dermal irritation and Guinea Pig Maximization (Buehler) tests.
Technological Characteristics
Material: Nitrile rubber. Powder-free. Dimensions and physical properties meet ASTM D6319-05. Chemotherapy resistance tested per ASTM D6978-05. Biocompatibility: Passed primary skin irritation and sensitization tests. Non-sterile.
Indications for Use
Indicated for use as a disposable patient examination glove worn by examiners to prevent cross-contamination between patient and examiner. Suitable for use with specific chemotherapy drugs, excluding Carmustine and Thiotepa.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
# YTY INDUSTRY (MANJUNG) SDN. BHD.
(Company No : 380830-P) Lot 1422-1424, Batu 10 Lekir, 32020 Sitiawan, Perak Darul Ridzuan, Malaysia. Tel : 05-6792288 (Hunting Line), 6792443 & 6792445 Fax : 05-6791188
DEC - 1 2011
| 2 | - | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
|---|---|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | | |
| | | | |
510 (K) SUMMARY SHEETS
| 1.0 | | SMDA 510 (K) SUMMARY |
|-----|--------------------------|--------------------------------------------------------------------------------------------------------------------------|
| 2.0 | Submitter | YTY INDUSTRY (MANJUNG) SDN. BHD.,<br>Lot 1422-1424, Batu 10 Lekir<br>32020 Sitiawan, Perak<br>Malaysia |
| | Tel | 605-6792288 |
| | Fax | 605-6791188 |
| | Name of Contact Person | 1. Mr. Roger Moh Ung Nang<br>2. Mr. Arivalagan |
| | Date of Summary Prepared | November 1, 2010 |
| 3.0 | Name of Device | |
| | Trade Name: | Non-Sterile, Powder-Free Nitrile Examination Gloves (Original Blue, Cobalt Blue) Tested for use with Chemotherapy Drugs. |
| | Common Name: | Synthetic Rubber Examination Gloves |
Classification Name: Patient Examination Glove, Powder Free
Identification of The Legally Marketed Devices 4.0
> The legally marketed device to which we claim substantial equivalence is K101822, Blue Nitrile Examination Gloves, Tested for Use With Chemotherapy Drugs Labeling Claim, (Non-Sterle), which is a Class 1 device with Product Codes LZA/LZC. The predicate device was approved Nov 19, 2010, Except for the color differences, and some different chemotherapy chemicals broakthroughtesting times, we do claim this submission is essentially equivalent.
#### 5.0 Description of The Device
Non-Sterile, Powder-Free Nitrile Examination Gloves (Original Blue, Cobalt Blue) Tested for use with Chemotherapy Drugs, meets all the current specifications listed under ASTM Specifications D6319-05 , Standard Specification for Nitrile Examination Gloves for Medical Application and D6978-05, Standard Practice for Assessment of Resistance to Permeation by Chemotherapy Drugs.
Page 1 of 1
{1}------------------------------------------------
#### 6.0 The Intended Use of Glove
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. This device is for over-the counter use.
### Summary of Performance Data: 7.0
| Performance data of gloves based on ASTM D6319-10 and FDA 1000ML watertight test. | | | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------|-------------------------------------------|---------------|
| Test | FDA 1000ml Water Leak<br>Test | Powder Free Nitrile<br>Examination Gloves | |
| | Multiple Normal | Original Blue | Cobalt Blue |
| 1. Watertight (1000ml) | GI<br>AQL = 2.5 | Pass | Pass |
| | | | |
| Test | ASTM D6319-10 | | |
| 2. Length (mm)<br>Size<br>M<br>L<br>XL | Min 230 (chemo claim min 270)<br>Min 230 (chemo claim min 270) | 273 - 278 | 276 - 278 |
| 3. Palm width (mm)<br>Size<br>M | 95 ± 10 | 97 - 99 | 98 - 99 |
| 4. Thickness (mm)<br>(Single Layer) | | | |
| Finger | Min 0.05 (chemo claim min0.10) | 0.15-0.21 | 0.16-0.20 |
| Palm | Min 0.05 (chemo claim min0.10) | 0.11-0.12 | 0.11-0.12 |
| 5. Physical Properties | | | |
| Before Aging | | | |
| Tensile Strength (MPa) | Min 14 | 20.3 - 25.9 | 19.6 - 25.3 |
| Ultimate Elongation (%) | Min 500 | 580 - 620 | 560 - 580 |
| After Aging | | | |
| Tensile Strength (MPa) | Min 14 | 26.5 - 35.6 | 25.0 - 28.4 |
| Ultimate Elongation (%) | Min 400 | 500 - 520 | 460 - 520 |
| 6. Powder Content | Max 2.0mg/glove | 0.34 mg/glove | 0.39 mg/glove |
| 7. Moisture Content | Max 2.0% | 0.92% | 0.84% |
Page 2 of 4
510 (k) Summary
Page 2 of 4 revised 10-20-2011
Joanna Rucker
{2}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE COMPARISONS
12111248
| TESTED PERFORMANCE DATA FOR BLUE | | | SGMP K101822 |
|-----------------------------------------------------------|-------------|-------------|--------------|
| YTY POWDER FREE NITRILE EXAM GLOVES | | | BLUE NITRILE |
| WITH CHEMO DRUG LABELING CLAIMS | | | CHEMO TESTED |
| ASTM D6319-05 NITRILE SPECIFICATION | | | |
| | ORIGINAL | COBALT | K101822 |
| FDA 1000ML Water Tight AOL | 2.5 | 2.5 | 2.5 |
| Length (mm) | 244-249 | 247-250 | 295 |
| Width (mm) | 97-99 | 98-99 | 95 |
| Thickness (mm) | | | |
| Finger: | 0.12 - 0.14 | 0.12 - 0.13 | 0.18 - 0.21 |
| Palm: | 0.06 - 0.07 | 0.07 - 0.08 | 0.16-0.19 |
| Tensile Strength (Mpa) | | | |
| Before Aging | 17.8 - 20.0 | 23/0 - 27.0 | 14 |
| After Aging | 21.3 - 23.6 | 30.2 - 34.6 | 14 |
| Ultimate Elongation (%) | | | |
| Before Aging | 680 - 720 | 660 - 680 | 500 |
| After Aging | 600 - 620 | 560 - 580 | 400 |
| Finished Total Powder Level (mg/glove | 0.48 | 0.22 | <2.0 |
| Biocompatibility Testing (CPTC) | | | |
| Pass Primary Skin Irritation in Rabbits | Yes | Yes | Yes |
| Pass Guinea Pig Maximization | Yes | Yes | Yes |
| CHEMOTHERAPY TESTING BY ARDL CHEMICAL ANALYTICAL SERVICES | | | |
| Minimum Breakthrough Detection Time in Minutes: | | | |
| Carboplatin, 10 mg/ml | >240 | >240 | Not tested |
| Carmustine (BCNU), 3.3 mg ml | 1.84 | 1.82 | 0.49 |
| Cisplatin, 1.0 mg/ml | >240 | >240 | >240 |
| Cyclophosphamide (Cytoxan), 20.0 mg/ml | >240 | >240 | >240 |
| Dacarbazine (DTIC), 10.0 mg/ml | >240 | >240 | >240 |
| Doxorubicin Hydrochloride, 2.0 mg/ml | >240 | >240 | >240 |
| Etoposide (Toposar), 20.0 mg/ml | >240 | >240 | >240 |
| Fluorouracil, 50.0 mg/ml | >240 | >240 | >240 |
| Ifosfamide, 50,0 mgs/ml | >240 | >240 | Not tested |
| Methotrexate, 25 mg/ml | >240 | >240 | >240 |
| Mitomycin C. 0.5 mg/ml | >240 | >240 | Not tested |
| Mitoxantrone, 2 mg/ml | >240 | >240 | Not tested |
| Paclitaxel (Taxol), 6.0 mg/ml | >240 | >240 | >240 |
| Thiotepa, 10.0 mg/ml | 0.76 | 0.93 | 2.61 |
| Vincristine Sulfate, 1.0 mg/ml | >240 | >240 | >240 |
* WARNING: DO NOT USE WITH CARMUSTINE OR THIOTEPA´ is included as warnings on all three products (YTY and SGMP labeling claims)
{3}------------------------------------------------
- The performance data of the glove as shown above meet the ASTM D6319-05 8.0 Standard and FDA's requirement. Standard and I DA 3 requirement.
Powder content is below 2 mg per glove which meet the FDA Requirements.
- The Bio-compatibility Test consists of Primary Dermal Irritation Test and Guinea Pig 9.0 Sensitization (Buehler) test. For both colors The gloves pass the Bio-compatibility Test.
- Conclusion 10.0
.
- We conclude that the Non-Sterile, Powder-Free Nitrile Examination Gloves (Original Blue, We Concided that the Non Sterth, Chemotherapy Drugs does meet:
- ASTM D6319-05 Standard
- FDA pinhole requirements -
- PDA philose requirements
Are below the maximum Powder Residual Content as specified in ASTM -D6319-02
- ASTM D5151 (which equates to the FDA Water Leak Test) -
- ASTM D6978-05 (Chemotherapy Testing Standards)
**Page**
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes, possibly representing human figures or waves, stacked on top of each other.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC = 1 2011
YTY Industry (Manjung) Sdn. Bhd. C/O Ms. Janna P. Tucker Consultant Tucker & Associates 198 Avenue De La D'Emerald Sparks, Nevada 89434
Re: K111248
Trade/Device Name: Non-Sterile, Powder Free Nitrile Examination Gloves (Original Blue, Cobalt Blue). Tested for Use with The Below Chemotherapy Drugs
Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LZA, LZC Dated: November 10, 2011 Received: November 21, 2011
Dear Ms. Tucker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Ms. Tucker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices /Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D, man
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure .
{6}------------------------------------------------
## INDICATIONS FOR USE
| APPLICANT: | YTY INDUSTRY (MANJUNG) SDN BHD |
|----------------|-----------------------------------------------------|
| 510(K) NUMBER: | K111248 |
| DEVICE NAME: | Non-Sterile, Powder Free Nitrile Examination Gloves |
Non-Sterile, Powder Free Nitrile Examination Gloves (Original Blue, Cobalt Blue). Tested for use with the below chemotherapy drugs.
INDICATIONS FOR USE:
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient an examiner.
## Chemotherapy Drug Permeation (breakthrough detection time) in minutes. The following chemicals have been tested with these gloves:
| TESTED CHEMOTHERAPY DRUG | Original Blue Average<br>Breakthrough Detection<br>Time in Minutes | Original Blue Average<br>Breakthrough Detection<br>Time in Minutes |
|----------------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|
| Carboplatin, 10 mg/ml | >240 | >240 |
| Carmustine (BCNU), 3.3 mg/ml) | **<br>1.84 | **<br>1.82 |
| Cisplatin, 1.0 mg/ml | >240 | >240 |
| Cyclophosphamide (Cytoxan), 20.0 mg/ml | >240 | >240 |
| Dacarbazine (DTIC), 10.0 mg/ml | >240 | >240 |
| Doxorubicin Hydrochloride, 2.0 mg/ml | >240 | >240 |
| Etoposide (Toposar) 20.0 mg/ml | >240 | >240 |
| Fluorouracil, 50.0 mg/ml | >240 | >240 |
| Ifosfamide, 50.0 mg/ml | >240 | >240 |
| Methotrexate, 25 mg/ml | >240 | >240 |
| Mitomycin C, 0.5 mg/ml | >240 | >240 |
| Mitoxantrone, 2 mg/ml | >240 | >240 |
| Paclitaxel (Taxol) 6.0 mg/ml | >240 | >240 |
| Thiotepa, 10.0 mg/ml | **<br>0.76 | **<br>0.93 |
| Vincristine Sulfate, 1.0 mg/ml | >240 | >240 |
** Please note that Carmustine (BCNU)and Thiotepa have extremely low permeation times for both glove colors. These two chemicals are not recommended for use with either color of the gloves.
**Prescription Use**
(Part 21 CFR 801.Subpart D
and/or
**Over-The-Counter**
21 CFR 801.Subpart C
Concurrence of CDRH, Office of Device Evaluation (ODE)
Adhip Ganguly
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
-10(k) Number: K111248
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.