K111170 · Smith & Nephew, Inc. · OMP · Dec 15, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111170
Device Name
PICO
Applicant
Smith & Nephew, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Dec 15, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. Product Codes: 66800862, 66800863, 66800865 and 66800866 are indicated for use on low and moderately exuding wounds. Examples of appropriate wound types include: Chronic . Acute . Traumatic . Subacute and dehisced wounds . Partial-thickness burns . Ulcers (such as diabetic or pressure) . Flaps and grafts . Closed surgical incisions . PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting. PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low levels of exudate and infectious materials. Product Codes: 66800864, 66800867, 66800868 and 66800869 are indicated for use on low exuding wounds, Examples of appropriate wound types include: Chronic . Acute . Traumatic . Subacute and dehisced wounds . Partial-thickness burns . Ulcers (such as diabetic or pressure) . Flaps and grafts . Closed surgical incisions . PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.
Device Story
PICO is a portable, single-use negative pressure wound therapy (NPWT) system. It consists of an electric motor-driven, twin-diaphragm vacuum pump connected to a super-absorbent, adhesive dressing. The pump is battery-operated (two AA lithium batteries) and provides up to 168 hours (7 days) of therapy. Operation involves applying the dressing to the wound, securing it with fixation strips, and activating the pump to create negative pressure, which draws excess fluid into the dressing. The device is designed to automatically stop after 168 hours and will not restart. Used in hospital and homecare settings by clinicians or patients. The system provides a non-invasive method to manage wound exudate, potentially promoting healing. Healthcare providers monitor the wound and dressing status to inform clinical decisions.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-5 and ISO 10993-10. Electrical safety and electromagnetic compatibility testing performed per IEC 60601-1 and IEC 60601-1-2. Software validation performed per IEC 62304. Wound healing study referenced for the wound contact layer (ALLEVYN Gentle Border).
Technological Characteristics
Portable, battery-operated (2x AA lithium) twin-diaphragm vacuum pump. System includes super-absorbent adhesive dressing and secondary fixation strips. Materials assessed for biocompatibility per ISO 10993. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC). Single-use, non-sterile pump; sterile dressing/strips. Software validated per IEC 62304.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to promote healing via removal of low to moderate exudate and infectious materials. Applicable to chronic, acute, traumatic, subacute, dehisced wounds, partial-thickness burns, ulcers (diabetic/pressure), flaps, grafts, and closed surgical incisions. Suitable for hospital and homecare settings.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
NPD 1000 Negative Pressure Wound Therapy System (K080275)
{0}------------------------------------------------
K111170
DEC 1 5 2011
pg 10+6
## 510(k) SUMMARY
### PICO Single Use Negative Pressure Wound Therapy System
| 1. Submitter: | Smith & Nephew, Inc. |
|---------------|------------------------------------|
| | 970 Lake Carillon Drive, Suite 110 |
| | St. Petersburg, FL 33716 |
- 2. Contact: Terry McMahon Director of Regulatory Affairs & Quality, North America 727-399-3785 Email: terry.mcmahon@smith-nephew.com
- 3. Device name: PICO Single Use Negative Pressure Wound Therapy System Common name: Negative Pressure Wound Therapy powered suction pump Classification name: Powered suction pump (21 CFR 878.4780) Product Classification/Code: Class II, OMP
- 4. Date summary prepared: December 15, 2011
- 5. Predicate Device Information: NPD 1000 Negative Pressure Wound Therapy System (K080275) Kalypto Medical 6393 Oakgreen Avenue Hastings, MN 55033
Renasys Go (K083375) Smith & Nephew Inc, St. Petesburg, FL
{1}------------------------------------------------
KIIII70
Prevena Incision Management System (K100821) KCI USA, INC. 6203 Farinon Dr. San Antonio, TX 78249
#### 6. Device Description:
PICO Single Use Negative Pressure Wound Therapy System is a small, lightweight, portable suction device consisting of a electric motor driven, twindiaphraqm, vacuum pump connected to a super-absorbent, gentle adhesive dressing. The pump is supplied non-sterile and single use, the dressing and secondary fixations strips are supplied sterile and single use. The dressing is applied to the wound and secondary fixation strips are placed over the outside edges to help hold the dressing in place. When the suction pump is turned on, air is pulled out of the dressing creating negative pressure and drawing excess fluid from the wound into the dressing. The pump is battery operated and is supplied with two AA lithium batteries which provide up to 168 hours (7 days) of battery life depending upon leak rate. The batteries can be replaced if required. The pump is designed to stop working after 168 hours (7 days) of use and will not re-start after this time, even with new batteries. Negative pressure will not be applied at this point. PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting. A carry bag is not provided with the system however will be available to order separately.
{2}------------------------------------------------
KIIII70
pg 3 of fo
### 7. Intended Use
# 7a. Low and moderately exuding wounds
PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.
Product Codes: 66800862, 66800863, 66800865 and 66800866 are indicated for use on low and moderately exuding wounds.
Examples of appropriate wound types include:
- Chronic .
- Acute .
- Traumatic .
- Subacute and dehisced wounds .
- Partial-thickness burns .
- Ulcers (such as diabetic or pressure) .
- Flaps and grafts .
- Closed surgical incisions .
PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.
{3}------------------------------------------------
K111170
#### 7b. Low exuding wounds
PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low levels of exudate and infectious materials.
Product Codes: 66800864, 66800867, 66800868 and 66800869 are indicated for use on low exuding wounds,
Examples of appropriate wound types include:
- Chronic .
- Acute .
- Traumatic .
- Subacute and dehisced wounds .
- Partial-thickness burns . �
- Ulcers (such as diabetic or pressure) .
- . Flaps and grafts
- Closed surgical incisions .
PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.
#### 8. Summary of Non-Clinical Testing
The biocompatibility of the PICO Single Use Negative Pressure Wound Therapy System has been demonstrated through assessment according to ISO 10993-1: 2003 and appropriate in vivo and in vitro tests have been conducted using product that has been packaged and sterilised. These tests include cytotoxicity, sensitization and irritation and cover the PICO dressing and secondary fixation strips. These studies indicated that the PICO Single Use Negative Pressure Wound Therapy System is safe for its intended use.
{4}------------------------------------------------
KIIII70.
pg 5 of 6
| Test | Description of Method | Result |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Cytotoxicity | PICO Dressing<br>In order to determine the cytotoxic potential<br>of the dressing, a Minimum Essential<br>Medium (MEM) elution assay was<br>conducted in accordance with ISO 10993-5. | N to 1:16, all non-cytotoxic<br>(Grade 0) |
| | Secondary Fixation Strips<br>Cytotoxicity testing has been performed on<br>the secondary fixation strips, tested in<br>accordance with ISO 10993-5. The test<br>article extract was diluted in supplemented<br>EMEM 1X and placed onto triplicate<br>confluent monolayers of L-929 mouse<br>fibroblast cells at the 100%, 50%, 25% and<br>5% solutions (v/v). | N to 1:16, all non-cytotoxic<br>(Grade 0) |
| Sensitisation | PICO Dressing<br>A Guinea Pig Maximisation Test was<br>performed on the dressing in accordance<br>with ISO 10993-10. | Non-sensitising |
| | Secondary Fixation Strips<br>A guinea pig maximisation test was<br>performed on the secondary fixation strips<br>in accordance with ISO 10993-10. | Non-sensitising |
| Irritation | PICO Dressing<br>The dressing was evaluated for irritation in<br>accordance with ISO 10993-10. | PII SC = 0.0<br>PII SC = 0.1<br>Negligible irritant |
| | Secondary Fixation Strips<br>Irritation testing has been performed on the<br>secondary fixation strips in accordance<br>with ISO 10993-10. | 0.9% NaCl extract: Primary<br>Irritation Index: 0.11<br>(Negligible) |
{5}------------------------------------------------
Additionally, the effects of the dressing wound contact layer have been assessed during a wound healing study on ALLEVYN Gentle Border dressings, report reference which have the same wound contact layer as the dressing included in the PICO Single Use Negative Pressure Wound Therapy System.
604 6
ုးရှိ
Testing has been completed in accordance with IEC / UL 60601-1; General Requirements for Safety of Electrical Medical Equipment and IEC 60601-1-2; Electromagnetic Safety Requirements For Electrical Medical Equipment. Software validation has been completed in accordance with IEC 62304 Software for Medical Devices.
#### 9. Conclusions Drawn
Based on results of testing performed the subject device is substantially equivalent to the predicate devices NPD 1000 Negative Pressure Wound Therapy System (K080275), Renasys Go (K083375) and Prevena Incision Management System (K100821).
- PICO has similar design characteristics and provides similar functions to NPD 1000 Negative Pressure Wound Therapy System (K080275), Renasys Go (K083375) and Prevena Incision Management System (K100821). The intended use, indications and instructions for use for the subject and predicate devices are similar.
PICO does not raise any new issues of safety and effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 1 5 2011
Smith and Nephew, Inc. % Mr. Terry McMahon 970 Lake Carillon Drive, Suite 110 St. Petersburg, Florida 33716
Re: K111170
Trade/Device Name: PICO Single Use Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: December 12, 2011 Received: December 13, 2011
Dear Mr. McMahon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{7}------------------------------------------------
Page 2 - Mr. Terry McMahon
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Sincerely yours,
F-n Patinir C.C., A
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K111170
Device Name: PICO Single Use Negative Pressure Wound Therapy System
Indications for Use:
PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.
Product Codes: 66800862, 66800863,66800865 and 66800866 are indicated for use on low and moderately exuding wounds.
Examples of appropriate wound types include:
- Chronic .
- Acute .
- Traumatic .
- Subacute and dehisced wounds �
- Partial-thickness burns .
- Ulcers (such as diabetic or pressure) ●
- Flaps and grafts .
- Closed surqical incisions .
PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.
AND/OR Prescription Use X
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David Kane for MKM
(Division Sign Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111170
{9}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K111170
Device Name: PICO Single Use Negative Pressure Wound Therapy System
Indications for Use:
PICO is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low levels of exudate and infectious materials.
Product Codes: 66800864, 66800867,66800868 and 66800869 are indicated for use on low exuding wounds,
Examples of appropriate wound types include:
- Chronic .
- Acute .
- Traumatic .
- Subacute and dehisced wounds .
- Partial-thickness burns .
- Ulcers (such as diabetic or pressure) .
- Flaps and grafts .
- Closed surgical incisions .
PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.
Prescription Use X AND/OR Over-The-Counter Use
(21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Keane for MXM
(Division Sign-On)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K11170
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.