← Product Code [LLZ](/productcode/LLZ) · K111142

# ECHOSOFT(TM) (K111142)

_Echometrix, LLC · LLZ · Sep 21, 2012 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K111142

## Device Facts

- **Applicant:** Echometrix, LLC
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Sep 21, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

EchoSoft™ is a software calculation package that is used with diagnostic ultrasound images to provide mechanical information about tendon and tendon-like tissues (such as ligaments) that may be used by a physician, along with other medical data, to assist in clinical diagnosis of various types of physical conditions or injuries.

## Device Story

EchoSoft is a standalone software package that processes diagnostic ultrasound images to analyze mechanical properties of tendon and tendon-like tissues. It utilizes an acoustoelastography technique to track motion, deformation, and ultrasonic echo magnitude changes within a region of interest. The software is intended for use by physicians to gather qualitative mechanical data, which is then integrated with other clinical information to assist in diagnosing injuries or conditions. It operates by ingesting ultrasound images via standard communication protocols like DICOM. The output provides a visual or quantitative representation of tissue mechanical characteristics. The device is used in clinical settings where ultrasound imaging is performed. It benefits patients by providing non-invasive assessment of soft tissue mechanical and functional properties, aiding in more informed clinical decision-making.

## Clinical Evidence

Bench testing only. Performance was verified using laboratory phantoms with varying elastic properties to assess resolution, sensitivity, and precision. Additional testing was conducted on bovine tendons with known defects to demonstrate performance on tissue samples. Clinical images of previously diagnosed tendon injuries were used to demonstrate software performance. Reproducibility was verified across multiple ultrasound systems (GE Logiqe, Siemens S2000, Terason t3000).

## Technological Characteristics

Standalone software package; utilizes acoustoelastography technique; processes diagnostic ultrasound images; inputs via DICOM communication protocols; algorithm performs motion, deformation, and echo magnitude tracking; intended for use with standard diagnostic ultrasound systems.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Philips iU22 Diagnostic Ultrasound System ([K093563](/device/K093563.md))
- Toshiba SSA-790A Aplio XG v4.0R001 Diagnostic Ultrasound System ([K092179](/device/K092179.md))
- Hitachi Hi VISION Preirus Diagnostic Ultrasound System ([K093466](/device/K093466.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

|||||42

SEP 2 1 BOR

Echometrix, LLC 437 S. Yellowstone Drive Madison, WI 53719 USA

# 610 K SUMMARY OF SAFETY AND EFFECTIVENESS

- o This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
### o Identification of Submitter

Echometrix, LLC 437 S. Yellowstone Drive, Suite 210 Madison, WI 53719 Contact: Larry A. Kroger, Ph.D., Regulatory Consultant Ph. 262-549-6646 Revised July 12, 2012

Identification of the Product

Proprietary name: EchoSoft™ Classification name: Picture archiving and communication system Product Code: LLZ

### 0 Marketed Devices_

EchoSoft™ is a new software package which utilizes an acoustoelastography technique with diagnostic ultrasound images to analyze the mechanical properties of tendon and tendon-like tissues. It performs functions similar to software currently available on the following marketed diagnostic ultrasound devices:

- Philips iU22 Diagnostic Ultrasound System intended for diagnostic A ultrasound imaging in B-mode (or 2-D), M-mode (including Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging, Harmonics (Tissue and Contrast) and Elastography modes, PMN# K093563, classification number and name 21 CFR 892.1550 Ultrasonic pulsed doppler imaging system, product code IYN.
- > Toshiba SSA-790A Aplio XG v4.0R001 Diagnostic Ultrasound System intended for diagnostic imaging in B mode, M mode, PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler), Color Doppler, Dynamic Flow, 3D Imaging, and indicated for determining the relative elasticity of a lesion, PMN# K092179, classification number and name 21 CFR 892.1550 Ultrasonic
- pulsed doppler imaging system, product code IYN.
- A Hitachi Hi VISION Preirus Diagnostic Ultrasound System intended for diagnostic imaging in B mode, M mode, PW mode (Pulsed

{1}------------------------------------------------

- Wave Doppler), CW mode (Continuous Wave Doppler), Color Doppler, Amplitude Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, 4D Imaging, Real Time Tissue Elastography, and Real Time Virtual Sonography, PMN# K093466, classification number and name 21 CFR 892.1560 Ultrasonic pulsed echo imaging system, product code IYO.
### 0 Device Description

The Echometrix, LLC software, EchoSoft, is an image processing algorithm that, when used with diagnostic ultrasound images, provides qualitative information about the mechanical characteristics of the deforming material by tracking motion, deformation, and ultrasonic echo magnitude change within a given region. This technology can be used by a physician to gather information about the mechanical and functional properties of soft tissues which, in conjunction with standard medical data, can be used to assist in clinical diagnosis.

# Intended Uses

EchoSoft™ is a software calculation package that is used with diagnostic ultrasound images to provide mechanical information about tendon and tendon-like tissues (such as ligaments) that may be used by a physician, along with other medical data, to assist in clinical diagnosis of various types of physical conditions or injuries.

# Comparison with Predicate

EchoSoft™ is a software calculation package that will be marketed as stand-alone software. Ultrasound images are accessed and incorporated into the analysis package via standard communication protocols such as DICOM. For the predicate devices similar software capabilities are incorporated into the system and are marketed as part of a total diagnostic ultrasound system.

### 0 Summary of Studies

The EchoSoft 100 software was evaluated with multiple studies to verify the ability of the algorithm to distinguish between materials with different mechanical properties. The studies consisted of laboratory measurements on phantoms containing materials with varying elastic properties to verify performance parameters such as resolution, sensitivity and precision. Several studies were conducted on bovine tendons to demonstrate performance on tissue samples with known defects. Sample clinical images were also provided to demonstrate performance of the software on tendons with previously diagnosed injuries. Some of the studies were conducted on ultrasound systems from different

{2}------------------------------------------------

manufacturers to verify reproducibility of the results. The ultrasound systems used in the studies were produced by GE (model Logiqe), Siemens (model S2000), and Terason (model t3000).

# Conclusions

o

It is the opinion of Echometrix that EchoSoft™ software is substantially equivalent to the elastography options available for the predicate devices. EchoSoft™ does not include any new indications for use with regards to evaluating the mechanical properties of tendon and tendon-like tissues. nor does use of this software result in any new potential hazards.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol of an eagle or bird-like figure, represented by three curved lines that suggest wings and a body.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

SEP 21 2012

Echometrix, LLC % Mr. Larry A. Kroger Regulatory Consultant 2210 Woodhiil Way WAUKESHA WI 53189

Re: K111142

Trade/Device Name: EchoSoft™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 30, 2012 Received: July 31, 2012

# Dear Mr. Kroger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Director Pivision of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

# 510(k) Number (if known): K111142

Device Name: EchoSoft™

Indications For Use: EchoSoft™ is a software calculation package that is used with diagnostic ultrasound images to provide mechanical information about tendon-like tissues of interest that may be used by a physician, along with other medical data, to assist in clinical diagnosis. The software is intended to be used by trained professionals only.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

### Concurrence of CDRH, Office of Device Evaluation (ODE)

Aluph D. O'thim.

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K. K///42

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K111142](https://fda-staging.innolitics.com/device/K111142)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
