K111100 · C.R. Bard, Inc. · KNW · Jun 16, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K111100
Device Name
ENCOR ENSPIRE BREAST BIOPSY SYSTEM
Applicant
C.R. Bard, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Jun 16, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The EnCor Enspire™ Breast Biopsy System is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. It is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality. The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g. fibroadenoma, fibrocystic lesion), the EnCor Enspire™ Breast Biopsy Systems may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Device Story
Automated directional vacuum-assisted breast biopsy system; integrates control module, vacuum system, and touch screen display into single console. Inputs: user selections via touch screen, driver/probe activation, foot pedal signals. Operation: vacuum system creates negative pressure at probe tip to draw tissue into acquisition chamber; control module adjusts motor speed for drivers/probes based on user input. Output: breast tissue samples for histology; system status displayed on LCD. Used in clinical settings by physicians. Benefits: enables minimally invasive diagnostic tissue sampling; allows partial/complete removal of abnormalities for histologic examination.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through non-clinical testing including vacuum level, joint tensile strength, cassette functionality, rinse volume, reliability, packaging validation, user validation, and software verification/validation.
Technological Characteristics
Integrated console with vacuum system and touch screen display. Disposable vacuum/rinse tubing sets housed in ABS plastic. Motorized driver/probe control. Software-based GUI for system status and operation. Electrical compatibility maintained. Sterilization method not specified.
Indications for Use
Indicated for patients requiring diagnostic breast tissue sampling of imaged or palpable abnormalities. Used for partial or complete removal of imaged abnormalities and partial removal of palpable benign lesions (e.g., fibroadenoma, fibrocystic lesion).
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
EnCor Enspire™ Breast Biopsy System JUN 16 2011
## 510(k) Summary of Safety and Effectiveness
21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
## Submitter Information:
Applicant: SenoRx, Inc. A Business Unit of Bard Peripheral Vascular, Inc. 3 Morgan Irvine, CA 92618
Phone: 1-480-638-2915
1-480-449-2546 Fax:
Contact: Justin Lovelace, Regulatory Affairs Specialist
Date: April 18, 2011
# Subject Device Name:
| Device Trade Name: | EnCor Enspire <sup>TM</sup> Breast Biopsy System |
|-----------------------|----------------------------------------------------------------------------------|
| Common or Usual Name: | Biopsy Instrument |
| Classification: | Class II (21 CFR 876.1075) |
| Classification Panel: | Gastroenterology & Urology |
| Product Code: | KNW |
| Predicate Device: | EnCor <sup> ® </sup> Breast Biopsy System (K093512; Clearance November 20, 2009) |
Bard Peripheral Vascular, Inc.
Image /page/0/Picture/15 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, with a white fill. The word is slightly distorted, with the letters appearing to be stretched horizontally. The background is white.
(1 of 5)
{1}------------------------------------------------
#### Device Description:
The subject device is the EnCor Enspire™ Breast Biopsy System, an automated directional vacuum-assisted biopsy device consisting of an integrated control module and vacuum system with a detachable touch screen display. The key functional components of the EnCor Enspire™ Breast Biopsy System are integrated into a single console.
Control Module - The control module is designed to provide control operations and serve as an interface for drivers, probes and foot pedals.
- Inputs are delivered to the control module via either a driver or foot pedal. o
- Based on the input received, the control module adjusts the motor speed of o drivers and probes.
- System status is provided to the physician via an integrated LCD screen. o
Vacuum System - During a breast biopsy procedure, the vacuum system draws the target tissue into the biopsy probe tissue acquisition chamber by creating a negative pressure differential at the tip of the probe being used. A disposable vacuum canister and disposable vacuum tubing cassette (with or without optional rinse tubing) is installed in the vacuum system at the time of each breast biopsy procedure.
Touch Screen Display – The touch Screen Display provides an intuitive user interface for the EnCor Enspire™ Breast Biopsy System. The display provides the system status and initiates operation of the installed driver and probe based on information and options entered or selected by the user.
#### Indications for Use of Device:
The EnCor Enspire™ Breast Biopsy System is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities.
It is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
It is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy.
Bard Peripheral Vascular, Inc.
BARD
{2}------------------------------------------------
When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g. fibroadenoma, fibrocystic lesion), the EnCor Enspire™ Breast Biopsy Systems may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
#### Technological Comparison to Predicate Devices:
The predicate device is the EnCor® Breast Biopsy System (K093512, cleared on November 20, 2009). It is referenced as the predicate device for the EnCor Enspire™ Breast Biopsy System because the two are substantially equivalent in terms of:
Intended Use Indications for Use Contraindications Patient Population Mechanics of Action Mode of Action Electrical Compatibility Patient Contacting Materials Available Accessories and Disposable Components
The subject device and the predicate device are different in the following manner:
The subject device integrates the major components of the predicate device (control module, vacuum system and cart) into a single assembly.
The subject device adds a large touch screen display with integrated graphical user interface software which replaces the small LCD screen display with integrated into the control module of the predicate device.
The subject device adds an adjustable tray to the body of the console.
Bard Peripheral Vascular, Inc.
Image /page/2/Picture/12 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, with a white fill. The word is horizontally oriented and appears to be a logo or brand name.
{3}------------------------------------------------
The design of the single-use vacuum and rinse tubing sets has been changes to contain them in an ABS plastic housing.
#### Performance Testing Summary:
To demonstrate substantial equivalence of the subject device, the EnCor Enspire™ Breast Biopsy System to the predicate device, the technological characteristics and performance criterion were evaluated. Using the FDA draft guidance document, "Design Control Guidance for Medica! Device Manufacturers" dated March 11, 1997, FDA guidance document "General Principles of Software Validation: Final Guidance for Industry and FDA Staff," issued January 11, 2002 and internal Risk Assessment procedures, the following non-clinical tests were performed:
- . Vacuum Level
- Joint Tensile Strength .
- . Cassette Functionality
- ◆ Rinse Volume
- . Cassette and Vacuum Reliability
- . Packaging Validation
- User Validation .
- . Software Verification and Validation
The results from these tests demonstrate that the technological characteristics and performance criteria of the EnCor Enspire™ Breast Biopsy System are comparable to the predicate device and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
#### Conclusions:
The outcome of the risk management activities demonstrate the EnCor Enspire™ Breast Biopsy System presents an acceptable level of risk when used within its intended use. The RA, DFMEA, and design verification and validation activities demonstrate the design outputs of the subject device meet the device inputs and user need requirements.
In summary, the subject devise, the EnCor Enspire™ Breast Biopsy System, is substantially equivalent to the legally marketed predicate device, the EnCor® Breast
Bard Peripheral Vascular, Inc.
Image /page/3/Picture/17 description: The image shows the word "BARD" in a bold, outlined font. The letters are all capitalized and evenly spaced. The font style gives the word a blocky, geometric appearance.
Page 23
{4}------------------------------------------------
Biopsy System (K093512).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
C. R. Bard, Inc. % Bard Peripheral Vascular, Inc. Mr. Justin Lovelace 1415 West Third Street Tempe, Arizona 85281-1740
JUN 1 6 2011
Re: K111100
Trade/Device Name: EnCor Enspire™ Breast Biopsy System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: May 18, 2011 Received: May 19, 2011
Dear Mr. Lovelace:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{6}------------------------------------------------
Page 2 - Mr. Justin Lovelace
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Eric Keith
or Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement
EnCor Enspire™ Breast Biopsy System
K111100 510(k) Number (if known):
Device Name:
Indications for Use:
The EnCor Enspire™ Breast Biopsy System is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities.
- . It is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
- lt is intended to provide breast tissue for histologic examination . with partial removal of a palpable abnormality. The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the EnCor Enspire™Breast Biopsy System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
:AND/OR
Over the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.