K111064 · Laax, Inc. · GDW · May 18, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111064
Device Name
TIGERPAW SYSTEM II
Applicant
Laax, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
May 18, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TigerPaw® System II is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, without assistance from a camera, endoscope, or any other viewing technology. This includes procedures performed by sternotomy (full or partial) as well as thoracotomy (single or multiple).
Device Story
The TigerPaw® System II is a mechanical implantable fastener and delivery tool used for left atrial appendage (LAA) occlusion during open cardiac surgery. The system comprises a fastener made of linearly spaced connectors overmolded with soft silicone to conform to anatomy, and a hand-held, disposable, pistol-grip delivery tool. The surgeon operates the tool under direct visualization; two triggers on the handle are actuated to close opposing jaws, mate the fastener sides, and release the implant. The device includes a visual indicator for redundant confirmation of first-trigger closure. The system is used by cardiac surgeons in an operating room setting to exclude the LAA, potentially reducing the risk of thromboembolic events. The device is single-use and sterile.
Clinical Evidence
Bench testing only. Evidence includes in vitro bench evaluations (connector tensile testing, pressure leak testing) and biocompatibility testing. No clinical data provided.
Technological Characteristics
Implantable fastener composed of linearly spaced connectors overmolded with soft silicone. Hand-held, disposable, pistol-grip delivery tool with two-trigger actuation. Features a 15° angled jaw and a visual indicator for first-trigger closure. Sterile, single-use device. Mechanical operation; no software or electronic components.
Indications for Use
Indicated for occlusion of the left atrial appendage in patients undergoing open cardiac surgical procedures requiring direct visualization (sternotomy or thoracotomy).
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
K111064
Special 510(k): Device Modification TigerPaw® System II
# 510(K) SUMMARY
MAY 18 2011
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### 510(k) Number: TBD
#### Applicant Information:
Date Prepared: April 15, 2011
| Name: | LAAx Inc. |
|----------|----------------------------------------------------|
| Address: | 151 Lindbergh Ave., Suite I<br>Livermore, CA 94551 |
| Phone: | 925-449-4108 |
| Contact Persons (Primary/Secondary): | Michael A. Daniel | / | Christopher R. Timmons |
|--------------------------------------|-------------------|---|------------------------|
| Office Phone Numbers: | 925-254-5228 | / | 925-449-4108 |
| Cell Phone Numbers: | 415-407-0223 | / | 925-699-4072 |
| Facsimile Numbers: | 925-254-5187 | / | 925-449-1845 |
#### Device Information:
Classification: Class II Trade Name: TigerPaw® System II Common Name: Implantable Fastener and Accessories Classification Name: Staple, Implantable, GDW / Staple Line Reinforcement Material, FTL Regulation Number: 21 CFR 878.4750 21 CFR 870.3470
# Predicate Devices:
The TigerPaw® System II is substantially equivalent in intended use and method of operation to the following predicate device:
LAAx, Inc. TigerPaw® System (K101961)
#### Device Description:
The TigerPaw® System II consists of an implantable Fastener and Delivery Tool designed for occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures. The Fastener is composed of linearly spaced connectors overmolded with a soft silicone. The spacing between the connectors along with the silicone gives the Fastener an ability to conform to the anatomy and exclude the LAA. The TigerPaw System II also includes a hand-held disposable Delivery Tool. The Delivery Tool consists of a pistol grip handle with two triggers that when separately actuated, close opposing jaws thus mating the
{1}------------------------------------------------
two sides of the permanently implanted Fastener, and, then allowing release of the Fastener and opening of the jaws for Delivery Tool removal.
# Indications for Use:
The TigerPaw® System II is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, without assistance from a camera, endoscope, or any other viewing technology. This includes procedures performed by sternotomy (full or partial) as well as thoracotomy (single or multiple).
#### Comparison to Predicate Device(s):
The design of the TigerPaw® System II is essentially the same as the predicate device with design modifications to increase manufacturing efficiencies, make the system less expensive to produce, and add a 1st Trigger Closure visual indicator for redundant confirmation of first trigger closure. This continues to be a device designed to occlude the left atrial appendage under direct visualization. The products share common features such as identical operating principles, same scissor type jaw actuation, same fastener closure distance, and the same 15° angled jaw. The devices are packaged sterile and are for single use.
The TigerPaw® System II and the predicate device have the same intended use: Occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.
# Device Evaluation:
Verification testing provided proof that the modifications met all design specifications. Extensive bench evaluations were performed on the TigerPaw® II System. This testing consisted of in virro bench testing, including connector tensile testing and pressure leak testing of fasteners, Biocompatibility testing provided evidence of continued biocompatibility of the revised device.
# Summary:
Based upon the intended use, product technical information, performance testing and biocompatibility information provided in this pre-market notification, the TigerPaw System II has been shown to be substantially equivalent to the currently cleared predicate device in terms of design, performance and intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
LAAx, Inc. c/o Mr. Michael A. Daniel Regulatory and Clinical Affairs 151 Lindbergh Avenue, Suite I Livermore, CA 94551
MAY 18 2011
Re: K111064 TigerPaw® System II Regulation Number: 21 CFR 870.4750 Regulation Name: Staple, Implantable Regulatory Class: Class II (two) Product Code: GDW/FTL Dated: April 15, 2011 Received: April 18, 2011
Dear Mr. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Michael Daniel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
B. D. Zuker, M.P.
Bram D. Zuckeman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K11064
Device Name:
TigerPaw® System II
Indications For Use:
The TigerPaw® System II is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, without assistance from a camera, endoscope, or any other viewing technology. This includes procedures performed by sternotomy (full or partial) as well as thoracotomy (single or multiple).
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K110641 |
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.