ESOPHAGEAL / RECTAL TEMPERATURE PROBE, 9FR, 12FR, ESOPHAGEAL STETHOSCOPE WITH TEMPERATURE SENSOR, 9FR, 12FR, 18FR, 24FR
Applicant
Starboard Medical, LLC
Product Code
BZT · Anesthesiology
Decision Date
Apr 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1920
Device Class
Class 2
Indications for Use
Esophageal / Rectal temperature probe (4009-ER and 40012-ER): The Starboard Medical Esophageal / Rectal temperature probe is indicated for continuous patient temperature monitoring, when these placement sites are clinically recommended. The probe is designed for insertion into the esophagus, nasopharynx, or rectum. Esophageal Stethoscope with temperature sensor (4009-ES, 40012-ES, 40018-ES, and 40024-ES): The Starboard Medical Esophageal Stethoscope with temperature sensor is indicated for continuous monitoring of patient temperature along with auscultation of the heart and lung sounds. The probe is designed for insertion into the esophagus.
Device Story
Single-use, sterile temperature probes and esophageal stethoscopes; compatible with YSI 400 series monitoring instrumentation. Probes inserted into esophagus, nasopharynx, or rectum for continuous temperature sensing; stethoscope models include acoustic channel for heart/lung auscultation. Used in clinical settings by healthcare providers to monitor patient thermal status and vital sounds. Output displayed on external monitoring equipment; assists clinicians in real-time physiological assessment and patient management.
Clinical Evidence
Bench testing only. Includes raw materials biocompatibility testing, accuracy testing, and electrical testing/comparison to predicate devices.
Technological Characteristics
Compatible with YSI 400 series monitoring instrumentation. Single-use, sterile, disposable probes. Available in various French (FR) sizes (9FR, 12FR, 18FR, 24FR). Sensing principle based on electrical temperature transduction.
Indications for Use
Indicated for continuous patient temperature monitoring in clinical settings requiring esophageal, nasopharyngeal, or rectal placement. Esophageal stethoscopes also indicated for concurrent heart and lung sound auscultation. No specific age or gender restrictions noted.
Regulatory Classification
Identification
An esophageal stethoscope with electrical conductors is a device that is inserted into the esophagus to listen to a patient's heart and breath sounds and to monitor electrophysiological signals. The device may also incorporate a thermistor for temperature measurement.
Predicate Devices
SMITHS LEVEL 1 Esophageal / Rectal temperature probe (K863646)
SMITHS LEVEL 1 Esophageal Stethoscope with temperature sensor (K854212)
Submission Summary (Full Text)
{0}------------------------------------------------
#### ട്. 510(k) Summarv
Date prepared: March 2, 2011
This 510(k) is being submitted by Anthony Beran on behalf of Starboard Medical, LLC.
# Contact:
Anthony Beran President Starboard Medical, LLC 22845-H Savi Ranch Parkway Yorba Linda, CA 92887 Tel: 714 283 3099 Fax: 714 283 3033 Email: aberan@cox.net FDA Establishment Registration #: 3006845683
# Contract Manufacturer:
Starboard Medical Jiangsu Tel: 86-515-82306811 FDA Establishment Registration#: 3006845687
## Trade Name:
The device trade name is Disposable Temperature probes/ sensors.
# Device common, usual, or classification names:
Esophageal / Rectal temperature probe Esophageal Stethoscope with temperature sensor
# Classification:
Class II, product code BZT Classification Panel: Anesthesiology, Reg# 21CFR 868.1920
# Predicate Device:
The following devices have been identified as predicate devices: SMITHS LEVEL 1 Esophageal / Rectal temperature probe - K863646
SMITHS LEVEL 1 Esophageal Stethoscope with temperature sensor - K854212
# Description of device:
The Starboard Medical temperature probes are intended for use in clinical situations where continuous monitoring of patient's body temperatures is required. The probes are compatible with all monitoring instrumentation designed to accept YSI 400 series temperature probes or equivalent.
# This 510(k) includes the following probes:
4009-ER (Esophageal / Rectal temperature probe - 9FR) 40012-ER (Esophageal / Rectal temperature probe - 12FR) 4009-ES (Esophageal Stethoscope with temperature sensor - 9FR)
{1}------------------------------------------------
40012-ES (Esophageal Stethoscope with temperature sensor - 12FR) 40018-ES (Esophageal Stethoscope with temperature sensor - 18FR) 40024-ES (Esophageal Stethoscope with temperature sensor - 24FR)
The probes are single use, and they are sterile.
# Intended Use:
The Starboard Medical Esophageal / Rectal temperature probe is indicated for continuous patient temperature monitoring, when these placement sites are clinically recommended. The sensor is designed for insertion into the esophaqus, nasopharynx, or rectum.
The Starboard Medical Esophageal Stethoscope with temperature sensor is indicated for continuous monitoring of patient temperature along with auscultation of the heart and lung sounds.
# Technology Characteristics:
Both devices are substantially equivalent to the predicate devices based on material, technology, manufacturing processes, and performance.
# Performance Data:
Both devices have been subjected to raw materials bio-compatibility testing, accuracy testing, and electrical testing and comparison.
# Conclusion:
We believe the differences between the Starboard Medical devices and the predicate devices are minor, and conclude that the subject devices are as safe and effective as the predicate devices.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Anthony Beran President Starboard Medical, LLC 22845 Savi Ranch Parkway, Suite H Yorda Linda, California 92887
APR 1 2 2012
Re: K111050
Trade/Device Name: Esophageal/ Rectal Temperature Probe, Catalog Numbers 4009-ER and 40012-ER. Esophageal Stethoscope with Temperature Sensor, Catalog
Numbers 4009-ES, 40012-ES, 40018-ES, and 40024-ES.
Regulation Number: 21 CFR 868.1920
Regulation Name: Esophageal stethoscope with electrical conductors
Regulatory Class: II
Product Code: BZT
Dated: March 30, 2012
Received: April 5, 2012
Dear Mr. Beran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{3}------------------------------------------------
# Page 2- Mr. Beran
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Antony B. nratin
Anthony D. Watson, B.S., M.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
### Enclosure
{4}------------------------------------------------
#### Indications for Use Statement 4.
# 510(k) Number (if known): Unknown.
## Device Name(s):
Esophageal / Rectal Temperature Probe, Catalog Numbers 4009-ER and 40012-ER. Esophageal Stethoscope with Temperature Sensor, Catalog Numbers 4009-ES, 40012-ES, 40018-ES, and 40024-ES.
## Indications for Use:
Esophageal / Rectal temperature probe (4009-ER and 40012-ER):
The Starboard Medical Esophageal / Rectal temperature probe is indicated for continuous patient temperature monitoring, when these placement sites are clinically recommended. The probe is designed for insertion into the esophagus, nasopharynx, or rectum.
Esophageal Stethoscope with temperature sensor (4009-ES, 40012-ES, 40018-ES, and 40024-ES):
The Starboard Medical Esophageal Stethoscope with temperature sensor is indicated for oontinuous monitoring of patient temperature along with auscultation of the heart and lung sounds. The probe is designed for insertion into the esophagus.
Prescription Use: XXXXX (Part 21CFR801 Subpart D) and/or
Over-the-Counter Use: (Part 21CFR801 Subpart C)
111050
Hild C. Chapman 4/12/12
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.